Speckle tracking echocardiography

Last updated
Speckle tracking echocardiography
Postoperative circumferential LV strain rate 2-D speckle tracking
Purposeanalyzes the motion of tissues in the heart

In the fields of cardiology and medical imaging, speckle tracking echocardiography (STE) is an echocardiographic imaging technique. It analyzes the motion of tissues in the heart by using the naturally occurring speckle pattern in the myocardium (or motion of blood when imaged by ultrasound).

Contents

This method of documentation of myocardial motion is a noninvasive method of definition for both vectors and velocity. When compared to other technologies seeking noninvasive definition of ischemia, speckle tracking seems a valuable endeavor. The speckle pattern is a mixture of interference patterns and natural acoustic reflections. [1] These reflections are also described as speckles or markers.

The pattern being random, each region of the myocardium has a unique speckle pattern (also called patterns, features, or fingerprints) that allows the region to be tracked. The speckle pattern is relatively stable, at least from one frame to the next. [2] [3] In post processing this can be tracked consecutively frame to frame and ultimately resolved into angle-independent two-dimensional (2D) and three-dimensional strain-based sequences (3D). [3] [4] [5] These sequences provide both quantitative and qualitative information regarding tissue deformation and motion.

Basic principles

As the speckle pattern is random, any region of the myocardium has a unique speckle pattern: Within the picture, a defined area "kernel" can be defined, and as this speckle pattern is relatively stable, the kernel can be recognised in the next frame, within a larger search area, by a "best match" search algorithm. There are different search algorithms, the most commonly used is "sum of absolute differences", [3] shown to be similarly accurate as cross-correlation, which is an alternative. [6] [7] The movement of the kernel across the image can thus be tracked, in principle independent of the beam angle, as opposed to tissue Doppler. Speckle tracking can thus track in two dimensions. However, as the axial (in the direction of the beam) resolution of the ultrasound is far better than the transverse, the tracking ability is less in the transverse direction. Also, the transverse resolution (and hence, tracking ability) decreases with depth, in a sector scan were ultrasound beams diverge.

Different commercial and non commercial operators then use different approaches to derive motion and deformation parameters. The motion of a single kernel can be resolved into displacement curves, and the distance between two kernels into strain (deformation). [8] [9] Strain rate will then be time derivative of strain. In some commercial applications, the acoustic markers are tracked more individually, calculating the velocity from the motion and the sampling interval (inverse of frame rate) generating a velocity field. [4] Unlike tissue Doppler, this velocity field in not limited to the beam direction. Strain rate and strain are then calculated from the velocities. Speckle tracking has been shown to be comparable to tissue Doppler derived strain, [10] and has been validated against MR. [9] [11] [12]

Strain

Strain is defined as the fractional or percentage change in an objects dimension in comparison to the object’s original dimension. [13] Similarly, strain rate can be defined as the speed at which deformation occurs. Mathematically, three components of normal strain (εx, εy, and εz) and three components of shear strain (εxy, εxz, and εyz) are recognized. Congruently, when applied to the left ventricle, left ventricular deformation is defined by the three normal strains (longitudinal, circumferential, and radial) and three shear strains (circumferential-longitudinal, circumferential-radial, and longitudinal-radial). The principal benefit of LV shear strains is amplification of the 15% shortening of myocytes into 40% radial LV wall thickening, which ultimately translates into a >60% change in LV ejection fraction. Left ventricular shearing increases towards the subendocardium, resulting in a subepicardial to subendocardial thickening strain gradient. Similar to MRI, STE utilizes "Lagrangian strain" which defines motion around a particular point in tissue as it revolves through time and space. [14] Throughout the cardiac cycle, the end-diastolic tissue dimension represents the unstressed initial material length. Speckle tracking is one of two methods for Strain rate imaging, the other being Tissue Doppler.

Twist or torsional deformation define the base-to-apex gradient and is the result of myocardial shearing in the circumferential-longitudinal planes such that, when viewed from the apex, the base rotates in a counterclockwise direction. Likewise the LV apex concomitantly rotates in a clockwise direction. During ejection, LV torsion results in the storage of potential energy into the deformed myofibers. This stored energy is released with the onset of relaxation similar to a spring uncoiling and results in suction forces. These forces are then used for rapid early diastolic restoration.

