Pirani gauge

Last updated
Pirani probe, opened Pirani 02.jpg
Pirani probe, opened

The Pirani gauge is a robust thermal conductivity gauge used for the measurement of the pressures in vacuum systems. [1] It was invented in 1906 by Marcello Pirani. [2]

Contents

Marcello Stefano Pirani was a German physicist working for Siemens & Halske which was involved in the vacuum lamp industry. In 1905 their product was tantalum lamps which required a high vacuum environment for the filaments. The gauges that Pirani was using in the production environment were some fifty McLeod gauges, each filled with 2 kg of mercury in glass tubes. [3]

Pirani was aware of the gas thermal conductivity investigations of Kundt and Warburg [4] (1875) published thirty years earlier and the work of Marian Smoluchowski [5] (1898). In 1906 he described his "directly indicating vacuum gauge" that used a heated wire to measure vacuum by monitoring the heat transfer from the wire by the vacuum environment. [2]

Structure

The Pirani gauge consists of a metal sensor wire (usually gold plated tungsten or platinum) suspended in a tube which is connected to the system whose vacuum is to be measured. The wire is usually coiled to make the gauge more compact. The connection is usually made either by a ground glass joint or a flanged metal connector, sealed with an o-ring. The sensor wire is connected to an electrical circuit from which, after calibration, a pressure reading may be taken.

Mode of operation

Block diagram of Pirani gauge Blockbild-Pirani-Vakuumeter.png
Block diagram of Pirani gauge
Curves to convert air readings to other gases Nomogramm pirani-Vakuummeter.jpg
Curves to convert air readings to other gases

In order to understand the technology, consider that in a gas filled system there are four ways that a heated wire transfers heat to its surroundings.

  1. Gas conduction at high pressure (r representing the distance from the heated wire)
  2. Gas transport at low pressure
  3. Thermal radiation
  4. End losses through the support structures

A heated metal wire (sensor wire, or simply sensor) suspended in a gas will lose heat to the gas as its molecules collide with the wire and remove heat. If the gas pressure is reduced, the number of molecules present will fall proportionately and the wire will lose heat more slowly. Measuring the heat loss is an indirect indication of pressure.

There are three possible schemes that can be done. [2]

  1. Keep the bridge voltage constant and measure the change in resistance as a function of pressure
  2. Keep the current constant and measure the change in resistance as a function of pressure
  3. Keep the temperature of the sensor wire constant and measure the voltage as a function of pressure

Note that keeping the temperature constant implies that the end losses(4.) and the thermal radiation losses (3.) are constant. [3]

The electrical resistance of a wire varies with its temperature, so the resistance indicates the temperature of wire. In many systems, the wire is maintained at a constant resistance R by controlling the current I through the wire. The resistance can be set using a bridge circuit. The current required to achieve this balance is therefore a measure of the vacuum.

The gauge may be used for pressures between 0.5 Torr to 1×104 Torr. Below 5×104 Torr, a Pirani gauge has only one significant digit of resolution. The thermal conductivity and heat capacity of the gas affects the readout from the meter, and therefore the apparatus may need calibrating before accurate readings are obtainable. For lower pressure measurement, the thermal conductivity of the gas becomes increasingly smaller and more difficult to measure accurately, and other instruments such as a Penning gauge or Bayard–Alpert gauge are used instead.

Pulsed Pirani gauge

A special form of the Pirani gauge is the pulsed Pirani vacuum gauge where the sensor wire is not operated at a constant temperature, but is cyclically heated up to a certain temperature threshold by an increasing voltage ramp. When the threshold is reached, the heating voltage is switched off and the sensor cools down again. The required heat-up time is used as a measure of pressure.

For adequately low pressure, the following first-order dynamic thermal response model relating supplied heating power and sensor temperature T(t) applies: [6]

where and are specific heat and emissivity of the sensor wire (material properties), and are surface area and mass of the sensor wire, and and are constants determined for each sensor in calibration.

Advantages and disadvantages of the pulsed gauge

Advantages
Disadvantages

Alternative

An alternative to the Pirani gauge is the thermocouple gauge, which works on the same principle of detecting thermal conductivity of the gas by a change in temperature. In the thermocouple gauge, the temperature is sensed by a thermocouple rather than by the change in resistance of the heated wire.

Related Research Articles

<span class="mw-page-title-main">Electric current</span> Flow of electric charge

An electric current is a stream of charged particles, such as electrons or ions, moving through an electrical conductor or space. It is measured as the net rate of flow of electric charge through a surface or into a control volume. The moving particles are called charge carriers, which may be one of several types of particles, depending on the conductor. In electric circuits the charge carriers are often electrons moving through a wire. In semiconductors they can be electrons or holes. In an electrolyte the charge carriers are ions, while in plasma, an ionized gas, they are ions and electrons.

