Ion beam analysis

Last updated

Ion beam analysis (IBA) is an important family of modern analytical techniques involving the use of MeV ion beams to probe the composition and obtain elemental depth profiles in the near-surface layer of solids. All IBA methods are highly sensitive and allow the detection of elements in the sub-monolayer range. The depth resolution is typically in the range of a few nanometers to a few ten nanometers. Atomic depth resolution can be achieved, but requires special equipment. The analyzed depth ranges from a few ten nanometers to a few ten micrometers. IBA methods are always quantitative with an accuracy of a few percent. Channeling allows to determine the depth profile of damage in single crystals.

Contents

The quantitative evaluation of IBA methods requires the use of specialized simulation and data analysis software. SIMNRA and DataFurnace are popular programs for the analysis of RBS, ERD and NRA, while GUPIX is popular for PIXE. A review of IBA software [2] was followed by an intercomparison of several codes dedicated to RBS, ERD and NRA, organized by the International Atomic Energy Agency. [3]

IBA is an area of active research. The last major Nuclear Microbeam conference in Debrecen (Hungary) was published in NIMB 267(12-13).

Overview

Ion beam analysis works on the basis that ion-atom interactions are produced by the introduction of ions to the sample being tested. Major interactions result in the emission of products that enable information regarding the number, type, distribution and structural arrangement of atoms to be collected. To use these interactions to determine sample composition a technique must be selected along with irradiation conditions and the detection system that will best isolate the radiation of interest providing the desired sensitivity and detection limits. The basic layout of an ion beam apparatus is an accelerator which produces an ion beam that is feed through an evacuated beam-transport tube to a beam handling device. This device isolates the ion species and charge of interest which then are transported through an evacuated beam-transport tube into the target chamber. This chamber is where the refined ion beam will come into contact with the sample and thus the resulting interactions can be observed. The configuration of the ion beam apparatus can be changed and made more complex with the incorporation of additional components. The techniques for ion beam analysis are designed for specific purposes. Some techniques and ion sources are shown in table 1. Detector types and arrangements for ion beam techniques are shown in table 2.

Table 1: Techniques and Ion Sources [4]
Ion SourceCurrentIon BeamsTechnique
Low EnergyRadio frequency1 mAH, He, N, OLEIS
Duoplasmatron10 mAH, He, N, O
Colutron
Penning5 mAC, N, Ne, Kr
CaeDsium sputterMost SolidSIMS
Freeman10 mAMost Solid
Electron impact
LMISGa, In, Au, Bi
High EnergyPositive ions
Radio-frequency1 mAH, He, N, ORBS, PIXE, NRA
Duoplamatron10 mAH, He, N, O
Negative ions
Duoplasmatron

(off-axis extraction)

100 mAH, ORBS, PIXE, NRA
Penning2 mAH, 2H
Sputter-sourceMost
RF with charge exchange100 mAH, He, N, O
Table 2: Detector Types and Arrangements for Ion Beam Techniques [4]
ProductDetectorConfigurationVacuum
LEISScattered IonsChanneltronVacuum, movable advantageous

Energy measurement requires

Electrostatic/magnetic analyser

10 nPa
SIMSSecondary IonsChanneltronVacuum, fixed geometry

Low mass resolution with ESA, QMA

High mass resolution with Sector Field Analyser

< 1mPa
SIPSOptical PhotonsSpectrophotometerExternal to chamber, Fixed geometry, High wavelength resolution.< 1mPa
PIXEX-RaysSi (Li)

IG

Vacuum or external. Filters Thin Window. Liquid N cooling< 1mPa
RBSIonsSurf.barrierVacuum, movable geometry

Small and simple arrangement

RBS-CIonsSurf.barrier< 100 mPa
ERAIonsSurf.barrierGlancing angle geometry for improved depth resolution
NRAIonsSurf.barrier
PIGMEGamma-raysGe (Li)

