Profilometer

Last updated
Non-Contact Optical Profilometer NANOVEA PS50 Non-Contact Optical Profilometer.jpg
Non-Contact Optical Profilometer
A contact profilometer at LAAS technological facility in Toulouse, France. Contact profilometer at LAAS 0419.jpg
A contact profilometer at LAAS technological facility in Toulouse, France.

A profilometer is a measuring instrument used to measure a surface's profile, in order to quantify its roughness. Critical dimensions as step, curvature, flatness are computed from the surface topography.

Contents

While the historical notion of a profilometer was a device similar to a phonograph that measures a surface as the surface is moved relative to the contact profilometer's stylus, this notion is changing with the emergence of numerous non-contact profilometry techniques.

Non-scanning technologies measure the surface topography within a single camera acquisition, XYZ scanning is no longer needed. As a consequence, dynamic changes of topography are measured in real-time. Contemporary profilometers are not only measuring static topography, but now also dynamic topography – such systems are described as time-resolved profilometers.

Types

Original 1940s Taylor-Hobson Talysurf surface profile measuring machine Talysurf, Snibston.jpg
Original 1940s Taylor-Hobson Talysurf surface profile measuring machine

Optical methods [1] [2] include interferometry based methods such as digital holographic microscopy, vertical scanning interferometry/white light interferometry, phase shifting interferometry, and differential interference contrast microscopy (Nomarski microscopy); focus detection methods such as intensity detection, focus variation, differential detection, critical angle method, astigmatic method, Foucault method, and confocal microscopy; pattern projection methods such as fringe projection, Fourier profilometry, Moire, and pattern reflection methods.

Contact and pseudo-contact methods [1] [2] include stylus profilometer (mechanical profilometer), [3] atomic force microscopy, [4] and scanning tunneling microscopy

Contact profilometers

A diamond stylus is moved vertically in contact with a sample and then moved laterally across the sample for a specified distance and specified contact force. A profilometer can measure small surface variations in vertical stylus displacement as a function of position. A typical profilometer can measure small vertical features ranging in height from 10 nanometres to 1 millimetre. The height position of the diamond stylus generates an analog signal which is converted into a digital signal, stored, analyzed, and displayed. The radius of diamond stylus ranges from 20 nanometres to 50 μm, and the horizontal resolution is controlled by the scan speed and data signal sampling rate. The stylus tracking force can range from less than 1 to 50 milligrams.

Advantages of contact profilometers include acceptance, surface independence, resolution, it is a direct technique with no modeling required. Most of the world's surface finish standards are written for contact profilometers. To follow the prescribed methodology, this type of profilometer is often required. Contacting the surface is often an advantage in dirty environments where non-contact methods can end up measuring surface contaminants instead of the surface itself. Because the stylus is in contact with the surface, this method is not sensitive to surface reflectance or color. The stylus tip radius can be as small as 20 nanometres, significantly better than white-light optical profiling. Vertical resolution is typically sub-nanometer as well.

Non-contact profilometers

An optical profilometer is a non-contact method for providing much of the same information as a stylus based profilometer. There are many different techniques which are currently being employed, such as laser triangulation (triangulation sensor), confocal microscopy (used for profiling very small objects), coherence scanning interferometry, and digital holography.

Advantages of optical profilometers are speed, reliability and spot size. For small steps and requirements to do 3D scanning, because the non-contact profilometer does not touch the surface the scan speeds are dictated by the light reflected from the surface and the speed of the acquisition electronics. For doing large steps, a 3D scan on an optical profiler can be much slower than a 2D scan on a stylus profiler. Optical profilometers do not touch the surface and therefore cannot be damaged by surface wear or careless operators. Many non-contact profilometers are solid-state which tends to reduce the required maintenance significantly. The spot size, or lateral resolution, of optical methods ranges from a few micrometres down to sub micrometre.

Time-resolved profilometers

Self-Healing-Polymer from Tosoh Corporation (Japan) measured by Digital Holographic Microscope Self-Healing-Polymer-DHM-Digital-Holographic-Microscopy-lyncee-Tosoh-Corporation.gif
Self-Healing-Polymer from Tosoh Corporation (Japan) measured by Digital Holographic Microscope
MEMS Ultrasonic-Transducers measured at 8 MHz in stroboscopic mode Ultrasonic-Transducers-MUT-IPMS-Digital-Holographic-Microscopy.gif
MEMS Ultrasonic-Transducers measured at 8 MHz in stroboscopic mode

Non-scanning technologies as digital holographic microscopy enable 3D topography measurement in real-time. 3D topography is measured from a single camera acquisition as a consequence the acquisition rate is only limited by the camera acquisition rate, some systems measure topography at a frame rate of 1000 fps. Time-resolved systems enable measurement of topography changes as healing of smart materials or measurement of moving specimens. Time-resolved profilometers can be combined with a stroboscopic unit to measure MEMS vibrations in the MHz range. The stroboscopic unit provides excitation signal to the MEMS and provides trigger signal to light source and camera.

