Micellar liquid chromatography (MLC) is a form of reversed phase liquid chromatography that uses an aqueous micellar solutions as the mobile phase. [1]

Contents

Theory

The use of micelles in high performance liquid chromatography was first introduced by Armstrong and Henry in 1980. [2] [3] The technique is used mainly to enhance retention and selectivity of various solutes that would otherwise be inseparable or poorly resolved. Micellar liquid chromatography (MLC) has been used in a variety of applications including separation of mixtures of charged and neutral solutes, direct injection of serum and other physiological fluids, analysis of pharmaceutical compounds, separation of enantiomers, analysis of inorganic organometallics, and a host of others.

One of the main drawbacks of the technique is the reduced efficiency that is caused by the micelles. Despite the sometimes poor efficiency, MLC is a better choice than ion-exchange LC or ion-pairing LC for separation of charged molecules and mixtures of charged and neutral species. [1] Some of the aspects which will be discussed are the theoretical aspects of MLC, the use of models in predicting retentive characteristics of MLC, the effect of micelles on efficiency and selectivity, and general applications of MLC.

Reverse phase high-performance liquid chromatography (RP-HPLC) involves a non-polar stationary phase, often a hydrocarbon chain, and a polar mobile or liquid phase. The mobile phase generally consists of an aqueous portion with an organic addition, such as methanol or acetonitrile. When a solution of analytes is injected into the system, the components begin to partition out of the mobile phase and interact with the stationary phase. Each component interacts with the stationary phase in a different manner depending upon its polarity and hydrophobicity. In reverse phase HPLC, the solute with the greatest polarity will interact less with the stationary phase and spend more time in the mobile phase. As the polarity of the components decreases, the time spent in the column increases. Thus, a separation of components is achieved based on polarity. [4] The addition of micelles to the mobile phase introduces a third phase into which the solutes may partition.

Micelles

Micelles are composed of surfactant, or detergent, monomers with a hydrophobic moiety, or tail, on one end, and a hydrophilic moiety, or head group, on the other. The polar head group may be anionic, cationic, zwitterionic, or non-ionic. When the concentration of a surfactant in solution reaches its critical micelle concentration (CMC), it forms micelles which are aggregates of the monomers. The CMC is different for each surfactant, as is the number of monomers which make up the micelle, termed the aggregation number (AN). [5] Table 1 lists some common detergents used to form micelles along with their CMC and AN where available.

Table 1 [1] [5]
TypeNameCMC (mM)AN
Anionic Cholic acid, sodium salt142-4
Deoxycholic acid, sodium salt54-10
Glycocholic acid, sodium salt132
Sodium dodecyl sulfate (SDS)8.2762
Taurocholic acid, sodium salt10-154
Sodium tetradecyl sulfate 2.1
Cationic Cetyltrimethylammonium chloride 1
Cetyltrimethylammonium bromide (CTAB)1.378
Dodecyltrimethlyammonium bromide (DTAB)1450
Hexadecyltrimethylammonium bromide 0.026169
Zwitterionic3-[(3-cholamidopropyl)dimethylammonio]-1-propanesulfonate (CHAPS)810
3-[(3-cholamidopropyl)dimethylammonio]-2-hydroxy-1-propanesulfonate (CHAPSO)811
N-Dodecyl-N,N-dimethylammonio-3-propane sulfonate3.3
Nonionicn-Decyl-b-D-glucopyranoside2.2
Triton X-100 0.24140
Polyoxyethylene (23) dodecanol (BRIJ 35)0.1
Polyoxyethylene [20]-sorbitane monooleate (Tween 80)0.01
Polyoxyethylene [20]-sorbitane monolaurate (Tween 20)0.059

Many of the characteristics of micelles differ from those of bulk solvents. For example, the micelles are, by nature, spatially heterogeneous with a hydrocarbon, nearly anhydrous core and a highly solvated, polar head group. They have a high surface-to-volume ratio due to their small size and generally spherical shape. Their surrounding environment (pH, ionic strength, buffer ion, presence of a co-solvent, and temperature) has an influence on their size, shape, critical micelle concentration, aggregation number and other properties. [6]

