Analytical nebulizer

Last updated
Many of today's current analytical nebulizers, as of 2017 Analytical Nebulizers 2017.jpg
Many of today's current analytical nebulizers, as of 2017

The general term nebulizer refers to an apparatus that converts liquids into a fine mist. Nozzles also convert liquids into a fine mist, but do so by pressure through small holes. Nebulizers generally use gas flows to deliver the mist. The most common form of nebulizers are medical appliances such as asthma inhalers or paint spray cans. Analytical nebulizers are a special category in that their purpose is to deliver a fine mist to spectrometric instruments for elemental analysis. They are necessary parts of inductively coupled plasma atomic emission spectroscopy (ICP-AES), inductively coupled plasma mass spectrometry (ICP-MS), and atomic absorption spectroscopy (AAS).

Contents

Applications

Analytical nebulizers are used in trace element analysis. This type of work plays an important role in areas of pharmaceutical and clinical study, biological, environmental and agricultural assessment and petroleum testing. They also have nuclear applications.

Nebulizer designs

Most analytical pneumatic nebulizers use the same essential principle (induction) to atomize the liquid: When gas at a higher pressure exits from a small hole (the orifice) into gas at a lower pressure, it forms a gas jet into the lower pressure zone, and pushes the lower pressure gas away from the orifice. This creates a current in the lower pressure gas zone, and draws some of the lower pressure gas into the higher pressure gas jet. At the orifice, the draw of the lower pressure gas creates considerable suction, the extent depending on the differential pressures, the size of the orifice, and the shape of the orifice and surrounding apparatus. In all pneumatic induction nebulizers, the suction near the orifice is utilized to draw the liquid into the gas jet. The liquid is broken into small droplets in the process.

Present induction pneumatic nebulizer designs fit into 5 categories: 1. Concentric: Liquid flow surrounded by a Gas flow or Gas flow surrounded by a Liquid flow; 2. Cross Flow: Gas flow at right angles to the Liquid flow; 3. Entrained: Gas and Liquid mixed in the system and emitted as a combined flow. 4. Babington and V Groove: Liquid is spread over a surface to decrease the surface tension, and passed over a gas orifice; 5. Parallel Path: Liquid is delivered beside a gas orifice and induction pulls the liquid into the gas stream.

Newer non-induction nebulizers include 3 more categories: 6. Enhanced Parallel Path: Liquid is delivered beside a gas orifice and drawn into the gas stream by surface tension along a spout; 7: Flow Blurring: liquid is injected by pressure into a gas stream; 8. Vibrating Mesh: liquid is pushed through tiny holes by a vibrating ultrasonic plate.

Induction nebulizers

Concentric nebulizers

A teflon PFA concentric nebulizer. PFA nebulizer.JPG
A teflon PFA concentric nebulizer.
A Meinhard glass concentric nebulizer. Meinhard nebulizer.jpg
A Meinhard glass concentric nebulizer.

Concentric nebulizers have a central capillary with the liquid and an outer capillary with the gas. The gas draws the liquid into the gas stream through induction, and the liquid is broken into a fine mist as it moves into the gas stream. In theory, the gas and liquid may be switched with the gas in the center and the liquid in the outer capillary, but generally they work better with the gas outside and the liquid inside. [1] The first Canadian concentric patent was Canadian Patent #2405 of April 18, 1873. It was designed to deliver a better spray of oil into a burner. The design is larger but essentially the same as modern analytical nebulizers. The first one developed for spectrometers was a glass design developed by Dr. Meinhard of California in 1973. [2] His design enabled early ICP users to have a consistent sample introduction nebulizer, but it plugged easily. Today many companies produce glass concentrics, and since 1997, Teflon concentrics have become available.

Cross flow nebulizers

Cross flow nebulizers have a gas capillary set at right angles to the liquid capillary. The gas is blown across the liquid capillary and this produces a low pressure that draws the liquid into the gas stream. Generally the suction is similar to what is produced in a concentric nebulizer. The benefit of a cross flow is that the liquid capillary have a larger inside diameter allowing for more particles to pass through without plugging the nebulizer. The disadvantage is that the mist is usually not as fine or as consistent. [3]

Entrained nebulizers

There are no analytical nebulizers at present using this technique, but some oil burners do. Mainly used in much older designs as newer concentrics and cross flows are much better and easier to make.

V-groove nebulizers

V Groove nebulizers are similar to a cross flow in that the liquid is delivered in a capillary at right angles to the gas capillary, but the liquid is poured down a vertically orientated groove that flows past a gas orifice. The gas pulls the liquid into the gas flow and forms a fine mist. These allow for very large ID liquid capillaries, but have no suction and require a pump to feed the liquid to the device. They must be correctly orientated or they do not allow the liquid to flow past the gas stream. And their mist usually produces larger droplets than with concentrics or cross flows.

