A transition-edge sensor (TES) is a type of cryogenic energy sensor or cryogenic particle detector that exploits the strongly temperature-dependent resistance of the superconducting phase transition.
The first demonstrations of the superconducting transition's measurement potential appeared in the 1940s, 30 years after Onnes's discovery of superconductivity. D. H. Andrews demonstrated the first transition-edge bolometer, a current-biased tantalum wire which he used to measure an infrared signal. Subsequently he demonstrated a transition-edge calorimeter made of niobium nitride which was used to measure alpha particles. [1] However, the TES detector did not gain popularity for about 50 years, due primarily to the difficulty in stabilizing the temperature within the narrow superconducting transition region, especially when more than one pixel was operated at the same time, and also due to the difficulty of signal readout from such a low-impedance system. Joule heating in a current-biased TES can lead to thermal runaway that drives the detector into the normal (non-superconducting) state, a phenomenon known as positive electrothermal feedback. The thermal runaway problem was solved in 1995 by K. D. Irwin by voltage-biasing the TES, establishing stable negative electrothermal feedback, and coupling them to superconducting quantum interference devices (SQUID) current amplifiers. [2] This breakthrough has led to widespread adoption of TES detectors. [3]
The TES is voltage-biased by driving a current source Ibias through a load resistor RL (see figure). The voltage is chosen to put the TES in its so-called "self-biased region" where the power dissipated in the device is constant with the applied voltage. When a photon is absorbed by the TES, this extra power is removed by negative electrothermal feedback: the TES resistance increases, causing a drop in TES current; the Joule power in turn drops, cooling the device back to its equilibrium state in the self-biased region. In a common SQUID readout system, the TES is operated in series with the input coil L, which is inductively coupled to a SQUID series-array. Thus a change in TES current manifests as a change in the input flux to the SQUID, whose output is further amplified and read by room-temperature electronics.
Any bolometric sensor employs three basic components: an absorber of incident energy, a thermometer for measuring this energy, and a thermal link to base temperature to dissipate the absorbed energy and cool the detector. [4]
The simplest absorption scheme can be applied to TESs operating in the near-IR, optical, and UV regimes. These devices generally utilize a tungsten TES as its own absorber, which absorbs up to 20% of the incident radiation. [5] If high-efficiency detection is desired, the TES may be fabricated in a multi-layer optical cavity tuned to the desired operating wavelength and employing a backside mirror and frontside anti-reflection coating. Such techniques can decrease the transmission and reflection from the detectors to negligibly low values; 95% detection efficiency has been observed. [4] At higher energies, the primary obstacle to absorption is transmission, not reflection, and thus an absorber with high photon stopping power and low heat capacity is desirable; a bismuth film is often employed. [3] Any absorber should have low heat capacity with respect to the TES. Higher heat capacity in the absorber will contribute to noise and decrease the sensitivity of the detector (since a given absorbed energy will not produce as large of a change in TES resistance). For far-IR radiation into the millimeter range, the absorption schemes commonly employ antennas or feedhorns. [3]
The TES operates as a thermometer in the following manner: absorbed incident energy increases the resistance of the voltage-biased sensor within its transition region, and the integral of the resulting drop in current is proportional to the energy absorbed by the detector. [5] The output signal is proportional to the temperature change of the absorber, and thus for maximal sensitivity, a TES should have low heat capacity and a narrow transition. Important TES properties including not only heat capacity but also thermal conductance are strongly temperature dependent, so the choice of transition temperature Tc is critical to the device design. Furthermore, Tc should be chosen to accommodate the available cryogenic system. Tungsten has been a popular choice for elemental TESs as thin-film tungsten displays two phases, one with Tc ~15 mK and the other with Tc ~1–4 K, which can be combined to finely tune the overall device Tc. [6] Bilayer and multilayer TESs are another popular fabrication approach, where thin films of different materials are combined to achieve the desired Tc. [3]
Finally, it is necessary to tune the thermal coupling between the TES and the bath of cooling liquid; a low thermal conductance is necessary to ensure that incident energy is seen by the TES rather than being lost directly to the bath. However, the thermal link must not be too weak, as it is necessary to cool the TES back to bath temperature after the energy has been absorbed. Two approaches to control the thermal link are by electron–phonon coupling and by mechanical machining. At cryogenic temperatures, the electron and phonon systems in a material can become only weakly coupled. The electron–phonon thermal conductance is strongly temperature-dependent, and hence the thermal conductance can be tuned by adjusting Tc. [3] [4] Other devices use mechanical means of controlling the thermal conductance such as building the TES on a sub-micrometre membrane over a hole in the substrate or in the middle of a sparse "spiderweb" structure. [7]
TES detectors are attractive to the scientific community for a variety of reasons. Among their most striking attributes are an unprecedented high detection efficiency customizable to wavelengths from the millimeter regime to gamma rays [3] [4] and a theoretical negligible background dark count level (less than 1 event in 1000 s from intrinsic thermal fluctuations of the device [5] ). (In practice, although only a real energy signal will create a current pulse, a nonzero background level may be registered by the counting algorithm or the presence of background light in the experimental setup. Even thermal blackbody radiation may be seen by a TES optimized for use in the visible regime.)
