Ultrasound-modulated optical tomography | |
---|---|
Purpose | imaging of soft tissue |
Ultrasound-modulated optical tomography (UOT), also known as Acousto-Optic Tomography (AOT), is a hybrid imaging modality that combines light and sound; it is a form of tomography involving ultrasound. [1] It is used in imaging of biological soft tissues and has potential applications for early cancer detection. [2] As a hybrid modality which uses both light and sound, UOT provides some of the best features of both: the use of light provides strong contrast and sensitivity (both molecular and functional); these two features are derived from the optical component of UOT. The use of ultrasound allows for high resolution, as well as a high imaging depth. However, the difficulty of tackling the two fundamental problems with UOT (low SNR in deep tissue and short speckle decorrelation time) have caused UOT to evolve relatively slowly; most work in the field is limited to theoretical simulations or phantom / sample studies. [3]
In UOT, ultrasound transducers are used to apply ultrasounds wave into a medium, usually some biological tissue. Applying these ultrasound waves, or an ultrasound field, to a region of tissue will change the optical properties of the tissue in time and space. This region of ultrasound-modulated tissue is the region of interest (ROI) which will be analyzed. Photons are then sent into the tissue from some source, such as a laser. Eventually, despite the strength of optical scattering in tissue, some of these photons will pass through the ROI. The photons that pass through the ROI will change according to the modulation of the tissue; this causes the photons to be "tagged". Typically, this tagging will cause the frequency of the light to shift by the frequency of the US field. [4]
Sufficiently coherent light traveling through a medium creates a speckle pattern (Fig 1 of next citation). [5] Modulating the ultrasound field applied to the ROI will cause the speckle patterns to change, due to the 3 modulation mechanisms which are explained below. The changes of these speckle patterns are used to derive various properties of the tissue during reconstruction and analysis. Optical properties that can be derived include the optical absorption coefficient, the optical scattering coefficient, and the fluence in the region of interest; UOT can also be used to derive mechanical properties as well. [5]
Optical imaging modalities typically rely on ballistic photons to collect and convey information. However, as a result of strong optical scattering in tissue, conventional imaging modalities struggle to image deeper into tissue, past the optical diffusion limit (typically about 1 mm into tissue). [6] Various imaging modalities have been developed to peer deeper into tissue, such as diffuse optical tomography (DOT) and optical coherence tomography (OCT). While OCT has fantastic spatial resolution (3.5 and 7 micrometer resolutions, axially and lateral respectively), its imaging depth is limited to the millimeter range (e.g. 2.5 mm). [7] DOT has fantastic penetration, on the scale of centimeters, but suffers from inferior resolution (~1cm). [8]
To address the difficulty of deep optical imaging, hybrid ultrasound and optical imaging modalities have been developed, namely UOT and photoacoustic imaging (PAI). Both imaging modalities use diffuse photons, which typically cannot be used to transmit information from deep within the tissue. This is because strong optical scattering means that it is incredibly difficult to determine where the photons have traveled and how many scattering events they might have gone through. UOT and PAI have different ways to effectively transfer information from the diffuse photons within the tissue back to the system, which allows for centimeter imaging depths (deeper than DOT) while retaining high spatial resolutions (millimeters to 100s of micrometers). [9]
Photoacoustic imaging sends a pulse of light into tissue; as photons are absorbed by the tissue, the resulting heating rise results in thermoelastic expansion. This leads to a pressure wave propagation, which is collected by ultrasound transducers. [10] Thus, PAI uses photons to acquire information deep inside tissue and uses ultrasound to transmit that information to the system. In contrast, UOT relies on photons for information transmission and uses ultrasound to acquire information. Ultrasound is used in UOT to "tag" or identify photons which have passed through the region of interest; these photons can be trusted to carry information back to the system regarding the region of interest. [4] The modulation of the ultrasound will accordingly change the optical properties of the tissue, which can be used to derive both optical and mechanical properties.
