Centrifugal partition chromatography

Last updated

Centrifugal partition chromatography is a special chromatographic technique where both stationary and mobile phase are liquid, and the stationary phase is immobilized by a strong centrifugal force. Centrifugal partition chromatography consists of a series-connected network of extraction cells, which operates as elemental extractors, and the efficiency is guaranteed by the cascade. [1]

Contents

History

In the 1940s Craig invented the first apparatus to conduct countercurrent partitioning; he called this Countercurrent Distribution Craig apparatus consists of a series of glass tubes that are designed and arranged such that the lighter liquid phase is transferred from one tube to the next. The next major milestone was Droplet countercurrent chromatography (DCCC). It uses only gravity to move the mobile phase through the stationary phase which is held in long vertical tubes connected in series. The modern era of CCC began with the development of the planetary centrifuge by Ito which was first introduced in 1966 as a closed helical tube which was rotated on a "planetary" axis as is turned on a "sun" axis. Centrifugal partition chromatography was introduced in Japan in 1982; the first instrument was built at Sanki Eng. Ltd. in Kyoto. The first instrument consisted of twelve cartridges arranged around the rotor of a centrifuge; the inner volume of each cartridge was about 15 mL for 50 channels. In 1999 Kromaton was developed of the first FCPC with Radial Cells. During cell development, the Z cell was completed in 2005 and the twin cell in 2009. In 2017 RotaChrom designed its top performing CPC cells through Computed Fluid Dynamic simulation software. After thousands of simulations, this tool revealed the drawbacks of conventional CPC cell designs and highlighted the unparallel load capacity and scalable cell design of RotaChrom. [2]

Operation

The extraction cells consist of hollow bodies with inlets and outlets of liquid connection. The cells are first filled with the liquid chosen to be the stationary phase. Under rotation, the pumping of the mobile phase is started, which enters the cells from the inlet. When entering the flow of mobiles phase forms small droplets according to the Stokes' law, which is called atomization. These droplets fall through the stationary phase, creating a high interface area, which is called the extraction. At the end of the cells, these droplets unite due to the surface tension, which is called settling.

When a sample mixture is injected as a plug into the flow of mobile phase the compounds of the mixtures elute according to their partition coefficients:

Centrifugal partition chromatography requires only a biphasic mixture of solvents, so by varying the constitution of the solvent system it is possible to tune the partition coefficients of different compounds so that separation is guaranteed by the high selectivity.

Comparison with countercurrent chromatography

Countercurrent chromatography and centrifugal partition chromatography are two different instrumental realization of the same liquid–liquid chromatographic theory. Countercurrent chromatography usually uses a planetary gear motion without rotary seals, while centrifugal partition chromatography uses circular rotation with rotary seals for liquid connection. CCC has interchanging mixing and settling zones in the coil tube, so atomization, extraction and settling are time and zone separated. Inside centrifugal partition chromatography, all three steps happen continuously in one time, inside the cells.

Advantages of centrifugal partition chromatography:

Disadvantages of centrifugal partition chromatography:

Laboratory scale

Centrifugal partition chromatography has been extensively used for isolation and purification of natural products for 40 years. [4] Due to the ability to get very high selectivity, and the ability to tolerate samples containing particulated matter, it is possible to work with direct extracts of biomass, opposed to traditional liquid chromatography, where impurities degrade the solid stationary phase so that separation become impossible.

There are numerous laboratory scale centrifugal partition chromatography manufacturers around the world, like Gilson (Armen Instrument), Kromaton (Rousselet Robatel), and AECS-QUIKPREP. These instruments operate at flow rates of 1–500 mL/min. with stationary phase retentions of 40–80%.

Production scale

Centrifugal partition chromatography does not uses any solid stationary phase, so it guarantees a cost-effective separation for the highest industrial levels. As opposed to countercurrent chromatography, it is possible to get very high flow rates (for example 10 liters / min) with active stationary phase ratio of >80%, which guarantees good separation and high productivity. As in centrifugal partition chromatography, material is dissolved, and loaded the column in mass / volume units, loading capability can be much higher than standard solid-liquid chromatographic techniques, where material is loaded to the active surface area of the stationary phase, which takes up less than 10% of the column.

Industrial instrument like Gilson (Armen Instrument), Kromaton (Rousselet Robatel) and RotaChrom Technologies (RotaChrom) differ from laboratory scale instruments by the applicable flow rate with satisfactory stationary phase retention (70-90%). Industrial instruments have flow rates of multiple liter / minutes, while able to purify materials from 10 kg to tonnes per month.

