Karl Fischer titration

Last updated
A Karl Fischer titrator KF Titrator.jpg
A Karl Fischer titrator

In analytical chemistry, Karl Fischer titration is a classic titration method that uses coulometric or volumetric titration to determine trace amounts of water in a sample. It was invented in 1935 by the German chemist Karl Fischer. [1] [2] Today, the titration is done with an automated Karl Fischer titrator.

Contents

Chemical principle

The elementary reaction responsible for water quantification in the Karl Fischer titration is oxidation of sulfur dioxide (SO2) with iodine:

H2O + SO2 + I2 → SO3 + 2 HI

This elementary reaction consumes exactly two molar equivalent of water vs. iodine. Iodine is added to the solution until it is present in excess, marking the end point of the titration, which can be detected by potentiometry. The reaction is run in an alcohol solution containing a base, which consumes the sulfur trioxide and hydroiodic acid produced.

Coulometric titration

The main compartment of the titration cell contains the anode solution plus the analyte. The anode solution consists of an alcohol (ROH), a base (B), sulfur dioxide (SO2) and KI. Typical alcohols that may be used include ethanol, diethylene glycol monoethyl ether, or methanol, sometimes referred to as Karl Fischer grade. A common base is imidazole.

The titration cell also consists of a smaller compartment with a cathode immersed in the anode solution of the main compartment. The two compartments are separated by an ion-permeable membrane.

The Pt anode generates I2 from the KI when current is provided through the electric circuit. The net reaction as shown below is oxidation of SO2 by I2. One mole of I2 is consumed for each mole of H2O. In other words, 2 moles of electrons are consumed per mole of water.

2 I → I2 + 2 e
B·I2 + B·SO2 + B + H2O → 2 BH+I + BSO3
BSO3 + ROH → BHRSO4

The end point is detected most commonly by a bipotentiometric titration method. A second pair of Pt electrodes is immersed in the anode solution. The detector circuit maintains a constant current between the two detector electrodes during titration. Prior to the equivalence point, the solution contains I but little I2. At the equivalence point, excess I2 appears and an abrupt voltage drop marks the end point. The amount of charge needed to generate I2 and reach the end point can then be used to calculate the amount of water in the original sample.

Volumetric titration

The volumetric titration is based on the same principles as the coulometric titration, except that the anode solution above now is used as the titrant solution. The titrant consists of an alcohol (ROH), base (B), SO2 and a known concentration of I2. Pyridine has been used as the base in this case.

One mole of I2 is consumed for each mole of H2O. The titration reaction proceeds as above, and the end point may be detected by a bipotentiometric method as described above.

Disadvantages and advantages

The popularity of the Karl Fischer titration (henceforth referred to as KF) is due in large part to several practical advantages that it holds over other methods of moisture determination, such as accuracy, speed and selectivity.

KF is selective for water, because the titration reaction itself consumes water. In contrast, measurement of mass loss on drying will detect the loss of any volatile substance. However, the strong redox chemistry (SO2/I2) means that redox-active sample constituents may react with the reagents. For this reason, KF is unsuitable for solutions containing e.g. dimethyl sulfoxide.

KF has a high accuracy and precision, typically within 1% of available water, e.g. 3.00% appears as 2.97–3.03%. Although KF is a destructive analysis, the sample quantity is small and is typically limited by the accuracy of weighing. For example, in order to obtain an accuracy of 1% using a scale with the typical accuracy of 0.2 mg, the sample must contain 20 mg water, which is e.g. 200 mg for a sample with 10% water. For coulometers, the measuring range is from 1–5 ppm to about 5%. Volumetric KF readily measures samples up to 100%, but requires impractically large amounts of sample for analytes with less than 0.05% water. [3] The KF response is linear. Therefore, single-point calibration using a calibrated 1% water standard is sufficient and no calibration curves are necessary.

Little sample preparation is needed: a liquid sample can usually be directly injected using a syringe. The analysis is typically complete within a minute. However, KF suffers from an error called drift, which is an apparent water input that can confuse the measurement. The glass walls of the vessel adsorb water, and if any water leaks into the cell, the slow release of water into the titration solution can continue for a long time. Therefore, before measurement, it is necessary to carefully dry the vessel and run a 10–30-minute "dry run" in order to calculate the rate of drift. The drift is then subtracted from the result.

KF is suitable for measuring liquids and, with special equipment, gases. The major disadvantage with solids is that the water has to be accessible and easily brought into methanol solution. Many common substances, especially foods such as chocolate, release water slowly and with difficulty, requiring additional efforts to reliably bring the total water content into contact with the Karl Fischer reagents. For example, a high-shear mixer may be installed to the cell in order to break the sample. KF has problems with compounds with strong binding to water, as in water of hydration, for example with lithium chloride, so KF is unsuitable for the special solvent LiCl/DMAc.