The difference in the use of ultrasound between PAI and UOT means that different types of information can be derived from the two different modalities. PAI is proficient in delivering information regarding optical absorption; UOT can provide information regarding both optical absorption and optical scattering. Thus, UOT has a strong advantage over PAI in that UOT can provide information about tissue/organ structure as well as tissue metabolism. [3]
UOT utilizes all photons within deep tissue; as long as they have not been absorbed and they pass through the ROI, those photons can still be used to convey information in UOT. Thus, UOT can achieve imaging depths (exploiting diffuse photons) deeper than 9 cm into the body while retaining high spatial resolutions (from the dimensions of the ultrasound focus), ~mm scale, as of 2017. [5] [11] UOT can be used to derive mechanical and optical properties of the tissue; compared to PAI, which can only derive optical absorption, UOT can derive absorption and scattering properties of the tissue.
There are two fundamental problems with UOT:
UOT is built upon the modulation of light due to the effects of ultrasonic waves on the optical properties of the testing medium. The target within the testing medium will be irradiated by a laser beam and a focused ultrasonic wave. The reemitted light propagating through the ultrasonic field or ultrasonic focal zone will then carry information of local optical and acoustic properties of this zone. These properties will be used to reconstruct images showing the inside view of the medium.
There are three important mechanisms behind UOT technology.
1 Incoherent Modulation of Light Due to Ultrasound-Induced Variations in Optical Properties of Medium. [5] [13] As the ultrasonic wave propagates through the medium, the mass density of the medium will be changed due to the vibration. This variation in mass density will then influence the local optical properties. For example, the local absorption coefficient, scattering coefficient, and the index of refraction will all be modulated. With different optical properties, the reemitted light features (like intensity) will be modified.
2 Variations in Optical Phase in Response to Ultrasound-Induced Displacements of Scatterers. [5] [13] This mechanism mainly describes the effect in a microcosmic view. With the vibration from focused ultrasound, the local scatterers within the medium will be moved. When coherent light passes through such region, the displacement of scatterers will then cause optical phases change and then further modulate the light's free-path lengths. In the end, the reemitted light will form speckle pattern.
3 Variations in Optical Phase in Response to Ultrasonic Modulation of Index of Refraction of Background Medium. [5] [13] Similar to the second mechanism, the mass density modulation caused by ultrasonic wave vibration will also modulate the medium's index of refraction. This effect can further influence the free-path phases when light passes through the ultrasonic region and form a speckle pattern.
In conclusion, these three mechanisms describe how the modulation of the ultrasonic wave can modulate and fluctuate the light intensity (in mechanism 1), the light phase and form speckle pattern (mechanism 2 and 3). Mechanisms 2 and 3 require a coherent light source; mechanism 1 does not. When coherent light sources are applied, mechanism 1 can be disregarded, as its effect compared to mechanisms 2 and 3 are effectively negligible. [5] These three mechanisms are the fundamental building blocks required to design a UOT system.
In analytic model, two approximations are made. (1) the optical wavelength is much shorter than the mean free path (weak-scattering approximation) and (2) the ultrasound-induced change in the optical path length is much less than the optical wavelength (weak-modulation approximation). [13]
With the first approximation, we can assume the ensemble-averaged correlations between electric fields are the same. Since the differences between correlation results from different paths are negligible. With this assumption, the correlations for electric fields G1(ꚍ) can be written as the following:
In this equation the parenthesis represents the ensemble and time averaging. Es demsontrates the unit-amplitude electric field of the scattered light of a path of length s, and p(s) denotes the probability density function of s. [13] In analytic UOT model, we treat the light source as an optical plane wave. We assume the plane wave light normally hit a slab of thickness d. The transmitted light will be captured by a point detector. After applying diffusion theory, the original G1 equation can be further modified as:
Where
In these equations, ωa represents the acoustic angular frequency, and the n0 is the background index of refraction; k0 is the magnitude of the optical wave vector in vacuo; A is the acoustic amplitude, which is proportional to the acoustic pressure; ka is the magnitude of the acoustic wavevector; lt' is the optical transport mean free path; η is the elasto-optical coefficient; ρ is the mass density; [14] represents how ultrasound averagely change the light per free path via index of refraction and displacementare respectively.
