Photon counting

Last updated
A prototype single-photon detector that was used on the 200-inch Hale Telescope. The Hubble Space Telescope has a similar detector. Prototype photon counting system, c 1980s. (9660571969).jpg
A prototype single-photon detector that was used on the 200-inch Hale Telescope. The Hubble Space Telescope has a similar detector.

Photon counting is a technique in which individual photons are counted using a single-photon detector (SPD). A single-photon detector emits a pulse of signal for each detected photon. The counting efficiency is determined by the quantum efficiency and the system's electronic losses.

Contents

Many photodetectors can be configured to detect individual photons, each with relative advantages and disadvantages. [1] [2] Common types include photomultipliers, geiger counters, single-photon avalanche diodes, superconducting nanowire single-photon detectors, transition edge sensors, and scintillation counters. Charge-coupled devices can be used.

Advantages

Photon counting eliminates gain noise, where the proportionality constant between analog signal out and number of photons varies randomly. Thus, the excess noise factor of a photon-counting detector is unity, and the achievable signal-to-noise ratio for a fixed number of photons is generally higher than the same detector without photon counting. [3]

Photon counting can improve temporal resolution. In a conventional detector, multiple arriving photons generate overlapping impulse responses, limiting temporal resolution to approximately the fall time of the detector. However, if it is known that a single photon was detected, the center of the impulse response can be evaluated to precisely determine its arrival time. Using time-correlated single-photon counting (TCSPC), temporal resolution of less than 25 ps has been demonstrated using detectors with a fall time more than 20 times greater. [4]

Disadvantages

Single-photon detectors are typically limited to detecting one photon at a time and may require time between detection events to reset. Photons that arrive during this interval may not be detected. Therefore, the maximum light intensity that can be accurately measured is typically low. Measurements composed of small numbers of photons intrinsically have a low signal-to-noise ratio caused by the randomly varying numbers of emitted photons. This effect is less pronounced in conventional detectors that can concurrently detect large numbers of photons. Because of the lower maximum signal level, either the signal-to-noise ratio will be lower or the exposure time longer than for conventional detection.

Applications

Single-photon detection is useful in fields such as: [1]

Medicine

In radiology, one of the major disadvantages of X-ray imaging modalities is the negative effects of ionising radiation. Although the risk from small exposures (as used in most medical imaging) is thought to be small, the radiation protection principle of "as low as reasonably practicable" (ALARP) is always applied. One way of reducing exposures is to make X-ray detectors as efficient as possible, so that lower doses can be used for a given diagnostic image quality. Photon counting detectors could help, due to their ability to reject noise more easily. [5] [6] Photon counting is analogous to color photography, where each photon's differing energy affects the output, as compared to charge integration, which considers only the intensity of the signal, as in black and white photography. [7]

Photon-counting mammography was introduced commercially in 2003. Although such systems are not widespread, some evidence supports their ability to produce comparable images at an approximately 40% lower dose than other digital mammography systems with flat panel detectors. [8] [9] Spectral imaging technology was subsequently developed to discriminate between photon energies, [10] [6] with the possibility to further improve image quality [11] and to distinguish tissue types. [12] Photon-counting computed tomography is another interest area, which is rapidly evolving and is approaching clinical feasibility. [13] [14] [15] [16]

Fluorescence-lifetime imaging microscopy

Time-correlated single-photon counting (TCSPC) precisely records the arrival times of individual photons, enabling measurement of picosecond time-scale differences in the arrival times of photons generated by fluorescent, phosphorescence or other chemical processes that emit light, providing additional molecular information about samples. The use of TCSPC enables relatively slow detectors to measure extremely minute time differences that would be obscured by overlapping impulse responses if multiple photons were incident concurrently.

LIDAR

Some pulse LIDAR systems operate in single photon counting mode using TCSPC to achieve higher resolution. Infrared photon-counting technologies for LIDAR are advancing rapidly. [17]

Measured quantities

The number of photons observed per unit time is the photon flux. The photon flux per unit area is the photon irradiance if the photons are incident on a surface, or photon exitance if the emission of photons from a broad-area source is being considered. The flux per unit solid angle is the photon intensity. The flux per unit source area per unit solid angle is photon radiance. SI units for these quantities are summarized in the table below.

