Infrared open-path detector

Last updated

Infrared open-path gas detectors send out a beam of infrared light, detecting gas anywhere along the path of the beam. This linear 'sensor' is typically a few metres up to a few hundred metres in length. Open-path detectors can be contrasted with infrared point sensors.

Contents

They are widely used in the petroleum and petrochemical industries, mostly to achieve very rapid gas leak detection for flammable gases at concentrations comparable to the lower flammable limit (typically a few percent by volume). They are also used, but so far to a lesser extent, in other industries where flammable concentrations can occur, such as in coal mining and water treatment. In principle the technique can also be used to detect toxic gases, for instance hydrogen sulfide, at the necessary parts-per-million concentrations, but the technical difficulties involved have so far prevented widespread adoption for toxic gases.

Usually, there are separate transmitter and receiver units at either end of a straight beam path. Alternatively, the source and receiver are combined, and the beam bounced off a retroreflector at the far end of the measurement path. For portable use, detectors have also been made which use the natural albedo of surrounding objects in place of the retroreflector. The presence of a chosen gas (or class of gases) is detected from its absorption of a suitable infrared wavelength in the beam. Rain, fog etc. in the measurement path can also reduce the strength of the received signal, so it is usual to make a simultaneous measurement at one or more reference wavelengths. The quantity of gas intercepted by the beam is then inferred from the ratio of the signal losses at the measurement and reference wavelengths. The calculation is typically carried out by a microprocessor which also carries out various checks to validate the measurement and prevent false alarms.

The measured quantity is the sum of all the gas along the path of the beam, sometimes termed the path-integral concentration of the gas. Thus the measurement has a natural bias (desirable in many applications) towards the total size of an unintentional gas release, rather than the concentration of the gas that has reached any particular point. Whereas the natural units of measurement for an Infrared point sensor are parts-per-million (ppm) or the percentage of the lower flammable limit (%LFL), the natural units of measurement for an open path detector are ppm.metres (ppm.m) or LFL.metres (LFL.m). For instance, the fire and gas safety system on an offshore platform in the North Sea typically has detectors set to a full-scale reading of 5LFL.m, with low and high alarms triggered at 1LFL.m and 3LFL.m respectively.

Advantages and disadvantages versus fixed-point detectors

An open path detector usually costs more than a single point detector, so there is little incentive for applications that play to a point detector's strengths: where the point detector can be placed at the known location of the highest gas concentration, and a relatively slow response is acceptable. The open path detector excels in outdoor situations where, even if the likely source of the gas release is known, the evolution of the developing cloud or plume is unpredictable. Gas will almost certainly enter an extended linear beam before finding its way to any single chosen point. Also, point detectors in exposed outdoor locations require weather shields to be fitted, increasing the response time significantly. Open path detectors can also show a cost advantage in any application where a row of point detectors would be required to achieve the same coverage, for instance monitoring along a pipeline, or around the perimeter of a plant. Not only will one detector replace several, but the costs of installation, maintenance, cabling etc. are likely to be lower.

Component parts

In principle any source of infrared radiation could be used, together with an optical system of lenses or mirrors to form the transmitted beam. In practice the following sources have been used, always with some form of modulation to aid the signal processing at the receiver:

An incandescent light bulb, modulated by pulsing the current powering the filament or by a mechanical chopper. For systems used outdoors, it is difficult for an incandescent source to compete with the intensity of sunlight when the sun shines directly into the receiver. Also, it is difficult to achieve modulation frequencies distinguishable from those that can be produced naturally, for instance by heat shimmer or by sunlight reflecting off waves at sea.

A gas-discharge lamp is capable of exceeding the spectral power of direct sunlight in the infrared, especially when pulsed. Modern open path systems typically use a xenon flashtube powered by a capacitor discharge. Such pulsed sources are inherently modulated.

A semiconductor laser provides a relatively weak source, but one that can be modulated at high frequency in wavelength as well as amplitude. This property permits various signal processing schemes based on Fourier analysis, of use when the absorption of the gas is weak but narrow in spectral linewidth.