After deriving the autocorrelation equation, Wiener-Khinchin theorem is applied. With this theorem, we can further connect G1 with the spectral density of the modulated speckle. Their relationship in frequency space is shown as the following Fourier transformation equation.
For simplicity, the Fourier transformed term exp(-iω0t) is dropped, and the ω here represents relative angular frequency of unmodulated light. For example, if ω=0 this equation is calculating the spectral density with absolute angular frequency ω0. Since G1 is an even autocorrelation function, the spectral intensity at ωa can be written as:
Here, the n and Ta represent the acoustic period. Since the frequency spectrum is symmetric about ω0, the one side modulation depth is defined as:
Then we can consider the condition under the second approximation (weak-modulation approximation). In this situation, the term is much smaller than 1. By applying the feature of sinh function, which is the main component of G1 function, the original autocorrelation G1 can be further simplified as:
Therefore, the one side modulation depth can be simplified as . From previous equation, we can see . Therefore, in conclusion, A and M1 has quadratic relationship. Such quadratic modulation can be captured by a Fabry-Perot interferometer. Or, we can calculate the ratio between observed AC signal and he observed DC signal (also named as apparent modulation depth) which can carry enough information to represent such modulation as well.
In conclusion, in UOT analytic model, with the help of weak-scattering approximation, weak-modulation approximation, diffusion theory, and Wiener-Khinchin theorem, the relationship between acoustic amplitude and modulated light can be successfully observed.
For single-frequency UOT, the axial resolution along ultrasonic axis is always limited by the elongated ultrasonic focal zone. To improve the axial resolution, Ultrasonic frequency-swept UOT model is designed. In this system, the object is placed in a tank full of UOT scattering medium. There will also be an ultrasound absorber at the bottom of the tank to avoid rebound of ultrasound. Basically, a function generator will produce a frequency signal relating to time. After passing through a power amplifier and a transformer, such frequency command will be sent to the ultrasonic transducer to generate ultrasonic beam with different frequencies. After a brief calculation, a focused ultrasonic beam will be sent to the medium and the target. Meanwhile, a laser beam which is perpendicular to the ultrasonic beam will also illuminate the scattering medium. Then, on the other side of the light source, the PMT, modulated by frequency signal sent through the first function generator will detect transmitted light signal within the tank and transfer the optical signal to electrical signal. The electrical signal will then pass through an amplifier, an Oscilloscope and be stored in the data base.
With such data base, spectral intensity vs frequency plots at multiple points can be generated (The first spectrum is generated as a reference, produced by optical signal far from the object.). Each of the spectrum can then be further converted to a 1D image showing the interior of the medium in the direction perpendicular to the tank (z direction). In the end, all the 1D image will be pieced together to generate a full view inside the medium.
In summary, a frequency-swept (chirped) ultrasonic wave can encode laser light traversing the acoustic axis with various frequencies. Decoding the transmitted light provides resolution along the acoustic axis. This scheme is analogous to MRI. [16]
Was first proposed as a method for virus detection in 2013. [17]
Recent advances in UOT (2020 onwards) include 1) the development of Coded Ultrasound Transmissions for SNR gain in AOI, [18] 2) the development of Homodyne Time of Flight AOI, [9] 3) the use of super-resolution techniques to improve UOT beyond the acoustic diffraction limit, [11] and 4) the use of coaxial interferometry to better enable modern high-performance cameras for parallel detection of UOT signals. [19]
In physics, attenuation is the gradual loss of flux intensity through a medium. For instance, dark glasses attenuate sunlight, lead attenuates X-rays, and water and air attenuate both light and sound at variable attenuation rates.
Optical coherence tomography (OCT) is an imaging technique that uses interferometry with short-coherence-length light to obtain micrometer-level depth resolution and uses transverse scanning of the light beam to form two- and three-dimensional images from light reflected from within biological tissue or other scattering media. Short-coherence-length light can be obtained using a superluminescent diode (SLD) with a broad spectral bandwidth or a broadly tunable laser with narrow linewidth. The first demonstration of OCT imaging was published by a team from MIT and Harvard Medical School in a 1991 article in the journal Science. The article introduced the term "OCT" to credit its derivation from optical coherence-domain reflectometry, in which the axial resolution is based on temporal coherence. The first demonstrations of in vivo OCT imaging quickly followed.
