Synthetic aperture ultrasound

Last updated
Example of 2-D Slices of 3-D Synthetic Aperture Ultrasound Imaging From Field-II Simulation UltrasoundImage.PNG
Example of 2-D Slices of 3-D Synthetic Aperture Ultrasound Imaging From Field-II Simulation

Synthetic aperture ultrasound (SAU) imaging is an advanced form of imaging technology used to form high-resolution images in biomedical ultrasound systems. Ultrasound imaging has become an important and popular medical imaging method, as it is safer and more economical than computer tomography (CT) and magnetic resonance imaging (MRI).

Contents

Compared with the conventional ultrasound image formation where one transducer or linear array is used, SAU imaging has achieved higher lateral resolution and deeper penetration, which will enable a more accurate diagnosis in medical applications, with no obvious loss in frame rate and without a large burden in computational complexities. [2]

Background

SAU is based on synthetic beamforming and focusing. These are processes whereby the pulse-echo responses from individual pairs of elements are synthesized to reconstruct the formation and focusing, relying on the rule of linear superposition. The working principle of a basic SAU imaging system is one that transmits an ultrasound wave by each small group of all array elements in turn, receive a signal using all possible receive array units, and then reconstruct the image by weighted summation of all demodulated received recordings. [3]

The important advantage of SAU is its ability to achieve dynamic focusing at both transmission and reception without repeating the transmission process with all array elements for each focal zone. It thus saves large computational requirements for delay-and-sum beamforming and leaves space for increasing the frame rate. Also the single transmission and receiving requirement for each firing significantly reduces the hardware complexity for system implementations. [3]

Motivation and history

The basic idea of SAU originated from synthetic aperture radar (SAR) and sonar, where the motion of antenna is used over a region around the target to generate a higher resolution image of the object. In 1972, Burckhardt first put forward the idea that uses synthetic aperture sonar in medical applications and proved that a synthetic aperture system gives a significantly higher lateral resolution than a conventional B-scan. [4]

In 1980, Corl proposed a real-time system implementation for synthetic aperture in medical use. [5] However, at that time, SAU implementations were restricted due to the lack of powerful computational machines at a reasonable cost and size. It was during the 1990s that the development of SAU imaging really began to take off due to the availability of fast computers. [6] [7] [8]

Imaging theory

Theory

One typical implementation of the SAU is the one proposed by Jensen. [9] In his model, only one array element in the transducer is used each time to emit a nearly spherical wave that will cover the whole region of interest. All elements are used to receive the back-scattered signal at the same time, obtaining a low-resolution for each firing. The back-scattered signal contains information required to form the image in all directions and the specific direction can be acquired by applying different delays to signals received by different elements, thus this process works as dynamic focusing at the receiver. By summing all the low-resolution images acquired each time with a determined weight, the final high resolution image is formed and dynamic focusing at transmission is synthesized. [9]

Equations representations

Assuming the lth focal line in the low-resolution image acquired from the ith element's firing is represented as:

Here, represents the jth receive element's apodization factor. represents the received signal and represents the delay applied to the jth receive element when the element receives signal transmitted from ith element, to beamform at a specific direction. So the low resolution image can be represented by

By summing up all the low-resolution images, the region of interest for the high resolution image can be represented as

The high resolution image is dynamically focused both at the transmission end and the receive end.

Imaging efficiency

The major concern for the medical ultrasound imaging is the frame rate, which is determined by the number of lines that are formed by the focal points required to reconstruct the image, and the pulse repletion frequency. By considering these two parameters, and assuming the depth of the expected image is 150 and the propagation speed of sound at 1500 , each line will require 200 to receive the back-scatter information from the farthest focal point. It can be seen that to satisfy the frame with 200 lines each, the frame rate of SAU can achieve a peak at 25 Hz, reflecting the appreciable potential in imaging speed and economy in implementation. [3]