Another important property of micelles is the Krafft point, the temperature at which the solubility of the surfactant is equal to its CMC. For HPLC applications involving micelles, it is best to choose a surfactant with a low Krafft point and CMC. A high CMC would require a high concentration of surfactant which would increase the viscosity of the mobile phase, an undesirable condition. Additionally, a Krafft point should be well below room temperature to avoid having to apply heat to the mobile phase. To avoid potential interference with absorption detectors, a surfactant should also have a small molar absorptivity at the chosen wavelength of analysis. Light scattering should not be a concern due to the small size, a few nanometers, of the micelle. [1]

The effect of organic additives on micellar properties is another important consideration. A small amount of organic solvent is often added to the mobile phase to help improve efficiency and to improve separations of compounds. Care needs to be taken when determining how much organic to add. Too high a concentration of the organic may cause the micelle to disperse, as it relies on hydrophobic effects for its formation. The maximum concentration of organic depends on the organic solvent itself, and on the micelle. This information is generally not known precisely, but a generally accepted practice is to keep the volume percentage of organic below 15–20%. [1]

Research

Fischer and Jandera [7] studied the effect of changing the concentration of methanol on CMC values for three commonly used surfactants. Two cationic, hexadecyltrimethylammonium bromide (CTAB), and N-(a-carbethoxypentadecyl) trimethylammonium bromide (Septonex), and one anionic surfactant, sodium dodecyl sulphate (SDS) were chosen for the experiment. Generally speaking, the CMC increased as the concentration of methanol increased. It was then concluded that the distribution of the surfactant between the bulk mobile phase and the micellar phase shifts toward the bulk as the methanol concentration increases. For CTAB, the rise in CMC is greatest from 010% methanol, and is nearly constant from 1020%. Above 20% methanol, the micelles disaggregate and do not exist. For SDS, the CMC values remain unaffected below 10% methanol, but begin to increase as the methanol concentration is further increased. Disaggregation occurs above 30% methanol. Finally, for Septonex, only a slight increase in CMC is observed up to 20%, with disaggregation occurring above 25%. [7]

As has been asserted, the mobile phase in MLC consists of micelles in an aqueous solvent, usually with a small amount of organic modifier added to complete the mobile phase. A typical reverse phase alkyl-bonded stationary phase is used. The first discussion of the thermodynamics involved in the retention mechanism was published by Armstrong and Nome in 1981. [8] In MLC, there are three partition coefficients which must be taken into account. The solute will partition between the water and the stationary phase (KSW), the water and the micelles (KMW), and the micelles and the stationary phase (KSM).

Armstrong and Nome derived an equation describing the partition coefficients in terms of the retention factor, formally capacity factor, k¢. In HPLC, the capacity factor represents the molar ratio of the solute in the stationary phase to the mobile phase. The capacity factor is easily measure based on retention times of the compound and any unretained compound. The equation rewritten by Guermouche et al. [9] is presented here:

1/k¢ = [n • (KMW-1)/(f • KSW)] • CM +1/(f • KSW)

Where:

A plot of 1/k¢ verses CM gives a straight line in which KSW can be calculated from the intercept and KMW can be obtained from the ratio of the slope to the intercept. Finally, KSM can be obtained from the ratio of the other two partition coefficients:

KSM = KSW/ KMW [8]

As can be observed from Figure 1, KMW is independent of any effects from the stationary phase, assuming the same micellar mobile phase. [9]

The validity of the retention mechanism proposed by Armstrong and Nome has been successfully, and repeated confirmed experimentally. However, some variations and alternate theories have also been proposed. Jandera and Fischer [10] developed equations to describe the dependence of retention behavior on the change in micellar concentrations. They found that the retention of most compounds tested decreased with increasing concentrations of micelles. From this, it can be surmised that the compounds associate with the micelles as they spend less time associated with the stationary phase. [10]

Foley proposed a similar retentive model to that of Armstrong and Nome which was a general model for secondary chemical equilibria in liquid chromatography. [11] While this model was developed in a previous reference, and could be used for any secondary chemical equilibria such as acid-base equilibria, and ion-pairing, Foley further refined the model for MLC. When an equilibrant (X), in this case surfactant, is added to the mobile phase, a secondary equilibria is created in which an analyte will exist as free analyte (A), and complexed with the equilibrant (AX). The two forms will be retained by the stationary phase to different extents, thus allowing the retention to be varied by adjusting the concentration of equilibrant (micelles). [11]

The resulting equation solved for capacity factor in terms of partition coefficients is much the same as that of Armstrong and Nome:

1/k¢ = (KSM/k¢S) • [M] + 1/k¢S

Where:

Foley used the above equation to determine the solute-micelle association constants and free solute retention factors for a variety of solutes with different surfactants and stationary phases. From this data, it is possible to predict the type and optimum surfactant concentrations needed for a given solute or solutes. [11]

Foley has not been the only researcher interested in determining the solute-micelle association constants. A review article by Marina and Garcia with 53 references discusses the usefulness of obtaining solute-micelle association constants. [12] The association constants for two solutes can be used to help understand the retention mechanism. The separation factor of two solutes, a, can be expressed as KSM1/KSM2. If the experimental a coincides with the ratio of the two solute-micelle partition coefficients, it can be assumed that their retention occurs through a direct transfer from the micellar phase to the stationary phase. In addition, calculation of a would allow for prediction of separation selectivity before the analysis is performed, provided the two coefficients are known. [12]

The desire to predict retention behavior and selectivity has led to the development of several mathematical models. [13] Changes in pH, surfactant concentration, and concentration of organic modifier play a significant role in determining the chromatographic separation. Often one or more of these parameters need to be optimized to achieve the desired separation, yet the optimum parameters must take all three variables into account simultaneously. The review by Garcia-Alvarez-Coque et al. mentioned several successful models for varying scenarios, a few of which will be mentioned here. The classic models by Armstrong and Nome and Foley are used to describe the general cases. Foley's model applies to many cases and has been experimentally verified for ionic, neutral, polar and nonpolar solutes; anionic, cationic, and non-ionic surfactants, and C8, C¬18, and cyano stationary phases. The model begins to deviate for highly and lowly retained solutes. Highly retained solutes may become irreversibly bound to the stationary phase, where lowly retained solutes may elute in the column void volume. [13]

Other models proposed by Arunyanart and Cline-Love and Rodgers and Khaledi describe the effect of pH on the retention of weak acids and bases. These authors derived equations relating pH and micellar concentration to retention. As the pH varies, sigmoidal behavior is observed for the retention of acidic and basic species. This model has been shown to accurately predict retention behavior. [13] Still other models predict behavior in hybrid micellar systems using equations or modeling behavior based on controlled experimentation. Additionally, models accounting for the simultaneous effect of pH, micelle and organic concentration have been suggested. These models allow for further enhancement of the optimization of the separation of weak acids and bases. [13]

One research group, Rukhadze, et al. [14] derived a first order linear relationship describing the influence of micelle and organic concentration, and pH on the selectivity and resolution of seven barbiturates. The researchers discovered that a second order mathematical equation would more precisely fit the data. The derivations and experimental details are beyond the scope of this discussion. The model was successful in predicting the experimental conditions necessary to achieve a separation for compounds which are traditionally difficult to resolve. [14]

Jandera, Fischer, and Effenberger approached the modeling problem in yet another way. [15] The model used was based on lipophilicity and polarity indices of solutes. The lipophilicity index relates a given solute to a hypothetical number of carbon atoms in an alkyl chain. It is based and depends on a given calibration series determined experimentally. The lipophilicity index should be independent of the stationary phase and organic modifier concentration. The polarity index is a measure of the polarity of the solute-solvent interactions. It depends strongly on the organic solvent, and somewhat on the polar groups present in the stationary phase. 23 compounds were analyzed with varying mobile phases and compared to the lipophilicity and polarity indices. The results showed that the model could be applied to MLC, but better predictive behavior was found with concentrations of surfactant below the CMC, sub-micellar. [15]

A final type of model based on molecular properties of a solute is a branch of quantitative structure-activity relationships (QSAR). QSAR studies attempt to correlate biological activity of drugs, or a class of drugs, with structures. The normally accepted means of uptake for a drug, or its metabolite, is through partitioning into lipid bilayers. The descriptor most often used in QSAR to determine the hydrophobicity of a compound is the octanol-water partition coefficient, log P. [16] MLC provides an attractive and practical alternative to QSAR. When micelles are added to a mobile phase, many similarities exist between the micellar mobile phase/stationary phase and the biological membrane/water interface. In MLC, the stationary phase become modified by the adsorption of surfactant monomers which are structurally similar to the membranous hydrocarbon chains in the biological model. Additionally, the hydrophilic/hydrophobic interactions of the micelles are similar to that in the polar regions of a membrane. Thus, the development of quantitative structure-retention relationships (QRAR) has become widespread. [17]