The precise wavelength passbands used must be isolated from the broad infrared spectrum. In principle any conventional spectrometer technique is possible, but the NDIR technique with multilayer dielectric filters and beamsplitters is most often used. These wavelength-defining components are usually located in the receiver, although one design has shared the task with the transmitter.

At the receiver, the infrared signal strengths are measured by some form of infrared detector. Generally photodiode detectors are preferred, and are essential for the higher modulation frequencies, whereas slower photoconductive detectors may be required for longer wavelength regions. The signals are fed to low-noise amplifiers, then invariably subject to some form of digital signal processing. The absorption coefficient of the gas will vary across the passband, so the simple Beer–Lambert law cannot be applied directly. For this reason the processing usually employs a calibration table, applicable for a particular gas, type of gas, or gas mixture, and sometimes configurable by the user.

Operating wavelengths

The choice of infrared wavelengths used for the measurement largely defines the detector's suitability for a particular applications. Not only must the target gas (or gases) have a suitable absorption spectrum, the wavelengths must lie within a spectral window so the air in the beam path is itself transparent. These wavelength regions have been used:

History

The first open-path detector offered for routine industrial use, as distinct from research instruments built in small numbers, was the Wright and Wright 'Pathwatch' in the US, 1983. Acquired by Det-Tronics (Detector Electronics Corporation) in 1992, the detector operated in the 3.4 μm region with a powerful incandescent source and a mechanical chopper. It did not achieve large volume sales, mainly because of cost and doubts about long-term reliability with moving parts. Beginning in 1985, Shell Research in UK was funded by Shell Natural Gas to develop an open-path detector with no moving parts. The advantages of the 2.3 μm wavelength were identified, and a research prototype was demonstrated. This design had a combined transmitter-receiver with a corner-cube retroreflector at 50 m. It used a pulsed incandescent lamp, PbS photoconductive detectors in the gas and reference channels, and an Intel 8031 microprocessor for signal processing. In 1987 Shell licensed this technology to Sieger-Zellweger (later Honeywell) who designed and marketed their industrial version as the 'Searchline', using a retro-reflective panel made up of multiple corner-cubes. This was the first open-path detector to be certified for use in hazardous areas and to have no moving parts. Later work by Shell Research used two alternately pulsed incandescent sources in the transmitter and a single PbS detectors in the receiver, avoiding zero drifts caused by the variable responsivity of PbS detectors. This technology was offered to Sieger-Zellweger, and later licensed to PLMS. a company part-owned by Shell Ventures UK. The PLMS GD4001/2 in 1991 were the first detectors to achieve a truly stable zero without moving parts or software compensation of slow drifts. They were also the first infrared gas detectors of any kind to be certified intrinsically safe. The Israeli company Spectronix (also Spectrex) made an important advance in 1996 with their SafEye, the first to use a flash tube source, followed by Sieger-Zellweger with their Searchline Excel in 1998. In 2001 the PLMS Pulsar, soon afterwards acquired by Dräger as their Polytron Pulsar, was the first detector to incorporate sensing to monitor the mutual alignment of the transmitter and receiver during both installation and routine operation.

Related Research Articles

Atomic absorption spectroscopy

Atomic absorption spectroscopy (AAS) and atomic emission spectroscopy (AES) is a spectroanalytical procedure for the quantitative determination of chemical elements using the absorption of optical radiation (light) by free atoms in the gaseous state. Atomic absorption spectroscopy is based on absorption of light by free metallic ions.