Extension developments

Virtual sources focusing

Passman and Ermert proposed using transmit focal points as virtual sources and this was further investigated in SAU by later researchers. [10] This method treats the focus of a transducer as a virtual element besides the actual source points. SAU imaging can be performed without regard to whether or not the element actually exists, after the aperture angle of the transducer was determined. By introducing the virtual sources focusing method, the resolution of imaging and depth of penetration were increased. It was also suggested that the focusing in azimuth and elevation can be treated separately and that the concept of virtual sources focusing SAU can be applied in two planes. [11] [12]

Sequential beamforming

Kortbek has put forward a sequential beamforming method to reduce the complexity and improve the simplicity of the hardware implementation. The basic idea is to separate the beamforming process into two-stage procedures by using two independent beamformers. The beamformer in the first stage produces scan lines by using a single focal point in both the transmit and receive processes. The second stage beamformer creates a set of high resolution image points by combining information from multiple focused scan lines acquired in the first stage. For a linear array transducer with multiple elements, the lateral resolution of sequential beamforming SAU can be made more range independent and significantly improved compared to conventional dynamic transmit and receive focusing. [13] [14]

Bi-directional pixel-based focusing

The bi-directional pixel-based focusing (BiPBF) method was proposed to solve the problem that SAU imaging suffers from low SNR as the transmission is done by a small part of the array. [15] In BiPBF, the same firing sequence is used in the transmission process as in the traditional array imaging, but the radio frequency (RF) data collected using adjacent groups of array elements are compounded in the receive process. The pixel-based time delays used for compounding are calculated using the distances between pixels and virtual sources located at the successive lateral positions of the transmission focus. Experiments have been done in both phantom and in-vivo experiments; the image quality of the SAU-BiPBF method was considerably improved compared to imaging by conventional dynamic focusing. [16] [17]

3D developments

With the development of very large scale integrated circuit technologies, large workload of computations became achievable which made real-time 3D ultrasound imaging possible. Much research showed that 3D imaging which generates volumetric images, will result in better medical interpretations and more accurate diagnosis than the conventional 2D method. [18] However, the benefits of 3-D ultrasound imaging come with challenges concerning the computational complexities.

3-D imaging requires a 2-D array transducer in which the number of elements can be ten times more than the linear 1-D array transducer. Moreover, the number of scanlines required in a 3-D volume is at least one order of magnitude higher than in a 2-D image in order to form the volumetric display. To reduce the computational complexities, researchers proposed several methods to support the implementation. To replace the traditional 2-D array, 2-D sparse arrays were put forward to reduce the spatial redundancy. [19] [20] The method of preliminary beamforming for reconfigurable arrays was proposed to reduce the number of A/D converters and the front-end complexities. To improve the imaging resolution, separable beamforming for SAU 3D imaging was used to decompose the 2D beamforming as a series of 1-D array processing, which was shown to produce images with comparable quality as the non-separable method. [1] A better performance of these methods and devices will be achieved as the technology is improving in terms of higher computation speed and smaller size.

See also

Related Research Articles

<span class="mw-page-title-main">Phased array</span> Array of antennas creating a steerable beam

In antenna theory, a phased array usually means an electronically scanned array, a computer-controlled array of antennas which creates a beam of radio waves that can be electronically steered to point in different directions without moving the antennas. The general theory of an electromagnetic phased array also finds applications in ultrasonic and medical imaging application and in optics optical phased array.

<span class="mw-page-title-main">Medical ultrasound</span> Diagnostic and therapeutic technique

Medical ultrasound includes diagnostic techniques using ultrasound, as well as therapeutic applications of ultrasound. In diagnosis, it is used to create an image of internal body structures such as tendons, muscles, joints, blood vessels, and internal organs, to measure some characteristics or to generate an informative audible sound. The usage of ultrasound to produce visual images for medicine is called medical ultrasonography or simply sonography, or echography. The practice of examining pregnant women using ultrasound is called obstetric ultrasonography, and was an early development of clinical ultrasonography. The machine used is called an ultrasound machine, a sonograph or an echograph. The visual image formed using this technique is called an ultrasonogram, a sonogram or an echogram.

