Stationary-wave integrated Fourier-transform spectrometry

Last updated

Stationary-wave integrated Fourier-transform spectrometry (SWIFTS), or standing-wave integrated Fourier-transform spectrometry, is an analytical technique used for measuring the distribution of light across an optical spectrum. SWIFTS technology is based on a near-field Lippmann architecture. An optical signal is injected into a waveguide and ended by a mirror (true Lippman configuration). The input signal interferes with the reflected signal, creating a standing, or stationary, wave.

Contents

In a counter-propagative architecture, the two optical signals are injected at the opposite ends of the waveguide. The evanescent waves propagating within the waveguide are then sampled by optical probes. This results in an interferogram. A mathematical function known as a Lippmann transform, similar to a Fourier transform, is later used to give the spectrum of the light.

History

In 1891, at the Académie des Sciences in Paris, Gabriel Lippmann presented a colour photograph of the Sun's spectrum obtained with his new photographic plate. [1] Later, in 1894, he published an article on how his plate was able to record colour information in the depth of photographic grainless gelatin and how the same plate after processing could restore the original colour image merely through light reflection. [2] He was thus the inventor of true interferential colour photography. He received the Nobel Prize in Physics in 1908 for this breakthrough. Unfortunately, this principle was too complex to use. The method was abandoned a few years after its discovery.

One aspect of the Lippmann concept that was ignored at that time relates to spectroscopic applications. Early in 1933, Herbert E. Ives proposed to use a photoelectric device to probe stationary waves to make spectrometric measurements. [3] In 1995, P. Connes [4] proposed to use the emerging new technology of detectors for three-dimensional Lippmann-based spectrometry. Following this, a first realization of a very compact spectrometer based on a microoptoelectromechanical system (MOEMS) was reported by Knipp et al. in 2005, [5] but it had a very limited spectral resolution. In 2004, two French researchers, Etienne Le Coarer from Joseph Fourier University and Pierre Benech from INP Grenoble, coupled sensing elements to the evanescent part of standing waves within a single-mode waveguide. In 2007, those two researchers reported a near-field method to probe the interferogram within a waveguide. [6] The first SWIFTS-based spectrometers appeared in 2011 based on a SWIFTS linear configuration.

Technology principle

The technology works by probing an optical standing wave, or the sum of the standing waves in the case of polychromatic light, created by a light to be analyzed. In a SWIFTS linear configuration (true Lippman configuration), the stationary wave is created by a single-mode waveguide ended by a fixed mirror. The stationary wave is regularly sampled on one side of a waveguide using nano-scattering dots. These dots are located in the evanescent field. These nanodots are characterized by an optical index difference with the medium in which the evanescent field is located. The light is then scattered around an axis perpendicular to the waveguide. For each dot, this scattered light is detected by a pixel aligned with this axis. The intensity detected is therefore proportional to the intensity inside the waveguide at the exact location of the dot. This results in a linear image of the interferogram. No moving parts are used. A mathematical function known as a Lippmann transform, similar to a Fourier transform, is then applied to this linear image and gives the spectrum of the light.

The interferogram is truncated. Only the frequencies corresponding to the zero optical path difference at the mirror, up to the farthest dots are sampled. Higher frequencies are rejected. This interferogram’s truncation determines the spectral resolution. The interferogram is undersampled. A consequence of this under-sampling is a limitation of the wavelength bandwidth to which the mathematical function is applied.

SWIFTS technology displays the Fellgett's advantage, which is derived from the fact that an interferometer measures wavelengths simultaneously with the same elements of the detector, whereas a dispersive spectrometer measures them successively. Fellgett's advantage also states that when collecting a spectrum whose measurement noise is dominated by detector noise, a multiplex spectrometer such as a Fourier-transform spectrometer will produce a relative improvement in the signal-to-noise ratio, with respect to an equivalent scanning monochromator, that is approximately equal to the square root of the number of sample points comprising the spectrum. The Connes advantage states that the wavenumber scale of an interferometer, derived from a helium–neon laser, is more accurate and boasts better long-term stability than the calibration of dispersive instruments.

Related Research Articles

<span class="mw-page-title-main">Spectroscopy</span> Study involving matter and electromagnetic radiation

Spectroscopy is the field of study that measures and interprets the electromagnetic spectra that result from the interaction between electromagnetic radiation and matter as a function of the wavelength or frequency of the radiation. Matter waves and acoustic waves can also be considered forms of radiative energy, and recently gravitational waves have been associated with a spectral signature in the context of the Laser Interferometer Gravitational-Wave Observatory (LIGO).

