Fluorescence anisotropy

Last updated

Fluorescence anisotropy or fluorescence polarization is the phenomenon where the light emitted by a fluorophore has unequal intensities along different axes of polarization. Early pioneers in the field include Aleksander Jablonski, Gregorio Weber, [1] and Andreas Albrecht. [2] The principles of fluorescence polarization and some applications of the method are presented in Lakowicz's book. [3]

Contents

Definition of fluorescence anisotropy

Fluorescence Polarization Anisotropy.png

The anisotropy (r) of a light source is defined as the ratio of the polarized component to the total intensity (): [3]

When the excitation is polarized along the z-axis, emission from the fluorophore is symmetric around the z-axis(Figure). Hence statistically we have . As , and , we have

.

Principle – Brownian motion and photoselection

Brownian Motion of a Nanoparticle Brownian motion large.gif
Brownian Motion of a Nanoparticle

In fluorescence, [4] a molecule absorbs a photon and gets excited to a higher energy state. After a short delay (the average represented as the fluorescence lifetime ), it comes down to a lower state by losing some of the energy as heat and emitting the rest of the energy as another photon. The excitation and de-excitation involve the redistribution of electrons about the molecule. Hence, excitation by a photon can occur only if the electric field of the light is oriented in a particular axis about the molecule. Also, the emitted photon will have a specific polarization with respect to the molecule.

The first concept to understand for anisotropy measurements is the concept of Brownian motion. Although water at room temperature contained in a glass to the eye may look very still, on the molecular level each water molecule has kinetic energy and thus there are many collisions between water molecules in any amount of time. A nanoparticle (yellow dot in the figure) suspended in solution will undergo a random walk due to the summation of these underlying collisions. The rotational correlation time (Φr), the time it takes for the molecule to rotate 1 radian, is dependent on the viscosity (η), temperature (T), Boltzmann constant (kB) and volume (V) of the nanoparticle: [5]

The second concept is photoselection by use of a polarized light. When polarized light is applied to a group of randomly oriented fluorophores, most of the excited molecules will be those oriented within a particular range of angles to the applied polarization. If they do not move, the emitted light will also be polarized within a particular range of angles to the applied light.

For single-photon excitation the intrinsic anisotropy r0 has a maximum theoretical value of 0.4 when the excitation and emission dipoles are parallel and a minimum value of -0.2 when the excitation and emission dipoles are perpendicular.

where β is the angle between the excitation and emission dipoles. For steady-state fluorescence measurements it is usually measured by embedding the fluorophore in a frozen polyol.

Taking the idealistic simplest case a subset of dye molecules suspended in solution that have a mono-exponential fluorescence lifetime and r0=0.4 (rhodamine 6g in ethylene glycol made to have an absorbance of ~0.05 is a good test sample). If the excitation is unpolarized then the measured fluorescence emission should likewise be unpolarized. If however the excitation source is vertically polarized using an excitation polarizer then polarization effects will be picked up in the measured fluorescence. These polarization artifacts can be combated by placing an emission polarizer at the magic angle of 54.7º. If the emission polarizer is vertically polarized there will be an additional loss of fluorescence as Brownian motion results in dye molecules moving from an initial vertical polarized configuration to an unpolarized configuration. On the other hand, if the emission polarizer is horizontally polarized there will be an additional introduction of excited molecules that were initially vertically polarized and became depolarized via Brownian motion. The fluorescence sum and difference can be constructed by addition of the intensities and subtraction of the fluorescence intensities respectively:

Dividing the difference by the sum gives the anisotropy decay:

The grating factor G is an instrumental preference of the emission optics for the horizontal orientation to the vertical orientation. It can be measured by moving the excitation polarizer to the horizontal orientation and comparing the intensities when the emission polarizer is vertically and horizontally polarized respectively.

G is emission wavelength dependent. Note G in literature is defined as the inverse shown.