TES single-photon detectors suffer nonetheless from a few disadvantages as compared to their avalanche photodiode (APD) counterparts. APDs are manufactured in small modules, which count photons out-of-the-box with a dead time of a few nanoseconds and output a pulse corresponding to each photon with a jitter of tens of picoseconds. In contrast, TES detectors must be operated in a cryogenic environment, output a signal that must be further analyzed to identify photons, and have a jitter of approximately 100 ns. [4] Furthermore, a single-photon spike on a TES detector lasts on the order of microseconds.
TES arrays are becoming increasingly common in physics and astronomy experiments such as SCUBA-2, the HAWC+ instrument on the Stratospheric Observatory for Infrared Astronomy, the Atacama Cosmology Telescope, the Cryogenic Dark Matter Search, the Cryogenic Observatory for Signatures Seen in Next-Generation Underground Searches, the Cryogenic Rare Event Search with Superconducting Thermometers, the E and B Experiment, the South Pole Telescope, the Spider polarimeter, the X-IFU instrument of the Advanced Telescope for High Energy Astrophysics satellite, the future LiteBIRD Cosmic Microwave Background polarization experiment, the Simons Observatory, and the CMB Stage-IV Experiment.
Superconductivity is a set of physical properties observed in superconductors: materials where electrical resistance vanishes and magnetic fields are expelled from the material. Unlike an ordinary metallic conductor, whose resistance decreases gradually as its temperature is lowered, even down to near absolute zero, a superconductor has a characteristic critical temperature below which the resistance drops abruptly to zero. An electric current through a loop of superconducting wire can persist indefinitely with no power source.
A SQUID is a very sensitive magnetometer used to measure extremely weak magnetic fields, based on superconducting loops containing Josephson junctions.
A bolometer is a device for measuring radiant heat by means of a material having a temperature-dependent electrical resistance. It was invented in 1878 by the American astronomer Samuel Pierpont Langley.
Infrared thermography (IRT), thermal video or thermal imaging, is a process where a thermal camera captures and creates an image of an object by using infrared radiation emitted from the object in a process, which are examples of infrared imaging science. Thermographic cameras usually detect radiation in the long-infrared range of the electromagnetic spectrum and produce images of that radiation, called thermograms. Since infrared radiation is emitted by all objects with a temperature above absolute zero according to the black body radiation law, thermography makes it possible to see one's environment with or without visible illumination. The amount of radiation emitted by an object increases with temperature; therefore, thermography allows one to see variations in temperature. When viewed through a thermal imaging camera, warm objects stand out well against cooler backgrounds; humans and other warm-blooded animals become easily visible against the environment, day or night. As a result, thermography is particularly useful to the military and other users of surveillance cameras.
A scintillator is a material that exhibits scintillation, the property of luminescence, when excited by ionizing radiation. Luminescent materials, when struck by an incoming particle, absorb its energy and scintillate. Sometimes, the excited state is metastable, so the relaxation back down from the excited state to lower states is delayed. The process then corresponds to one of two phenomena: delayed fluorescence or phosphorescence. The correspondence depends on the type of transition and hence the wavelength of the emitted optical photon.
Energy-dispersive X-ray spectroscopy, sometimes called energy dispersive X-ray analysis or energy dispersive X-ray microanalysis (EDXMA), is an analytical technique used for the elemental analysis or chemical characterization of a sample. It relies on an interaction of some source of X-ray excitation and a sample. Its characterization capabilities are due in large part to the fundamental principle that each element has a unique atomic structure allowing a unique set of peaks on its electromagnetic emission spectrum. The peak positions are predicted by the Moseley's law with accuracy much better than experimental resolution of a typical EDX instrument.
Photodetectors, also called photosensors, are sensors of light or other electromagnetic radiation. There are a wide variety of photodetectors which may be classified by mechanism of detection, such as photoelectric or photochemical effects, or by various performance metrics, such as spectral response. Semiconductor-based photodetectors typically use a p–n junction that converts photons into charge. The absorbed photons make electron–hole pairs in the depletion region. Photodiodes and photo transistors are a few examples of photo detectors. Solar cells convert some of the light energy absorbed into electrical energy.