Applications and limitations

The utilities of STE are increasingly recognized. Strain results derived from STE have been validated using sonomicrometry and tagged MRI and results correlate significantly with Tissue Doppler–derived measurements. [15] [16] [17] Tissue Doppler technology, the alternative method for strain rate imaging to speckle tracking technology, requires achieving sufficient parallel orientation between the direction of motion and the ultrasound beam. Its use has remained limited due to angle dependency, substantial intraobserver and interobserver variability and noise interference. Speckle tracking technology has to a certain degree overcome these limitations.

In order to achieve sufficient tracking quality when single markers are used, however commercial algorithms very often resort to varieties of spline smoothing using available information from the strongest echoes, very oft the mitral annulus, so the regional measurements are not pure regional, but rather to a degree, spline functions of the global average. AS the method uses B-mode, frame rate of speckle tracking is limited to the relatively low frame rate of B-mode. If the frame rate is too low, the tracking quality becomes reduced, due to frame-to-frame decorrelation. This may also be a problem if the heart rate is high, (which in fact is a relative decrease in frame rate - fewer frames per heart cycle).

Increasing frame rate in B-mode is done by reducing line density, i.e. lateral resolution, and thus making the method more angle dependent. Finally, the method on some applications is dependent on the ROI (Region Of Interest) size and shape. In principle Speckle tracking is available for deformation measurement in all directions, however, due to the limitation of lateral resolution in apical images, measuring circumferential and transmural deformation needs parasternal cross sectional views. [11] On the other hand, compared to Tissue Doppler, that method is mainly only available for longitudinal measures from the apical position. [11]

In the study by Cho et al, [11] both TVI derived and speckle tracking derived longitudinal strain showed modest correlation with MRI derived strain. The ROC analysis showed significantly higher AUC for speckle tracking for detecting dysfunctional segments. However, this study only included patients with coronary disease. The lower frame rate has been seen to be a problem in stress echo, as the peak stress shows a fairly high frame rate. [18]

The main problem with speckle tracking, however, is increasingly recognised: The lack of standardisation. Each vendor of ultrasound equipment, or analysis software, has different algorithms, that will perform differently during analysis. In head to head comparisons, biases between analysis may be substantial, especially when compared to an external reference. [19] Thus, measurements, normal limits and cut off values are only vendor specific. Due to industrial secrecy, the details of the different algorithms may also be largely unavailable, so a detailed investigation in modelling is difficult.

Clinical Applications of Speckle Tracking Technology:

See also

Related Research Articles

<span class="mw-page-title-main">Medical ultrasound</span> Diagnostic and therapeutic technique

Medical ultrasound includes diagnostic techniques using ultrasound, as well as therapeutic applications of ultrasound. In diagnosis, it is used to create an image of internal body structures such as tendons, muscles, joints, blood vessels, and internal organs, to measure some characteristics or to generate an informative audible sound. The usage of ultrasound to produce visual images for medicine is called medical ultrasonography or simply sonography, or echography. The practice of examining pregnant women using ultrasound is called obstetric ultrasonography, and was an early development of clinical ultrasonography. The machine used is called an ultrasound machine, a sonograph or an echograph. The visual image formed using this technique is called an ultrasonogram, a sonogram or an echogram.

<span class="mw-page-title-main">Cardiac output</span> Measurement of blood pumped by the heart

In cardiac physiology, cardiac output (CO), also known as heart output and often denoted by the symbols , , or , is the volumetric flow rate of the heart's pumping output: that is, the volume of blood being pumped by a single ventricle of the heart, per unit time. Cardiac output (CO) is the product of the heart rate (HR), i.e. the number of heartbeats per minute (bpm), and the stroke volume (SV), which is the volume of blood pumped from the left ventricle per beat; thus giving the formula:

<span class="mw-page-title-main">Echocardiography</span> Medical imaging technique of the heart

An echocardiography, also known as a cardiac echography, is an ultrasound of the heart. It is a type of medical imaging of the heart, using standard ultrasound or Doppler ultrasound. The visual image formed using this technique is called an echocardiogram.