Escuder-Gilabert et al. [18] tested three different QRAR retention models on ionic compounds. Several classes of compounds were tested including catecholamines, local anesthetics, diuretics, and amino acids. The best model relating log K and log P was found to be one in which the total molar charge of a compound at a given pH is included as a variable. This model proved to give fairly accurate predictions of log P, R  > 0.9. [18] Other studies have been performed which develop predictive QRAR models for tricyclic antidepressants [17] and barbiturates. [16]

Efficiency

The main limitation in the use of MLC is the reduction in efficiency (peak broadening) that is observed when purely aqueous micellar mobile phases are used. [19] Several explanations for the poor efficiency have been theorized. Poor wetting of the stationary phase by the micellar aqueous mobile phase, slow mass transfer between the micelles and the stationary phase, and poor mass transfer within the stationary phase have all been postulated as possible causes. To enhance efficiency, the most common approaches have been the addition of small amounts of isopropyl alcohol and increase in temperature. A review by Berthod [19] studied the combined theories presented above and applied the Knox equation to independently determine the cause of the reduced efficiency. The Knox equation is commonly used in HPLC to describe the different contributions to overall band broadening of a solute. The Knox equation is expressed as:

h = An^(1/3)+ B/n + Cn

Where:

Berthod's use of the Knox equation to experimentally determine which of the proposed theories was most correct led him to the following conclusions. The flow anisotropy in micellar phase seems to be much greater than in traditional hydro-organic mobile phases of similar viscosity. This is likely due to the partial clogging of the stationary phase pores by adsorbed surfactant molecules. Raising the column temperature served to both decrease viscosity of the mobile phase and the amount of adsorbed surfactant. Both results reduce the A term and the amount of eddy diffusion, and thereby increase efficiency. [19]

The increase in the B term, as related to longitudinal diffusion, is associated with the decrease in the solute diffusion coefficient in the mobile phase, DM, due to the presence of the micelles, and an increase in the capacity factor, k¢. Again, this is related to surfactant adsorption on the stationary phase causing a dramatic decrease in the solute diffusion coefficient in the stationary phase, DS. Again an increase in temperature, now coupled with an addition of alcohol to the mobile phase, drastically decreases the amount of the absorbed surfactant. In turn, both actions reduce the C term caused by a slow mass transfer from the stationary phase to the mobile phase. Further optimization of efficiency can be gained by reducing the flow rate to one closely matched to that derived from the Knox equation. Overall, the three proposed theories seemed to have contributing effects of the poor efficiency observed, and can be partially countered by the addition of organic modifiers, particularly alcohol, and increasing the column temperature. [19]

Applications

Despite the reduced efficiency verses reversed phase HPLC, hundreds of applications have been reported using MLC. One of the most advantageous is the ability to directly inject physiological fluids. Micelles have an ability to solubilize proteins which enables MLC to be useful in analyzing untreated biological fluids such as plasma, serum, and urine. [1] Martinez et al. [20] found MLC to be highly useful in analyzing a class of drugs called b-antagonists, so called beta-blockers, in urine samples. The main advantage of the use of MLC with this type of sample, is the great time savings in sample preparation. Alternative methods of analysis including reversed phase HPLC require lengthy extraction and sample work up procedures before analysis can begin. With MLC, direct injection is often possible, with retention times of less than 15 minutes for the separation of up to nine b-antagonists. [20]

Another application compared reversed phase HPLC with MLC for the analysis of desferrioxamine in serum. [21] Desferrioxamine (DFO) is a commonly used drug for removal of excess iron in patients with chronic and acute levels. The analysis of DFO along with its chelated complexes, Fe(III) DFO and Al(III) DFO has proven to be difficult at best in previous attempts. This study found that direct injection of the serum was possible for MLC, verses an ultrafiltration step necessary in HPLC. This analysis proved to have difficulties with the separation of the chelated DFO compounds and with the sensitivity levels for DFO itself when MLC was applied. The researcher found that, in this case, reverse phase HPLC, was a better, more sensitive technique despite the time savings in direct injection. [21]

