Rotating ring-disk electrode

Last updated

In analytical chemistry, a rotating ring-disk electrode (RRDE) [1] is a double working electrode used in hydrodynamic voltammetry, very similar to a rotating disk electrode (RDE). [2] The electrode rotates during experiments inducing a flux of analyte to the electrode. This system used in electrochemical studies when investigating reaction mechanisms related to redox chemistry and other chemical phenomena.

Contents

Structure

The difference between a rotating ring-disk electrode and a rotating disk electrode is the addition of a second working electrode in the form of a ring around the central disk of the first working electrode. To operate such an electrode, it is necessary to use a potentiostat, such as a bipotentiostat, capable of controlling a four-electrode system. The two electrodes are separated by a non-conductive barrier and connected to the potentiostat through different leads. This rotating hydrodynamic electrode motif can be extended to rotating double-ring electrodes, rotating double-ring-disk electrodes, and even more esoteric constructions, as suited to the experiment.

Function

The RRDE takes advantage of the laminar flow created during rotation. As the system is rotated, the solution in contact with the electrode is driven to its side, similar to the situation of a rotating disk electrode. As the solution flows to the side, it crosses the ring electrode and flows back into the bulk solution. If the flow in the solution is laminar, the solution is brought in contact with the disk and with the ring quickly afterward, in a very controlled manner. The resulting currents depend on the potential, area, and spacing of the electrodes, as well as the rotation speed and the substrate.

This design makes a variety of experiments possible, for example a complex could be oxidized at the disk and then reduced back to the starting material at the ring. It is easy to predict what the ring/disk current ratios is if this process is entirely controlled by the flow of solution. If it is not controlled by the flow of the solution the current will deviate. For example, if the first oxidation is followed by a chemical reaction, an EC mechanism, to form a product that can not be reduced at the ring then the magnitude of the ring current would be reduced. By varying the rate of rotation it is possible to determine the rate of the chemical reaction if it is in the proper kinetic regime.

Applications

The RRDE setup allows for many additional experiments well beyond the capacity of a RDE. For example, while one electrode conducts linear sweep voltammetry the other can be kept at a constant potential or also swept in a controlled manner. Step experiments with each electrode acting independently can be conducted. These as well as many other extremely elegant experiments are possible, including those tailored to the needs of a given system. Such experiments are useful in studying multi-electrons processes, the kinetics of a slow electron transfer, adsorption/desorption steps, and electrochemical reaction mechanisms.

The RRDE is an important tool for characterizing the fundamental properties of electrocatalysts used in fuel cells. For example, in a proton exchange membrane (PEM) fuel cell, dioxygen reduction at the cathode is often enhanced by an electrocatalyst comprising platinum nanoparticles. When oxygen is reduced using an electrocatalyst, an unwanted and harmful by-product, hydrogen peroxide, may be produced. Hydrogen peroxide can damage the internal components of a PEM fuel cell, so oxygen-reduction electrocatalysts are engineered in such a way as to limit the amount of peroxide formed. An RRDE "collection experiment" can be used to probe the peroxide generating tendencies of an electrocatalyst. [3] In this experiment, the disk is coated with a thin layer bearing the electrocatalyst, and the disk electrode is poised at a potential which reduces the oxygen. Any products generated at the disk electrode are then swept past the ring electrode. The potential of the ring electrode is poised to detect any hydrogen peroxide that may have been generated at the disk.

Design considerations

In general, narrowing the gap between the disk outer diameter and the ring inner diameter allows probing of systems with faster kinetics. A narrow gap reduces the "transit time" necessary for an intermediate species generated at the disk to successfully reach the ring electrode and be detected. Using precision machining techniques, it is possible to make gaps between 0.1 and 0.5 millimeters, and narrower gaps have been created using microlithography techniques.

Another important parameter for an RRDE is the "collection efficiency". This parameter is a measure of the percentage of the material generated at the disk electrode which is detected at the ring electrode. For any given set of RRDE dimensions (disk OD, ring ID, and ring OD), the collection efficiency can be computed using formulae derived from fluid dynamics first principles. One useful aspect of the theoretical collection efficiency is that it is only a function of the RRDE dimensions. That is, it is independent of the rotation rate over a wide range of rotation rates.

It is desirable for an RRDE to have a large collection efficiency if only to assure that the current signal measured at the ring electrode is detectable. On the other hand, it also desirable for an RRDE to have a small transit time so that short-lived (unstable) intermediate products generated at the disk survive long enough to be detected at the ring. The choice of actual RRDE dimensions is often a trade-off between a large collection efficiency or a short transit time.

See also

Related Research Articles

<span class="mw-page-title-main">Electrochemistry</span> Branch of chemistry

Electrochemistry is the branch of physical chemistry concerned with the relationship between electrical potential difference, as a measurable and quantitative phenomenon, and identifiable chemical change, with the potential difference as an outcome of a particular chemical change, or vice versa. These reactions involve electrons moving via an electronically-conducting phase between electrodes separated by an ionically conducting and electronically insulating electrolyte.