Operating the production scale equipment requires industrial volume solvent preparation (mixer/settler) and solvent recovery equipment.

See also

Related Research Articles

In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent called the mobile phase, which carries it through a system on which a material called the stationary phase is fixed. Because the different constituents of the mixture tend to have different affinities for the stationary phase and are retained for different lengths of time depending on their interactions with its surface sites, the constituents travel at different apparent velocities in the mobile fluid, causing them to separate. The separation is based on the differential partitioning between the mobile and the stationary phases. Subtle differences in a compound's partition coefficient result in differential retention on the stationary phase and thus affect the separation.

<span class="mw-page-title-main">High-performance liquid chromatography</span> Technique used to separate components of a liquid mixture

High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify each component in a mixture. It relies on pumps to pass a pressurized liquid solvent containing the sample mixture through a column filled with a solid adsorbent material. Each component in the sample interacts slightly differently with the adsorbent material, causing different flow rates for the different components and leading to the separation of the components as they flow out of the column.

<span class="mw-page-title-main">Gas chromatography</span> Type of chromatography

Gas chromatography (GC) is a common type of chromatography used in analytical chemistry for separating and analyzing compounds that can be vaporized without decomposition. Typical uses of GC include testing the purity of a particular substance, or separating the different components of a mixture. In preparative chromatography, GC can be used to prepare pure compounds from a mixture.

<span class="mw-page-title-main">Paper chromatography</span> Separation of coloured chemicals on paper

Paper chromatography is an analytical method used to separate coloured chemicals or substances. It is now primarily used as a teaching tool, having been replaced in the laboratory by other chromatography methods such as thin-layer chromatography (TLC).

<span class="mw-page-title-main">Column chromatography</span> Method to isolate a compound in a mixture

Column chromatography in chemistry is a chromatography method used to isolate a single chemical compound from a mixture. Chromatography is able to separate substances based on differential adsorption of compounds to the adsorbent; compounds move through the column at different rates, allowing them to be separated into fractions. The technique is widely applicable, as many different adsorbents can be used with a wide range of solvents. The technique can be used on scales from micrograms up to kilograms. The main advantage of column chromatography is the relatively low cost and disposability of the stationary phase used in the process. The latter prevents cross-contamination and stationary phase degradation due to recycling. Column chromatography can be done using gravity to move the solvent, or using compressed gas to push the solvent through the column.

<span class="mw-page-title-main">Liquid chromatography–mass spectrometry</span> Analytical chemistry technique

Liquid chromatography–mass spectrometry (LC–MS) is an analytical chemistry technique that combines the physical separation capabilities of liquid chromatography with the mass analysis capabilities of mass spectrometry (MS). Coupled chromatography - MS systems are popular in chemical analysis because the individual capabilities of each technique are enhanced synergistically. While liquid chromatography separates mixtures with multiple components, mass spectrometry provides spectral information that may help to identify each separated component. MS is not only sensitive, but provides selective detection, relieving the need for complete chromatographic separation. LC–MS is also appropriate for metabolomics because of its good coverage of a wide range of chemicals. This tandem technique can be used to analyze biochemical, organic, and inorganic compounds commonly found in complex samples of environmental and biological origin. Therefore, LC–MS may be applied in a wide range of sectors including biotechnology, environment monitoring, food processing, and pharmaceutical, agrochemical, and cosmetic industries. Since the early 2000s, LC–MS has also begun to be used in clinical applications.

<span class="mw-page-title-main">Solid-phase extraction</span> Process to separate compounds by properties

Solid-phase extraction (SPE) is a solid-liquid extractive technique by which compounds that are dissolved or suspended in a liquid mixture are separated from other compounds in the mixture according to their physical and chemical properties. Analytical laboratories use solid phase extraction to concentrate and purify samples for analysis. Solid phase extraction can be used to isolate analytes of interest from a wide variety of matrices, including urine, blood, water, beverages, soil, and animal tissue.