KF is suitable for automation. Generally, KF is conducted using a separate KF titrator, or for volumetric titration, a KF titration cell installed into a general-purpose titrator. There are also oven attachments that can be used for materials that have problems being analyzed normally in the cell. The important aspect about the oven attachment is that the material doesn't decompose into water when heated to release the water. The oven attachment also supports automation of samples.

Using volumetric titration with visual detection of a titration endpoint is also possible with coloured samples by UV/VIS spectrophotometric detection. [4]

See also

Literature

Related Research Articles

<span class="mw-page-title-main">Electrochemistry</span> Branch of chemistry

Electrochemistry is the branch of physical chemistry concerned with the relationship between electrical potential difference and identifiable chemical change. These reactions involve electrons moving via an electronically conducting phase between electrodes separated by an ionically conducting and electronically insulating electrolyte.

<span class="mw-page-title-main">Sulfuric acid</span> Chemical compound (H₂SO₄)

Sulfuric acid or sulphuric acid, known in antiquity as oil of vitriol, is a mineral acid composed of the elements sulfur, oxygen, and hydrogen, with the molecular formula H2SO4. It is a colorless, odorless, and viscous liquid that is soluble with water.

<span class="mw-page-title-main">Titration</span> Laboratory method for determining the concentration of an analyte

Titration is a common laboratory method of quantitative chemical analysis to determine the concentration of an identified analyte. A reagent, termed the titrant or titrator, is prepared as a standard solution of known concentration and volume. The titrant reacts with a solution of analyte to determine the analyte's concentration. The volume of titrant that reacted with the analyte is termed the titration volume.

The Winkler test is used to determine the concentration of dissolved oxygen in water samples. Dissolved oxygen (D.O.) is widely used in water quality studies and routine operation of water reclamation facilities to analyze its level of oxygen saturation.

<span class="mw-page-title-main">Nitrous acid</span> Chemical compound

Nitrous acid is a weak and monoprotic acid known only in solution, in the gas phase, and in the form of nitrite salts. It was discovered by Carl Wilhelm Scheele, who called it "phlogisticated acid of niter". Nitrous acid is used to make diazonium salts from amines. The resulting diazonium salts are reagents in azo coupling reactions to give azo dyes.

A redox titration is a type of titration based on a redox reaction between the analyte and titrant. It may involve the use of a redox indicator and/or a potentiometer. A common example of a redox titration is the treatment of a solution of iodine with a reducing agent to produce iodide using a starch indicator to help detect the endpoint. Iodine (I2) can be reduced to iodide (I) by, say, thiosulfate (S2O2−3), and when all the iodine is consumed, the blue colour disappears. This is called an iodometric titration.

<span class="mw-page-title-main">Acid–base titration</span> Method of chemical quantitative analysis

An acid–base titration is a method of quantitative analysis for determining the concentration of Brønsted-Lowry acid or base (titrate) by neutralizing it using a solution of known concentration (titrant). A pH indicator is used to monitor the progress of the acid–base reaction and a titration curve can be constructed.

In analytical electrochemistry, coulometry determines the amount of matter transformed during an electrolysis reaction by measuring the amount of electricity consumed or produced. It can be used for precision measurements of charge, and the amperes even used to have a coulometric definition. However, today coulometry is mainly used for analytical applications. It is named after Charles-Augustin de Coulomb.

In chemistry, acid value is a number used to quantify the acidity of a given chemical substance. It is the quantity of base, expressed as milligrams of KOH required to neutralize the acidic constituents in 1 gram of a sample. The acid value measures the acidity of water-insoluble substances like oils, fats, waxes and resins, which do not have a pH value.

<span class="mw-page-title-main">Hydrogen iodide</span> Chemical compound

Hydrogen iodide (HI) is a diatomic molecule and hydrogen halide. Aqueous solutions of HI are known as hydroiodic acid or hydriodic acid, a strong acid. Hydrogen iodide and hydroiodic acid are, however, different in that the former is a gas under standard conditions, whereas the other is an aqueous solution of the gas. They are interconvertible. HI is used in organic and inorganic synthesis as one of the primary sources of iodine and as a reducing agent.

Iodometry, known as iodometric titration, is a method of volumetric chemical analysis, a redox titration where the appearance or disappearance of elementary iodine indicates the end point.

In analytical chemistry, a standard solution is a solution containing an accurately known concentration. Standard solutions are generally prepared by dissolving a solute of known mass into a solvent to a precise volume, or by diluting a solution of known concentration with more solvent.