Parallel path nebulizers

This design was developed by John Burgener of Burgener Research Inc. Here, the gas stream and sample run through the nebulizer in parallel capillaries. At the tip of the nebulizer, the liquid is pulled into the gas stream and then dispersed into the chamber as a mist.

Non-induction nebulizers

Enhanced parallel path nebulizers

This design was developed by John Burgener of Burgener Research Inc. [4] Here, the gas stream and sample run through the nebulizer in parallel capillaries. At the tip of the nebulizer, the liquid is pulled into the gas stream by surface tension along a spout dipping into the gas stream. This allows the gas to impact the liquid, and has the liquid interact in the center of the gas flow where the gas flow speed is highest, producing a better transfer of energy from the gas to the liquid, and producing a finer droplet size. The Burgener Mira Mist nebulizers are the main products using the Enhanced Parallel Path method.

Flow blurring nebulizers

This is a new type of nebulizer which does not use induction to mix the sample and gas. Instead, pneumatic atomization is employed here, which results in the micro-mixing of fluids using a reflux cell. [5] This means that there is a turbulent mixing of the liquid and gas which results in great sensitivity and is very efficient. The OneNeb is the only example of this sort.

Piezoelectric vibrating mesh

Since 2011, this variation on ultrasonic nebulizers has been available. There is a vibrating membrane which has micro holes in it. The sample enters through the back and is pushed through the holes as the membrane vibrates. This makes a fine mist with a droplet size proportional to the hole size. This method requires no gas flow, and is used in conjunction with a chamber. If the droplets are less than 5μm then they are too small to stick to the chamber walls and the chamber remains dry while 90–100% of the sample makes it to the torch.

Chronology of analytical nebulizer development

The early history of medical nebulizers can be read here. The development of analytical nebulizers since the introduction of the ICP / ICP-MS is seen below: [6]

1970s Adjustable Cross flow (US patent #4,344,574) [7]

1974 Meinhard Concentric

1978 V-groove (by Suddendorf and Boyer) (US Patent #4,206,160) [8]

1980 Pillar and Post (by Garbarino and Taylor)

1983 GMK Nebulizer: Glass Babington V-groove

1983 Meinhard C-type nebulizer

1983 Precision glassblowing (similar to Minehard A-type)

1983 Jarrell Ash (Thermo) Sapphire V-groove

1983 Meddings' MAK: glass fixed cross flow

1984 Meinhard K-type: recessed inner capillary

1984 Glass Expansion begins making ICP glassware

1985 Burgener-Legere – first commercial teflon nebulizer – V-groove – no adjustable parts

1986 Direct injection micro nebulizer by Fassel, Rice & Lawrence (US patent #4,575,609) [9]

1986 Hildebrand Grid nebulizer

Late 1980s Perkin Elmer Gem Tip cross flow

1988 CETAC Ultrasonic Nebs

1980s Cyclonic chambers

1987 Glass Expansion's first neb – the VeeSpray (ceramic V-groove)

1989 Glass Expansion first concentric – the Conikal (machined instead of glass blown)

1989 Noordermeer Glass V Groove (US patent #4,880,164) [10]

1992 Glass Expansion – non salting Sea Spray

1993 Modified Lichte Glass V-Groove

1993 Burgener BTF – first Parallel Path Neb (US patent #5,411,208) [11]

1994–1995 Main Burgener Parallel Path Nebs – BTS 50, BTN & T2002

Mid 1990s Perkin Elmer GemCone: Miniature V-Groove

With the introduction of the ICP-MS to the laboratory, the creation of micro nebulizers became a priority in order to deliver smaller amounts of sample at lower flow rates.

1993 The Meinhard HEN (high efficiency nebulizer) was produced which handled very low flow rates but salted and plugged easily as a result. (25 times less sample than a standard Meinhard)

1997 Cetac Microconcentric Nebulizer – first Teflon concentric 50, 100, 200 or 400 µL/min

1997 Meinhard Direct Injection HEN – (DIHEN) (US Patent #6,166,379) [12]

1999 Elemental Scientific – PFA Concentric Nebs 20, 50, 100 or 400 µL/min

1999 Burgener Micro 1: Parallel Path

2000 Burgener Micro 3: Parallel Path

2001 Burgener Mira Mist: First Enhanced Parallel Path Nebulizer (US patent #6,634,572) [13]

2004 Epond Typhoon: Glass Concentric

2005 Ingeniatrics OneNeb: Flow Blurring Technology

2010 Epond Lucida: Teflon Micro Concentric

2012 Burgener PFA 250: PFA Micro flow Enhanced Parallel Path Nebulizer

2010 – 2013 Meinhard and Glass Expansion: Significant improvements in attachments and designs of glass concentrics.