<span class="mw-page-title-main">Fast protein liquid chromatography</span>

Fast protein liquid chromatography (FPLC) is a form of liquid chromatography that is often used to analyze or purify mixtures of proteins. As in other forms of chromatography, separation is possible because the different components of a mixture have different affinities for two materials, a moving fluid (the mobile phase) and a porous solid (the stationary phase). In FPLC the mobile phase is an aqueous buffer solution. The buffer flow rate is controlled by a positive-displacement pump and is normally kept constant, while the composition of the buffer can be varied by drawing fluids in different proportions from two or more external reservoirs. The stationary phase is a resin composed of beads, usually of cross-linked agarose, packed into a cylindrical glass or plastic column. FPLC resins are available in a wide range of bead sizes and surface ligands depending on the application.

Countercurrent distribution is an analytical chemistry technique which was developed by Lyman C. Craig in the 1940s. Countercurrent distribution is a separation process that is founded on the principles of liquid–liquid extraction where a chemical compound is distributed (partitioned) between two immiscible liquid phases according to its relative solubility in the two phases. The simplest form of liquid-liquid extraction is the partitioning of a mixture of compounds between two immiscible liquid phases in a separatory funnel. This occurs in five steps: 1) preparation of the separatory funnel with the two phase solvent system, 2) introduction of the compound mixture into the separatory funnel, 3) vigorous shaking of the separatory funnel to mix the two layers and allow for mass transfer of compounds in and out of the phases, 4) The contents of the separatory funnel are allowed to settle back into two distinct phases and 5) the two phases are separated from each other by draining out the bottom phase. If a compound is insoluble in the lower phase it will distribute into the upper phase and stay in the separatory funnel. If a compound is insoluble in the upper phase it will distribute into the lower phase and be removed from the separatory funnel. If the mixture contains one or more compounds that are soluble in the upper phase and one or more compounds that are soluble in the lower phase, then an extraction has occurred. Often, an individual compound is soluble to a certain extent in both phases and the extraction is, therefore, incomplete. The relative solubility of a compound in two phases is known as the partition coefficient.

In manufacturing, the simulated moving bed (SMB) process is a highly engineered process for implementing chromatographic separation. It is used to separate one chemical compound or one class of chemical compounds from one or more other chemical compounds to provide significant quantities of the purified or enriched material at a lower cost than could be obtained using simple (batch) chromatography. It cannot provide any separation or purification that cannot be done by a simple column purification. The process is rather complicated. The single advantage which it brings to a chromatographic purification is that it allows the production of large quantities of highly purified material at a dramatically reduced cost. The cost reductions come about as a result of: the use of a smaller amount of chromatographic separation media stationary phase, a continuous and high rate of production, and decreased solvent and energy requirements. This improved economic performance is brought about by a valve-and-column arrangement that is used to lengthen the stationary phase indefinitely and allow very high solute loadings to the process.

<span class="mw-page-title-main">Hydrophilic interaction chromatography</span> Type of chromatography

Hydrophilic interaction chromatography is a variant of normal phase liquid chromatography that partly overlaps with other chromatographic applications such as ion chromatography and reversed phase liquid chromatography. HILIC uses hydrophilic stationary phases with reversed-phase type eluents. The name was suggested by Andrew Alpert in his 1990 paper on the subject. He described the chromatographic mechanism for it as liquid-liquid partition chromatography where analytes elute in order of increasing polarity, a conclusion supported by a review and re-evaluation of published data.

Micellar liquid chromatography (MLC) is a form of reversed phase liquid chromatography that uses an aqueous micellar solutions as the mobile phase.

<span class="mw-page-title-main">Elution</span> Extraction of a material by washing with a solvent

In analytical and organic chemistry, elution is the process of extracting one material from another by washing with a solvent; as in washing of loaded ion-exchange resins to remove captured ions.

Partition chromatography theory and practice was introduced through the work and publications of Archer Martin and Richard Laurence Millington Synge during the 1940s. They would later receive the 1952 Nobel Prize in Chemistry "for their invention of partition chromatography".

A monolithic HPLC column, or monolithic column, is a column used in high-performance liquid chromatography (HPLC). The internal structure of the monolithic column is created in such a way that many channels form inside the column. The material inside the column which separates the channels can be porous and functionalized. In contrast, most HPLC configurations use particulate packed columns; in these configurations, tiny beads of an inert substance, typically a modified silica, are used inside the column. Monolithic columns can be broken down into two categories, silica-based and polymer-based monoliths. Silica-based monoliths are known for their efficiency in separating smaller molecules while, polymer-based are known for separating large protein molecules.