QuantityUnitDimensionNotes
NameSymbol [nb 1] NameSymbol
Photon energy n1count of photons n with energy Qp = hc. [nb 2]
Photon flux Φqcount per second s −1T−1photons per unit time, dn/d t with n = photon number.
also called photon power.
Photon intensity Icount per steradian per second sr −1⋅s−1T−1dn/d ω
Photon radiance Lqcount per square metre per steradian per second m−2⋅sr−1⋅s−1L−2T−1d2n/(d A cos(θ) dω)
Photon irradiance Eqcount per square metre per secondm−2⋅s−1L−2T−1dn/dA
Photon exitance Mcount per square metre per secondm−2⋅s−1L−2T−1dn/dA
See also:
  1. Standards organizations recommend that photon quantities be denoted with a suffix "q" (for "quantum") to avoid confusion with radiometric and photometric quantities.
  2. The energy of a single photon at wavelength λ is Qp = h⋅c/λ with h = Planck's constant and c = velocity of light.

See also

Related Research Articles

<span class="mw-page-title-main">CT scan</span> Medical imaging procedure using X-rays to produce cross-sectional images

A computed tomography scan is a medical imaging technique used to obtain detailed internal images of the body. The personnel that perform CT scans are called radiographers or radiology technologists.

<span class="mw-page-title-main">Radiography</span> Imaging technique using ionizing and non-ionizing radiation

Radiography is an imaging technique using X-rays, gamma rays, or similar ionizing radiation and non-ionizing radiation to view the internal form of an object. Applications of radiography include medical and industrial radiography. Similar techniques are used in airport security,. To create an image in conventional radiography, a beam of X-rays is produced by an X-ray generator and it is projected towards the object. A certain amount of the X-rays or other radiation are absorbed by the object, dependent on the object's density and structural composition. The X-rays that pass through the object are captured behind the object by a detector. The generation of flat two-dimensional images by this technique is called projectional radiography. In computed tomography, an X-ray source and its associated detectors rotate around the subject, which itself moves through the conical X-ray beam produced. Any given point within the subject is crossed from many directions by many different beams at different times. Information regarding the attenuation of these beams is collated and subjected to computation to generate two-dimensional images on three planes which can be further processed to produce a three-dimensional image.

<span class="mw-page-title-main">Mammography</span> Process of using low-energy X-rays to examine the human breast for diagnosis and screening

Mammography is the process of using low-energy X-rays to examine the human breast for diagnosis and screening. The goal of mammography is the early detection of breast cancer, typically through detection of characteristic masses or microcalcifications.

In experimental and applied particle physics, nuclear physics, and nuclear engineering, a particle detector, also known as a radiation detector, is a device used to detect, track, and/or identify ionizing particles, such as those produced by nuclear decay, cosmic radiation, or reactions in a particle accelerator. Detectors can measure the particle energy and other attributes such as momentum, spin, charge, particle type, in addition to merely registering the presence of the particle.

<span class="mw-page-title-main">Photodetector</span> Sensors of light or other electromagnetic energy

Photodetectors, also called photosensors, are sensors of light or other electromagnetic radiation. There are a wide variety of photodetectors which may be classified by mechanism of detection, such as photoelectric or photochemical effects, or by various performance metrics, such as spectral response. Semiconductor-based photodetectors typically use a p–n junction that converts photons into charge. The absorbed photons make electron–hole pairs in the depletion region. Photodiodes and photo transistors are a few examples of photo detectors. Solar cells convert some of the light energy absorbed into electrical energy.

<span class="mw-page-title-main">PILATUS (detector)</span>

PILATUS is the name of a series of x-ray detectors originally developed by the Paul Scherrer Institute at the Swiss Light Source and further developed and commercialized by DECTRIS. The PILATUS detectors are based on hybrid photon counting (HPC) technology, by which X-rays are converted to electrical signals by the photoelectric effect in a semiconductor sensor layer—either silicon or cadmium telluride—which is subject to a substantial bias voltage. The electric signals are counted directly by a series of cells in an ASIC bonded to the sensor. Each cell—or pixel—is a complete detector in itself, equipped with an amplifier, discriminator and counter circuit. This is possible thanks to contemporary CMOS integrated circuit technology.