The advantage of time-resolved profilometers is that they are robust against vibrations. Unlike scanning methods, time-resolved profilometer acquisition time is in the milliseconds range. There is no need of vertical calibration: vertical measurement does not depend on a scanning mechanism, digital holographic microscopy vertical measurement has an intrinsic vertical calibration based on laser source wavelength. Samples are not static and there is response of the specimen topography to external stimulus. With on-flight measurement the topography of a moving sample is acquired with short exposure time. MEMS vibrations measurement can be accomplished when the system is combined with a stroboscopic unit.

Fiber-based optical profilometers

Optical fiber-based optical profilometers scan surfaces with optical probes which send light interference signals back to the profilometer detector via an optical fiber. Fiber-based probes can be physically located hundreds of meters away from the detector enclosure, without signal degradation. The additional advantages of using fiber-based optical profilometers are flexibility, long profile acquisition, ruggedness, and ease of incorporating into industrial processes. With the small diameter of certain probes, surfaces can be scanned even inside hard-to-reach spaces, such as narrow crevices or small-diameter tubes. [5]

Because these probes generally acquire one point at a time and at high sample speeds, acquisition of long (continuous) surface profiles is possible. Scanning can take place in hostile environments, including very hot or cryogenic temperatures, or in radioactive chambers, while the detector is located at a distance, in a human-safe environment. [6] Fiber-based probes are easily installed in-process, such as above moving webs or mounted onto a variety of positioning systems.

Applications

A furrow profilometer is used for the measurement of the cross-sectional geometry of furrows and corrugations, and is important in furrow assessments. [7]

See also

Related Research Articles

<span class="mw-page-title-main">Microscopy</span> Viewing of objects which are too small to be seen with the naked eye

Microscopy is the technical field of using microscopes to view objects and areas of objects that cannot be seen with the naked eye. There are three well-known branches of microscopy: optical, electron, and scanning probe microscopy, along with the emerging field of X-ray microscopy.

<span class="mw-page-title-main">Interferometry</span> Measurement method using interference of waves

Interferometry is a technique which uses the interference of superimposed waves to extract information. Interferometry typically uses electromagnetic waves and is an important investigative technique in the fields of astronomy, fiber optics, engineering metrology, optical metrology, oceanography, seismology, spectroscopy, quantum mechanics, nuclear and particle physics, plasma physics, biomolecular interactions, surface profiling, microfluidics, mechanical stress/strain measurement, velocimetry, optometry, and making holograms.

<span class="mw-page-title-main">Atomic force microscopy</span> Type of microscopy

Atomic force microscopy (AFM) or scanning force microscopy (SFM) is a very-high-resolution type of scanning probe microscopy (SPM), with demonstrated resolution on the order of fractions of a nanometer, more than 1000 times better than the optical diffraction limit.

<span class="mw-page-title-main">Kelvin probe force microscope</span> Noncontact variant of atomic force microscopy

Kelvin probe force microscopy (KPFM), also known as surface potential microscopy, is a noncontact variant of atomic force microscopy (AFM). By raster scanning in the x,y plane the work function of the sample can be locally mapped for correlation with sample features. When there is little or no magnification, this approach can be described as using a scanning Kelvin probe (SKP). These techniques are predominantly used to measure corrosion and coatings.

Surface metrology is the measurement of small-scale features on surfaces, and is a branch of metrology. Surface primary form, surface fractality, and surface finish are the parameters most commonly associated with the field. It is important to many disciplines and is mostly known for the machining of precision parts and assemblies which contain mating surfaces or which must operate with high internal pressures.

<span class="mw-page-title-main">Near-field scanning optical microscope</span>

Near-field scanning optical microscopy (NSOM) or scanning near-field optical microscopy (SNOM) is a microscopy technique for nanostructure investigation that breaks the far field resolution limit by exploiting the properties of evanescent waves. In SNOM, the excitation laser light is focused through an aperture with a diameter smaller than the excitation wavelength, resulting in an evanescent field on the far side of the aperture. When the sample is scanned at a small distance below the aperture, the optical resolution of transmitted or reflected light is limited only by the diameter of the aperture. In particular, lateral resolution of 6 nm and vertical resolution of 2–5 nm have been demonstrated.