<span class="mw-page-title-main">Countercurrent chromatography</span>

Countercurrent chromatography is a form of liquid–liquid chromatography that uses a liquid stationary phase that is held in place by inertia of the molecules composing the stationary phase accelerating toward the center of a centrifuge due to centripetal force and is used to separate, identify, and quantify the chemical components of a mixture. In its broadest sense, countercurrent chromatography encompasses a collection of related liquid chromatography techniques that employ two immiscible liquid phases without a solid support. The two liquid phases come in contact with each other as at least one phase is pumped through a column, a hollow tube or a series of chambers connected with channels, which contains both phases. The resulting dynamic mixing and settling action allows the components to be separated by their respective solubilities in the two phases. A wide variety of two-phase solvent systems consisting of at least two immiscible liquids may be employed to provide the proper selectivity for the desired separation.

A separation process is a method that converts a mixture or a solution of chemical substances into two or more distinct product mixtures, a scientific process of separating two or more substance in order to obtain purity. At least one product mixture from the separation is enriched in one or more of the source mixture's constituents. In some cases, a separation may fully divide the mixture into pure constituents. Separations exploit differences in chemical properties or physical properties between the constituents of a mixture.

<span class="mw-page-title-main">Capillary electrochromatography</span> Method of separating components of a mixture via electro-osmosis

In chemical analysis, capillary electrochromatography (CEC) is a chromatographic technique in which the mobile phase is driven through the chromatographic bed by electro-osmosis. Capillary electrochromatography is a combination of two analytical techniques, high-performance liquid chromatography and capillary electrophoresis. Capillary electrophoresis aims to separate analytes on the basis of their mass-to-charge ratio by passing a high voltage across ends of a capillary tube, which is filled with the analyte. High-performance liquid chromatography separates analytes by passing them, under high pressure, through a column filled with stationary phase. The interactions between the analytes and the stationary phase and mobile phase lead to the separation of the analytes. In capillary electrochromatography capillaries, packed with HPLC stationary phase, are subjected to a high voltage. Separation is achieved by electrophoretic migration of solutes and differential partitioning.

Droplet countercurrent chromatography was introduced in 1970 by Tanimura, Pisano, Ito, and Bowman. DCCC is considered to be a form of liquid-liquid separation, which includes countercurrent distribution and countercurrent chromatography, that employs a liquid stationary phase held in a collection of vertical glass columns connected in series. The mobile phase passes through the columns in the form of droplets. The DCCC apparatus may be run with the lower phase stationary and the upper phase being introduced to the bottom of each column. Or it may be run with the upper phase stationary and the lower phase being introduced from the top of the column. In both cases, the work of gravity is allowed influence the two immiscible liquids of different densities to form the signature droplets that rise or descend through the column. The mobile phase is pumped at a rate that will allow droplets to form that maximize the mass transfer of a compound between the upper and lower phases. Compounds that are more soluble in the upper phase will travel quickly through the column, while compounds that are more soluble in the stationary phase will linger. Separation occurs because different compounds distribute differently, in a ratio called the partition coefficient, between the two phases.

An evaporative light scattering detector (ELSD) is a detector used in conjunction with high-performance liquid chromatography (HPLC), Ultra high-performance liquid chromatography (UHPLC), Purification liquid chromatography such as flash or preparative chromatography, countercurrent or centrifugal partition chromatographies and Supercritical Fluid chromatography (SFC). It is commonly used for analysis of compounds where UV detection might be a restriction and therefore used where compounds do not efficiently absorb UV radiation, such as sugars, antivirals, antibiotics, fatty acids, lipids, oils, phospholipids, polymers, surfactants, terpenoids and triglycerides. ELSDs is related to the charged aerosol detector (CAD) and like the CAD, falls under the category of destructive detectors.

References

  1. P, Foucault, Alain (1995). Centrifugal Partition Chromatography. New York: Marcel Dekker, Inc. ISBN   0-8247-9257-2.
  2. "US Patent Application for Extraction cell for a centrifugal partition chromatograph, a centrifugal partition chromatograph containing such a cell, and a method for producing such an extraction cell Patent Application (Application #20180280830 issued October 4, 2018) - Justia Patents Search".
  3. Laszlo, Lorantfy. "Development of Industrial Scale CPC". ResearchGate. Laszlo Lorantfy. Retrieved 2016-03-21.
  4. Guido, F. Pauli (2008). "Countercurrent Separation of Natural Products". Journal of Natural Products. 71 (8): 1489–508. doi:10.1021/np800144q. PMID   18666799.