An ejection fraction (EF) is the volumetric fraction of fluid ejected from a chamber with each contraction. It can refer to the cardiac atrium, ventricle, gall bladder, or leg veins, although if unspecified it usually refers to the left ventricle of the heart. EF is widely used as a measure of the pumping efficiency of the heart and is used to classify heart failure types. It is also used as an indicator of the severity of heart failure, although it has recognized limitations.

A transthoracic echocardiogram (TTE) is the most common type of echocardiogram, which is a still or moving image of the internal parts of the heart using ultrasound. In this case, the probe is placed on the chest or abdomen of the subject to get various views of the heart. It is used as a non-invasive assessment of the overall health of the heart, including a patient's heart valves and degree of heart muscle contraction. The images are displayed on a monitor for real-time viewing and then recorded.

<span class="mw-page-title-main">Doppler echocardiography</span> Medical imaging technique of the heart

Doppler echocardiography is a procedure that uses Doppler ultrasonography to examine the heart. An echocardiogram uses high frequency sound waves to create an image of the heart while the use of Doppler technology allows determination of the speed and direction of blood flow by utilizing the Doppler effect.

<span class="mw-page-title-main">3D ultrasound</span> Rendering technique in medical imaging

3D ultrasound is a medical ultrasound technique, often used in fetal, cardiac, trans-rectal and intra-vascular applications. 3D ultrasound refers specifically to the volume rendering of ultrasound data. When involving a series of 3D volumes collected over time, it can also be referred to as 4D ultrasound or real-time 3D ultrasound.

Aortic valve area calculation is an indirect method of determining the area of the aortic valve. The calculated aortic valve orifice area is currently one of the measures for evaluating the severity of aortic stenosis. A valve area of less than 1.0 cm2 is considered to be severe aortic stenosis.

The photoacoustic Doppler effect is a type of Doppler effect that occurs when an intensity modulated light wave induces a photoacoustic wave on moving particles with a specific frequency. The observed frequency shift is a good indicator of the velocity of the illuminated moving particles. A potential biomedical application is measuring blood flow.

The E/A ratio is a marker of the function of the left ventricle of the heart. It represents the ratio of peak velocity blood flow from left ventricular relaxation in early diastole to peak velocity flow in late diastole caused by atrial contraction. It is calculated using Doppler echocardiography, an ultrasound-based cardiac imaging modality. Abnormalities in the E/A ratio suggest that the left ventricle, which pumps blood into the systemic circulation, cannot fill with blood properly in the period between contractions. This phenomenon is referred to as diastolic dysfunction and can eventually lead to the symptoms of heart failure.

Tissue Doppler echocardiography (TDE) is a medical ultrasound technology, specifically a form of echocardiography that measures the velocity of the heart muscle (myocardium) through the phases of one or more heartbeats by the Doppler effect of the reflected ultrasound. The technique is the same as for flow Doppler echocardiography measuring flow velocities. Tissue signals, however, have higher amplitude and lower velocities, and the signals are extracted by using different filter and gain settings. The terms tissue Doppler imaging (TDI) and tissue velocity imaging (TVI) are usually synonymous with TDE because echocardiography is the main use of tissue Doppler.

Harmonic phase (HARP) algorithm is a medical image analysis technique capable of extracting and processing motion information from tagged magnetic resonance image (MRI) sequences. It was initially developed by N. F. Osman and J. L. Prince at the Image Analysis and Communications Laboratory at Johns Hopkins University. The method uses spectral peaks in the Fourier domain of tagged MRI, calculating the phase images of their inverse Fourier transforms, which are called harmonic phase (HARP) images. The motion of material points through time is then tracked, under the assumption that the HARP value of a fixed material point is time-invariant. The method is fast and accurate, and has been accepted as one of the most popular tagged MRI analysis methods in medical image processing.

In cardiology, ventricular dyssynchrony is a difference in the timing, or lack of synchrony, of contractions in different ventricles in the heart. Large differences in timing of contractions can reduce cardiac efficiency and is correlated with heart failure.

<span class="mw-page-title-main">Left ventricular thrombus</span> Medical condition

Left ventricular thrombus is a blood clot (thrombus) in the left ventricle of the heart. LVT is a common complication of acute myocardial infarction (AMI). Typically the clot is a mural thrombus, meaning it is on the wall of the ventricle. The primary risk of LVT is the occurrence of cardiac embolism, in which the thrombus detaches from the ventricular wall and travels through the circulation and blocks blood vessels. Blockage can be especially damaging in the heart or brain (stroke).

