Resonance frequency analysis (RFA) is a method used to determine stability (the level of osseointegration) in dental implants. [1] [2] The stability is presented as an implant stability quotient (ISQ) value. The higher the ISQ value the higher the stability.

Contents

Utilizing RFA involves sending magnetic pulses to a small metal rod temporarily attached to the implant. As the rod vibrates, the probe reads its resonance frequency and translates it into an ISQ value.

RFA measurements are used to assess the stability of the implant immediately after placement, as well as to measure the stability during the healing time. This helps the dentist determine if further healing time (osseointegration) is needed before the prosthetic tooth is attached, as well as to identify at-risk patients with compromised bone tissue, or other risk factors. [3]

History

Resonance frequency analysis was first suggested as an alternative method of analyzing peri-implant bone in a scientific paper by Meredith N et al in 1996. [4] As stated in the paper’s abstract, in measuring implant stability and osseointegration, “radiographs are of value, but a standardised technique is necessary to ensure repeatability.” [4] The new technique tested involved connecting a small transducer (aluminum rod) to implants. Measurements showed that the resonance frequency increased in direct relation to the increase in the stiffness of the bone-implant interface, [4] thus demonstrating a repeatable and quantitative method of assessing stability.

The method underwent further research and in 1999 the Sweden-based company Osstell AB formed to commercialize the new technique. [5] Osstell developed a device that transmitted vibrational frequencies to a metal peg inserted in the implant and measured the frequency at which resonance was reached. Whereas Meredith et al. measured in the range of 3500–8500 kHz, Osstell developed the Implant Stability Quotient (ISQ) that translated this kHz range to a score 1–100. [6]

Resonance frequency analysis has considerable scientific interest since its advent, largely owing to the increasing number of patients demanding dental implants as the technology improves. [7] As it is a non-invasive and objective way to evaluate short- and long-term implant viability, RFA is an increasingly utilized method. [8]

Scientific foundations

The method that preceded RFA, percussion or “tapping,” may be used to understand the underlying functionality of RFA devices as the same principles are at work. When an implant was percussed with a blunt instrument, the nature of the sound elicited would qualitatively indicate the level of the implant’s stability. [9] A low pitched, dull sound (low frequency) indicated a loose bond with the bone, as the vibrations moved slower across the distance between the implant and surrounding tissue. [8] A high pitched, crystalline sound indicated a tight connection along the implant-bone interface, with vibrations moving quicker across a more restricted area. The dentist would make a qualitative assessment of the level of stability based on the sound heard. [10]

With RFA, vibrations are being used to determine stability, but on a micro scale and in a non-invasive manner. A metal peg (transducer) with a magnet top is attached to the implant. Magnetic pulses (alternating sine waves of uniform amplitude) cause the peg to vibrate, increasing steadily in pitch until the implant resonates. [11] The higher the resonant frequency, the more stable the implant.

Measurement

The frequency readings, translated to an Implant Stability Quotient (ISQ), are used as an assessment and ongoing monitoring tool. [12] Medical interpretations of ISQ values may then be used to inform treatment plans, as analyzed and documented in hundreds of clinical studies. [13]

A reading of 55 or below indicates that too much lateral movement is possible, and the implant needs to reach secondary stability (greater bonding with the bone) before the prosthesis may be attached. [14] If the resonance frequency reading increases, it signals that osseointegration is occurring. [14] Along with other diagnostic tools, measurements over time can be used to indicate the rate of osseointegration, and treatment plans may be assigned accordingly. If the rate is initially low and does not increase, it signals the implant is not viable. [10]

Related Research Articles

<span class="mw-page-title-main">Bridge (dentistry)</span> Dental restoration for missing teeth

A bridge is a fixed dental restoration used to replace one or more missing teeth by joining an artificial tooth definitively to adjacent teeth or dental implants.

<span class="mw-page-title-main">Dental implant</span> Surgical component that interfaces with the bone of the jaw

A dental implant is a prosthesis that interfaces with the bone of the jaw or skull to support a dental prosthesis such as a crown, bridge, denture, or facial prosthesis or to act as an orthodontic anchor. The basis for modern dental implants is a biological process called osseointegration, in which materials such as titanium or zirconia form an intimate bond to the bone. The implant fixture is first placed so that it is likely to osseointegrate, then a dental prosthetic is added. A variable amount of healing time is required for osseointegration before either the dental prosthetic is attached to the implant or an abutment is placed which will hold a dental prosthetic/crown.

Osseointegration is the direct structural and functional connection between living bone and the surface of a load-bearing artificial implant. A more recent definition defines osseointegration as "functional ankylosis ", where new bone is laid down directly on the implant surface and the implant exhibits mechanical stability. Osseointegration has enhanced the science of medical bone and joint replacement techniques as well as dental implants and improving prosthetics for amputees.

<span class="mw-page-title-main">Toothlessness</span> Lacking teeth

Toothlessness, or edentulism, is the condition of having no teeth. In organisms that naturally have teeth, it is the result of tooth loss. Organisms that never possessed teeth can also be described as edentulous. Examples are the members of the former zoological classification order of Edentata, which included anteaters and sloths, as they possess no anterior teeth and no or poorly developed posterior teeth.