Analysis of pharmaceuticals by MLC is also gaining popularity. The selectivity and peak shape of MLC over commonly used ion-pair chromatography is much enhanced. [22] MLC mimics, yet enhances, the selectivity offered by ion-pairing reagents for the separation of active ingredients in pharmaceutical drugs. For basic drugs, MLC improves the excessive peak tailing frequently observed in ion-pairing. Hydrophilic drugs are often unretained using conventional HPLC, are retained by MLC due to solubilization into the micelles. Commonly found drugs in cold medications such as acetaminophen, L-ascorbic acid, phenylpropanolamine HCL, tipepidine hibenzate, and chlorpheniramine maleate have been successfully separated with good peak shape using MLC. Additional basic drugs like many narcotics, such as codeine and morphine, have also been successfully separated using MLC. [22]

Another novel application of MLC involves the separation and analysis of inorganic compounds, mostly simple ions. This is a relatively new area for MLC, but has seen some promising results. [23] MLC has been observed to provide better selectivity of inorganic ions that ion-exchange or ion-pairing chromatography. While this application is still in the beginning stages of development, the possibilities exist for novel, much enhanced separations of inorganic species. [23]

Since the technique was first reported on in 1980, micellar liquid chromatography has been used in hundreds of applications. This micelle controlled technique provides for unique opportunities for solving complicated separation problems. Despite the poor efficiency of MLC, it has been successfully used in many applications. The use of MLC in the future appears to be extremely advantages in the areas of physiological fluids, pharmaceuticals, and even inorganic ions. The technique has proven to be superior over ion-pairing and ion-exchange for many applications. As new approaches are developed to combat the poor efficiency of MLC, its application is sure to spread and gain more acceptance.

Related Research Articles

In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent called the mobile phase, which carries it through a system on which a material called the stationary phase is fixed. Because the different constituents of the mixture tend to have different affinities for the stationary phase and are retained for different lengths of time depending on their interactions with its surface sites, the constituents travel at different apparent velocities in the mobile fluid, causing them to separate. The separation is based on the differential partitioning between the mobile and the stationary phases. Subtle differences in a compound's partition coefficient result in differential retention on the stationary phase and thus affect the separation.

<span class="mw-page-title-main">High-performance liquid chromatography</span> Technique in analytical chemistry

High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify specific components in mixtures. The mixtures can originate from food, chemicals, pharmaceuticals, biological, environmental and agriculture, etc., which have been dissolved into liquid solutions.

In the physical sciences, a partition coefficient (P) or distribution coefficient (D) is the ratio of concentrations of a compound in a mixture of two immiscible solvents at equilibrium. This ratio is therefore a comparison of the solubilities of the solute in these two liquids. The partition coefficient generally refers to the concentration ratio of un-ionized species of compound, whereas the distribution coefficient refers to the concentration ratio of all species of the compound.

Micellar electrokinetic chromatography (MEKC) is a chromatography technique used in analytical chemistry. It is a modification of capillary electrophoresis (CE), extending its functionality to neutral analytes, where the samples are separated by differential partitioning between micelles and a surrounding aqueous buffer solution.

<span class="mw-page-title-main">Column chromatography</span> Method to isolate a compound in a mixture

Column chromatography in chemistry is a chromatography method used to isolate a single chemical compound from a mixture. Chromatography is able to separate substances based on differential absorption of compounds to the adsorbent; compounds move through the column at different rates, allowing them to be separated into fractions. The technique is widely applicable, as many different adsorbents can be used with a wide range of solvents. The technique can be used on scales from micrograms up to kilograms. The main advantage of column chromatography is the relatively low cost and disposability of the stationary phase used in the process. The latter prevents cross-contamination and stationary phase degradation due to recycling. Column chromatography can be done using gravity to move the solvent, or using compressed gas to push the solvent through the column.

<span class="mw-page-title-main">Ion chromatography</span> Separates ions and polar molecules

Ion chromatography is a form of chromatography that separates ions and ionizable polar molecules based on their affinity to the ion exchanger. It works on almost any kind of charged molecule—including small inorganic anions, large proteins, small nucleotides, and amino acids. However, ion chromatography must be done in conditions that are one pH unit away from the isoelectric point of a protein.

In colloidal and surface chemistry, the critical micelle concentration (CMC) is defined as the concentration of surfactants above which micelles form and all additional surfactants added to the system will form micelles.