<span class="mw-page-title-main">Synthetic-aperture radar</span> Form of radar used to create images of landscapes

Synthetic-aperture radar (SAR) is a form of radar that is used to create two-dimensional images or three-dimensional reconstructions of objects, such as landscapes. SAR uses the motion of the radar antenna over a target region to provide finer spatial resolution than conventional stationary beam-scanning radars. SAR is typically mounted on a moving platform, such as an aircraft or spacecraft, and has its origins in an advanced form of side looking airborne radar (SLAR). The distance the SAR device travels over a target during the period when the target scene is illuminated creates the large synthetic antenna aperture. Typically, the larger the aperture, the higher the image resolution will be, regardless of whether the aperture is physical or synthetic – this allows SAR to create high-resolution images with comparatively small physical antennas. For a fixed antenna size and orientation, objects which are further away remain illuminated longer – therefore SAR has the property of creating larger synthetic apertures for more distant objects, which results in a consistent spatial resolution over a range of viewing distances.

<span class="mw-page-title-main">Sensor array</span> Group of sensors used to increase gain or dimensionality over a single sensor

A sensor array is a group of sensors, usually deployed in a certain geometry pattern, used for collecting and processing electromagnetic or acoustic signals. The advantage of using a sensor array over using a single sensor lies in the fact that an array adds new dimensions to the observation, helping to estimate more parameters and improve the estimation performance. For example an array of radio antenna elements used for beamforming can increase antenna gain in the direction of the signal while decreasing the gain in other directions, i.e., increasing signal-to-noise ratio (SNR) by amplifying the signal coherently. Another example of sensor array application is to estimate the direction of arrival of impinging electromagnetic waves. The related processing method is called array signal processing. A third examples includes chemical sensor arrays, which utilize multiple chemical sensors for fingerprint detection in complex mixtures or sensing environments. Application examples of array signal processing include radar/sonar, wireless communications, seismology, machine condition monitoring, astronomical observations fault diagnosis, etc.

Beamforming or spatial filtering is a signal processing technique used in sensor arrays for directional signal transmission or reception. This is achieved by combining elements in an antenna array in such a way that signals at particular angles experience constructive interference while others experience destructive interference. Beamforming can be used at both the transmitting and receiving ends in order to achieve spatial selectivity. The improvement compared with omnidirectional reception/transmission is known as the directivity of the array.

<span class="mw-page-title-main">Synthetic-aperture sonar</span> Form of sonar using post-processing of sonar data

Synthetic-aperture sonar (SAS) is a form of sonar in which sophisticated post-processing of sonar data is used in ways closely analogous to synthetic-aperture radar. Synthetic-aperture sonars combine a number of acoustic pings to form an image with much higher along-track resolution than conventional sonars. The along-track resolution can approach half the length of one sonar element, though is downward limited by 1/4 wavelength. The principle of synthetic-aperture sonar is to move the sonar while illuminating the same spot on the sea floor with several pings. When moving along a straight line, those pings that have the image position within the beamwidth constitute the synthetic array. By coherent reorganization of the data from all the pings, a synthetic-aperture image is produced with improved along-track resolution. In contrast to conventional side-scan sonar, SAS processing provides range-independent along-track resolution. At maximum range the resolution can be magnitudes better than that of side-scan sonars.

<span class="mw-page-title-main">Photoacoustic imaging</span> Imaging using the photoacoustic effect

Photoacoustic imaging or optoacoustic imaging is a biomedical imaging modality based on the photoacoustic effect. Non-ionizing laser pulses are delivered into biological tissues and part of the energy will be absorbed and converted into heat, leading to transient thermoelastic expansion and thus wideband ultrasonic emission. The generated ultrasonic waves are detected by ultrasonic transducers and then analyzed to produce images. It is known that optical absorption is closely associated with physiological properties, such as hemoglobin concentration and oxygen saturation. As a result, the magnitude of the ultrasonic emission, which is proportional to the local energy deposition, reveals physiologically specific optical absorption contrast. 2D or 3D images of the targeted areas can then be formed.