<span class="mw-page-title-main">Sinus lift</span>

Maxillary sinus floor augmentation is a surgical procedure which aims to increase the amount of bone in the posterior maxilla, in the area of the premolar and molar teeth, by lifting the lower Schneiderian membrane and placing a bone graft.

Armin Abron is a periodontist, who practices in Washington, D.C. He is also an adjunct assistant professor of periodontology at UNC School of Dentistry in Chapel Hill.

<span class="mw-page-title-main">Crown lengthening</span> Dental procedure

Crown lengthening is a surgical procedure performed by a dentist, or more frequently a periodontist, where more tooth is exposed by removing some of the gingival margin (gum) and supporting bone. Crown lengthening can also be achieved orthodontically by extruding the tooth.

In dentistry, the hydrodynamic or fluid movement theory is one of three main theories developed to explain dentine hypersensitivity, which is a sharp, transient pain arising from stimuli exposure. It states that different types of stimuli act on exposed dentine, causing increased fluid flow through the dentinal tubules. In response to this movement, mechanoreceptors on the pulp nerves trigger the acute, temporary pain of dentine hypersensitivity.

The implant stability quotient (ISQ) is the value on a scale that indicates the level of stability and osseointegration in dental implants. The scale ranges from 1 to 100, with higher values indicating greater stability. The acceptable stability range lies between 55 and 85 ISQ. ISQ values are obtained using resonance frequency analysis (RFA).

<span class="mw-page-title-main">Peri-implantitis</span> Inflammatory disease

Peri-implantitis is a destructive inflammatory process affecting the soft and hard tissues surrounding dental implants. The soft tissues become inflamed whereas the alveolar bone, which surrounds the implant for the purposes of retention, is lost over time.

Peri-implant mucositis is defined as an inflammatory lesion of the peri-implant mucosa in the absence of continuing marginal bone loss.

Platelet-rich fibrin (PRF) or leukocyte- and platelet-rich fibrin (L-PRF) is a derivative of PRP where autologous platelets and leukocytes are present in a complex fibrin matrix to accelerate the healing of soft and hard tissue and is used as a tissue-engineering scaffold for endodontics. PRF falls under FDA Product Code KST, labeling it as a blood draw/Hematology product classifying it as 510(k) exempt.

<span class="mw-page-title-main">Bicon Dental Implants</span>

Bicon Dental Implants is a privately owned company located in Boston, MA. The company specializes in short dental implants that use a locking taper or cold welding connection to secure the abutment to the implant. Bicon is notable and worthy of mention for the following three reasons: First, Bicon implants are extremely short in length. The size of Bicon implants allow them to be placed in regions that are crowded with natural teeth and/or implants, or in regions that would otherwise require bone grafting. Second, the implants do not have the screw-form design typical of other available implants. Third, the abutments are connected to the implant via a locking taper. This is notable from both a medical and engineering standpoint as no other implant company offers an implant with a biological seal at the implant/abutment interface; almost all other implants possess an internal screw to connect their abutments.

<span class="mw-page-title-main">Daniel Y. Sullivan</span>

Daniel Y. Sullivan was an American dentist, prosthodontist, author, and teacher credited with helping to bring the practice of osseointegrated dental implants, or the fusion of bone and titanium inside the mouth, to the United States. He worked alongside Swedish pioneer Per-Ingvar Brånemark to insert the U.S.’s first osseointegrated dental implants in 1982. In later years, he taught the technique to thousands of dentists, served as president of two prestigious dental membership organizations, and co-wrote one of the first textbooks in the United States on the subject of esthetic dental implants.

Osstell AB is a company headquartered in Gothenburg, Sweden that develops, manufactures, and sells devices and accessories used to measure dental implant stability. It was founded in 1999 with the aim of developing and commercializing a device that utilized resonance frequency analysis (RFA) to determine the level of stability of a dental implant.

<span class="mw-page-title-main">Overdenture</span> Removable dental prosthesis

Overdenture is any removable dental prosthesis that covers and rests on one or more remaining natural teeth, the roots of natural teeth, and/or dental implants. It is one of the most practical measures used in preventive dentistry. Overdentures can be either tooth supported or implant supported. It is found to help in the preservation of alveolar bone and delay the process of complete edentulism.

Alloplasty is a surgical procedure performed to substitute and repair defects within the body with the use of synthetic material. It can also be performed in order to bridge wounds. The process of undergoing alloplasty involves the construction of an alloplastic graft through the use of computed tomography (CT), rapid prototyping and "the use of computer-assisted virtual model surgery." Each alloplastic graft is individually constructed and customised according to the patient's defect to address their personal health issue. Alloplasty can be applied in the form of reconstructive surgery. An example where alloplasty is applied in reconstructive surgery is in aiding cranial defects. The insertion and fixation of alloplastic implants can also be applied in cosmetic enhancement and augmentation. Since the inception of alloplasty, it has been proposed that it could be a viable alternative to other forms of transplants. The biocompatibility and customisation of alloplastic implants and grafts provides a method that may be suitable for both minor and major medical cases that may have more limitations in surgical approach. Although there has been evidence that alloplasty is a viable method for repairing and substituting defects, there are disadvantages including suitability of patient bone quality and quantity for long term implant stability, possibility of rejection of the alloplastic implant, injuring surrounding nerves, cost of procedure and long recovery times. Complications can also occur from inadequate engineering of alloplastic implants and grafts, and poor implant fixation to bone. These include infection, inflammatory reactions, the fracture of alloplastic implants and prostheses, loosening of implants or reduced or complete loss of osseointegration.