The degree of decorrelation in the polarization of the incident and emitted light depends on how quickly the fluorophore orientation gets scrambled (the rotational lifetime ) compared to the fluorescence lifetime (). The scrambling of orientations can occur by the whole molecule tumbling or by the rotation of only the fluorescent part. The rate of tumbling is related to the measured anisotropy by the Perrin equation:

Where r is the observed anisotropy, r0 is the intrinsic anisotropy of the molecule, is the fluorescence lifetime and is the rotational correlation time. [6]

This analysis is valid only if the fluorophores are relatively far apart. If they are very close to another, they can exchange energy by FRET and because the emission can occur from one of many independently moving (or oriented) molecules this results in a lower than expected anisotropy or a greater decorrelation. This type of homotransfer Förster resonance energy transfer is called energy migration FRET or emFRET.

Steady-state fluorescence anisotropy only give an "average" anisotropy. Much more information can be obtained with time-resolved fluorescence anisotropy [7] where the decay time, residual anisotropy and rotational correlation time can all be determined from fitting the anisotropy decay. Typically a vertically pulsed laser source is used for excitation and timing electronics are added between the start pulses of the laser (start) and the measurement of the fluorescence photons (stop). The technique Time-Correlated Single Photon Counting (TCSPC) is typically employed.

Again using the idealistic simplest case a subset of dye molecules suspended in solution that have a mono-exponential fluorescence lifetime and an initial anisotropy r0=0.4. If the sample is excited with a pulsed vertically orientated excitation source then a single decay time should be measured when the emission polarizer is at the magic angle. If the emission polarizer is vertically polarized instead two decay times will be measured both with positive pre-exponential factors, the first decay time should be equivalent to measured with the unpolarized emission set-up and the second decay time will be due to the loss of fluorescence as Brownian motion results in dye molecules moving from an initial vertical polarized configuration to an unpolarized configuration. On the other hand, if the emission polarizer is horizontally polarized, two decay times will again be recovered the first one with a positive pre-exponential factor and will be equivalent to but the second one will have a negative pre-exponential factor resulting from the introduction of excited molecules that were initially vertically polarized and became depolarized via Brownian motion. The fluorescence sum and difference can be constructed by addition of the decays and subtraction of the fluorescence decays respectively:

Dividing the difference by the sum gives the anisotropy decay:

In the simplest case for only one species of spherical dye:

Applications

Fluorescence anisotropy can be used to measure the binding constants and kinetics of reactions that cause a change in the rotational time of the molecules. If the fluorophore is a small molecule, the rate at which it tumbles can decrease significantly when it is bound to a large protein. If the fluorophore is attached to the larger protein in a binding pair, the difference in polarization between bound and unbound states will be smaller (because the unbound protein will already be fairly stable and tumble slowly to begin with) and the measurement will be less accurate. The degree of binding is calculated by using the difference in anisotropy of the partially bound, free and fully bound (large excess of protein) states measured by titrating the two binding partners.

If the fluorophore is bound to a relatively large molecule like a protein or an RNA, the change in the mobility accompanying folding can be used to study the dynamics of folding. This provides a measure of the dynamics of how the protein achieves its final, stable 3D shape. In combination with fluorophores which interact via Förster resonance energy transfer(FRET), fluorescence anisotropy can be used to detect the oligomeric state of complex-forming molecules ("How many of the molecules are interacting?"). [8]

Fluorescence anisotropy is also applied to microscopy, with use of polarizers in the path of the illuminating light and also before the camera. This can be used to study the local viscosity of the cytosol or membranes, with the latter giving information about the membrane microstructure and the relative concentrations of various lipids. This technique has also been used to detect the binding of molecules to their partners in signaling cascades in response to certain cues.

The phenomenon of emFRET and the associated decrease in anisotropy when close interactions occur between fluorophores has been used to study the aggregation of proteins in response to signaling.

See also

Related Research Articles

<span class="mw-page-title-main">Fluorescence</span> Emission of light by a substance that has absorbed light

Fluorescence is the emission of light by a substance that has absorbed light or other electromagnetic radiation. It is a form of luminescence. In most cases, the emitted light has a longer wavelength, and therefore a lower photon energy, than the absorbed radiation. A perceptible example of fluorescence occurs when the absorbed radiation is in the ultraviolet region of the electromagnetic spectrum, while the emitted light is in the visible region; this gives the fluorescent substance a distinct color that can only be seen when the substance has been exposed to UV light. Fluorescent materials cease to glow nearly immediately when the radiation source stops, unlike phosphorescent materials, which continue to emit light for some time after.