Infrared spectroscopy Interaction of infrared radiation with matter

Infrared spectroscopy is the measurement of the interaction of infrared radiation with matter by absorption, emission, or reflection. It is used to study and identify chemical substances or functional groups in solid, liquid, or gaseous forms. The method or technique of infrared spectroscopy is conducted with an instrument called an infrared spectrometer which produces an infrared spectrum. An IR spectrum can be visualized in a graph of infrared light absorbance on the vertical axis vs. frequency or wavelength on the horizontal axis. Typical units of frequency used in IR spectra are reciprocal centimeters, with the symbol cm−1. Units of IR wavelength are commonly given in micrometers, symbol μm, which are related to wave numbers in a reciprocal way. A common laboratory instrument that uses this technique is a Fourier transform infrared (FTIR) spectrometer. Two-dimensional IR is also possible as discussed below.

Fourier-transform spectroscopy is a measurement technique whereby spectra are collected based on measurements of the coherence of a radiative source, using time-domain or space-domain measurements of the electromagnetic radiation or other type of radiation. It can be applied to a variety of types of spectroscopy including optical spectroscopy, infrared spectroscopy, nuclear magnetic resonance (NMR) and magnetic resonance spectroscopic imaging (MRSI), mass spectrometry and electron spin resonance spectroscopy. There are several methods for measuring the temporal coherence of the light, including the continuous wave Michelson or Fourier-transform spectrometer and the pulsed Fourier-transform spectrograph.

Ultraviolet–visible spectroscopy Range of spectroscopic analysis

Ultraviolet–visible spectroscopy or ultraviolet–visible spectrophotometry refers to absorption spectroscopy or reflectance spectroscopy in part of the ultraviolet and the full, adjacent visible regions of the electromagnetic spectrum. This means it uses light in the visible and adjacent ranges. The absorption or reflectance in the visible range directly affects the perceived color of the chemicals involved. In this region of the spectrum, atoms and molecules undergo electronic transitions. Absorption spectroscopy is complementary to fluorescence spectroscopy, in that fluorescence deals with transitions of electrons from the excited state to the ground state, while absorption measures transitions from the ground state to the excited state.

X-ray fluorescence Physical phenomenom

X-ray fluorescence (XRF) is the emission of characteristic "secondary" X-rays from a material that has been excited by being bombarded with high-energy X-rays or gamma rays. The phenomenon is widely used for elemental analysis and chemical analysis, particularly in the investigation of metals, glass, ceramics and building materials, and for research in geochemistry, forensic science, archaeology and art objects such as paintings

Absorption spectroscopy Spectroscopic techniques that measure the absorption of radiation

Absorption spectroscopy refers to spectroscopic techniques that measure the absorption of radiation, as a function of frequency or wavelength, due to its interaction with a sample. The sample absorbs energy, i.e., photons, from the radiating field. The intensity of the absorption varies as a function of frequency, and this variation is the absorption spectrum. Absorption spectroscopy is performed across the electromagnetic spectrum.

Tunable diode laser absorption spectroscopy is a technique for measuring the concentration of certain species such as methane, water vapor and many more, in a gaseous mixture using tunable diode lasers and laser absorption spectrometry. The advantage of TDLAS over other techniques for concentration measurement is its ability to achieve very low detection limits. Apart from concentration, it is also possible to determine the temperature, pressure, velocity and mass flux of the gas under observation. TDLAS is by far the most common laser based absorption technique for quantitative assessments of species in gas phase.

Spectrophotometry

Spectrophotometry is a branch of electromagnetic spectroscopy concerned with the quantitative measurement of the reflection or transmission properties of a material as a function of wavelength. Spectrophotometry uses photometers, known as spectrophotometers, that can measure the intensity of a light beam at different wavelengths. Although spectrophotometry is most commonly applied to ultraviolet, visible, and infrared radiation, modern spectrophotometers can interrogate wide swaths of the electromagnetic spectrum, including x-ray, ultraviolet, visible, infrared, and/or microwave wavelengths.

Photometer Scientific instrument

A photometer is an instrument that measures the strength of electromagnetic radiation in the range from ultraviolet to infrared and including the visible spectrum. Most photometers convert light into an electric current using a photoresistor, photodiode, or photomultiplier.