<span class="mw-page-title-main">Focused ultrasound</span> Non-invasive therapeutic technique

High-intensity focused ultrasound (HIFU) is a non-invasive therapeutic technique that uses non-ionizing ultrasonic waves to heat or ablate tissue. HIFU can be used to increase the flow of blood or lymph or to destroy tissue, such as tumors, via thermal and mechanical mechanisms. Given the prevalence and relatively low cost of ultrasound generation mechanisms, The premise of HIFU is that it is expect a non-invasive and low-cost therapy that can at minimum outperform operating room care.

<span class="mw-page-title-main">Phased array ultrasonics</span> Testing method

Phased array ultrasonics (PA) is an advanced method of ultrasonic testing that has applications in medical imaging and industrial nondestructive testing. Common applications are to noninvasively examine the heart or to find flaws in manufactured materials such as welds. Single-element probes, known technically as monolithic probes, emit a beam in a fixed direction. To test or interrogate a large volume of material, a conventional probe must be physically scanned to sweep the beam through the area of interest. In contrast, the beam from a phased array probe can be focused and swept electronically without moving the probe. The beam is controllable because a phased array probe is made up of multiple small elements, each of which can be pulsed individually at a computer-calculated timing. The term phased refers to the timing, and the term array refers to the multiple elements. Phased array ultrasonic testing is based on principles of wave physics, which also have applications in fields such as optics and electromagnetic antennae.

<span class="mw-page-title-main">Doppler echocardiography</span> Medical imaging technique of the heart

Doppler echocardiography is a procedure that uses Doppler ultrasonography to examine the heart. An echocardiogram uses high frequency sound waves to create an image of the heart while the use of Doppler technology allows determination of the speed and direction of blood flow by utilizing the Doppler effect.

D. Jackson Coleman is a professor of clinical ophthalmology at NewYork-Presbyterian Hospital at The Edward S. Harkness Eye Institute of Columbia University. He is the former John Milton McLean Professor of Ophthalmology and chairman emeritus at Weill Cornell Medical Center where he served as chairman from 1979 to 2006. His specialties are retinal diseases and ultrasound, working with patients at Columbia University Medical Center. Coleman is also engaged in research involving ultrasound, which he has pursued throughout his career with colleague Ronald Silverman in the Department of Ophthalmology at the Columbia University Medical Center.

<span class="mw-page-title-main">Ultrasonic transducer</span> Acoustic sensor

Ultrasonic transducers and ultrasonic sensors are devices that generate or sense ultrasound energy. They can be divided into three broad categories: transmitters, receivers and transceivers. Transmitters convert electrical signals into ultrasound, receivers convert ultrasound into electrical signals, and transceivers can both transmit and receive ultrasound.

Acoustic microscopy is microscopy that employs very high or ultra high frequency ultrasound. Acoustic microscopes operate non-destructively and penetrate most solid materials to make visible images of internal features, including defects such as cracks, delaminations and voids.