Quantum imaging is a new sub-field of quantum optics that exploits quantum correlations such as quantum entanglement of the electromagnetic field in order to image objects with a resolution or other imaging criteria that is beyond what is possible in classical optics. Examples of quantum imaging are quantum ghost imaging, quantum lithography, imaging with undetected photons, sub-shot-noise imaging, and quantum sensing. Quantum imaging may someday be useful for storing patterns of data in quantum computers and transmitting large amounts of highly secure encrypted information. Quantum mechanics has shown that light has inherent “uncertainties” in its features, manifested as moment-to-moment fluctuations in its properties. Controlling these fluctuations—which represent a sort of “noise”—can improve detection of faint objects, produce better amplified images, and allow workers to more accurately position laser beams.

<span class="mw-page-title-main">Tomosynthesis</span>

Tomosynthesis, also digital tomosynthesis (DTS), is a method for performing high-resolution limited-angle tomography at radiation dose levels comparable with projectional radiography. It has been studied for a variety of clinical applications, including vascular imaging, dental imaging, orthopedic imaging, mammographic imaging, musculoskeletal imaging, and chest imaging.

<span class="mw-page-title-main">Automatic exposure control</span>

Automatic Exposure Control (AEC) is an X-ray exposure termination device. A medical radiographic exposure is always initiated by a human operator but an AEC detector system may be used to terminate the exposure when a predetermined amount of radiation has been received. The intention of AEC is to provide consistent x-ray image exposure, whether to film, a digital detector or a CT scanner. AEC systems may also automatically set exposure factors such as the X-ray tube current and voltage in a CT.

The detective quantum efficiency is a measure of the combined effects of the signal and noise performance of an imaging system, generally expressed as a function of spatial frequency. This value is used primarily to describe imaging detectors in optical imaging and medical radiography.

<span class="mw-page-title-main">Flat-panel detector</span> Class of solid-state x-ray digital radiography devices

Flat-panel detectors are a class of solid-state x-ray digital radiography devices similar in principle to the image sensors used in digital photography and video. They are used in both projectional radiography and as an alternative to x-ray image intensifiers (IIs) in fluoroscopy equipment.

<span class="mw-page-title-main">Phase-contrast X-ray imaging</span> Imaging systems using changes in phase

Phase-contrast X-ray imaging or phase-sensitive X-ray imaging is a general term for different technical methods that use information concerning changes in the phase of an X-ray beam that passes through an object in order to create its images. Standard X-ray imaging techniques like radiography or computed tomography (CT) rely on a decrease of the X-ray beam's intensity (attenuation) when traversing the sample, which can be measured directly with the assistance of an X-ray detector. However, in phase contrast X-ray imaging, the beam's phase shift caused by the sample is not measured directly, but is transformed into variations in intensity, which then can be recorded by the detector.

<span class="mw-page-title-main">X-ray detector</span> Instrument that can measure properties of X-rays

X-ray detectors are devices used to measure the flux, spatial distribution, spectrum, and/or other properties of X-rays.

<span class="mw-page-title-main">Breast imaging</span>

In medicine, breast imaging is a sub-speciality of diagnostic radiology that involves imaging of the breasts for screening or diagnostic purposes. There are various methods of breast imaging using a variety of technologies as described in detail below. Traditional screening and diagnostic mammography uses x-ray technology and has been the mainstay of breast imaging for many decades. Breast tomosynthesis is a relatively new digital x-ray mammography technique that produces multiple image slices of the breast similar to, but distinct from, computed tomography (CT). Xeromammography and galactography are somewhat outdated technologies that also use x-ray technology and are now used infrequently in the detection of breast cancer. Breast ultrasound is another technology employed in diagnosis and screening that can help differentiate between fluid filled and solid lesions, an important factor to determine if a lesion may be cancerous. Breast MRI is a technology typically reserved for high-risk patients and patients recently diagnosed with breast cancer. Lastly, scintimammography is used in a subgroup of patients who have abnormal mammograms or whose screening is not reliable on the basis of using traditional mammography or ultrasound.

<span class="mw-page-title-main">Kinetic imaging</span>

Kinetic imaging is an imaging technology developed by Szabolcs Osváth and Krisztián Szigeti in the Department of Biophysics and Radiation Biology at Semmelweis University. The technology allows the visualization of motion; it is based on an altered data acquisition and image processing algorithm combined with imaging techniques that use penetrating radiation. Kinetic imaging has the potential for use in a wide variety of areas including medicine, engineering, and surveillance. For example, physiological movements, such as the circulation of blood or motion of organs can be visualized using kinetic imaging. Because of the reduced noise and the motion-based image contrast, kinetic imaging can be used to reduce X-ray dose and/or amount of required contrast agent in medical imaging. In fact, clinical trials are underway in the fields of vascular surgery and interventional radiology. Non-medical applications include non-destructive testing of products and port security scanning for stowaway pests.

