Disappearing-filament pyrometer

Last updated
Technician measuring the temperature of molten silicon at 2650degF with a disappearing-filament pyrometer, in Czochralski crystal growing equipment at Raytheon transistor plant in 1956. The knob she is turning (bottom right) controls current through the filament. Silicon grown by Czochralski process 1956.jpg
Technician measuring the temperature of molten silicon at 2650°F with a disappearing-filament pyrometer, in Czochralski crystal growing equipment at Raytheon transistor plant in 1956. The knob she is turning (bottom right) controls current through the filament.

The disappearing-filament pyrometer is an optical pyrometer, in which the temperature of a glowing incandescent object is measured by comparing it to the light of a heated filament. [1] [2] [3] Invented independently in 1901 by Ludwig Holborn and Ferdinand Kurlbaum in Germany and Harmon Northrup Morse in the United States, [1] it was the first device which could measure temperatures above 1000 °C. [3] Disappearing filament pyrometers have been used to measure temperatures between about 600 °C and 3000 °C. [1] Like other optical pyrometers they are used to measure the temperature of objects too hot for contact thermometers, such as molten metals. Widely used in the steel and ceramics industries as well as for research, they have been almost totally superseded by electronic spectral-band pyrometers. [3]

Temperature physical property of matter that quantitatively expresses the common notions of hot and cold

A temperature expresses hot and cold, as measured with a thermometer. In physics, hotness is a body's ability to impart energy as heat to another body that is colder.

Ferdinand Kurlbaum was a German physicist.

Steel alloy made by combining iron and other elements

Steel is an alloy of iron and carbon, and sometimes other elements. Because of its high tensile strength and low cost, it is a major component used in buildings, infrastructure, tools, ships, automobiles, machines, appliances, and weapons.

The simplest design has optics like a Keplerian telescope. A thin wire (filament), placed at the focal plane of the objective lens, is heated by electric current. [1] [2] [3] When seen through the eyepiece, the wire appears silhouetted in front of the hot luminous object under investigation. The user compares the brightness of the glowing filament with the object behind, and adjusts the current through the filament until it seems to "disappear" in front of the glowing object. [1] [3] At that point the filament and object are at the same temperature. The user then reads the temperature off the filament current control dial, which is calibrated by the filament's current-vs-temperature curve. [3] or in some instruments from a current-vs-temperature table.

The filament seems to "disappear" against the background of the object because two objects at the same temperature have the same black-body spectrum. [2]

In other designs the current through the filament is kept constant, and the radiation allowed through from the target object is varied with calibrated attenuating wedges in the optical path, [3] or a prism is used to place the images of the target object and a calibrated glowing surface next to each other, e.g. as a disk inside a ring.

Many disappearing-filament pyrometers use a red filter. [2] [3] The combination of the filter and the human eye's response only allows through a narrow band of red wavelengths, so the luminosity comparison is made over only a narrow band of wavelengths. [2] This reduces errors due to the target and filament not having identical emission spectra. [2] [3] For very hot objects, additional filters can be used to protect the eye from excessive light. The resolution of the instrument depends somewhat on the operator, but with a skilled operator a resolution of 10 °C for temperatures up to 2000 °C can be achieved. [3]

Disappearing-filament pyrometers can be used only if the object under study emits visible light similar to a hot black body; [3] this means that its temperature must be high enough (around 600 °C and up) and the object must not be fully transparent or highly reflective. For good accuracy, the object should appear dark gray or black when cold. [2]

Black-body radiation thermal electromagnetic radiation

Black-body radiation is the thermal electromagnetic radiation within or surrounding a body in thermodynamic equilibrium with its environment, emitted by a black body. It has a specific spectrum of wavelengths, inversely related to intensity that depend only on the body's temperature, which is assumed for the sake of calculations and theory to be uniform and constant.

See also

Related Research Articles

Apparent magnitude measure of brightness for celestial objects

Apparent magnitude is a measure of the brightness of a star or other astronomical object as seen from the Earth's location. An object's apparent magnitude depends on its intrinsic luminosity, its distance from Earth, and any extinction of the object's light by interstellar dust along the line of sight to the observer.