Terahertz time-domain spectroscopy

In physics, terahertz time-domain spectroscopy (THz-TDS) is a spectroscopic technique in which the properties of matter are probed with short pulses of terahertz radiation. The generation and detection scheme is sensitive to the sample's effect on both the amplitude and the phase of the terahertz radiation. By measuring in the time-domain, the technique can provide more information than conventional Fourier-transform spectroscopy, which is only sensitive to the amplitude.

Ultrafast laser spectroscopy is a spectroscopic technique that uses ultrashort pulse lasers for the study of dynamics on extremely short time scales. Different methods are used to examine the dynamics of charge carriers, atoms, and molecules. Many different procedures have been developed spanning different time scales and photon energy ranges; some common methods are listed below.

Nondispersive infrared sensor

A nondispersive infrared sensor is a simple spectroscopic sensor often used as a gas detector. It is non-dispersive in the fact that no dispersive element is used to separate out the broadband light into a narrow spectrum suitable for gas sensing. The majority of NDIR sensors use a broadband lamp source and an optical filter to select a narrow band spectral region that overlaps with the absorption region of the gas of interest. In this context narrow may be 50-300nm bandwidth. Modern NDIR sensors may use Microelectromechanical systems (MEMs) or mid IR LED sources, with or without an optical filter.

Planar laser-induced fluorescence

Planar laser-induced fluorescence (PLIF) is an optical diagnostic technique widely used for flow visualization and quantitative measurements. PLIF has been shown to be used for velocity, concentration, temperature and pressure measurements.

A gas detector is a device that detects the presence of gases in an area, often as part of a safety system. A gas detector can sound an alarm to operators in the area where the leak is occurring, giving them the opportunity to leave. This type of device is important because there are many gases that can be harmful to organic life, such as humans or animals.

Infrared lamps are electrical devices which emit infrared radiation. Infrared lamps are commonly used in radiant heating for industrial processes and building heating. Infrared LEDs are used for communication over optical fibers and in remote control devices. Infrared lamps are also used for some night vision devices where visible light would be objectionable. Infrared lamp sources are used in certain scientific and industrial instrument for chemical analysis of liquids and gases; for example, the pollutant sulfur dioxide in air can be measured using its infrared absorption characteristics.

A flame detector is a sensor designed to detect and respond to the presence of a flame or fire, allowing flame detection. Responses to a detected flame depend on the installation, but can include sounding an alarm, deactivating a fuel line, and activating a fire suppression system. When used in applications such as industrial furnaces, their role is to provide confirmation that the furnace is working properly; it can be used to turn off the ignition system though in many cases they take no direct action beyond notifying the operator or control system. A flame detector can often respond faster and more accurately than a smoke or heat detector due to the mechanisms it uses to detect the flame.

An infrared point sensor is a point gas detector based on the nondispersive infrared sensor technology.

Fourier-transform infrared spectroscopy technique to analyze the infrared spectrum of matter

Fourier-transform infrared spectroscopy (FTIR) is a technique used to obtain an infrared spectrum of absorption or emission of a solid, liquid or gas. An FTIR spectrometer simultaneously collects high-resolution spectral data over a wide spectral range. This confers a significant advantage over a dispersive spectrometer, which measures intensity over a narrow range of wavelengths at a time.

Infrared gas analyzer

An infrared gas analyzer measures trace gases by determining the absorption of an emitted infrared light source through a certain air sample. Trace gases found in the Earth's atmosphere become excited under specific wavelengths found in the infrared range. The concept behind the technology can be understood as testing how much of the light is absorbed by the air. Different molecules in the air absorb different frequencies of light. Air with much of a certain gas will absorb more of a certain frequency, allowing the sensor to report a high concentration of the corresponding molecule.

Optical beam smoke detector

An optical beam smoke detector is a device that uses a projected beam of light to detect smoke across large areas, typically as an indicator of fire. They are used to detect fires in buildings where standard point smoke detectors would either be uneconomical or restricted for use by the height of the building. Optical beam smoke detectors are often installed in warehouses as a cost-effective means of protecting large open spaces.

References