Related Research Articles

In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent called the mobile phase, which carries it through a system on which a material called the stationary phase is fixed. Because the different constituents of the mixture tend to have different affinities for the stationary phase and are retained for different lengths of time depending on their interactions with its surface sites, the constituents travel at different apparent velocities in the mobile fluid, causing them to separate. The separation is based on the differential partitioning between the mobile and the stationary phases. Subtle differences in a compound's partition coefficient result in differential retention on the stationary phase and thus affect the separation.

Inductively coupled plasma mass spectrometry

Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry that uses an inductively coupled plasma to ionize the sample. It atomizes the sample and creates atomic and small polyatomic ions, which are then detected. It is known and used for its ability to detect metals and several non-metals in liquid samples at very low concentrations. It can detect different isotopes of the same element, which makes it a versatile tool in isotopic labeling.

Electrospray ionization Technique used in mass spectroscopy

Electrospray ionization (ESI) is a technique used in mass spectrometry to produce ions using an electrospray in which a high voltage is applied to a liquid to create an aerosol. It is especially useful in producing ions from macromolecules because it overcomes the propensity of these molecules to fragment when ionized. ESI is different from other ionization processes since it may produce multiple-charged ions, effectively extending the mass range of the analyser to accommodate the kDa-MDa orders of magnitude observed in proteins and their associated polypeptide fragments.

Nebulizer drug delivery device

In medicine, a nebulizer or nebuliser is a drug delivery device used to administer medication in the form of a mist inhaled into the lungs. Nebulizers are commonly used for the treatment of asthma, cystic fibrosis, COPD and other respiratory diseases or disorders. They use oxygen, compressed air or ultrasonic power to break up solutions and suspensions into small aerosol droplets that are inhaled from the mouthpiece of the device. An aerosol is a mixture of gas and solid or liquid particles.

Originally the gas flow computer was a mechanical or later a pneumatic or hydraulic computing module, subsequently superseded in most applications by an electronic module, as the primary elements switched from transmitting the measured variables from pneumatic or hydraulic pressure signals to electric current as explosion-proof ) and then intrinsically safe transmitters became available, that simply provided a dedicated gas flow computer function. Today "gas flow computers" as such have become uncommon, since gas flow computing is a subfunction of a data acquisition and control program implemented with programmable logic controller (PLCs) and remote terminal unit (RTUs); with the rise of smart transmitters in the early 1980s, these functions have also been incorporated within the field transmitters themselves.

Liquid chromatography–mass spectrometry Analytical chemistry technique

Liquid chromatography–mass spectrometry (LC–MS) is an analytical chemistry technique that combines the physical separation capabilities of liquid chromatography with the mass analysis capabilities of mass spectrometry (MS). Coupled chromatography - MS systems are popular in chemical analysis because the individual capabilities of each technique are enhanced synergistically. While liquid chromatography separates mixtures with multiple components, mass spectrometry provides spectral information that may help to identify each separated component. MS is not only sensitive, but provides selective detection, relieving the need for complete chromatographic separation. LC-MS is also appropriate for Metabolomics because of its good coverage of a wide range of chemicals. This tandem technique can be used to analyze biochemical, organic, and inorganic compounds commonly found in complex samples of environmental and biological origin. Therefore, LC-MS may be applied in a wide range of sectors including biotechnology, environment monitoring, food processing, and pharmaceutical, agrochemical, and cosmetic industries.

Atmospheric-pressure chemical ionization

Atmospheric pressure chemical ionization (APCI) is an ionization method used in mass spectrometry which utilizes gas-phase ion-molecule reactions at atmospheric pressure (105 Pa), commonly coupled with high-performance liquid chromatography (HPLC). APCI is a soft ionization method similar to chemical ionization where primary ions are produced on a solvent spray. The main usage of APCI is for polar and relatively less polar thermally stable compounds with molecular weight less than 1500 Da. The application of APCI with HPLC has gained a large popularity in trace analysis detection such as steroids, pesticides and also in pharmacology for drug metabolites.