NaI

External with window, cryostat

High Resolution, Low efficiency

Poor Resolution, high efficiency

< 100 mPa
NRANeutronsBF3

Li glass

Scintillator

External, low efficiency

Detection only

Broad resolution by unfolding

Applications

Ion beam analysis has found use in a number of variable applications, ranging from biomedical uses to studying ancient artifacts. The popularity of this technique stems from the sensitive data that can be collected without significant distortion to the system on which it is studying. The unparalleled success found in using ion beam analysis has been virtually unchallenged over the past thirty years until very recently with new developing technologies. Even then, the use of ion beam analysis has not faded, and more applications are being found that take advantage of its superior detection capabilities. In an era where older technologies can become obsolete at an instant, ion beam analysis has remained a mainstay and only appears to be growing as researchers are finding greater use for the technique.

Biomedical elemental analysis

Gold nanoparticles have been recently used as a basis for a count of atomic species, especially with studying the content of cancer cells. [5] Ion beam analysis is a great way to count the amount of atomic species per cell. Scientists have found an effective way to make accurate quantitative data available by using ion beam analysis in conjunction with elastic backscattering spectrometry (EBS). [5] The researchers of a gold nanoparticle study were able to find much greater success using ion beam analysis in comparison to other analytical techniques, such as PIXE or XRF. [5] This success is due to the fact that the EBS signal can directly measure depth information using ion beam analysis, whereas this cannot be done with the other two methods. The unique properties of ion beam analysis make great use in a new line of cancer therapy.

Cultural heritage studies

Ion beam analysis also has a very unique application in the use of studying archaeological artifacts, also known as archaeometry. [6] For the past three decades, this has been the much preferred method to study artifacts while preserving their content. What many have found useful in using this technique is its offering of excellent analytical performance and non-invasive character. More specifically, this technique offers unparalleled performance in terms of sensitivity and accuracy. Recently however, there have been competing sources for archaeometry purposes using X-ray based methods such as XRF. Nonetheless, the most preferred and accurate source is ion beam analysis, which is still unmatched in its analysis of light elements and chemical 3D imaging applications (i.e. artwork and archaeological artifacts). [6] [7]

Forensic analysis

A third application of ion beam analysis is in forensic studies, particularly with gunshot residue characterization. Current characterization is done based on heavy metals found in bullets, however, manufacturing changes are slowly making these analyses obsolete. The introduction of techniques such as ion beam analysis are believed to alleviate this issue. Researchers are currently studying the use of ion beam analysis in conjunction with a scanning electron microscope and an Energy Dispersive X-ray spectrometer (SEM-EDS). [8] The hope is that this setup will detect the composition of new and old chemicals that older analyses could not efficiently detect in the past. [8] The greater amount of analytical signal used and more sensitive lighting found in ion beam analysis gives great promise to the field of forensic science.

Lithium battery development

The spatially resolved detection of light elements, for example lithium, remains challenging for most techniques based on the electronic shell of the target atoms such as XRF or SEM-EDS. For lithium and lithium ion batteries, the quantification of the lithium stoichiometry and its spatial distribution are important to understand the mechanisms behind dis-/charging and aging. Through ion beam focussing and a combination of methods, ion beam analysis offers the unique possibility for measuring the local state of charge (SoC) on the µm-scale. [9]

Iterative IBA

Ion beam-based analytical techniques represent a powerful set of tools for non-destructive, standard-less, depth-resolved and highly accurate elemental composition analysis in the depth regime from several nm up to few μm. [10] By changing type of incident ion, the geometry of experiment, particle energy, or by acquiring different products originating from ion-solid interaction, complementary information can be extracted. However, analysis is often challenged either in terms of mass resolution—when several comparably heavy elements are present in the sample—or in terms of sensitivity—when light species are present in heavy matrices. Hence, a combination of two or more ion beam-based techniques can overcome the limitations of each individual method and provide complementary information about the sample. [4] [5]