Jeffrey Harold Siewerdsen is an American physicist and biomedical engineer who is a Professor of Imaging Physics at The University of Texas MD Anderson Cancer Center as well as Biomedical Engineering, Computer Science, Radiology, and Neurosurgery at Johns Hopkins University.He is among the original inventors of cone-beam CT-guided radiotherapy as well as weight-bearing cone-beam CT for musculoskeletal radiology and orthopedic surgery. His work also includes the early development of flat-panel detectors on mobile C-arms for intraoperative cone-beam CT in image-guided surgery. He developed early models for the signal and noise performance of flat-panel detectors and later extended such analysis to dual-energy imaging and 3D imaging performance in cone-beam CT. He founded the ISTAR Lab in the Department of Biomedical Engineering, the Carnegie Center for Surgical Innovation at Johns Hopkins Hospital, and the Surgical Data Science Program at the Institute for Data Science in Oncology at The University of Texas MD Anderson Cancer Center.

Photon-counting computed tomography (PCCT) is a form of X-ray computed tomography (CT) in which X-rays are detected using a photon-counting detector (PCD) which registers the interactions of individual photons. By keeping track of the deposited energy in each interaction, the detector pixels of a PCD each record an approximate energy spectrum, making it a spectral or energy-resolved CT technique. In contrast, more conventional CT scanners use energy-integrating detectors (EIDs), where the total energy deposited in a pixel during a fixed period of time is registered. These EIDs thus register only photon intensity, comparable to black-and-white photography, whereas PCDs register also spectral information, similar to color photography.

Hybrid pixel detectors are a type of ionizing radiation detector consisting of an array of diodes based on semiconductor technology and their associated electronics. The term “hybrid” stems from the fact that the two main elements from which these devices are built, the semiconductor sensor and the readout chip, are manufactured independently and later electrically coupled by means of a bump-bonding process. Ionizing particles are detected as they produce electron-hole pairs through their interaction with the sensor element, usually made of doped silicon or cadmium telluride. The readout ASIC is segmented into pixels containing the necessary electronics to amplify and measure the electrical signals induced by the incoming particles in the sensor layer.

Spectral imaging is an umbrella term for energy-resolved X-ray imaging in medicine. The technique makes use of the energy dependence of X-ray attenuation to either increase the contrast-to-noise ratio, or to provide quantitative image data and reduce image artefacts by so-called material decomposition. Dual-energy imaging, i.e. imaging at two energy levels, is a special case of spectral imaging and is still the most widely used terminology, but the terms "spectral imaging" and "spectral CT" have been coined to acknowledge the fact that photon-counting detectors have the potential for measurements at a larger number of energy levels.

Photon-counting mammography was introduced commercially in 2003 and was the first widely available application of photon-counting detector technology in medical x-ray imaging. Photon-counting mammography improves dose efficiency compared to conventional technologies, and enables spectral imaging.