Inductively coupled plasma atomic emission spectroscopy

Inductively coupled plasma atomic emission spectroscopy (ICP-AES), also referred to as inductively coupled plasma optical emission spectrometry (ICP-OES), is an analytical technique used for the detection of chemical elements. It is a type of emission spectroscopy that uses the inductively coupled plasma to produce excited atoms and ions that emit electromagnetic radiation at wavelengths characteristic of a particular element. The plasma is a high temperature source of ionised source gas. The plasma is sustained and maintained by inductive coupling from cooled electrical coils at megahertz frequencies. The source temperature is in the range from 6000 to 10,000 K. The intensity of the emissions from various wavelengths of light are proportional to the concentrations of the elements within the sample.

Dust collector

A dust collector is a system used to enhance the quality of air released from industrial and commercial processes by collecting dust and other impurities from air or gas. Designed to handle high-volume dust loads, a dust collector system consists of a blower, dust filter, a filter-cleaning system, and a dust receptacle or dust removal system. It is distinguished from air purifiers, which use disposable filters to remove dust.

In mass spectrometry, direct analysis in real time (DART) is an ion source that produces electronically or vibronically excited-state species from gases such as helium, argon, or nitrogen that ionize atmospheric molecules or dopant molecules. The ions generated from atmospheric or dopant molecules undergo ion-molecule reactions with the sample molecules to produce analyte ions. Analytes with low ionization energy may be ionized directly. The DART ionization process can produce positive or negative ions depending on the potential applied to the exit electrode.

Spray nozzle Device that facilitates dispersion of liquid into a spray

A spray nozzle is a precision device that facilitates dispersion of liquid into a spray. Nozzles are used for three purposes: to distribute a liquid over an area, to increase liquid surface area, and create impact force on a solid surface. A wide variety of spray nozzle applications use a number of spray characteristics to describe the spray.

Thermospray

Thermospray is a soft ionization source by which a solvent flow of liquid sample passes through a very thin heated column to become a spray of fine liquid droplets. As a form of atmospheric pressure ionization in mass spectrometry these droplets are then ionized via a low-current discharge electrode to create a solvent ion plasma. A repeller then directs these charged particles through the skimmer and acceleration region to introduce the aerosolized sample to a mass spectrometer. It is particularly useful in liquid chromatography-mass spectrometry (LC-MS).

The term separator in oilfield terminology designates a pressure vessel used for separating well fluids produced from oil and gas wells into gaseous and liquid components. A separator for petroleum production is a large vessel designed to separate production fluids into their constituent components of oil, gas and water. A separating vessel may be referred to in the following ways: Oil and gas separator, Separator, Stage separator, Trap, Knockout vessel, Flash chamber, Expansion separator or expansion vessel, Scrubber, Filter. These separating vessels are normally used on a producing lease or platform near the wellhead, manifold, or tank battery to separate fluids produced from oil and gas wells into oil and gas or liquid and gas. An oil and gas separator generally includes the following essential components and features:

Desorption electrospray ionization

Desorption electrospray ionization (DESI) is an ambient ionization technique that can be coupled to mass spectrometry (MS) for chemical analysis of samples at atmospheric conditions. Coupled ionization sources-MS systems are popular in chemical analysis because the individual capabilities of various sources combined with different MS systems allow for chemical determinations of samples. DESI employs a fast-moving charged solvent stream, at an angle relative to the sample surface, to extract analytes from the surfaces and propel the secondary ions toward the mass analyzer. This tandem technique can be used to analyze forensics analyses, pharmaceuticals, plant tissues, fruits, intact biological tissues, enzyme-substrate complexes, metabolites and polymers. Therefore, DESI-MS may be applied in a wide variety of sectors including food and drug administration, pharmaceuticals, environmental monitoring, and biotechnology.

Respiratory gas humidification is a method of artificially conditioning respiratory gas for the patient during therapy, and involves humidification, warming, and occasionally filtration of the gas being delivered. If these three measures are not performed to compensate for the natural conditioning of air by the respiratory system, lung infections and lung tissue damage may occur. This is particularly problematic in high gas-flow therapies such as [mechanical ventilation], in patient populations with highly sensitive respiratory tracts, or among those requiring ventilation for longer periods of time. The two methods currently available for this purpose are active or passive respiratory gas humidification.

Atomizer nozzle Heart of a tool to generate and spray a liquid mist

An atomizer nozzle is the heart of a tool that generates a spray of liquid mist. It can take many forms. The first atomizer nozzle, also called an aspirator nozzle, was invented by Dr. Thomas DeVilbiss of Toledo, Ohio, in the late 19th century for producing a fine spray of a liquid based on the Venturi effect. His device was used for spraying medicine on the back of his patients' throats. Atomizer nozzles can create atomization from a variety of mechanical means, which includes but is not limited to electrostatic processes, ultrasonic nozzle and centrifugal forces.