<span class="mw-page-title-main">Doppler ultrasonography</span> Ultrasound imaging of the movement of tissues and body fluids using the Doppler effect

Doppler ultrasonography is medical ultrasonography that employs the Doppler effect to perform imaging of the movement of tissues and body fluids, and their relative velocity to the probe. By calculating the frequency shift of a particular sample volume, for example, flow in an artery or a jet of blood flow over a heart valve, its speed and direction can be determined and visualized.

The assessment of regional function of the heart is a good tool for early detection of deterioration in certain parts of the heart wall before a cardiac arrest is diagnosed. One of the most accurate measures of changes in regional function is the use of strain as a measure of regional function of cardiac muscle.

Numerical manipulation of Doppler parameters obtain during routine Echocardiography has been extensively utilized to non-invasively estimate intra-cardiac pressures, in many cases removing the need for invasive cardiac catheterization.

The Relative Atrial Index (RAI) is a numeric parameter used to assess for cardiac shunt defects. It is calculated from the standard transthoracic Doppler echocardiogram measurements of the right atrial area divided by the left atrial area. RAI = right atrial area / left atrial area. These measurements are made from the apical four chamber view.

<span class="mw-page-title-main">Strain rate imaging</span>

Strain rate imaging is a method in echocardiography for measuring regional or global deformation of the myocardium. The term "deformation" refers to the myocardium changing shape and dimensions during the cardiac cycle. If there is myocardial ischemia, or there has been a myocardial infarction, in part of the heart muscle, this part is weakened and shows reduced and altered systolic function. Also in regional asynchrony, as in bundle branch block, there is regional heterogeneity of systolic function. By strain rate imaging, the simultaneous function of different regions can be displayed and measured. The method was first based on colour tissue Doppler. by using the longitudinal myocardial velocity gradient, already in use transmurally. Later, the regional deformation has also been available by speckle tracking echocardiography, both methods having some, but different methodological weaknesses. Both methods, however, will acquire the same data, and also can be displayed by the same type of display.

In clinical cardiology the term "diastolic function" is most commonly referred as how the heart fills. Parallel to "diastolic function", the term "systolic function" is usually referenced in terms of the left ventricular ejection fraction (LVEF), which is the ratio of stroke volume and end-diastolic volume. Due to the epidemic of heart failure, particularly the cases determined as diastolic heart failure, it is increasingly urgent and crucial to understand the meaning of “diastolic function”. Unlike "systolic function", which can be simply evaluated by LVEF, there are no established dimensionless parameters for "diastolic function" assessment. Hence to further study "diastolic function" the complicated and speculative physiology must be taken into consideration.