<span class="mw-page-title-main">Pressure measurement</span> Analysis of force applied by a fluid on a surface

Pressure measurement is the measurement of an applied force by a fluid on a surface. Pressure is typically measured in units of force per unit of surface area. Many techniques have been developed for the measurement of pressure and vacuum. Instruments used to measure and display pressure mechanically are called pressure gauges,vacuum gauges or compound gauges. The widely used Bourdon gauge is a mechanical device, which both measures and indicates and is probably the best known type of gauge.

<span class="mw-page-title-main">Thermocouple</span> Electrical device for measuring temperature

A thermocouple, also known as a "thermoelectrical thermometer", is an electrical device consisting of two dissimilar electrical conductors forming an electrical junction. A thermocouple produces a temperature-dependent voltage as a result of the Seebeck effect, and this voltage can be interpreted to measure temperature. Thermocouples are widely used as temperature sensors.

<span class="mw-page-title-main">Vacuum</span> Space that is empty of matter

A vacuum is a space devoid of matter. The word is derived from the Latin adjective vacuus for "vacant" or "void". An approximation to such vacuum is a region with a gaseous pressure much less than atmospheric pressure. Physicists often discuss ideal test results that would occur in a perfect vacuum, which they sometimes simply call "vacuum" or free space, and use the term partial vacuum to refer to an actual imperfect vacuum as one might have in a laboratory or in space. In engineering and applied physics on the other hand, vacuum refers to any space in which the pressure is considerably lower than atmospheric pressure. The Latin term in vacuo is used to describe an object that is surrounded by a vacuum.

The thermal conductivity of a material is a measure of its ability to a particular material conduct heat. It is commonly denoted by , , or .

<span class="mw-page-title-main">Electrical resistance and conductance</span> Opposition to the passage of an electric current

The electrical resistance of an object is a measure of its opposition to the flow of electric current. Its reciprocal quantity is electrical conductance, measuring the ease with which an electric current passes. Electrical resistance shares some conceptual parallels with mechanical friction. The SI unit of electrical resistance is the ohm, while electrical conductance is measured in siemens (S).

Conduction is the process by which heat is transferred from the hotter end to the colder end of an object.

<span class="mw-page-title-main">Heat transfer</span> Transport of thermal energy in physical systems

Heat transfer is a discipline of thermal engineering that concerns the generation, use, conversion, and exchange of thermal energy (heat) between physical systems. Heat transfer is classified into various mechanisms, such as thermal conduction, thermal convection, thermal radiation, and transfer of energy by phase changes. Engineers also consider the transfer of mass of differing chemical species, either cold or hot, to achieve heat transfer. While these mechanisms have distinct characteristics, they often occur simultaneously in the same system.

<span class="mw-page-title-main">R-value (insulation)</span> Measure of how well an object, per unit of area, resists conductive flow of heat

In the context of construction, the R-value is a measure of how well a two-dimensional barrier, such as a layer of insulation, a window or a complete wall or ceiling, resists the conductive flow of heat. R-value is the temperature difference per unit of heat flux needed to sustain one unit of heat flux between the warmer surface and colder surface of a barrier under steady-state conditions.

<span class="mw-page-title-main">Thermostat</span> Component which maintains a setpoint temperature

A thermostat is a regulating device component which senses the temperature of a physical system and performs actions so that the system's temperature is maintained near a desired setpoint.

<span class="mw-page-title-main">Thermoelectric effect</span> Direct conversion of temperature differences to electric voltage and vice versa

The thermoelectric effect is the direct conversion of temperature differences to electric voltage and vice versa via a thermocouple. A thermoelectric device creates a voltage when there is a different temperature on each side. Conversely, when a voltage is applied to it, heat is transferred from one side to the other, creating a temperature difference. At the atomic scale, an applied temperature gradient causes charge carriers in the material to diffuse from the hot side to the cold side.

Resistance thermometers, also called resistance temperature detectors (RTDs), are sensors used to measure temperature. Many RTD elements consist of a length of fine wire wrapped around a heat-resistant ceramic or glass core but other constructions are also used. The RTD wire is a pure material, typically platinum (Pt), nickel (Ni), or copper (Cu). The material has an accurate resistance/temperature relationship which is used to provide an indication of temperature. As RTD elements are fragile, they are often housed in protective probes.