Atomic emission spectroscopy Analytical method using radiation to identify chemical elements in a sample

Atomic emission spectroscopy (AES) is a method of chemical analysis that uses the intensity of light emitted from a flame, plasma, arc, or spark at a particular wavelength to determine the quantity of an element in a sample. The wavelength of the atomic spectral line in the emission spectrum gives the identity of the element while the intensity of the emitted light is proportional to the number of atoms of the element. The sample may be excited by various methods.

Direct electron ionization liquid chromatography–mass spectrometry interface

A direct electron ionization liquid chromatography–mass spectrometry interface is a technique for coupling liquid chromatography and mass spectrometry (LC-MS) based on the direct introduction of the liquid effluent into an electron ionization (EI) source. Library searchable mass spectra are generated. Gas-phase EI has many applications for the detection of HPLC amenable compounds showing minimal adverse matrix effects. The direct-EI LC-MS interface provides access to well-characterized electron ionization data for a variety of LC applications and readily interpretable spectra from electronic libraries for environmental, food safety, pharmaceutical, biomedical, and other applications.

Flow focusing in fluid dynamics is a technology whose aim is the production of drops or bubbles by straightforward hydrodynamic means. The output is a dispersed liquid or gas, frequently in the form of a fine aerosol or an emulsion. No other driving force is required, apart from traditional pumping, a key difference with other comparable technologies, such as electrospray. Both flow focusing and electrospray working in their most extensively used regime produce high quality sprays composed by homogeneous and well-controlled-size droplets. Flow focusing was invented by Prof. Alfonso M. Ganan-Calvo in 1994, patented in 1996, and published for the first time in 1998.

Atmospheric pressure photoionization Soft ionization method

Atmospheric pressure photoionization (APPI) is a soft ionization method used in mass spectrometry (MS) usually coupled to liquid chromatography (LC). Molecules are ionized using a vacuum ultraviolet (VUV) light source operating at atmospheric pressure, either by direct absorption followed by electron ejection or through ionization of a dopant molecule that leads to chemical ionization of target molecules. The sample is usually a solvent spray that is vaporized by nebulization and heat. The benefit of APPI is that it ionizes molecules across a broad range of polarity and is particularly useful for ionization of low polarity molecules for which other popular ionization methods such as electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) are less suitable. It is also less prone to ion suppression and matrix effects compared to ESI and APCI and typically has a wide linear dynamic range. The application of APPI with LC/MS is commonly used for analysis of petroleum compounds, pesticides, steroids, and drug metabolites lacking polar functional groups and is being extensively deployed for ambient ionization particularly for explosives detection in security applications.

References

  1. "Glass Concentric Nebulizers" . Retrieved March 8, 2013.
  2. Coleman, Geoff. "Our History". Meinhard. Retrieved 5 March 2013.
  3. Robinson, James; et al. (2005). Undergraduate Instrumental Analysis, Sixth Edition. New York: Marcel Dekker. p. 493. ISBN   9780849306501.
  4. Burgener, John. "Enhanced Parallel Path" . Retrieved 5 March 2016.
  5. Ingeniatrics Tecnologías. "Flow Blurring Technology" . Retrieved 5 March 2013.
  6. Burgener, John. "Nebulizer History 2013" (PDF). Burgener Research Inc. Retrieved 5 March 2013.
  7. Basil Meddings, Heinz Kaiser "Cross-flow nebulizer" U.S. Patent 4,344,574 , Issue date: Aug 17, 1982
  8. Ronald F. Suddendorf, Kenneth W. Boyer "Mechanical device to produce a finely dispersed aerosol" U.S. Patent 4,206,160 , Issue date: Jun 3, 1980
  9. Velmer A. Fassel, Gary W. Rice, Kimberly E. Lawrence "Concentric micro-nebulizer for direct sample insertion" U.S. Patent 4,575,609 , Issue date: Mar 11, 1986
  10. Michael N. A. Noordermeer "One-piece high-solids nebulizer" U.S. Patent 4,880,164 , Issue date: Nov 14, 1989
  11. John A. Burgener "Parallel path induction pneumatic nebulizer" U.S. Patent 5,411,208 , Issue date: May 2, 1995
  12. Akbar Montaser, John A. McLean, Jerold M. Kacsir "Direct injection high efficiency nebulizer for analytical spectrometry" U.S. Patent 6,166,379 , Issue date: Dec 26, 2000
  13. John A. Burgener "Enhanced parallel path nebulizer with a large range of flow rates" U.S. Patent 6,634,572 , Issue date: Oct 21, 2003