Chiral column chromatography is a variant of column chromatography that is employed for the separation of chiral compounds, i.e. enantiomers, in mixtures such as racemates or related compounds. The chiral stationary phase (CSP) is made of a support, usually silica based, on which a chiral reagent or a macromolecule with numerous chiral centers is bonded or immobilized.

Partition equilibrium is a special case of chemical equilibrium wherein one or more solutes are in equilibrium between two immiscible solvents. The most common chemical equilibrium systems involve reactants and products in the same phase - either all gases or all solutions. However, it is also possible to get equilibria between substances in different phases, such a liquid and gas that do not mix. One example is gas-liquid partition equilibrium chromatography, where an analyte equilibrates between a gas and liquid phase. Partition equilibria are described by Nernst's distribution law. Partition equilibrium are most commonly seen and used for Liquid–liquid extraction.

Reversed-phase liquid chromatography (RP-LC) is a mode of liquid chromatography in which non-polar stationary phase and polar mobile phases are used for the separation of organic compounds. The vast majority of separations and analyses using high-performance liquid chromatography (HPLC) in recent years are done using the reversed phase mode. In the reversed phase mode, the sample components are retained in the system the more hydrophobic they are.

Mixed-mode chromatography (MMC), or multimodal chromatography, refers to chromatographic methods that utilize more than one form of interaction between the stationary phase and analytes in order to achieve their separation. What is distinct from conventional single-mode chromatography is that the secondary interactions in MMC cannot be too weak, and thus they also contribute to the retention of the solutes.

<span class="mw-page-title-main">Hydrophilic interaction chromatography</span> Type of chromatography

Hydrophilic interaction chromatography is a variant of normal phase liquid chromatography that partly overlaps with other chromatographic applications such as ion chromatography and reversed phase liquid chromatography. HILIC uses hydrophilic stationary phases with reversed-phase type eluents. The name was suggested by Andrew Alpert in his 1990 paper on the subject. He described the chromatographic mechanism for it as liquid-liquid partition chromatography where analytes elute in order of increasing polarity, a conclusion supported by a review and re-evaluation of published data.

Supercritical fluid chromatography (SFC) is a form of normal phase chromatography that uses a supercritical fluid such as carbon dioxide as the mobile phase. It is used for the analysis and purification of low to moderate molecular weight, thermally labile molecules and can also be used for the separation of chiral compounds. Principles are similar to those of high performance liquid chromatography (HPLC); however, SFC typically utilizes carbon dioxide as the mobile phase. Therefore, the entire chromatographic flow path must be pressurized. Because the supercritical phase represents a state whereby bulk liquid and gas properties converge, supercritical fluid chromatography is sometimes called convergence chromatography. The idea of liquid and gas properties convergence was first envisioned by Giddings.

A hydrotrope is a compound that solubilizes hydrophobic compounds in aqueous solutions by means other than micellar solubilization. Typically, hydrotropes consist of a hydrophilic part and a hydrophobic part, but the hydrophobic part is generally too small to cause spontaneous self-aggregation. Hydrotropes do not have a critical concentration above which self-aggregation spontaneously starts to occur. Instead, some hydrotropes aggregate in a step-wise self-aggregation process, gradually increasing aggregation size. However, many hydrotropes do not seem to self-aggregate at all, unless a solubilizate has been added. Examples of hydrotropes include urea, tosylate, cumenesulfonate and xylenesulfonate.

Aqueous normal-phase chromatography (ANP) is a chromatographic technique that involves the mobile phase compositions and polarities between reversed-phase chromatography (RP) and normal-phase chromatography (NP), while the stationary phases are polar.

Electrochromatography is a chemical separation technique in analytical chemistry, biochemistry and molecular biology used to resolve and separate mostly large biomolecules such as proteins. It is a combination of size exclusion chromatography and gel electrophoresis. These separation mechanisms operate essentially in superposition along the length of a gel filtration column to which an axial electric field gradient has been added. The molecules are separated by size due to the gel filtration mechanism and by electrophoretic mobility due to the gel electrophoresis mechanism. Additionally there are secondary chromatographic solute retention mechanisms.

Displacement chromatography is a chromatography technique in which a sample is placed onto the head of the column and is then displaced by a solute that is more strongly sorbed than the components of the original mixture. The result is that the components are resolved into consecutive "rectangular" zones of highly concentrated pure substances rather than solvent-separated "peaks". It is primarily a preparative technique; higher product concentration, higher purity, and increased throughput may be obtained compared to other modes of chromatography.