Fluorescence-lifetime imaging microscopy or FLIM is an imaging technique based on the differences in the exponential decay rate of the photon emission of a fluorophore from a sample. It can be used as an imaging technique in confocal microscopy, two-photon excitation microscopy, and multiphoton tomography.
An acousto-optic modulator (AOM), also called a Bragg cell or an acousto-optic deflector (AOD), uses the acousto-optic effect to diffract and shift the frequency of light using sound waves. They are used in lasers for Q-switching, telecommunications for signal modulation, and in spectroscopy for frequency control. A piezoelectric transducer is attached to a material such as glass. An oscillating electric signal drives the transducer to vibrate, which creates sound waves in the material. These can be thought of as moving periodic planes of expansion and compression that change the index of refraction. Incoming light scatters off the resulting periodic index modulation and interference occurs similar to Bragg diffraction. The interaction can be thought of as a three-wave mixing process resulting in sum-frequency generation or difference-frequency generation between phonons and photons.
Medical optical imaging is the use of light as an investigational imaging technique for medical applications, pioneered by American Physical Chemist Britton Chance. Examples include optical microscopy, spectroscopy, endoscopy, scanning laser ophthalmoscopy, laser Doppler imaging, and optical coherence tomography. Because light is an electromagnetic wave, similar phenomena occur in X-rays, microwaves, and radio waves.
Photoacoustic imaging or optoacoustic imaging is a biomedical imaging modality based on the photoacoustic effect. Non-ionizing laser pulses are delivered into biological tissues and part of the energy will be absorbed and converted into heat, leading to transient thermoelastic expansion and thus wideband ultrasonic emission. The generated ultrasonic waves are detected by ultrasonic transducers and then analyzed to produce images. It is known that optical absorption is closely associated with physiological properties, such as hemoglobin concentration and oxygen saturation. As a result, the magnitude of the ultrasonic emission, which is proportional to the local energy deposition, reveals physiologically specific optical absorption contrast. 2D or 3D images of the targeted areas can then be formed.
Acousto-optics is a branch of physics that studies the interactions between sound waves and light waves, especially the diffraction of laser light by ultrasound through an ultrasonic grating.
Ballistic light, also known as ballistic photons, is photons of light that have traveled through a scattering (turbid) medium in a straight line.
Electrophoretic light scattering is based on dynamic light scattering. The frequency shift or phase shift of an incident laser beam depends on the dispersed particles mobility. With dynamic light scattering, Brownian motion causes particle motion. With electrophoretic light scattering, oscillating electric field performs this function.
An ultrasonic grating is a type of diffraction grating produced by the interference of ultrasonic waves in a medium, which alters the physical properties of the medium in a grid-like pattern. The term acoustic grating is a more general term that includes operation at audible frequencies.
The photoacoustic Doppler effect is a type of Doppler effect that occurs when an intensity modulated light wave induces a photoacoustic wave on moving particles with a specific frequency. The observed frequency shift is a good indicator of the velocity of the illuminated moving particles. A potential biomedical application is measuring blood flow.
Diffusing-wave spectroscopy (DWS) is an optical technique derived from dynamic light scattering (DLS) that studies the dynamics of scattered light in the limit of strong multiple scattering. It has been widely used in the past to study colloidal suspensions, emulsions, foams, gels, biological media and other forms of soft matter. If carefully calibrated, DWS allows the quantitative measurement of microscopic motion in a soft material, from which the rheological properties of the complex medium can be extracted via the microrheology approach.