References

  1. 1 2 "High Efficiency in the Fastest Single-Photon Detector System" (Press release). National Institute of Standards and Technology. February 19, 2013. Retrieved 2018-10-11.
  2. Hadfield, RH (2009). "Single-photon detectors for optical quantum information applications". Nature Photonics. 3 (12): 696. Bibcode:2009NaPho...3..696H. doi:10.1038/nphoton.2009.230.
  3. K.K, Hamamatsu Photonics. "Detection Questions & Answers". hub.hamamatsu.com. Retrieved 2020-08-14.
  4. "Fast-Acquisition TCSPC FLIM System with sub-25 ps IRF Width" (PDF). Becker and Hickl. Retrieved 17 August 2020.
  5. Shikhaliev, M (2015). "Medical X-ray and CT Imaging with Photon-Counting Detectors". In Iwanczyk, Jan S. (ed.). Radiation Detectors for Medical Imaging. Boca Raton, FL: CRC Press. pp. 2–21. ISBN   9781498766821.
  6. 1 2 Taguchi, Katsuyuki; Iwanczyk, Jan S. (12 September 2013). "Vision 20/20: Single photon counting x-ray detectors in medical imaging". Medical Physics. 40 (10): 100901. Bibcode:2013MedPh..40j0901T. doi:10.1118/1.4820371. PMC   3786515 . PMID   24089889.
  7. "Photon Counting Explained". Direct Conversion. Retrieved 2022-02-10.
  8. McCullagh, J B; Baldelli, P; Phelan, N (November 2011). "Clinical dose performance of full field digital mammography in a breast screening programme". The British Journal of Radiology. 84 (1007): 1027–1033. doi:10.1259/bjr/83821596. PMC   3473710 . PMID   21586506.
  9. Weigel, Stefanie; Berkemeyer, Shoma; Girnus, Ralf; Sommer, Alexander; Lenzen, Horst; Heindel, Walter (May 2014). "Digital Mammography Screening with Photon-counting Technique: Can a High Diagnostic Performance Be Realized at Low Mean Glandular Dose?". Radiology. 271 (2): 345–355. doi: 10.1148/radiol.13131181 . PMID   24495234.
  10. Iwanczyk, Jan S; Barber, W C; Nygård, Einar; Malakhov, Nail; Hartsough, N E; Wessel, J C (2018). "Photon-Counting Energy-Dispersive Detector Arrays for X-Ray Imaging". In Iniewski, Krzysztof (ed.). Electronics for Radiation Detection. CRC Press. ISBN   9781439858844.
  11. Berglund, Johan; Johansson, Henrik; Lundqvist, Mats; Cederström, Björn; Fredenberg, Erik (2014-08-28). "Energy weighting improves dose efficiency in clinical practice: implementation on a spectral photon-counting mammography system". Journal of Medical Imaging. 1 (3): 031003. doi:10.1117/1.JMI.1.3.031003. ISSN   2329-4302. PMC   4478791 . PMID   26158045.
  12. Fredenberg, Erik; Willsher, Paula; Moa, Elin; Dance, David R; Young, Kenneth C; Wallis, Matthew G (2018-11-22). "Measurement of breast-tissue x-ray attenuation by spectral imaging: fresh and fixed normal and malignant tissue". Physics in Medicine & Biology. 63 (23): 235003. arXiv: 2101.02755 . Bibcode:2018PMB....63w5003F. doi:10.1088/1361-6560/aaea83. ISSN   1361-6560. PMID   30465547. S2CID   53717425.
  13. Yveborg, Moa; Xu, Cheng; Fredenberg, Erik; Danielsson, Mats (2009-02-26). "Photon-counting CT with silicon detectors: feasibility for pediatric imaging". In Samei, Ehsan; Hsieh, Jiang (eds.). Medical Imaging 2009: Physics of Medical Imaging. Vol. 7258. Lake Buena Vista, FL. pp. 704–709. arXiv: 2101.09439 . doi:10.1117/12.813733. S2CID   120218867.{{cite book}}: CS1 maint: location missing publisher (link)
  14. Pourmorteza, Amir; Symons, Rolf; Sandfort, Veit; Mallek, Marissa; Fuld, Matthew K.; Henderson, Gregory; Jones, Elizabeth C.; Malayeri, Ashkan A.; Folio, Les R.; Bluemke, David A. (April 2016). "Abdominal Imaging with Contrast-enhanced Photon-counting CT: First Human Experience". Radiology. 279 (1): 239–245. doi:10.1148/radiol.2016152601. ISSN   0033-8419. PMC   4820083 . PMID   26840654.
  15. "First 3D colour X-ray of a human using CERN technology". CERN. Retrieved 2020-11-23.
  16. "New 3D colour X-rays made possible with CERN technology". CERN. Retrieved 2020-11-23.
  17. Hadfield, Robert H.; Leach, Jonathan; Fleming, Fiona; Paul, Douglas J.; Tan, Chee Hing; Ng, Jo Shien; Henderson, Robert K.; Buller, Gerald S. (2023). "Single-photon detection for long-range imaging and sensing". Optica. 10 (9): 1124. doi: 10.1364/optica.488853 . hdl: 20.500.11820/4d60bb02-3c2c-4f86-a737-f985cb8613d8 . S2CID   259687483 . Retrieved 2023-08-29.