Overview of various ion-surface interactions. (1)-incoming ion; (2)-scattering; (3)-neutralization and scattering; (4)-sputtering or recoiling; (5)-electron emission; (6)-photon emission; (7)-adsorption; (8)-displacement, e.g. from sputtering event Ion surface interactions.gif
Overview of various ion-surface interactions. (1)-incoming ion; (2)-scattering; (3)-neutralization and scattering; (4)-sputtering or recoiling; (5)-electron emission; (6)-photon emission; (7)-adsorption; (8)-displacement, e.g. from sputtering event

An iterative and self-consistent analysis also enhances the accuracy of the information that can be obtained from each independent measurement. [11] [12] [13]

Software and simulation

Dating back to the 1960s the data collected via ion beam analysis has been analyzed through a multitude of computer simulation programs. Researchers who frequently use ion beam analysis in conjunction with their work require that this software be accurate and appropriate for describing the analytical process they are observing. [14] Applications of these software programs range from data analysis to theoretical simulations and modeling based on assumptions about the atomic data, mathematics and physics properties that detail the process in question. As the purpose and implementation of ion beam analysis has changed over the years, so has the software and codes used to model it. Such changes are detailed through the five classes by which the updated software are categorized. [15] [16]

Class-A

Includes all programs developed in the late 1960s and early 1970s. This class of software solved specific problems in the data; niy did not provide the full potential to analyze a spectrum of a full general case. The prominent pioneering program was IBA, developed by Ziegler and Baglin in 1971. At the time, the computational models only tackled the analysis associated with the back-scattering techniques of ion beam analysis and performed calculation based on a slab analysis. A variety of other programs arose during this time, such as RBSFIT, though due to the lack of in-depth knowledge on ion beam analysis, it became increasingly hard to develop programs that accurate.

Class-B

A new wave of programs sought to solve this accuracy problem in this next class of software. Developed during the 1980s, programs like SQEAKIE and BEAM EXPERT, afforded an opportunity to solve the complete general case by employing codes to perform direct analysis. This direct approach unfolds the produced spectrum with no assumptions made about the sample. Instead it calculates through separated spectrum signals and solves a set of linear equations for each layer. Problems still arise, though, and adjustments made to reduce noise in the measurements and room for uncertainty.

Class-C

In a trip back to square one, this third class of programs, created in the 1990s, take a few principles from Class A in accounting for the general case, however, now through the use of indirect methods. RUMP and SENRAS, for example, use an assumed model of the sample and simulate a comparative theoretical spectra, which afforded such properties as fine structure retention and uncertainty calculations. In addition to the improvement in software analysis tools came the ability to analyze other techniques aside from back-scattering; i.e. ERDA and NRA.

Class-D

Exiting the Class C era and into the early 2000s, software and simulation programs for ion beam analysis were tackling a variety of data collecting techniques and data analysis problems. Following along with the world's technological advancements, adjustments were made to enhance the programs into a state more generalized codes, spectrum evaluation, and structural determination. Programs produced like SIMNRA now account for the more complex interactions with the beam and sample; also providing a known database of scattering data.

Class-E

This most recently developed class, having similar characteristics to the previous, makes use of primary principles in the Monte Carlo computational techniques. [17] This class applies molecular dynamic calculations that are able to analyze both low and high energy physical interactions taking place in the ion beam analysis. A key and popular feature that accompanies such techniques is the possibility for the computations to be incorporated in real time with the ion beam analysis experiment itself.