ISO 25178: Geometrical Product Specifications (GPS) – Surface texture: areal is an International Organization for Standardization collection of international standards relating to the analysis of 3D areal surface texture.

Angle-resolved low-coherence interferometry (a/LCI) is an emerging biomedical imaging technology which uses the properties of scattered light to measure the average size of cell structures, including cell nuclei. The technology shows promise as a clinical tool for in situ detection of dysplastic, or precancerous tissue.

<span class="mw-page-title-main">White light scanner</span>

A white light scanner (WLS) is a device for performing surface height measurements of an object using coherence scanning interferometry (CSI) with spectrally-broadband, "white light" illumination. Different configurations of scanning interferometer may be used to measure macroscopic objects with surface profiles measuring in the centimeter range, to microscopic objects with surface profiles measuring in the micrometer range. For large-scale non-interferometric measurement systems, see structured-light 3D scanner.

<span class="mw-page-title-main">Piezoresponse force microscopy</span> Microscopy technique for piezoelectric materials

Piezoresponse force microscopy (PFM) is a variant of atomic force microscopy (AFM) that allows imaging and manipulation of piezoelectric/ferroelectric materials domains. This is achieved by bringing a sharp conductive probe into contact with a ferroelectric surface and applying an alternating current (AC) bias to the probe tip in order to excite deformation of the sample through the converse piezoelectric effect (CPE). The resulting deflection of the probe cantilever is detected through standard split photodiode detector methods and then demodulated by use of a lock-in amplifier (LiA). In this way topography and ferroelectric domains can be imaged simultaneously with high resolution.

<span class="mw-page-title-main">Digital holographic microscopy</span>

Digital holographic microscopy (DHM) is digital holography applied to microscopy. Digital holographic microscopy distinguishes itself from other microscopy methods by not recording the projected image of the object. Instead, the light wave front information originating from the object is digitally recorded as a hologram, from which a computer calculates the object image by using a numerical reconstruction algorithm. The image forming lens in traditional microscopy is thus replaced by a computer algorithm. Other closely related microscopy methods to digital holographic microscopy are interferometric microscopy, optical coherence tomography and diffraction phase microscopy. Common to all methods is the use of a reference wave front to obtain amplitude (intensity) and phase information. The information is recorded on a digital image sensor or by a photodetector from which an image of the object is created (reconstructed) by a computer. In traditional microscopy, which do not use a reference wave front, only intensity information is recorded and essential information about the object is lost.

Novacam Technologies Inc. specializes in designing and manufacturing advanced metrology and imaging systems for industrial and bio-medical applications. Novacam's fiber-based optical profilometers and Optical Coherence Tomography (OCT) systems are based on low coherence interferometry. The fiber-based nature of Novacam's detector probes is unique in the optical metrology industry.

The technique of vibrational analysis with scanning probe microscopy allows probing vibrational properties of materials at the submicrometer scale, and even of individual molecules. This is accomplished by integrating scanning probe microscopy (SPM) and vibrational spectroscopy. This combination allows for much higher spatial resolution than can be achieved with conventional Raman/FTIR instrumentation. The technique is also nondestructive, requires non-extensive sample preparation, and provides more contrast such as intensity contrast, polarization contrast and wavelength contrast, as well as providing specific chemical information and topography images simultaneously.

Scanning electrochemical microscopy (SECM) is a technique within the broader class of scanning probe microscopy (SPM) that is used to measure the local electrochemical behavior of liquid/solid, liquid/gas and liquid/liquid interfaces. Initial characterization of the technique was credited to University of Texas electrochemist, Allen J. Bard, in 1989. Since then, the theoretical underpinnings have matured to allow widespread use of the technique in chemistry, biology and materials science. Spatially resolved electrochemical signals can be acquired by measuring the current at an ultramicroelectrode (UME) tip as a function of precise tip position over a substrate region of interest. Interpretation of the SECM signal is based on the concept of diffusion-limited current. Two-dimensional raster scan information can be compiled to generate images of surface reactivity and chemical kinetics.

<span class="mw-page-title-main">White light interferometry</span> Measurement technique

As described here, white light interferometry is a non-contact optical method for surface height measurement on 3D structures with surface profiles varying between tens of nanometers and a few centimeters. It is often used as an alternative name for coherence scanning interferometry in the context of areal surface topography instrumentation that relies on spectrally-broadband, visible-wavelength light.

<span class="mw-page-title-main">Colloidal probe technique</span>

The colloidal probe technique is commonly used to measure interaction forces acting between colloidal particles and/or planar surfaces in air or in solution. This technique relies on the use of an atomic force microscope (AFM). However, instead of a cantilever with a sharp AFM tip, one uses the colloidal probe. The colloidal probe consists of a colloidal particle of few micrometers in diameter that is attached to an AFM cantilever. The colloidal probe technique can be used in the sphere-plane or sphere-sphere geometries. One typically achieves a force resolution between 1 and 100 pN and a distance resolution between 0.5 and 2 nm.

