Synthetic MRI is a simulation method in Magnetic Resonance Imaging (MRI), for generating contrast weighted images based on measurement of tissue properties. The synthetic (simulated) images are generated after an MR study, from parametric maps of tissue properties. It is thereby possible to generate several contrast weightings from the same acquisition. This is different from conventional MRI, where the signal acquired from the tissue is used to generate an image directly, often generating only one contrast weighting per acquisition. The synthetic images are similar in appearance to those normally acquired with an MRI scanner.
The parametric maps can be computed from a particular MRI acquisition designed to measure the tissue parameters, known as quantification. Using the maps, which contains the measured parameters for each voxel, virtual scanner settings that correspond to those used in conventional scan are given. These settings can be echo time (TE) and repetition time (TR) for a spin-echo (SE) sequence or TE, TR and inversion time (TI) for an inversion recovery (IR, FLAIR, STIR, PSIR, FSE-IR, TIRM) sequence. Using the signal equations for different types of MR acquisitions, it is then possible to calculate what a conventional image would look like. Calculating the images based on maps and scanner settings is called synthesizing the images.
Synthetic MRI was proposed as early as 1984 Bielke et al. [1] and 1985 by Bobman et al. [2]
Although scientifically interesting, the method was cumbersome for clinical use. The acquisition duration was too long for a patient to lie still, and the computations needed for quantification were too demanding for the standard commercial computers of the day.
For the computers, the problem was both overall computational speed and handling numbers over 16 bit. To perform the synthetization, Lee et al. suggested a purpose-built computer that would be able to synthesize images in about 600 ms. It would use a look-up table to save computations and be capable of handling number sizes up to 28 bit. [3] This device did however not solve the calculation needs for quantification, nor the long acquisition times.
MR Image Expert, a software to create synthetic magnetic resonance images, was introduced in the late 1980s. It was aimed at educational and research purposes, among them contrast agent applications. Since 1989, more than 12,000 licenses of this software have been distributed. [4] [5]
In 2004 the first rapid acquisition and quantification method for creating parametric maps was invented. This new acquisition method performs 8 acquisitions at 4 different excitation delays, giving 8 values to estimate T1, T2, PD and M0 for each imaged voxel. [6] [7] [8]
There are also other methods for creating the parametric maps being researched. Most notable is Magnetic Resonance Fingerprinting . This method uses a randomized acquisition that creates a unique response depending on T1-T2-PD values of a voxel. This unique response is then matched against a database of response from all possible T1-T2-PD combinations. [9] [10]
Creating a synthetic contrast weighted image is very quick and the scanner setting parameters can be changed interactively. This is different from acquiring the images directly on the MRI scanner with a pulse sequence where it might take several minutes to get new images with different settings. Synthetic MRI also enables the possibility to create new image contrast weightings after the scan has been completed and the patient has left the hospital. A prospective multi-reader, multi-case, multi-center clinical trial of overall image quality of synthetic MRI compared to conventional MRI in a general neuroimaging population has shown that synthetic MRI is non-inferior to conventional imaging. [11]
Synthetic MRI is based on parametric maps. These maps are generated by using the MR scanner not to acquire an image, but to measure magnetic properties in the tissue. [8]
The parametric maps can for example be longitudinal (T1) and transversal (T2) relaxation times as well as either proton density (PD) or apparent equilibrium magnetization (M0). It is conceivable to use other parameters depending on the acquisition sequence to be simulated, but these are sufficient for spin-echo and inversion recovery acquisitions.
The synthetic images are calculated using the signal equation of the type of image synthesized. The signal equation is a formula for calculating the signal intensity, i.e. the numerical value, of an image pixel. The signal intensity S in a pixel depends on the tissue properties T1, T2 and PD of the corresponding voxel, as well as the echo time TE and repetition time TR. [12]
The equation for synthesizing a fast spin-echo (FSE) image is: [13]
Here ETL is the echo train length, and ESP is the spacing of echoes in the echo train. If the acquisition simulated has the echo in the middle of ETL used for central k-space then . The scanner parameters TE and TR are the same over the entire slice/volume that is being synthesized but the parameters T1, T2, PD and the resulting signal S are different for different voxels. The parameters T1, T2, and PD correspond to physical properties of the tissue within the voxel.
For an inversion recovery (IR) sequence the equation is: [13]
Note that setting means that the signal equation becomes identical to the equation for FSE. This is because the inversion pulse is executed at the end of acquisition where the longitudinal magnetization is 0 due to excitation pulse being played out before acquisition.
For a double inversion recovery (DIR) sequence the equation is: [13]
Note that setting makes the signal equation identical to the IR equation, and setting gives the same equation as for FSE (since the inversion pulses cancel out). The equation is derived under the assumption that .
For IR and DIR the signal may be negative in some voxels. In a real MRI scanner the signal is complex and due to off-resonance effects and other scanner imperfections the phase of the complex value may vary between different voxels, this leads to it being difficult to distinguish between a positive and negative signal. It is common to only reconstruct the magnitude image. For the synthetic images it is easy to keep the sign of the synthesized signal and thereby effectively creating a contrast weight that corresponds to a phase-sensitive inversion recovery (PSIR) which may also be called IR Real or Corrected Real.