<span class="mw-page-title-main">Thermoacoustic imaging</span>

Thermoacoustic imaging was originally proposed by Theodore Bowen in 1981 as a strategy for studying the absorption properties of human tissue using virtually any kind of electromagnetic radiation. But Alexander Graham Bell first reported the physical principle upon which thermoacoustic imaging is based a century earlier. He observed that audible sound could be created by illuminating an intermittent beam of sunlight onto a rubber sheet. Shortly after Bowen's work was published, other researchers proposed methodology for thermoacoustic imaging using microwaves. In 1994 researchers used an infrared laser to produce the first thermoacoustic images of near-infrared optical absorption in a tissue-mimicking phantom, albeit in two dimensions (2D). In 1995 other researchers formulated a general reconstruction algorithm by which 2D thermoacoustic images could be computed from their "projections," i.e. thermoacoustic computed tomography (TCT). By 1998 researchers at Indiana University Medical Center extended TCT to 3D and employed pulsed microwaves to produce the first fully three-dimensional (3D) thermoacoustic images of biologic tissue [an excised lamb kidney ]. The following year they created the first fully 3D thermoacoustic images of cancer in the human breast, again using pulsed microwaves. Since that time, thermoacoustic imaging has gained widespread popularity in research institutions worldwide. As of 2008, three companies were developing commercial thermoacoustic imaging systems – Seno Medical, Endra, Inc. and OptoSonics, Inc.

<span class="mw-page-title-main">Schlieren imaging</span> Method to visualize density variations in transparent media

Schlieren imaging is a method to visualize density variations in transparent media.

Ultrasound computer tomography (USCT), sometimes also Ultrasound computed tomography, Ultrasound computerized tomography or just Ultrasound tomography, is a form of medical ultrasound tomography utilizing ultrasound waves as physical phenomenon for imaging. It is mostly in use for soft tissue medical imaging, especially breast imaging.

<span class="mw-page-title-main">High Resolution Wide Swath SAR imaging</span>

High Resolution Wide Swath (HRWS) imaging is an important branch in synthetic aperture radar (SAR) imaging, a remote sensing technique capable of providing high resolution images independent of weather conditions and sunlight illumination. This makes SAR very attractive for the systematic observation of dynamic processes on the Earth's surface, which is useful for environmental monitoring, earth resource mapping and military systems.

<span class="mw-page-title-main">Photoacoustic microscopy</span>

Photoacoustic microscopy is an imaging method based on the photoacoustic effect and is a subset of photoacoustic tomography. Photoacoustic microscopy takes advantage of the local temperature rise that occurs as a result of light absorption in tissue. Using a nanosecond pulsed laser beam, tissues undergo thermoelastic expansion, resulting in the release of a wide-band acoustic wave that can be detected using a high-frequency ultrasound transducer. Since ultrasonic scattering in tissue is weaker than optical scattering, photoacoustic microscopy is capable of achieving high-resolution images at greater depths than conventional microscopy methods. Furthermore, photoacoustic microscopy is especially useful in the field of biomedical imaging due to its scalability. By adjusting the optical and acoustic foci, lateral resolution may be optimized for the desired imaging depth.

<span class="mw-page-title-main">Functional ultrasound imaging</span>

Functional ultrasound imaging (fUS) is a medical ultrasound imaging technique of detecting or measuring changes in neural activities or metabolism, for example, the loci of brain activity, typically through measuring blood flow or hemodynamic changes. The method can be seen as an extension of Doppler imaging.

<span class="mw-page-title-main">Digital antenna array</span>

Digital antenna array(DAA) is a smart antenna with multi channels digital beamforming, usually by using fast Fourier transform (FFT). The development and practical realization of digital antenna arrays theory started in 1962 under the guidance of Vladimir Varyukhin (USSR).