<span class="mw-page-title-main">Fluorescence spectroscopy</span> Type of electromagnetic spectroscopy

Fluorescence spectroscopy is a type of electromagnetic spectroscopy that analyzes fluorescence from a sample. It involves using a beam of light, usually ultraviolet light, that excites the electrons in molecules of certain compounds and causes them to emit light; typically, but not necessarily, visible light. A complementary technique is absorption spectroscopy. In the special case of single molecule fluorescence spectroscopy, intensity fluctuations from the emitted light are measured from either single fluorophores, or pairs of fluorophores.

<span class="mw-page-title-main">Fluorophore</span> Agents that emit light after excitation by light

A fluorophore is a fluorescent chemical compound that can re-emit light upon light excitation. Fluorophores typically contain several combined aromatic groups, or planar or cyclic molecules with several π bonds.

Plate readers, also known as microplate readers or microplate photometers, are instruments which are used to detect biological, chemical or physical events of samples in microtiter plates. They are widely used in research, drug discovery, bioassay validation, quality control and manufacturing processes in the pharmaceutical and biotechnological industry and academic organizations. Sample reactions can be assayed in 1-1536 well format microtiter plates. The most common microplate format used in academic research laboratories or clinical diagnostic laboratories is 96-well with a typical reaction volume between 100 and 200 µL per well. Higher density microplates are typically used for screening applications, when throughput and assay cost per sample become critical parameters, with a typical assay volume between 5 and 50 µL per well. Common detection modes for microplate assays are absorbance, fluorescence intensity, luminescence, time-resolved fluorescence, and fluorescence polarization.

<span class="mw-page-title-main">Förster resonance energy transfer</span> Photochemical energy transfer mechanism

Förster resonance energy transfer (FRET), fluorescence resonance energy transfer, resonance energy transfer (RET) or electronic energy transfer (EET) is a mechanism describing energy transfer between two light-sensitive molecules (chromophores). A donor chromophore, initially in its electronic excited state, may transfer energy to an acceptor chromophore through nonradiative dipole–dipole coupling. The efficiency of this energy transfer is inversely proportional to the sixth power of the distance between donor and acceptor, making FRET extremely sensitive to small changes in distance.

<span class="mw-page-title-main">Franck–Condon principle</span> Quantum chemistry rule regarding vibronic transitions

The Franck–Condon principle is a rule in spectroscopy and quantum chemistry that explains the intensity of vibronic transitions. The principle states that during an electronic transition, a change from one vibrational energy level to another will be more likely to happen if the two vibrational wave functions overlap more significantly.

Fluorescence-lifetime imaging microscopy or FLIM is an imaging technique based on the differences in the exponential decay rate of the photon emission of a fluorophore from a sample. It can be used as an imaging technique in confocal microscopy, two-photon excitation microscopy, and multiphoton tomography.

In particle physics, the quantum yield of a radiation-induced process is the number of times a specific event occurs per photon absorbed by the system.

Fluorescence correlation spectroscopy (FCS) is a statistical analysis, via time correlation, of stationary fluctuations of the fluorescence intensity. Its theoretical underpinning originated from L. Onsager's regression hypothesis. The analysis provides kinetic parameters of the physical processes underlying the fluctuations. One of the interesting applications of this is an analysis of the concentration fluctuations of fluorescent particles (molecules) in solution. In this application, the fluorescence emitted from a very tiny space in solution containing a small number of fluorescent particles (molecules) is observed. The fluorescence intensity is fluctuating due to Brownian motion of the particles. In other words, the number of the particles in the sub-space defined by the optical system is randomly changing around the average number. The analysis gives the average number of fluorescent particles and average diffusion time, when the particle is passing through the space. Eventually, both the concentration and size of the particle (molecule) are determined. Both parameters are important in biochemical research, biophysics, and chemistry.