<span class="mw-page-title-main">Cyclic voltammetry</span> Method of analyzing electrochemical reactions

In electrochemistry, cyclic voltammetry (CV) is a type of potentiodynamic measurement. In a cyclic voltammetry experiment, the working electrode potential is ramped linearly versus time. Unlike in linear sweep voltammetry, after the set potential is reached in a CV experiment, the working electrode's potential is ramped in the opposite direction to return to the initial potential. These cycles of ramps in potential may be repeated as many times as needed. The current at the working electrode is plotted versus the applied voltage to give the cyclic voltammogram trace. Cyclic voltammetry is generally used to study the electrochemical properties of an analyte in solution or of a molecule that is adsorbed onto the electrode.

Coulometry determines the amount of matter transformed during an electrolysis reaction by measuring the amount of electricity consumed or produced. It can be used for precision measurements of charge, and the amperes even used to have a coulometric definition. However, today coulometry is mainly used for analytical applications. Coulometry is a group of techniques in analytical chemistry. It is named after Charles-Augustin de Coulomb.

<span class="mw-page-title-main">Potentiostat</span> Electronic system controlling a three electrode cell

A potentiostat is the electronic hardware required to control a three electrode cell and run most electroanalytical experiments. A Bipotentiostat and polypotentiostat are potentiostats capable of controlling two working electrodes and more than two working electrodes, respectively.

<span class="mw-page-title-main">Voltammetry</span>

Voltammetry is a category of electroanalytical methods used in analytical chemistry and various industrial processes. In voltammetry, information about an analyte is obtained by measuring the current as the potential is varied. The analytical data for a voltammetric experiment comes in the form of a voltammogram which plots the current produced by the analyte versus the potential of the working electrode.

<span class="mw-page-title-main">Chronoamperometry</span> Analytical method in electrochemistry

In electrochemistry, chronoamperometry is an analytical technique in which the electric potential of the working electrode is stepped and the resulting current from faradaic processes occurring at the electrode is monitored as a function of time. The functional relationship between current response and time is measured after applying single or double potential step to the working electrode of the electrochemical system. Limited information about the identity of the electrolyzed species can be obtained from the ratio of the peak oxidation current versus the peak reduction current. However, as with all pulsed techniques, chronoamperometry generates high charging currents, which decay exponentially with time as any RC circuit. The Faradaic current - which is due to electron transfer events and is most often the current component of interest - decays as described in the Cottrell equation. In most electrochemical cells, this decay is much slower than the charging decay-cells with no supporting electrolyte are notable exceptions. Most commonly a three-electrode system is used. Since the current is integrated over relatively longer time intervals, chronoamperometry gives a better signal-to-noise ratio in comparison to other amperometric techniques.

<span class="mw-page-title-main">Electrolysis of water</span> Electricity-induced chemical reaction

Electrolysis of water is using electricity to split water into oxygen and hydrogen gas by electrolysis. Hydrogen gas released in this way can be used as hydrogen fuel, or remixed with the oxygen to create oxyhydrogen gas, for use in welding and other applications.

In electrochemistry, overpotential is the potential difference (voltage) between a half-reaction's thermodynamically-determined reduction potential and the potential at which the redox event is experimentally observed. The term is directly related to a cell's voltage efficiency. In an electrolytic cell the existence of overpotential implies that the cell requires more energy than thermodynamically expected to drive a reaction. In a galvanic cell the existence of overpotential means less energy is recovered than thermodynamics predicts. In each case the extra/missing energy is lost as heat. The quantity of overpotential is specific to each cell design and varies across cells and operational conditions, even for the same reaction. Overpotential is experimentally determined by measuring the potential at which a given current density is achieved.

In electrochemistry, the working electrode is the electrode in an electrochemical system on which the reaction of interest is occurring. The working electrode is often used in conjunction with an auxiliary electrode, and a reference electrode in a three-electrode system. Depending on whether the reaction on the electrode is a reduction or an oxidation, the working electrode is called cathodic or anodic, respectively. Common working electrodes can consist of materials ranging from inert metals such as gold or platinum, to inert carbon such as glassy carbon, boron-doped diamond or pyrolytic carbon, and mercury drop and film electrodes. Chemically modified electrodes are employed for the analysis of both organic and inorganic samples.

In electrochemistry, an electrochemical reaction mechanism is the step-by-step sequence of elementary steps, involving at least one outer-sphere electron transfer, by which an overall electrochemical reaction occurs.

In analytical chemistry, a rotating disk electrode (RDE) is a working electrode used in three-electrode systems for hydrodynamic voltammetry. The electrode rotates during experiments, inducing a flux of analyte to the electrode. These working electrodes are used in electrochemical studies when investigating reaction mechanisms related to redox chemistry, among other chemical phenomena. The more complex rotating ring-disk electrode can be used as a rotating disk electrode if the ring is left inactive during the experiment.