Footnotes

  1. Huddle et al. (2007)
  2. Rauhala et al. (2006)
  3. Barradas et al. (2007)
  4. 1 2 3 WILLIAMS, J. S.; BIRD, J. R. (1989-01-01). 1 - Concepts and Principles of Ion Beam Analysis. San Diego: Academic Press. pp. 3–102. doi:10.1016/b978-0-08-091689-7.50006-9. ISBN   9780120997404.
  5. 1 2 3 4 Jeynes, J. Charles (September 26, 2013). "Measuring and Modeling Cell-to-Cell Variation in Uptake of Gold Nanoparticles". Analyst. 138 (23): 7070–4. Bibcode:2013Ana...138.7070J. doi: 10.1039/c3an01406a . PMID   24102065.
  6. 1 2 Dran, Jean-Claude (November 24, 2013). Ion Beam Analysis in Cultural Heritage Studies: Milestones and Perspectives. Multidisciplinary Applications of Nuclear Physics with Ion Beams. AIP Conference Proceedings. Vol. 1530. pp. 11–24. Bibcode:2013AIPC.1530...11D. doi:10.1063/1.4812900.
  7. "Applications of Ion Beam Analysis". www.surrey.ac.uk. Retrieved 2016-04-29.
  8. 1 2 Romolo, F.S. (May 2, 2013). "Integrated Ion Beam Analysis (IBA) in Gunshot Residue (GSR) Characterisation". Forensic Science International. 231 (1–3): 219–228. doi:10.1016/j.forsciint.2013.05.006. PMID   23890641.
  9. Möller, Sören; Satoh, Takahiro; Ishii, Yasuyuki; Teßmer, Britta; Guerdelli, Rayan; Kamiya, Tomihiro; Fujita, Kazuhisa; Suzuki, Kota; Kato, Yoshiaki; Wiemhöfer, Hans-Dieter; Mima, Kunioki (June 2021). "Absolute Local Quantification of Li as Function of State-of-Charge in All-Solid-State Li Batteries via 2D MeV Ion-Beam Analysis". Batteries. 7 (2): 41. doi: 10.3390/batteries7020041 .
  10. Handbook of modern ion beam materials analysis. Wang, Yongqiang., Nastasi, Michael Anthony, 1950- (2nd ed.). Warrendale, Pa.: Materials Research Society. 2009. ISBN   978-1-60511-217-6. OCLC   672203193.{{cite book}}: CS1 maint: others (link)
  11. Moro, M.V.; Holeňák, R.; Zendejas Medina, L.; Jansson, U.; Primetzhofer, D. (September 2019). "Accurate high-resolution depth profiling of magnetron sputtered transition metal alloy films containing light species: A multi-method approach". Thin Solid Films. 686: 137416. arXiv: 1812.10340 . Bibcode:2019TSF...686m7416M. doi:10.1016/j.tsf.2019.137416. S2CID   119415711.
  12. Comparotto, C.; Petter, S.; Donzel-Gargand, O.; Kubart, T.; Scragg, J.J.S. (April 2022). "Synthesis of BaZrS3 Perovskite Thin Films at a Moderate Temperature on Conductive Substrates". ACS Appl. Energy Mater. 5 (5): 6335–6343. doi: 10.1021/acsaem.2c00704 . S2CID   248359886.
  13. Jeynes, C.; Bailey, M.J.; Bright, N.J.; Christopher, M.E.; Grime, G.W.; Jones, B.N.; Palitsin, V.V.; Webb, R.P. (January 2012). ""Total IBA" – Where are we?" (PDF). Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 271: 107–118. Bibcode:2012NIMPB.271..107J. doi:10.1016/j.nimb.2011.09.020.
  14. Barradas, N.P. (2007). "International Atomic Energy Agency Intercomparison of Ion Beam Analysis Software". Nuclear Instruments and Methods in Physics Research B. 262 (2): 281. Bibcode:2007NIMPB.262..281B. doi:10.1016/j.nimb.2007.05.018. hdl: 11858/00-001M-0000-0027-0732-B .
  15. Rauhala, E. (2006). "Status of ion beam data analysis and simulation software". Nuclear Instruments and Methods in Physics Research B. 244 (2): 436. Bibcode:2006NIMPB.244..436R. doi:10.1016/j.nimb.2005.10.024. hdl: 11858/00-001M-0000-0027-0B1E-C .
  16. "Fullerene Simulations Introduction". www.surrey.ac.uk. Retrieved 2016-04-29.
  17. Schiettekatte, F (2008). "Fast Monte Carlo for ion beam analysis simulations". Nuclear Instruments and Methods in Physics Research B. 266 (8): 1880. Bibcode:2008NIMPB.266.1880S. doi:10.1016/j.nimb.2007.11.075.