In MRI and NMR spectroscopy, an observable nuclear spin polarization (magnetization) is created by a homogeneous magnetic field. This field makes the magnetic dipole moments of the sample precess at the resonance (Larmor) frequency of the nuclei. At thermal equilibrium, nuclear spins precess randomly about the direction of the applied field. They become abruptly phase coherent when they are hit by radiofrequency (RF) pulses at the resonant frequency, created orthogonal to the field. The RF pulses cause the population of spin-states to be perturbed from their thermal equilibrium value. The generated transverse magnetization can then induce a signal in an RF coil that can be detected and amplified by an RF receiver. The return of the longitudinal component of the magnetization to its equilibrium value is termed spin-latticerelaxation while the loss of phase-coherence of the spins is termed spin-spin relaxation, which is manifest as an observed free induction decay (FID).

<span class="mw-page-title-main">Two-photon absorption</span> Simultaneous absorption of two photons by a molecule

In atomic physics, two-photon absorption, also called two-photon excitation or non-linear absorption, is the simultaneous absorption of two photons of identical or different frequencies in order to excite a molecule from one state to a higher energy, most commonly an excited electronic state. Absorption of two photons with different frequencies is called non-degenerate two-photon absorption. Since TPA depends on the simultaneous absorption of two photons, the probability of TPA is proportional to the square of the light intensity; thus it is a nonlinear optical process. The energy difference between the involved lower and upper states of the molecule is equal or smaller than the sum of the photon energies of the two photons absorbed. Two-photon absorption is a third-order process, with absorption cross section typically several orders of magnitude smaller than one-photon absorption cross section.

<span class="mw-page-title-main">Dynamic light scattering</span> Technique for determining size distribution of particles

Dynamic light scattering (DLS) is a technique in physics that can be used to determine the size distribution profile of small particles in suspension or polymers in solution. In the scope of DLS, temporal fluctuations are usually analyzed using the intensity or photon autocorrelation function. In the time domain analysis, the autocorrelation function (ACF) usually decays starting from zero delay time, and faster dynamics due to smaller particles lead to faster decorrelation of scattered intensity trace. It has been shown that the intensity ACF is the Fourier transform of the power spectrum, and therefore the DLS measurements can be equally well performed in the spectral domain. DLS can also be used to probe the behavior of complex fluids such as concentrated polymer solutions.

Resonance fluorescence is the process in which a two-level atom system interacts with the quantum electromagnetic field if the field is driven at a frequency near to the natural frequency of the atom.

<span class="mw-page-title-main">Fluorescence in the life sciences</span> Scientific investigative technique

Fluorescence is used in the life sciences generally as a non-destructive way of tracking or analysing biological molecules. Some proteins or small molecules in cells are naturally fluorescent, which is called intrinsic fluorescence or autofluorescence. Alternatively, specific or general proteins, nucleic acids, lipids or small molecules can be "labelled" with an extrinsic fluorophore, a fluorescent dye which can be a small molecule, protein or quantum dot. Several techniques exist to exploit additional properties of fluorophores, such as fluorescence resonance energy transfer, where the energy is passed non-radiatively to a particular neighbouring dye, allowing proximity or protein activation to be detected; another is the change in properties, such as intensity, of certain dyes depending on their environment allowing their use in structural studies.

Single-molecule fluorescence resonance energy transfer is a biophysical technique used to measure distances at the 1-10 nanometer scale in single molecules, typically biomolecules. It is an application of FRET wherein a pair of donor and acceptor fluorophores are excited and detected at a single molecule level. In contrast to "ensemble FRET" which provides the FRET signal of a high number of molecules, single-molecule FRET is able to resolve the FRET signal of each individual molecule. The variation of the smFRET signal is useful to reveal kinetic information that an ensemble measurement cannot provide, especially when the system is under equilibrium with no ensemble/bulk signal change. Heterogeneity among different molecules can also be observed. This method has been applied in many measurements of intramolecular dynamics such as DNA/RNA/protein folding/unfolding and other conformational changes, and intermolecular dynamics such as reaction, binding, adsorption, and desorption that are particularly useful in chemical sensing, bioassays, and biosensing.

Within the scientific study of a molecule's color, Fluorescence intensity decay shape microscopy (FIDSAM) is a fluorescence microscope technique, which utilizes the time evolution of fluorescence emission after a pulsed excitation to analyse the decay statistics of an excited chromophore. The main application of FIDSAM is the discrimination of unspecific autofluorescent background signal from the target signal of a dedicated chromophore.