If apparent equilibrium magnetization (M0) is used instead of true PD there may be effects of coil sensitivity (central shading) and dielectric shading present in the images.
Dedicated acquisition for creating parametric maps is available on some scanners by GE and Philips, under the brand names MAGiC and SyntAc. [14] [15]
Post processing software for Synthetic MRI is available in the commercial product SyMRI from SyntheticMR AB which is included as an option called MAGiC on the MR console for GE's SIGNA Pioneer 3T MRI scanner. Since 2017, synthetic MRI is also available in Siemens machines. [16]
Olea Medical offers a product similar to Synthetic MRI named Olea Nova+. It offers a method for automatically computing conventional images from a standard protocol: T1 and T2 mapping. Users are able to create any image contrast-weighed in T1, T2 (...) or maps combining various TE, TR, TI. [17]
Magnetic resonance imaging (MRI) is a medical imaging technique used in radiology to form pictures of the anatomy and the physiological processes of the body. MRI scanners use strong magnetic fields, magnetic field gradients, and radio waves to generate images of the organs in the body. MRI does not involve X-rays or the use of ionizing radiation, which distinguishes it from computed tomography (CT) and positron emission tomography (PET) scans. MRI is a medical application of nuclear magnetic resonance (NMR) which can also be used for imaging in other NMR applications, such as NMR spectroscopy.
Functional magnetic resonance imaging or functional MRI (fMRI) measures brain activity by detecting changes associated with blood flow. This technique relies on the fact that cerebral blood flow and neuronal activation are coupled. When an area of the brain is in use, blood flow to that region also increases.
Diffusion-weighted magnetic resonance imaging is the use of specific MRI sequences as well as software that generates images from the resulting data that uses the diffusion of water molecules to generate contrast in MR images. It allows the mapping of the diffusion process of molecules, mainly water, in biological tissues, in vivo and non-invasively. Molecular diffusion in tissues is not random, but reflects interactions with many obstacles, such as macromolecules, fibers, and membranes. Water molecule diffusion patterns can therefore reveal microscopic details about tissue architecture, either normal or in a diseased state. A special kind of DWI, diffusion tensor imaging (DTI), has been used extensively to map white matter tractography in the brain.
Magnetic resonance angiography (MRA) is a group of techniques based on magnetic resonance imaging (MRI) to image blood vessels. Magnetic resonance angiography is used to generate images of arteries in order to evaluate them for stenosis, occlusions, aneurysms or other abnormalities. MRA is often used to evaluate the arteries of the neck and brain, the thoracic and abdominal aorta, the renal arteries, and the legs.
In MRI and NMR spectroscopy, an observable nuclear spin polarization (magnetization) is created by a homogeneous magnetic field. This field makes the magnetic dipole moments of the sample precess at the resonance (Larmor) frequency of the nuclei. At thermal equilibrium, nuclear spins precess randomly about the direction of the applied field. They become abruptly phase coherent when they are hit by radiofrequency (RF) pulses at the resonant frequency, created orthogonal to the field. The RF pulses cause the population of spin-states to be perturbed from their thermal equilibrium value. The generated transverse magnetization can then induce a signal in an RF coil that can be detected and amplified by an RF receiver. The return of the longitudinal component of the magnetization to its equilibrium value is termed spin-latticerelaxation while the loss of phase-coherence of the spins is termed spin-spin relaxation, which is manifest as an observed free induction decay (FID).
Steady-state free precession (SSFP) imaging is a magnetic resonance imaging (MRI) sequence which uses steady states of magnetizations. In general, SSFP MRI sequences are based on a gradient echo MRI sequence with a short repetition time which in its generic form has been described as the FLASH MRI technique. While spoiled gradient-echo sequences refer to a steady state of the longitudinal magnetization only, SSFP gradient-echo sequences include transverse coherences (magnetizations) from overlapping multi-order spin echoes and stimulated echoes. This is usually accomplished by refocusing the phase-encoding gradient in each repetition interval in order to keep the phase integral constant. Fully balanced SSFP MRI sequences achieve a phase of zero by refocusing all imaging gradients.
Fast low angle shot magnetic resonance imaging is a particular sequence of magnetic resonance imaging. It is a gradient echo sequence which combines a low-flip angle radio-frequency excitation of the nuclear magnetic resonance signal with a short repetition time. It is the generic form of steady-state free precession imaging.
Fluid-attenuated inversion recovery (FLAIR) is an MRI sequence with an inversion recovery set to null fluids. For example, it can be used in brain imaging to suppress cerebrospinal fluid (CSF) effects on the image, so as to bring out the periventricular hyperintense lesions, such as multiple sclerosis (MS) plaques. It was invented by Graeme Bydder. FLAIR can be used with both three-dimensional imaging or two dimensional imaging.