<span class="mw-page-title-main">Pressure sensor</span> Pressure measurement device

A pressure sensor is a device for pressure measurement of gases or liquids. Pressure is an expression of the force required to stop a fluid from expanding, and is usually stated in terms of force per unit area. A pressure sensor usually acts as a transducer; it generates a signal as a function of the pressure imposed. For the purposes of this article, such a signal is electrical.

<span class="mw-page-title-main">Marcello Pirani</span> German physicist (1880–1968)

Marcello Stefano Pirani was a German physicist known for his invention of the Pirani vacuum gauge, a vacuum gauge based on the principle of heat loss measurement. Throughout his career, he worked on advancing lighting technology and pioneered work on the physics of gas discharge.

Junction temperature, short for transistor junction temperature, is the highest operating temperature of the actual semiconductor in an electronic device. In operation, it is higher than case temperature and the temperature of the part's exterior. The difference is equal to the amount of heat transferred from the junction to case multiplied by the junction-to-case thermal resistance.

<span class="mw-page-title-main">Heat flux sensor</span>

A heat flux sensor is a transducer that generates an electrical signal proportional to the total heat rate applied to the surface of the sensor. The measured heat rate is divided by the surface area of the sensor to determine the heat flux.

There are a number of possible ways to measure thermal conductivity, each of them suitable for a limited range of materials, depending on the thermal properties and the medium temperature. Three classes of methods exist to measure the thermal conductivity of a sample: steady-state, time-domain, and frequency-domain methods.

<span class="mw-page-title-main">Measuring instrument</span> Device for measuring a physical quantity

A measuring instrument is a device to measure a physical quantity. In the physical sciences, quality assurance, and engineering, measurement is the activity of obtaining and comparing physical quantities of real-world objects and events. Established standard objects and events are used as units, and the process of measurement gives a number relating the item under study and the referenced unit of measurement. Measuring instruments, and formal test methods which define the instrument's use, are the means by which these relations of numbers are obtained. All measuring instruments are subject to varying degrees of instrument error and measurement uncertainty. These instruments may range from simple objects such as rulers and stopwatches to electron microscopes and particle accelerators. Virtual instrumentation is widely used in the development of modern measuring instruments.

<span class="mw-page-title-main">Thermopile laser sensor</span>

Thermopile laser sensors are used for measuring laser power from a few µW to several W. The incoming radiation of the laser is converted into heat energy at the surface. This heat input produces a temperature gradient across the sensor. Making use of the thermoelectric effect a voltage is generated by this temperature gradient. Since the voltage is directly proportional to the incoming radiation, it can be directly related to the irradiation power.

The transient hot wire method (THW) is a very popular, accurate and precise technique to measure the thermal conductivity of gases, liquids, solids, nanofluids and refrigerants in a wide temperature and pressure range. The technique is based on recording the transient temperature rise of a thin vertical metal wire with infinite length when a step voltage is applied to it. The wire is immersed in a fluid and can act both as an electrical heating element and a resistance thermometer. The transient hot wire method has advantage over the other thermal conductivity methods, since there is a fully developed theory and there is no calibration or single-point calibration. Furthermore, because of the very small measuring time there is no convection present in the measurements and only the thermal conductivity of the fluid is measured with very high accuracy.

References

  1. Ellett, A. (1931). "The Pirani Gauge for the Measurement of Small Changes of Pressure". Physical Review. 37 (9): 1102–1111. doi:10.1103/PhysRev.37.1102.
  2. 1 2 3 von Pirani, M (1906). "Selbstzeigendes Vakuum-Meßinstrument". Deutsche Physikalische Gesellschaft, Verh. 24 (8): 686–694.
  3. 1 2 Borichevsky (2017). Understanding Modern Vacuum Technology. p. 62. ISBN   9781974554461.
  4. Kundt, A.; Warburg, E. (1875). "Ueber Reibung und Wärmeleitung verdünnter Gase". Annalen der Physik und Chemie. 232 (10): 177–211. Bibcode:1875AnP...232..177K. doi:10.1002/andp.18752321002.
  5. Smoluchowski, Marian (1898). "Temperatursprung in verdünnten Gasen". Ann Phys Chem. 64: 101.
  6. DE 10115715,Plöchinger, Heinz,"Sensor and Method for Detecting Measurement Variables and Physical Parameters",published 2002-10-17, also description
  7. Jitschin, W.; Ludwig, S. (2004). "Gepulstes Heißdraht-Vakuummeter mit Pirani-Sensor". Vakuum in Forschung und Praxis (in German). 16: 23–29. doi:10.1002/vipr.200400015.