Pressure measurement Analysis of force applied by a fluid on a surface

Pressure measurement is the analysis of an applied force by a fluid on a surface. Pressure is typically measured in units of force per unit of surface area. Many techniques have been developed for the measurement of pressure and vacuum. Instruments used to measure and display pressure in an integral unit are called pressure meters or pressure gauges or vacuum gauges. A manometer is a good example, as it uses the surface area and weight of a column of liquid to both measure and indicate pressure. Likewise the widely used Bourdon gauge is a mechanical device, which both measures and indicates and is probably the best known type of gauge.

Optical spectrometer

An optical spectrometer is an instrument used to measure properties of light over a specific portion of the electromagnetic spectrum, typically used in spectroscopic analysis to identify materials. The variable measured is most often the light's intensity but could also, for instance, be the polarization state. The independent variable is usually the wavelength of the light or a unit directly proportional to the photon energy, such as reciprocal centimeters or electron volts, which has a reciprocal relationship to wavelength.

Melting point temperature at which a solid turns liquid

The melting point of a substance is the temperature at which it changes state from solid to liquid. At the melting point the solid and liquid phase exist in equilibrium. The melting point of a substance depends on pressure and is usually specified at a standard pressure such as 1 atmosphere or 100 kPa.

Ultraviolet–visible spectroscopy Range of spectroscopic analysis

Ultraviolet–visible spectroscopy or ultraviolet–visible spectrophotometry refers to absorption spectroscopy or reflectance spectroscopy in part of the ultraviolet and the full, adjacent visible spectral regions. This means it uses light in the visible and adjacent ranges. The absorption or reflectance in the visible range directly affects the perceived color of the chemicals involved. In this region of the electromagnetic spectrum, atoms and molecules undergo electronic transitions. Absorption spectroscopy is complementary to fluorescence spectroscopy, in that fluorescence deals with transitions from the excited state to the ground state, while absorption measures transitions from the ground state to the excited state.

Pyrometer remote-sensing thermometer used to measure high temperatures

A pyrometer is a type of remote-sensing thermometer used to measure the temperature of a surface. Various forms of pyrometers have historically existed. In the modern usage, it is a device that from a distance determines the temperature of a surface from the amount of the thermal radiation it emits, a process known as pyrometry and sometimes radiometry.

Photometry (astronomy) astronomy technique of measuring intensity of an astronomical objects emissions

Photometry, from Greek photo- ("light") and -metry ("measure"), is a technique used in astronomy that is concerned with measuring the flux or intensity of light radiated by astronomical objects. This light is measured through a telescope using a photometer, often made using electronic devices such as a CCD photometer or a photoelectric photometer that converts light into an electric current by the photoelectric effect. When calibrated against standard stars of known intensity and colour, photometers can measure the brightness or apparent magnitude of celestial objects.

Thermography infrared imaging

Infrared thermography (IRT), thermal imaging, and thermal video are examples of infrared imaging science. Thermographic cameras usually detect radiation in the long-infrared range of the electromagnetic spectrum and produce images of that radiation, called thermograms. Since infrared radiation is emitted by all objects with a temperature above absolute zero according to the black body radiation law, thermography makes it possible to see one's environment with or without visible illumination. The amount of radiation emitted by an object increases with temperature; therefore, thermography allows one to see variations in temperature. When viewed through a thermal imaging camera, warm objects stand out well against cooler backgrounds; humans and other warm-blooded animals become easily visible against the environment, day or night. As a result, thermography is particularly useful to the military and other users of surveillance cameras.

Observational astronomy part of astronomical science concerned with recording data

Observational astronomy is a division of astronomy that is concerned with recording data about the observable universe, in contrast with theoretical astronomy, which is mainly concerned with calculating the measurable implications of physical models. It is the practice and study of observing celestial objects with the use of telescopes and other astronomical instruments.

Monochromator optical device

A monochromator is an optical device that transmits a mechanically selectable narrow band of wavelengths of light or other radiation chosen from a wider range of wavelengths available at the input. The name is from the Greek roots mono-, "single", and chroma, "colour", and the Latin suffix -ator, denoting an agent.

Photometer scientific instrument

A photometer is an instrument that measures the strength of electromagnetic radiation in the range from ultraviolet to infrared and including the visible spectrum. Most photometers convert light into an electric current using a photoresistor, photodiode, or photomultiplier.