In physics, the spin–spin relaxation is the mechanism by which Mxy, the transverse component of the magnetization vector, exponentially decays towards its equilibrium value in nuclear magnetic resonance (NMR) and magnetic resonance imaging (MRI). It is characterized by the spin–spin relaxation time, known as T2, a time constant characterizing the signal decay. It is named in contrast to T1, the spin–lattice relaxation time. It is the time it takes for the magnetic resonance signal to irreversibly decay to 37% (1/e) of its initial value after its generation by tipping the longitudinal magnetization towards the magnetic transverse plane. Hence the relation
During nuclear magnetic resonance observations, spin–lattice relaxation is the mechanism by which the longitudinal component of the total nuclear magnetic moment vector (parallel to the constant magnetic field) exponentially relaxes from a higher energy, non-equilibrium state to thermodynamic equilibrium with its surroundings (the "lattice"). It is characterized by the spin–lattice relaxation time, a time constant known as T1.
The physics of magnetic resonance imaging (MRI) concerns fundamental physical considerations of MRI techniques and technological aspects of MRI devices. MRI is a medical imaging technique mostly used in radiology and nuclear medicine in order to investigate the anatomy and physiology of the body, and to detect pathologies including tumors, inflammation, neurological conditions such as stroke, disorders of muscles and joints, and abnormalities in the heart and blood vessels among others. Contrast agents may be injected intravenously or into a joint to enhance the image and facilitate diagnosis. Unlike CT and X-ray, MRI uses no ionizing radiation and is, therefore, a safe procedure suitable for diagnosis in children and repeated runs. Patients with specific non-ferromagnetic metal implants, cochlear implants, and cardiac pacemakers nowadays may also have an MRI in spite of effects of the strong magnetic fields. This does not apply on older devices, and details for medical professionals are provided by the device's manufacturer.
Real-time magnetic resonance imaging (RT-MRI) refers to the continuous monitoring ("filming") of moving objects in real time. Because MRI is based on time-consuming scanning of k-space, real-time MRI was possible only with low image quality or low temporal resolution. Using an iterative reconstruction algorithm these limitations have recently been removed: a new method for real-time MRI achieves a temporal resolution of 20 to 30 milliseconds for images with an in-plane resolution of 1.5 to 2.0 mm. Real-time MRI promises to add important information about diseases of the joints and the heart. In many cases MRI examinations may become easier and more comfortable for patients.
Quantitative susceptibility mapping (QSM) provides a novel contrast mechanism in magnetic resonance imaging (MRI) different from traditional susceptibility weighted imaging. The voxel intensity in QSM is linearly proportional to the underlying tissue apparent magnetic susceptibility, which is useful for chemical identification and quantification of specific biomarkers including iron, calcium, gadolinium, and super paramagnetic iron oxide (SPIO) nano-particles. QSM utilizes phase images, solves the magnetic field to susceptibility source inverse problem, and generates a three-dimensional susceptibility distribution. Due to its quantitative nature and sensitivity to certain kinds of material, potential QSM applications include standardized quantitative stratification of cerebral microbleeds and neurodegenerative disease, accurate gadolinium quantification in contrast enhanced MRI, and direct monitoring of targeted theranostic drug biodistribution in nanomedicine.
Magnetic resonance imaging of the brain uses magnetic resonance imaging (MRI) to produce high quality two-dimensional or three-dimensional images of the brain and brainstem as well as the cerebellum without the use of ionizing radiation (X-rays) or radioactive tracers.
Perfusion MRI or perfusion-weighted imaging (PWI) is perfusion scanning by the use of a particular MRI sequence. The acquired data are then post-processed to obtain perfusion maps with different parameters, such as BV, BF, MTT and TTP.
An MRI sequence in magnetic resonance imaging (MRI) is a particular setting of pulse sequences and pulsed field gradients, resulting in a particular image appearance.
Gradient echo is a magnetic resonance imaging (MRI) sequence that has wide variety of applications, from magnetic resonance angiography to perfusion MRI and diffusion MRI. Rapid imaging acquisition allows it to be applied to 2D and 3D MRI imaging. Gradient echo uses magnetic gradients to generate a signal, instead of using 180 degrees radiofrequency pulse like spin echo; thus leading to faster image acquisition time.
An MRI artifact is a visual artifact in magnetic resonance imaging (MRI). It is a feature appearing in an image that is not present in the original object. Many different artifacts can occur during MRI, some affecting the diagnostic quality, while others may be confused with pathology. Artifacts can be classified as patient-related, signal processing-dependent and hardware (machine)-related.
Inversion recovery is an MRI sequence that provides high contrast between tissue and lesion. It can be used to provide high T1 weighted image, high T2 weighted image, and to suppress the signals from fat, blood, or cerebrospinal fluid (CSF).
Magnetic resonance fingerprinting (MRF) is methodology in quantitative magnetic resonance imaging (MRI) characterized by a pseudo-randomized acquisition strategy. It involves creating unique signal patterns or 'fingerprints' for different materials or tissues after which a pattern recognition algorithm matches these fingerprints with a predefined dictionary of expected signal patterns. This process translates the data into quantitative maps, revealing information about the magnetic properties being investigated.