Clarke Error Grid

Last updated
The Clarke Error Grid Clarkeerrorgrid.gif
The Clarke Error Grid

The Clarke Error Grid Analysis (EGA) was developed in 1987 to quantify the clinical accuracy of patient estimates of their current blood glucose as provided by a blood glucose meter. [1] This technique was then applied to quantify the clinical accuracy of blood glucose estimates generated by personal meters as compared to values from a laboratory reference test. A description of the EGA appeared in Diabetes Care in 1987. [2] Eventually, the EGA became accepted as one of the “gold standard” techniques for determining the accuracy of blood glucose meters.

To perform the analysis, blood glucose samples are obtained and each sample is divided and measured by two meters: a "reference method" (a meter or laboratory test known to produce accurate results); and a new meter being evaluated. A scatterplot is generated, plotting values from the reference glucose meter (x-axis) against values from the new device (y-axis); thus, each point represents measurement of a single sample by two methods.

Grid lines are then drawn to indicate 5 different classifications (Regions A-E) based on clinical relevance:

See also

References

  1. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL: Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 10:622–628,1987
  2. Clarke, William L.; Cox, Daniel; Gonder-Frederick, Linda A.; Carter, William; Pohl, Stephen L. (1987). "Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood Glucose" . Diabetes Care. 10 (5): 622–628. doi:10.2337/diacare.10.5.622. PMID   3677983. S2CID   26388964.