A microbolometer is a specific type of bolometer used as a detector in a thermal camera. Infrared radiation with wavelengths between 7.5–14 μm strikes the detector material, heating it, and thus changing its electrical resistance. This resistance change is measured and processed into temperatures which can be used to create an image. Unlike other types of infrared detecting equipment, microbolometers do not require cooling.
Cryogenic particle detectors operate at very low temperature, typically only a few degrees above absolute zero. These sensors interact with an energetic elementary particle and deliver a signal that can be related to the type of particle and the nature of the interaction. While many types of particle detectors might be operated with improved performance at cryogenic temperatures, this term generally refers to types that take advantage of special effects or properties occurring only at low temperature.
SRON Netherlands Institute for Space Research (SRON) is the Dutch national institute for space research. It focuses on astrophysics, Earth observation, and exoplanetary research. SRON also develops new detection techniques for X-rays, infrared radiation, and visible light. As national expertise institute, SRON gives council to the Dutch government and coordinates national contributions to international space missions.
The Cryogenic Rare Event Search with Superconducting Thermometers (CRESST) is a collaboration of European experimental particle physics groups involved in the construction of cryogenic detectors for direct dark matter searches. The participating institutes are the Max Planck Institute for Physics (Munich), Technical University of Munich, University of Tübingen (Germany), University of Oxford, the Comenius University Bratislava (Slovakia) and the Istituto Nazionale di Fisica Nucleare.
The superconducting tunnel junction (STJ) – also known as a superconductor–insulator–superconductor tunnel junction (SIS) – is an electronic device consisting of two superconductors separated by a very thin layer of insulating material. Current passes through the junction via the process of quantum tunneling. The STJ is a type of Josephson junction, though not all the properties of the STJ are described by the Josephson effect.
A Quantum Well Infrared Photodetector (QWIP) is an infrared photodetector, which uses electronic intersubband transitions in quantum wells to absorb photons. In order to be used for infrared detection, the parameters of the quantum wells in the quantum well infrared photodetector are adjusted so that the energy difference between its first and second quantized states match the incoming infrared photon energy. QWIPs are typically made of gallium arsenide, a material commonly found in smartphones and high-speed communications equipment. Depending on the material and the design of the quantum wells, the energy levels of the QWIP can be tailored to absorb radiation in the infrared region from 3 to 20 μm.
Phonon noise, also known as thermal fluctuation noise, arises from the random exchange of energy between a thermal mass and its surrounding environment. This energy is quantized in the form of phonons. Each phonon has an energy of order , where is the Boltzmann constant and is the temperature. The random exchange of energy leads to fluctuations in temperature. This occurs even when the thermal mass and the environment are in thermal equilibrium, i.e. at the same time-average temperature. If a device has a temperature-dependent electrical resistance, then these fluctuations in temperature lead to fluctuations in resistance. Examples of devices where phonon noise is important include bolometers and calorimeters. The superconducting transition edge sensor (TES), which can be operated either as a bolometer or a calorimeter, is an example of a device for which phonon noise can significantly contribute to the total noise.
The kinetic inductance detector (KID) — also known as a microwave kinetic inductance detector (MKID) — is a type of superconducting photon detector capable of counting single photons whilst simultaneously measuring their energy and arrival time to high precision. They were first developed by scientists at the California Institute of Technology and the Jet Propulsion Laboratory in 2003. These devices operate at cryogenic temperatures, typically below 1 kelvin. They are being developed for high-sensitivity astronomical detection for frequencies ranging from the far-infrared to X-rays.
The superconducting nanowire single-photon detector is a type of optical and near-infrared single-photon detector based on a current-biased superconducting nanowire. It was first developed by scientists at Moscow State Pedagogical University and at the University of Rochester in 2001. The first fully operational prototype was demonstrated in 2005 by the National Institute of Standards and Technology (Boulder), and BBN Technologies as part of the DARPA Quantum Network.
In electronics, electrothermal feedback is the interaction of the electric current and the temperature in a device with a temperature-dependent electrical resistance. This interaction arises from Joule heating.
TMOS is a type of thermal sensor consisting in a micromachined thermally isolated transistor fabricated using CMOS-SOI(Silicon on Insulator) MEMS(Micro electro-mechanical system) technology. It has been developed in the last decade by the Technion - Israel Institute of Technology. A thermal sensor is a device able to detect the thermal radiation emitted by an object located in the FOV(Field Of View) of the sensor. Infrared radiation striking the sensor produces a change in the temperature of the device that as a consequence generates an electric output signal proportional to the incident IR power. The sensor is able to measure the temperature of the object radiating thanks to the information contained in the impinging radiation, exploiting in this sense Stefan - Boltzmann law. TMOS detector has two important characteristics that make it different from others: it's an active and uncooled sensor.