<span class="mw-page-title-main">Zinc iodide</span> Chemical compound

Zinc iodide is the inorganic compound with the formula ZnI2. It exists both in anhydrous form and as a dihydrate. Both are white and readily absorb water from the atmosphere. It has no major application.

<span class="mw-page-title-main">Iodine clock reaction</span> Experiment to show chemical kinetics in action

The iodine clock reaction is a classical chemical clock demonstration experiment to display chemical kinetics in action; it was discovered by Hans Heinrich Landolt in 1886. The iodine clock reaction exists in several variations, which each involve iodine species and redox reagents in the presence of starch. Two colourless solutions are mixed and at first there is no visible reaction. After a short time delay, the liquid suddenly turns to a shade of dark blue due to the formation of a triiodide–starch complex. In some variations, the solution will repeatedly cycle from colorless to blue and back to colorless, until the reagents are depleted.

<span class="mw-page-title-main">Thermometric titration</span>

A thermometric titration is one of a number of instrumental titration techniques where endpoints can be located accurately and precisely without a subjective interpretation on the part of the analyst as to their location. Enthalpy change is arguably the most fundamental and universal property of chemical reactions, so the observation of temperature change is a natural choice in monitoring their progress. It is not a new technique, with possibly the first recognizable thermometric titration method reported early in the 20th century. In spite of its attractive features, and in spite of the considerable research that has been conducted in the field and a large body of applications that have been developed; it has been until now an under-utilized technique in the critical area of industrial process and quality control. Automated potentiometric titration systems have pre-dominated in this area since the 1970s. With the advent of cheap computers able to handle the powerful thermometric titration software, development has now reached the stage where easy to use automated thermometric titration systems can in many cases offer a superior alternative to potentiometric titrimetry.

The Bunsen reaction is a chemical reaction that describes water, sulfur dioxide, and iodine reacting to form sulfuric acid and hydrogen iodide:

Base Number (BN) is a measurement of basicity that is expressed in terms of the number of milligrams of potassium hydroxide per gram of oil sample. BN is an important measurement in petroleum products, and the value varies depending on its application. BN generally ranges from 6–8 mg KOH/g in modern lubricants, 7–10 mg KOH/g for general internal combustion engine use and 10–15 mg KOH/g for diesel engine operations. BN is typically higher for marine grade lubricants, approximately 15-80 mg KOH/g, as the higher BN values are designed to increase the operating period under harsh operating conditions, before the lubricant requires replacement.

A chloridometer is a measuring instrument used to determine the concentration of chloride ions (Cl) in a solution. It uses a process known as coulometric titration or amperostatic coulometry, the accepted electrochemistry reference method to determine the concentration of chloride in biological fluids, including blood serum, blood plasma, urine, sweat, and cerebrospinal fluid. The coulometry process generates silver ions, which react with the chloride to form silver chloride (AgCl).

<span class="mw-page-title-main">Calconcarboxylic acid</span> Chemical compound

Calconcarboxylic acid is an azo dye that is used as an indicator for complexometric titrations of calcium with ethylenediaminetetraacetic acid (EDTA) in the presence of magnesium. Structurally, it is similar to eriochrome blue black R, which is obtained from calconcarboxylic acid by decarboxylation and reaction with sodium hydroxide.

The Ripper Method, developed in 1898, is an analytical chemistry technique used to determine the total amount of sulfur dioxide (SO2) in a solution. This technique uses iodine standard and a starch indicator to titrate the solution and determine the concentration of free SO2. The titration is done again with a new sample of the solution, but the sample is pretreated with sodium hydroxide (NaOH) to release bound SO2. The result of these two titrations can then be used to determine the bound, free, and total amount of SO2 in the solution. Instead of using a starch indicator, an electrode can be used to determine the presence of free iodine. This technique is widely used in wine making.

References

  1. Fischer, Karl (1935). "Neues Verfahren zur maßanalytischen Bestimmung des Wassergehaltes von Flüssigkeiten und festen Körpern". Angew. Chem. 48 (26): 394–396. doi:10.1002/ange.19350482605.
  2. Sella, Andrea. "Karl Fischer's Titrator". Chemistry World. Royal Society of Chemistry. Retrieved 3 January 2023.
  3. "ASTM E203 – 16 Standard Test Method for Water Using Volumetric Karl Fischer Titration". www.astm.org.
  4. Tavčar, E., Turk, E., Kreft, S. (2012). Simple Modification of Karl-Fischer Titration Method for Determination of Water Content in Colored Samples. Journal of Analytical Methods in Chemistry, Vol. 2012, Article ID 379724.