Wide-field multiphoton microscopy refers to an optical non-linear imaging technique tailored for ultrafast imaging in which a large area of the object is illuminated and imaged without the need for scanning. High intensities are required to induce non-linear optical processes such as two-photon fluorescence or second harmonic generation. In scanning multiphoton microscopes the high intensities are achieved by tightly focusing the light, and the image is obtained by beam scanning. In wide-field multiphoton microscopy the high intensities are best achieved using an optically amplified pulsed laser source to attain a large field of view (~100 µm). The image in this case is obtained as a single frame with a CCD without the need of scanning, making the technique particularly useful to visualize dynamic processes simultaneously across the object of interest. With wide-field multiphoton microscopy the frame rate can be increased up to a 1000-fold compared to multiphoton scanning microscopy. Wide-field multiphoton microscopes are not yet commercially available, but working prototypes exist in several optics laboratories.
Super-resolution photoacoustic imaging is a set of techniques used to enhance spatial resolution in photoacoustic imaging. Specifically, these techniques primarily break the optical diffraction limit of the photoacoustic imaging system. It can be achieved in a variety of mechanisms, such as blind structured illumination, multi-speckle illumination, or photo-imprint photoacoustic microscopy in Figure 1.
Photoacoustic microscopy is an imaging method based on the photoacoustic effect and is a subset of photoacoustic tomography. Photoacoustic microscopy takes advantage of the local temperature rise that occurs as a result of light absorption in tissue. Using a nanosecond pulsed laser beam, tissues undergo thermoelastic expansion, resulting in the release of a wide-band acoustic wave that can be detected using a high-frequency ultrasound transducer. Since ultrasonic scattering in tissue is weaker than optical scattering, photoacoustic microscopy is capable of achieving high-resolution images at greater depths than conventional microscopy methods. Furthermore, photoacoustic microscopy is especially useful in the field of biomedical imaging due to its scalability. By adjusting the optical and acoustic foci, lateral resolution may be optimized for the desired imaging depth.
Speckle variance optical coherence tomography (SV-OCT) is an imaging algorithm for functional optical imaging. Optical coherence tomography is an imaging modality that uses low-coherence interferometry to obtain high resolution, depth-resolved volumetric images. OCT can be used to capture functional images of blood flow, a technique known as optical coherence tomography angiography (OCT-A). SV-OCT is one method for OCT-A that uses the variance of consecutively acquired images to detect flow at the micron scale. SV-OCT can be used to measure the microvasculature of tissue. In particular, it is useful in ophthalmology for visualizing blood flow in retinal and choroidal regions of the eye, which can provide information on the pathophysiology of diseases.
Diffuse optical mammography, or simply optical mammography, is an emerging imaging technique that enables the investigation of the breast composition through spectral analysis. It combines in a single non-invasive tool the capability to implement breast cancer risk assessment, lesion characterization, therapy monitoring and prediction of therapy outcome. It is an application of diffuse optics, which studies light propagation in strongly diffusive media, such as biological tissues, working in the red and near-infrared spectral range, between 600 and 1100 nm.
Quantum optical coherence tomography (Q-OCT) is an imaging technique that uses nonclassical (quantum) light sources to generate high-resolution images based on the Hong-Ou-Mandel effect (HOM). Q-OCT is similar to conventional OCT but uses a fourth-order interferometer that incorporates two photodetectors rather than a second-order interferometer with a single photodetector. The primary advantage of Q-OCT over OCT is insensitivity to even-order dispersion in multi-layered and scattering media.
Ultrasound-switchable fluorescence (USF) imaging is a deep optics imaging technique. In last few decades, fluorescence microscopy has been highly developed to image biological samples and live tissues. However, due to light scattering, fluorescence microscopy is limited to shallow tissues. Since fluorescence is characterized by high contrast, high sensitivity, and low cost which is crucial to investigate deep tissue information, developing fluorescence imaging technique with high depth-to-resolution ratio would be promising.. Recently, ultrasound-switchable fluorescence imaging has been developed to achieve high signal-to-noise ratio (SNR) and high spatial resolution imaging without sacrificing image depth.
Diffuse correlation spectroscopy (DCS) is a type of medical imaging and optical technique that utilizes near-infrared light to directly and non-invasively measure tissue blood flow. The imaging modality was created by David Boas and Arjun Yodh in 1995.