Laser rangefinder range finding device using a laser beam to measure distance

A laser rangefinder is a rangefinder that uses a laser beam to determine the distance to an object. The most common form of laser rangefinder operates on the time of flight principle by sending a laser pulse in a narrow beam towards the object and measuring the time taken by the pulse to be reflected off the target and returned to the sender. Due to the high speed of light, this technique is not appropriate for high precision sub-millimeter measurements, where triangulation and other techniques are often used.

Polarimetry

Polarimetry is the measurement and interpretation of the polarization of transverse waves, most notably electromagnetic waves, such as radio or light waves. Typically polarimetry is done on electromagnetic waves that have traveled through or have been reflected, refracted or diffracted by some material in order to characterize that object.

Surface metrology is the measurement of small-scale features on surfaces, and is a branch of metrology. Surface primary form, surface fractality and surface roughness are the parameters most commonly associated with the field. It is important to many disciplines and is mostly known for the machining of precision parts and assemblies which contain mating surfaces or which must operate with high internal pressures.

Infrared thermometer

An infrared thermometer is a thermometer which infers temperature from a portion of the thermal radiation sometimes called black-body radiation emitted by the object being measured. They are sometimes called laser thermometers as a laser is used to help aim the thermometer, or non-contact thermometers or temperature guns, to describe the device's ability to measure temperature from a distance. By knowing the amount of infrared energy emitted by the object and its emissivity, the object's temperature can often be determined within a certain range of its actual temperature. Infrared thermometers are a subset of devices known as "thermal radiation thermometers".

Orange peel (effect) Kind of finish that may develop on painted and cast surfaces

Orange peel is a certain kind of finish that may develop on painted and cast surfaces, even screen protectors. The texture resembles the surface of the skin of an orange. Gloss paint sprayed on a smooth surface should also dry into a smooth surface. However, various factors can cause it to dry into a bumpy surface resembling the texture of an orange peel. The orange peel phenomenon can then be smoothed out with ultra-fine sandpaper, but it can be prevented altogether by changing the painting technique or the materials used. Orange peel is typically the result of improper painting technique, and is caused by the quick evaporation of thinner, incorrect spray gun setup, spraying the paint at an angle other than perpendicular, or applying excessive paint.

In astronomy, color–color diagrams are a means of comparing the apparent magnitudes of stars at different wavelengths. Astronomers typically observe at narrow bands around certain wavelengths, and objects observed will have different brightnesses in each band. The difference in brightness between two bands is referred to as color. On color–color diagrams, the color defined by two wavelength bands is plotted on the horizontal axis, and then the color defined by another brightness difference will be plotted on the vertical axis.

ISO 25178: Geometric Product Specifications (GPS) – Surface texture: areal is an International Organisation for Standardisation collection of international standards relating to the analysis of 3D areal surface texture.

An optical power meter (OPM) is a device used measure the power in an optical signal. The term usually refers to a device for testing average power in fiber optic systems. Other general purpose light power measuring devices are usually called radiometers, photometers, laser power meters, light meters or lux meters.

Measuring instrument device for measuring a physical quantity

A measuring instrument is a device for measuring a physical quantity. In the physical sciences, quality assurance, and engineering, measurement is the activity of obtaining and comparing physical quantities of real-world objects and events. Established standard objects and events are used as units, and the process of measurement gives a number relating the item under study and the referenced unit of measurement. Measuring instruments, and formal test methods which define the instrument's use, are the means by which these relations of numbers are obtained. All measuring instruments are subject to varying degrees of instrument error and measurement uncertainty.

References

  1. 1 2 3 4 5 Bud, Robert; Warner, Deborah Jean (1998). Instruments of Science: An Historical Encyclopedia. Taylor and Francis. p. 499. ISBN   0815315619.
  2. 1 2 3 4 5 6 7 Speyer, Robert (1993). Thermal Analysis of Materials. CRC Press. pp. 211–216. ISBN   0824789636.
  3. 1 2 3 4 5 6 7 8 9 10 11 Bentley, Robin E., Ed. (1998). Handbook of Temperature Measurement, Volume 1: Temperature and Humidity Measurement. Springer Science and Business Media. p. 75. ISBN   9814021091.