In analytical chemistry, hydrodynamic voltammetry is a form of voltammetry in which the analyte solution flows relative to a working electrode. In many voltammetry techniques, the solution is intentionally left still to allow diffusion-controlled mass transfer. When a solution is made to flow, through stirring or some other physical mechanism, it is very important to the technique to achieve a very controlled flux or mass transfer in order to obtain predictable results. These methods are types of electrochemical studies which use potentiostats to investigate reaction mechanisms related to redox chemistry among other chemical phenomenon.

Bulk electrolysis is also known as potentiostatic coulometry or controlled potential coulometry. The experiment is a form of coulometry which generally employs a three electrode system controlled by a potentiostat. In the experiment the working electrode is held at a constant potential (volts) and current (amps) is monitored over time (seconds). In a properly run experiment an analyte is quantitatively converted from its original oxidation state to a new oxidation state, either reduced or oxidized. As the substrate is consumed, the current also decreases, approaching zero when the conversion nears completion.

The Levich equation models the diffusion and solution flow conditions around a rotating disk electrode (RDE). It is named after Veniamin Grigorievich Levich who first developed an RDE as a tool for electrochemical research. It can be used to predict the current observed at an RDE, in particular, the Levich equation gives the height of the sigmoidal wave observed in rotating disk voltammetry. The sigmoidal wave height is often called the Levich current.

<span class="mw-page-title-main">Electrocatalyst</span> Catalyst participating in electrochemical reactions

An electrocatalyst is a catalyst that participates in electrochemical reactions. Electrocatalysts are a specific form of catalysts that function at electrode surfaces or, most commonly, may be the electrode surface itself. An electrocatalyst can be heterogeneous such as a platinized electrode. Homogeneous electrocatalysts, which are soluble, assist in transferring electrons between the electrode and reactants, and/or facilitate an intermediate chemical transformation described by an overall half reaction. Major challenges in electrocatalysts focus on fuel cells.

Electrochemical engineering is the branch of chemical engineering dealing with the technological applications of electrochemical phenomena, such as electrosynthesis of chemicals, electrowinning and refining of metals, flow batteries and fuel cells, surface modification by electrodeposition, electrochemical separations and corrosion.

A liquid metal electrode is an electrode that uses a liquid metal, such as mercury, Galinstan, and NaK. They can be used in electrocapillarity, voltammetry, and impedance measurements.

Scanning electrochemical microscopy (SECM) is a technique within the broader class of scanning probe microscopy (SPM) that is used to measure the local electrochemical behavior of liquid/solid, liquid/gas and liquid/liquid interfaces. Initial characterization of the technique was credited to University of Texas electrochemist, Allen J. Bard, in 1989. Since then, the theoretical underpinnings have matured to allow widespread use of the technique in chemistry, biology and materials science. Spatially resolved electrochemical signals can be acquired by measuring the current at an ultramicroelectrode (UME) tip as a function of precise tip position over a substrate region of interest. Interpretation of the SECM signal is based on the concept of diffusion-limited current. Two-dimensional raster scan information can be compiled to generate images of surface reactivity and chemical kinetics.

In electrochemistry, protein film voltammetry is a technique for examining the behavior of proteins immobilized on an electrode. The technique is applicable to proteins and enzymes that engage in electron transfer reactions and it is part of the methods available to study enzyme kinetics.

Gas-diffusion electrocrystallization (GDEx) is an electrochemical process consisting on the reactive precipitation of metal ions in solution with intermediaries produced by the electrochemical reduction of gases, at gas diffusion electrodes. It can serve for the recovery of metals or metalloids into solid precipitates or for the synthesis of libraries of nanoparticles.

References

  1. Albery W.J.; Hitchman M.L. Ring-Disc Electrodes Oxford: Clarendon Press 1971 ( ISBN   978-0198553496)
  2. Bard, A.J.; Faulkner, L.R. Electrochemical Methods: Fundamentals and Applications. New York: John Wiley & Sons, 2nd Edition, 2000.
  3. Schmidt, T.J.; Paulus, U.A.; Gasteiger, H.A.; Behm R.J. The oxygen reduction reaction on a Pt/carbon fuel cell catalyst in the presence of chloride anions Journal of Electroanalytical Chemistry 508 (2001) 41-47. doi : 10.1016/S0022-0728(01)00499-5

[1] [2]

  1. Guha S. Simulation studies on a rotating ring disk electrode: Effects of ionic migration with kinetic complication-disk results. AIChE J. 2021;e17505. doi:10.1002/aic.17505
  2. Guha S. Simulation studies on a rotating ring disk electrode system: role of supporting electrolyte in determination of relevance of ionic migration. AIChE J. 2013;59(4):1390-1399.DOI 10.1002/aic.13903