Time-resolved fluorescence energy transfer (TR-FRET) is the practical combination of time-resolved fluorometry (TRF) with Förster resonance energy transfer (FRET) that offers a powerful tool for drug discovery researchers. TR-FRET combines the low background aspect of TRF with the homogeneous assay format of FRET. The resulting assay provides an increase in flexibility, reliability and sensitivity in addition to higher throughput and fewer false positive/false negative results. FRET involves two fluorophores, a donor and an acceptor. Excitation of the donor by an energy source produces an energy transfer to the acceptor if the two are within a given proximity to each other. The acceptor in turn emits light at its characteristic wavelength.

<span class="mw-page-title-main">Fluorescence polarization immunoassay</span> Class of invitro biochemical test

Fluorescence polarization immunoassay (FPIA) is a class of in vitro biochemical test used for rapid detection of antibody or antigen in sample. FPIA is a competitive homogenous assay, that consists of a simple prepare and read method, without the requirement of separation or washing steps.

<span class="mw-page-title-main">Fluorescence imaging</span> Type of non-invasive imaging technique

Fluorescence imaging is a type of non-invasive imaging technique that can help visualize biological processes taking place in a living organism. Images can be produced from a variety of methods including: microscopy, imaging probes, and spectroscopy.

<span class="mw-page-title-main">Non-degenerate two-photon absorption</span> Simultaneous absorption of two photons of differing energies by a molecule

In atomic physics, non-degenerate two-photon absorption or two-color two-photon excitation is a type of two-photon absorption (TPA) where two photons with different energies are (almost) simultaneously absorbed by a molecule, promoting a molecular electronic transition from a lower energy state to a higher energy state. The sum of the energies of the two photons is equal to, or larger than, the total energy of the transition.

References

  1. Weber, G., 1953. Rotational Brownian motion and polarization of the fluorescence of solutions. Adv. Protein Chem. 8:415-459
  2. Albrecht, A., 1961. Polarizations and assignments of transitions: the method of photoselection. J. Mol. Spectrosc. 6:84-108.
  3. 1 2 Lakowicz, J.R., 2006. Principles of Fluorescence Spectroscopy (3rd ed., Springer. Chapter 10-12 deal with fluorescence polarization spectroscopy.)
  4. Miller, J. N., ed. (1981). Standards in Fluorescence Spectrometry - Ultraviolet Spectrometry | J. Miller | Springer. Techniques in Visible and Ultraviolet Spectrometry. Springer. doi:10.1007/978-94-009-5902-6. ISBN   9789400959040 . Retrieved 2016-01-16.
  5. Birch, David J.S.; Yip, Philip (2014-01-01). "Nanometrology" (PDF). In Engelborghs, Yves; Visser, Antonie J.W.G. (eds.). Fluorescence Spectroscopy and Microscopy. Methods in Molecular Biology. Vol. 1076. Humana Press. pp. 279–302. doi:10.1007/978-1-62703-649-8_11. ISBN   978-1-62703-648-1. PMID   24108630.
  6. Valeur, Bernard. 2001. Molecular Fluorescence: Principles and Applications Wiley-VCH, p.29
  7. Birch, David J. S.; Imhof, Robert E. (2002-01-01). "Time-Domain Fluorescence Spectroscopy Using Time-Correlated Single-Photon Counting". In Lakowicz, Joseph R. (ed.). Topics in Fluorescence Spectroscopy. Vol. 1. Springer US. pp. 1–95. doi:10.1007/0-306-47057-8_1. ISBN   978-0-306-43874-5.
  8. Heckmeier, Philipp J.; Agam, Ganesh; Teese, Mark G.; Hoyer, Maria; Stehle, Ralf; Lamb, Don C.; Langosch, Dieter (July 2020). "Determining the Stoichiometry of Small Protein Oligomers Using Steady-State Fluorescence Anisotropy". Biophysical Journal. 119 (1): 99–114. doi: 10.1016/j.bpj.2020.05.025 . PMC   7335908 .