Related Research Articles

<span class="mw-page-title-main">Ion implantation</span> Use of ions to cause chemical changes

Ion implantation is a low-temperature process by which ions of one element are accelerated into a solid target, thereby changing the physical, chemical, or electrical properties of the target. Ion implantation is used in semiconductor device fabrication and in metal finishing, as well as in materials science research. The ions can alter the elemental composition of the target if they stop and remain in the target. Ion implantation also causes chemical and physical changes when the ions impinge on the target at high energy. The crystal structure of the target can be damaged or even destroyed by the energetic collision cascades, and ions of sufficiently high energy can cause nuclear transmutation.

Particle-induced X-ray emission or proton-induced X-ray emission (PIXE) is a technique used for determining the elemental composition of a material or a sample. When a material is exposed to an ion beam, atomic interactions occur that give off EM radiation of wavelengths in the x-ray part of the electromagnetic spectrum specific to an element. PIXE is a powerful yet non-destructive elemental analysis technique now used routinely by geologists, archaeologists, art conservators and others to help answer questions of provenance, dating and authenticity.

A microprobe is an instrument that applies a stable and well-focused beam of charged particles to a sample.

<span class="mw-page-title-main">Beryllium-10</span> Isotope of beryllium

Beryllium-10 (10Be) is a radioactive isotope of beryllium. It is formed in the Earth's atmosphere mainly by cosmic ray spallation of nitrogen and oxygen. Beryllium-10 has a half-life of 1.39 × 106 years, and decays by beta decay to stable boron-10 with a maximum energy of 556.2 keV. It decays through the reaction 10Be→10B + e. Light elements in the atmosphere react with high energy galactic cosmic ray particles. The spallation of the reaction products is the source of 10Be (t, u particles like n or p):

<span class="mw-page-title-main">ISOLDE</span> Physics facility at CERN

The ISOLDE Radioactive Ion Beam Facility, is an on-line isotope separator facility located at the centre of the CERN accelerator complex on the Franco-Swiss border. Created in 1964, the ISOLDE facility started delivering radioactive ion beams (RIBs) to users in 1967. Originally located at the Synchro-Cyclotron (SC) accelerator, the facility has been upgraded several times most notably in 1992 when the whole facility was moved to be connected to CERN's ProtonSynchroton Booster (PSB). ISOLDE is currently the longest-running facility in operation at CERN, with continuous developments of the facility and its experiments keeping ISOLDE at the forefront of science with RIBs. ISOLDE benefits a wide range of physics communities with applications covering nuclear, atomic, molecular and solid-state physics, but also biophysics and astrophysics, as well as high-precision experiments looking for physics beyond the Standard Model. The facility is operated by the ISOLDE Collaboration, comprising CERN and sixteen (mostly) European countries. As of 2019, close to 1,000 experimentalists around the world are coming to ISOLDE to perform typically 50 different experiments per year.

Nuclear reaction analysis (NRA) is a nuclear method of nuclear spectroscopy in materials science to obtain concentration vs. depth distributions for certain target chemical elements in a solid thin film.

Elastic recoil detection analysis (ERDA), also referred to as forward recoil scattering, is an ion beam analysis technique in materials science to obtain elemental concentration depth profiles in thin films. This technique is known by several different names. These names are listed below. In the technique of ERDA, an energetic ion beam is directed at a sample to be characterized and there is an elastic nuclear interaction between the ions of beam and the atoms of the target sample. Such interactions are commonly of Coulomb nature. Depending on the kinetics of the ions, cross section area, and the loss of energy of the ions in the matter, ERDA helps determine the quantification of the elemental analysis. It also provides information about the depth profile of the sample.

<span class="mw-page-title-main">Stopping power (particle radiation)</span> Retarding force acting on charged particles due to interactions with matter

In nuclear and materials physics, stopping power is the retarding force acting on charged particles, typically alpha and beta particles, due to interaction with matter, resulting in loss of particle kinetic energy. Stopping power is also interpreted as the rate at which a material absorbs the kinetic energy of a charged particle. Its application is important in a wide range of thermodynamic areas such as radiation protection, ion implantation and nuclear medicine.