Fourier-transform spectroscopy is a measurement technique whereby spectra are collected based on measurements of the coherence of a radiative source, using time-domain or space-domain measurements of the radiation, electromagnetic or not. It can be applied to a variety of types of spectroscopy including optical spectroscopy, infrared spectroscopy, nuclear magnetic resonance (NMR) and magnetic resonance spectroscopic imaging (MRSI), mass spectrometry and electron spin resonance spectroscopy.

<span class="mw-page-title-main">Interferometry</span> Measurement method using interference of waves

Interferometry is a technique which uses the interference of superimposed waves to extract information. Interferometry typically uses electromagnetic waves and is an important investigative technique in the fields of astronomy, fiber optics, engineering metrology, optical metrology, oceanography, seismology, spectroscopy, quantum mechanics, nuclear and particle physics, plasma physics, biomolecular interactions, surface profiling, microfluidics, mechanical stress/strain measurement, velocimetry, optometry, and making holograms.

Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a mass spectrum, a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures.

This is an index of articles relating to electronics and electricity or natural electricity and things that run on electricity and things that use or conduct electricity.

<span class="mw-page-title-main">Michelson interferometer</span> Common configuration for optical interferometry

The Michelson interferometer is a common configuration for optical interferometry and was invented by the 19/20th-century American physicist Albert Abraham Michelson. Using a beam splitter, a light source is split into two arms. Each of those light beams is reflected back toward the beamsplitter which then combines their amplitudes using the superposition principle. The resulting interference pattern that is not directed back toward the source is typically directed to some type of photoelectric detector or camera. For different applications of the interferometer, the two light paths can be with different lengths or incorporate optical elements or even materials under test.

<span class="mw-page-title-main">X-ray spectroscopy</span> Technique to characterize materials using X-ray radiation

X-ray spectroscopy is a general term for several spectroscopic techniques for characterization of materials by using x-ray radiation.

Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or mass spectrometer) for determining the mass-to-charge ratio (m/z) of ions based on the cyclotron frequency of the ions in a fixed magnetic field. The ions are trapped in a Penning trap (a magnetic field with electric trapping plates), where they are excited (at their resonant cyclotron frequencies) to a larger cyclotron radius by an oscillating electric field orthogonal to the magnetic field. After the excitation field is removed, the ions are rotating at their cyclotron frequency in phase (as a "packet" of ions). These ions induce a charge (detected as an image current) on a pair of electrodes as the packets of ions pass close to them. The resulting signal is called a free induction decay (FID), transient or interferogram that consists of a superposition of sine waves. The useful signal is extracted from this data by performing a Fourier transform to give a mass spectrum.

<span class="mw-page-title-main">Near-field scanning optical microscope</span>

Near-field scanning optical microscopy (NSOM) or scanning near-field optical microscopy (SNOM) is a microscopy technique for nanostructure investigation that breaks the far field resolution limit by exploiting the properties of evanescent waves. In SNOM, the excitation laser light is focused through an aperture with a diameter smaller than the excitation wavelength, resulting in an evanescent field on the far side of the aperture. When the sample is scanned at a small distance below the aperture, the optical resolution of transmitted or reflected light is limited only by the diameter of the aperture. In particular, lateral resolution of 6 nm and vertical resolution of 2–5 nm have been demonstrated.

<span class="mw-page-title-main">Attenuated total reflectance</span> Infrared spectroscopy sampling technique

Attenuated total reflection (ATR) is a sampling technique used in conjunction with infrared spectroscopy which enables samples to be examined directly in the solid or liquid state without further preparation.

<span class="mw-page-title-main">Nondispersive infrared sensor</span> Infrared detector without diffracting element dispersing the IR wave lengths

A nondispersive infrared sensor is a simple spectroscopic sensor often used as a gas detector. It is non-dispersive in the fact that no dispersive element is used to separate out the broadband light into a narrow spectrum suitable for gas sensing. The majority of NDIR sensors use a broadband lamp source and an optical filter to select a narrow band spectral region that overlaps with the absorption region of the gas of interest. In this context narrow may be 50-300nm bandwidth. Modern NDIR sensors may use Microelectromechanical systems (MEMs) or mid IR LED sources, with or without an optical filter.

Fellgett's advantage or the multiplex advantage is an improvement in signal-to-noise ratio (SNR) that is gained when taking multiplexed measurements rather than direct measurements. The name is derived from P. B. Fellgett, who first made the observation as part of his PhD. When measuring a signal whose noise is dominated by detector noise, a multiplexed measurement such as the signal generated by a Fourier transform spectrometer can produce a relative improvement in SNR, compared to an equivalent scanning monochromator, of the order of the square root of m, where m is the number of sample points comprising the spectrum.