References

  1. Geyer, Holly; Caracciolo, Giuseppe; Abe, Haruhiko; Wilansky, Susan (2010), "Assessment of Myocardial Mechanics Using Speckle Tracking Echocardiography: Fundamentals and Clinical Applications", Journal of the American Society of Echocardiography, C.V. Mosby, 23 (4): 351–69, quiz 453-5, doi:10.1016/j.echo.2010.02.015, ISSN   0894-7317, OCLC   605144740, PMID   20362924
  2. Bohs LN, Trahey GE. A novel method for angle independent ultrasonic imaging of blood flow and tissue motion. IEEE Trans Biomed Eng. 1991 Mar;38(3):280-6.
  3. 1 2 3 Kaluzynski K, Chen X, Emelianov SY, Skovoroda AR, O'Donnell M. Strain rate imaging using two-dimensional speckle tracking. IEEE Trans Ultrason Ferroelectr Freq Control. 2001 Jul;48(4):1111-23.
  4. 1 2 Reisner, SA; Lysyansky, P; Agmon, Y; Mutlak, D (2004), "Global longitudinal strain: a novel index of left ventricular systolic function", Journal of the American Society of Echocardiography, 17 (6): 630–3, doi:10.1016/j.echo.2004.02.011, ISSN   0894-7317, OCLC   110737191, PMID   15163933
  5. Leitman M, Lysyansky P, Sidenko S, Shir V, Peleg E, Binenbaum M, et al.Two-dimensional strain-a novel software for real-time quantitative echocardiographic assessment of myocardial function. JAm Soc Echocardiogr 2004;17:1021-9.
  6. Insana MF, Wagner RF, Garra BS, Momenan R, Shawker TH. Pattern recognition methods for optimizing multivariate tissue signatures in diagnostic ultrasound. Ultrason Imaging. 1986 Jul;8(3):165-80
  7. Bohs LN, Friemel BH, Trahey GE. Experimental velocity profiles and volumetric flow via two-dimensional speckle tracking. Ultrasound Med Biol. 1995;21(7):885-98
  8. Ingul CB, Torp H, Aase SA, Berg S, Stoylen A, Slordahl SA. Automated analysis of strain rate and strain: feasibility and clinical implications. J Am Soc Echocardiogr. 2005 May;18(5):411-8.
  9. 1 2 Amundsen BH, Crosby J, Steen PA, Torp H, Slørdahl SA, Støylen A. Regional myocardial long-axis strain and strain rate measured by different tissue Doppler and speckle tracking echocardiography methods: a comparison with tagged magnetic resonance imaging. Eur J Echocardiogr. 2009 Mar;10(2):229-37
  10. Modesto KM, Cauduro S, Dispenzieri A, Khandheria B, Belohlavek M, Lysyansky P, Friedman Z, Gertz M, Abraham TP.Two-dimensional acoustic pattern derived strain parameters closely correlate with one-dimensional tissue Doppler derived strain measurements. Eur J Echocardiogr. 2006 Aug;7(4):315-21
  11. 1 2 3 4 Cho GY, Chan J, Leano R, Strudwick M, Marwick TH. Comparison of two-dimensional speckle and tissue velocity based strain and validation with harmonic phase magnetic resonance imaging. Am J Cardiol 2006; 97:1661-6
  12. Helle-Valle T, Crosby J, Edvardsen T, Lyseggen E, Amundsen BH, Smith HJ, Rosen BD, Lima JA, Torp H, Ihlen H, Smiseth OA. New noninvasive method for assessment of left ventricular rotation: speckle tracking echocardiography. Circulation. 2005 Nov 15;112(20):3149-56
  13. Abraham TP, Dimaano VL, Liang HY. Role of tissue Doppler and strain echocardiography in current clinical practice. Circulation 2007;116: 2597-609.
  14. D’Hooge J, Heimdal A, Jamal F, Kukulski T, Bijnens B, Rademakers F, et al. Regional strain and strain rate measurements by cardiac ultrasound: principles, implementation and limitations. Eur J Echocardiogr 2000;1: 154-70.
  15. Edvardsen T, Gerber BL, Garot J, Bluemke DA, Lima JA, Smiseth OA.Quantitative assessment of intrinsic regional myocardial deformation by Doppler strain rate echocardiography in humans: validation against three-dimensional tagged magnetic resonance imaging. Circulation 2002;106:50-6
  16. Amundsen BH, Helle-Valle T, Edvardsen T, Torp H, Crosby J, Lyseggen E,et al. Noninvasive myocardial strain measurement by speckle tracking echocardiography: validation against sonomicrometry and tagged magnetic resonance imaging. J Am Coll Cardiol 2006;47:789-93
  17. Roes SD, Mollema SA, Lamb HJ, van derWall EE, de Roos A, Bax JJ. Validation of echocardiographic two-dimensional speckle tracking longitudinal strain imaging for viability assessment in patients with chronic ischemic left ventricular dysfunction and comparison with contrastenhanced magnetic resonance imaging. Am J Cardiol 2009;104:312-7
  18. Hanekom L, Cho GY, Leano R, Jeffriess L, Marwick TH. Comparison of two-dimensional speckle and tissue Doppler strain measurement during dobutamine stress echocardiography: an angiographic correlation. Eur Heart J. 2007 Jul;28(14):1765-72.
  19. Costa SP, Beaver TA, Rollor JL, Vanichakarn P, Magnus PC, Palac RT.Quantification of the variability associated with repeat measurements of left ventricular two-dimensional global longitudinal strain in a real-world setting. J Am Soc Echocardiogr. 2014 Jan;27(1):50-4

Further reading