A radio-frequency quadrupole (RFQ) beam cooler is a device for particle beam cooling, especially suited for ion beams. It lowers the temperature of a particle beam by reducing its energy dispersion and emittance, effectively increasing its brightness (brilliance). The prevalent mechanism for cooling in this case is buffer-gas cooling, whereby the beam loses energy from collisions with a light, neutral and inert gas. The cooling must take place within a confining field in order to counteract the thermal diffusion that results from the ion-atom collisions.

Gas cluster ion beams (GCIB) is a technology for nano-scale modification of surfaces. It can smooth a wide variety of surface material types to within an angstrom of roughness without subsurface damage. It is also used to chemically alter surfaces through infusion or deposition.

Rutherford backscattering spectrometry (RBS) is an analytical technique used in materials science. Sometimes referred to as high-energy ion scattering (HEIS) spectrometry, RBS is used to determine the structure and composition of materials by measuring the backscattering of a beam of high energy ions (typically protons or alpha particles) impinging on a sample.

<span class="mw-page-title-main">Collision cascade</span> Series of collisions between nearby atoms, initiated by a single energetic atom

In condensed-matter physics, a collision cascade is a set of nearby adjacent energetic collisions of atoms induced by an energetic particle in a solid or liquid.

Stopping and Range of Ions in Matter (SRIM) is a group of computer programs which calculate interactions between ions and matter; the core of SRIM is a program called Transport of Ions in Matter (TRIM). SRIM is popular in the ion implantation research and technology community, and also used widely in other branches of radiation material science.

Swift heavy ions are the components of a type of particle beam with high enough energy that electronic stopping dominates over nuclear stopping. They are accelerated in particle accelerators to very high energies, typically in the MeV or GeV range and have sufficient energy and mass to penetrate solids on a straight line. In many solids swift heavy ions release sufficient energy to induce permanently modified cylindrical zones, so-called ion tracks. If the irradiation is carried out in an initially crystalline material, ion tracks consist of an amorphous cylinder. Ion tracks can be produced in many amorphizing materials, but not in pure metals, where the high electronic heat conductivity dissipates away the electronic heating before the ion track has time to form.

<span class="mw-page-title-main">Binary collision approximation</span> Heuristic used in simulations of ions passing through solids

In condensed-matter physics, the binary collision approximation (BCA) is a heuristic used to more efficiently simulate the penetration depth and defect production by energetic ions in solids. In the method, the ion is approximated to travel through a material by experiencing a sequence of independent binary collisions with sample atoms (nuclei). Between the collisions, the ion is assumed to travel in a straight path, experiencing electronic stopping power, but losing no energy in collisions with nuclei.

Particle-induced gamma emission (PIGE) is a form of nuclear reaction analysis, one of the ion beam analysis thin-film analytical techniques.

<span class="mw-page-title-main">Main Magnetic Focus Ion Source</span>

Main Magnetic Focus Ion Source (MaMFIS) is a compact ion source with extremely high electron current density. The device is designed for production of ions of arbitrary elements in any charge states, in particular, of highly charged ions of heavy elements.

<span class="mw-page-title-main">CRIS experiment</span>

The Collinear Resonance Ionization Spectroscopy (CRIS) experiment is located in the ISOLDE facility at CERN. The experiment aims to study ground-state properties of exotic nuclei and produce high purity isomeric beams used for decay studies. CRIS does this by using the high resolution technique of fast beam collinear laser spectroscopy, with the high efficiency technique of resonance ionization.

<span class="mw-page-title-main">MIRACLS experiment</span>

The Multi Ion Reflection Apparatus for CoLlinear Spectroscopy (MIRACLS) is a permanent experiment setup being constructed at the ISOLDE facility at CERN. The purpose of the experiment is to measure properties of exotic radioisotopes, from precise measurements of their hyperfine structure. MIRACLS will use laser spectroscopy for measurements, aiming to increase the sensitivity of the technique by trapping ion bunches in an ion trap.