Optical heterodyne detection is a method of extracting information encoded as modulation of the phase, frequency or both of electromagnetic radiation in the wavelength band of visible or infrared light. The light signal is compared with standard or reference light from a "local oscillator" (LO) that would have a fixed offset in frequency and phase from the signal if the latter carried null information. "Heterodyne" signifies more than one frequency, in contrast to the single frequency employed in homodyne detection.

<span class="mw-page-title-main">Fourier-transform infrared spectroscopy</span> Technique to analyze the infrared spectrum of matter

Fourier-transform infrared spectroscopy (FTIR) is a technique used to obtain an infrared spectrum of absorption or emission of a solid, liquid, or gas. An FTIR spectrometer simultaneously collects high-resolution spectral data over a wide spectral range. This confers a significant advantage over a dispersive spectrometer, which measures intensity over a narrow range of wavelengths at a time.

Angle-resolved low-coherence interferometry (a/LCI) is an emerging biomedical imaging technology which uses the properties of scattered light to measure the average size of cell structures, including cell nuclei. The technology shows promise as a clinical tool for in situ detection of dysplastic, or precancerous tissue.

<span class="mw-page-title-main">White light scanner</span>

A white light scanner (WLS) is a device for performing surface height measurements of an object using coherence scanning interferometry (CSI) with spectrally-broadband, "white light" illumination. Different configurations of scanning interferometer may be used to measure macroscopic objects with surface profiles measuring in the centimeter range, to microscopic objects with surface profiles measuring in the micrometer range. For large-scale non-interferometric measurement systems, see structured-light 3D scanner.

Incoherent broad band cavity enhanced absorption spectroscopy (IBBCEAS), sometimes called broadband cavity enhanced extinction spectroscopy (IBBCEES), measures the transmission of light intensity through a stable optical cavity consisting of high reflectance mirrors (typically R>99.9%). The technique is realized using incoherent sources of radiation e.g. Xenon arc lamps, LEDs or supercontinuum (SC) lasers, hence the name.

The operation of a photon scanning tunneling microscope (PSTM) is analogous to the operation of an electron scanning tunneling microscope, with the primary distinction being that PSTM involves tunneling of photons instead of electrons from the sample surface to the probe tip. A beam of light is focused on a prism at an angle greater than the critical angle of the refractive medium in order to induce total internal reflection within the prism. Although the beam of light is not propagated through the surface of the refractive prism under total internal reflection, an evanescent field of light is still present at the surface.

<span class="mw-page-title-main">Spectrum (physical sciences)</span> Concept relating to waves and signals

In the physical sciences, the term spectrum was introduced first into optics by Isaac Newton in the 17th century, referring to the range of colors observed when white light was dispersed through a prism. Soon the term referred to a plot of light intensity or power as a function of frequency or wavelength, also known as a spectral density plot.

<span class="mw-page-title-main">Anisotropic terahertz microspectroscopy</span> Spectroscopic technique

Anisotropic terahertz microspectroscopy (ATM) is a spectroscopic technique in which molecular vibrations in an anisotropic material are probed with short pulses of terahertz radiation whose electric field is linearly polarized parallel to the surface of the material. The technique has been demonstrated in studies involving single crystal sucrose, fructose, oxalic acid, and molecular protein crystals in which the spatial orientation of molecular vibrations are of interest.

References

  1. G. Lippmann: Compte Rendus de l'Académie des Sciences, Paris, 112 (1891), 274
  2. G. Lippmann: Compte Rendus de l'Académie des Sciences, Paris (1894), 92
  3. Herbert E. Ives, Standing light waves, repetition of an experiment by Wiener, using a photoelectric probe surface, JOSA, 1933, 23, pp. 73–83 doi : 10.1364/JOSA.23.000073
  4. P. Connes, E. le Coarer, 3-D spectroscopy: The historical and logical viewpoint. IAU Colloquium, vol. 149, pp. 38–49, Marseille, 22–25 Mars 1994
  5. D. Knipp, Spectrometers shrink down, Nature Photonics, 2007, pp. 1, 8, 444 and 445
  6. E. le Coarer, S. Blaize, P. Benech, I. Stefanon, A. Morand, G. Lérondel, G. Leblond, P. Kern, J.-M. Fedeli, P. Royer, Wavelength-scale stationary-wave integrated Fourier-transform spectrometry, Nature Photonics (2007), 1, 8, 473–478