<span class="mw-page-title-main">Coherence scanning interferometry</span>

Coherence scanning interferometry (CSI) is any of a class of optical surface measurement methods wherein the localization of interference fringes during a scan of optical path length provides a means to determine surface characteristics such as topography, transparent film structure, and optical properties. CSI is currently the most common interference microscopy technique for areal surface topography measurement. The term "CSI" was adopted by the International Organization for Standardization (ISO).

The operation of a photon scanning tunneling microscope (PSTM) is analogous to the operation of an electron scanning tunneling microscope, with the primary distinction being that PSTM involves tunneling of photons instead of electrons from the sample surface to the probe tip. A beam of light is focused on a prism at an angle greater than the critical angle of the refractive medium in order to induce total internal reflection within the prism. Although the beam of light is not propagated through the surface of the refractive prism under total internal reflection, an evanescent field of light is still present at the surface.

<span class="mw-page-title-main">Multi-tip scanning tunneling microscopy</span>

Multi-tip scanning tunneling microscopy extends scanning tunneling microscopy (STM) from imaging to dedicated electrical measurements at the nanoscale like a ″multimeter at the nanoscale″. In materials science, nanoscience, and nanotechnology, it is desirable to measure electrical properties at a particular position of the sample. For this purpose, multi-tip STMs in which several tips are operated independently have been developed. Apart from imaging the sample, the tips of a multi-tip STM are used to form contacts to the sample at desired locations and to perform local electrical measurements.

Scanning quantum dot microscopy (SQDM) is a scanning probe microscopy (SPM) that is used to image nanoscale electric potential distributions on surfaces. The method quantifies surface potential variations via their influence on the potential of a quantum dot (QD) attached to the apex of the scanned probe. SQDM allows, for example, the quantification of surface dipoles originating from individual adatoms, molecules, or nanostructures. This gives insights into surface and interface mechanisms such as reconstruction or relaxation, mechanical distortion, charge transfer and chemical interaction. Measuring electric potential distributions is also relevant for characterizing organic and inorganic semiconductor devices which feature electric dipole layers at the relevant interfaces. The probe to surface distance in SQDM ranges from 2 nm to 10 nm and therefore allows imaging on non-planar surfaces or, e.g., of biomolecules with a distinct 3D structure. Related imaging techniques are Kelvin Probe Force Microscopy (KPFM) and Electrostatic Force Microscopy (EFM).

References

  1. 1 2 Jean M. Bennett, Lars Mattsson, Introduction to Surface Roughness and Scattering, Optical Society of America, Washington, D.C.
  2. 1 2 W J Walecki, F Szondy, and M M Hilali, "Fast in-line surface topography metrology enabling stress calculation for solar cell manufacturing for throughput in excess of 2000 wafers per hour" 2008 Meas. Sci. Technol. 19 025302 (6pp) doi : 10.1088/0957-0233/19/2/025302
  3. Stout, K. J.; Blunt, Liam (2000). Three-Dimensional Surface Topography (2nd ed.). Penton Press. p. 22. ISBN   978-1-85718-026-8.
  4. Binnig, Gerd; Quate, Calvin F; Gerber, Ch (1986). "Atomic force microscope". Physical Review Letters. 56 (9): 930–933. doi: 10.1103/PhysRevLett.56.930 . PMID   10033323.
  5. Dufour, Marc; Lamouche, G.; Gauthier, B.; Padioleau, C.; Monchalin, J.P. (2006). "Inspectionofhard-to-reachindustrialpartsusingsmalldiameterprobes" (PDF). SPIE Newsroom. SPIE. doi:10.1117/2.1200610.0467. S2CID   120476700 . Retrieved December 15, 2010.
  6. Dufour, M. L.; Lamouche, G.; Detalle, V.; Gauthier, B.; Sammut, P. (April 2005). "Low-Coherence Interferometry, an Advanced Technique for Optical Metrology in Industry". Insight: Non-Destructive Testing and Condition Monitoring . 47 (4): 216–219. CiteSeerX   10.1.1.159.5249 . doi:10.1784/insi.47.4.216.63149. ISSN   1354-2575.
  7. Cahoon, Joel E. (January 1995). "Defining Furrow Cross Section". Journal of Irrigation and Drainage Engineering. 121 (1): 114–119. doi:10.1061/(ASCE)0733-9437(1995)121:1(114). ISSN   0733-9437.