References

  1. 1 2 M. Yang, R. Sampson, S. Wei, T. F. Wenisch, and C. Chakrabarti, "Separable Beamforming for 3D Medical Ultrasound Imaging," IEEE Transactions on Signal Processing, vol. 63, no. 2, pp. 279-290, 2015.
  2. J. A. Jensen, S. I. Nikolov, K. L. Gammelmark, and M. H. Pedersen, "Synthetic Aperture Ultrasound Imaging," Ultrasonics, vol. 44, pp. e5-e15, 2006.
  3. 1 2 3 R. S. Cobbold, Foundations of Biomedical Ultrasound. Oxford University Press on Demand, 2007.
  4. C. B. Burckhardt, P.A. Grandchamp, and H. Hoffmann, "An Experimental 2Mhz Synthetic Aperture Sonar System Intended for Medical Use," IEEE Transactions on Sonics and Ultrasonics, vol. 21, no. 1, pp. 1-6, 1974.
  5. P. Corl and G. Kino, "A Real-time Synthetic Aperture Imaging System," in Acoustical Imaging. Springer, 1980, pp. 341-355.
  6. M. Karaman, P.C. Li, and M. O'Donnell, "Synthetic Aperture Imaging for Small Scale Systems," IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 42, no. 3, pp. 429-442, 1995.
  7. C. Cooley and B. Robinson, "Synthetic Focus Imaging Using Partial Datasets," in IEEE Ultrasonics Symposium., vol. 3, 1994, pp. 1539-1542.
  8. G. Lockwood and F. S. Foster, "Design of Sparse Array Imaging Systems," in IEEE Ultrasonics Symposium, vol. 2, 1995, pp. 1237-1243.
  9. 1 2 S. Nikolov and J. A. Jensen, "Comparison Between Different Encoding Schemes for Synthetic Aperture Imaging," in Medical Imaging. International Society for Optics and Photonics, 2002, pp. 1-12.
  10. C. Passmann and H. Ermert, "A 100Mhz Ultrasound Imaging System for Dermatologic and Ophthalmologic Diagnostics," IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 43, no. 4, pp. 545-552, 1996.
  11. S. Nikolov and J. A. Jensen, "Virtual Ultrasound Sources in High-resolution Ultrasound Imaging," in Medical Imaging, International Society for Optics and Photonics, 2002, pp. 395-405.
  12. C. H. Frazier and W. Brien, "Synthetic Aperture Techniques with A Virtual Source Element," IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 45, no. 1, pp. 196-207, 1998.
  13. J. Kortbek, J. A. Jensen, and K. L. Gammelmark, "Synthetic Aperture Sequential Beamforming," in IEEE Ultrasonics Symposium, 2008, pp. 966-969.
  14. M. C. Hemmsen, J. M. Hansen, and J. A. Jensen, "Synthetic Aperture Sequential Beamformation Applied to Medical Imaging," in 9th European Conference on Synthetic Aperture Radar, 2012, pp. 34-37.
  15. M.-H. Bae and M.-K. Jeong, "Bidirectional Pixel Based Focusing in Conventional B-mode Ultrasound Imaging," Electron Lett, vol. 34, no. 22, pp. 2105-2107, 1998.
  16. C. Kim, C. Yoon, J.-H. Park, Y. Lee, W. H. Kim, J. M. Chang, B. I. Choi, T.-K. Song, and Y.-M. Yoo, "Evaluation of Ultrasound Synthetic Aperture Imaging Using Bidirectional Pixel-based Focusing: Preliminary Phantom and In-vivo Breast Study," IEEE Transactions on Biomedical Engineering, vol. 60, no. 10, pp. 2716-2724, 2013.
  17. K.-S. Kim and T. K. Song, "High Frame Rate and High Resolution Imaging with Synthetic Aperture," in Key Engineering Materials, vol. 270. Trans Tech Publications, 2004, pp. 168-173.
  18. S. Campbell, C. Lees, G. Moscoso, and P. Hall, "Ultrasound Antenatal Diagnosis of Cleft Palate by A New Technique: the 3D Reverse Faceview," Ultrasound in Obstetrics & Gynecology, vol. 25, no. 1, pp. 12-18, 2005.
  19. M. Karaman, I. O. Wygant, O. Oralkan, and B. T. Khuri-Yakub, "Minimally Redundant 2D Array Designs for 3D Medical Ultrasound Imaging," IEEE Transactions on Medical Imaging, vol. 28, no. 7, pp. 1051-1061, 2009.
  20. B. Diarra, H. Liebgott, M. Robini, P. Tortoli, and C. Cachard, "Optimized 2D Array Design for Ultrasound Imaging," in IEEE Proceedings of the 20th European Signal Processing Conference (EUSIPCO), 2012, pp. 2718-2722.