Michael A. Pikos is an American implant oral and maxillofacial surgeon known as the CEO/Founder of Coastal Jaw Surgery and Pikos Institute.

A root-analog dental implant (RAI) – also known as a truly anatomic dental implant, or an anatomical/custom implant – is a medical device to replace one or more roots of a single tooth immediately after extraction. In contrast to common titanium screw type implants, these implants are custom-made to exactly match the extraction socket of the specific patient. Thus there is usually no need for surgery.

Drug-induced gingival enlargement (DIGE), also referred to as drug-induced gingival hyperplasia (DIGH) or drug-induced gingival overgrowth (DIGO), is a side effect of many systemic medications for which the Gingervae are not the target receptor. It is normally resultant of medications including immunoregulators, calcium channel blockers and anticonvulsants. When allowed to progress assisted by routinely poor oral hygiene, DIGE can lead to pain and disfigurement, however there are great variations in its presentation and severity dependent on the case. It is suggested that enlargement is aided by genetic predispositions, tending to occur more frequently in the papillae of the anterior Gingivae in younger age groups.

References

  1. Bilbao A et al. Assessment of dental implant stability in osseodistraction-generated bone: a resonance frequency analysis. Clin Oral Implants Res Apr 28 2009
  2. Veltri M et al. Soft bone primary stability of 3 different miniscrews for orthodontic anchorage: a resonance frequency investigation. Am J Orthod Dentofacial Orthop. 2009 May;135(5):642-8.
  3. Valderrama, Pilar (2007). "Evaluation of Two Different Resonance Frequency Devices to Detect Implant Stability: A Clinical Trial" (PDF). Journal of Periodontology. 78 (2): 262–272. doi:10.1902/jop.2007.060143. PMID   17274715.
  4. 1 2 3 Meredith, N.; Alleyne, D.; Cawley, P. (September 1996). "Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis". Clinical Oral Implants Research. 7 (3): 261–267. doi:10.1034/j.1600-0501.1996.070308.x. ISSN   0905-7161. PMID   9151590.
  5. www.osstell.com, Osstell -. "About - Osstell – Implant Stability". Osstell – Implant Stability. Retrieved 2016-12-13.
  6. Konstantinovic Vitomir (2015). "Assessment of implant stability by resonant frequency analysis" (PDF). Vojnosanit Pregl. 72 (2): 169–174. doi: 10.2298/vsp130801063k . PMID   25831910.
  7. Group, Alexis Martino, Market Research Analyst, The Key (2012-03-21). "The continuing evolution of implant production". staging.dentalproductsreport.advanstar.com. Retrieved 2016-12-13.{{cite web}}: CS1 maint: multiple names: authors list (link)
  8. 1 2 Swami V, Vijayaraghavan V, Swami V (2016). "Current trends to measure implant stability". Journal of Indian Prosthodontic Society. 16 (2): 124–130. doi: 10.4103/0972-4052.176539 . PMC   4837777 . PMID   27141160.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  9. Mistry Gaurang (2014). "Measuring implant stability: A review of different methods". Journal of Dental Implants. 4 (2): 165–169. doi: 10.4103/0974-6781.140891 .
  10. 1 2 Sennerby Lars (2008). "Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications" (PDF). Periodontology 2000. 47: 51–66. doi:10.1111/j.1600-0757.2008.00267.x. PMID   18412573.
  11. Satwalekar, Parth; Nalla, Sandeep; Reddy, Ramaswamy; Chowdary, Sheeba Glory (2015-01-01). "Clinical evaluation of osseointegration using resonance frequency analysis". Journal of Indian Prosthodontic Society. 15 (3): 192–199. doi: 10.4103/0972-4052.165171 . ISSN   0972-4052. PMC   4762340 . PMID   26929512.
  12. Implant Dentistry - A Rapidly Evolving Practice. InTech. 2011. pp. 111–126. ISBN   978-953-307-658-4.
  13. www.osstell.com, Osstell -. "Scientific Database - Osstell – Implant Stability". Osstell – Implant Stability. Retrieved 2016-12-13.
  14. 1 2 Suzuki, Senichi; Kobayashi, Hiroyuki; Ogawa, Takahiro (2013-10-01). "Implant stability change and osseointegration speed of immediately loaded photofunctionalized implants" (PDF). Implant Dentistry. 22 (5): 481–490. doi:10.1097/ID.0b013e31829deb62. ISSN   1538-2982. PMID   24021973. S2CID   25708862.