Hydrophobicity scales are values that define the relative hydrophobicity or hydrophilicity of amino acid residues. The more positive the value, the more hydrophobic are the amino acids located in that region of the protein. These scales are commonly used to predict the transmembrane alpha-helices of membrane proteins. When consecutively measuring amino acids of a protein, changes in value indicate attraction of specific protein regions towards the hydrophobic region inside lipid bilayer.

<span class="mw-page-title-main">Countercurrent chromatography</span>

Countercurrent chromatography is a form of liquid–liquid chromatography that uses a liquid stationary phase that is held in place by inertia of the molecules composing the stationary phase accelerating toward the center of a centrifuge due to centripetal force and is used to separate, identify, and quantify the chemical components of a mixture. In its broadest sense, countercurrent chromatography encompasses a collection of related liquid chromatography techniques that employ two immiscible liquid phases without a solid support. The two liquid phases come in contact with each other as at least one phase is pumped through a column, a hollow tube or a series of chambers connected with channels, which contains both phases. The resulting dynamic mixing and settling action allows the components to be separated by their respective solubilities in the two phases. A wide variety of two-phase solvent systems consisting of at least two immiscible liquids may be employed to provide the proper selectivity for the desired separation.

<span class="mw-page-title-main">Capillary electrochromatography</span> Method of separating components of a mixture via electro-osmosis

In chemical analysis, capillary electrochromatography (CEC) is a chromatographic technique in which the mobile phase is driven through the chromatographic bed by electro-osmosis. Capillary electrochromatography is a combination of two analytical techniques, high-performance liquid chromatography and capillary electrophoresis. Capillary electrophoresis aims to separate analytes on the basis of their mass-to-charge ratio by passing a high voltage across ends of a capillary tube, which is filled with the analyte. High-performance liquid chromatography separates analytes by passing them, under high pressure, through a column filled with stationary phase. The interactions between the analytes and the stationary phase and mobile phase lead to the separation of the analytes. In capillary electrochromatography capillaries, packed with HPLC stationary phase, are subjected to a high voltage. Separation is achieved by electrophoretic migration of solutes and differential partitioning.

References

  1. 1 2 3 4 5 6 Khaledi, M. G. (1997-09-12). "Micelles as separation media in high-performance liquid chromatography and high-performance capillary electrophoresis: overview and perspective". Journal of Chromatography A. 780 (1): 3–40. doi:10.1016/S0021-9673(97)00610-9.
  2. Armstrong, D. W. & Henry, S. J. (1980). "Use of an Aqueous Micellar Mobile Phase for Separation of Phenols and Polynuclear Aromatic Hydrocarbons via HPLC". Journal of Liquid Chromatography & Related Technologies. 3 (5): 657–662. doi:10.1080/01483918008060181.
  3. D.W. Armstrong, Sep. Purif. Methods 14 (1985) 213
  4. Meyer, V. (1999). Practical High-Performance Liquid Chromatography (3 ed.). John Wiley & Sons. pp. 14–16. ISBN   978-0-471-98373-6.
  5. 1 2 Baker, D. (1995). Capillary Electrophoreses. New York: Wiley Interscience. pp. 56–57. ISBN   978-0-471-11763-6.
  6. Poole, C. Journal of Chromatography A, 1998, 807, 307–310
  7. 1 2 Fischer, J. & Jandera, P. (1996-05-31). "Chromatographic behaviour in reversed-phase high-performance liquid chromatography with micellar and submicellar mobile phases: effects of the organic modifier". Journal of Chromatography B. 681 (1): 3–19. doi:10.1016/0378-4347(95)00538-2. PMID   8798907.[ dead link ]
  8. 1 2 Armstrong, Daniel W. & Nome, Faruk (September 1981). "Partitioning behavior of solutes eluted with micellar mobile phases in liquid chromatography". Analytical Chemistry. 53 (11): 1662–1666. doi:10.1021/ac00234a026.
  9. 1 2 Guermouche, M. H.; Habel, D.; Guermouche, S. (June 1998). "Theoretical aspects of micellar liquid chromatography using C12DAPS surfactant". Fluid Phase Equilibria. 147 (1–2): 301–307. doi:10.1016/S0378-3812(98)00242-8.
  10. 1 2 Jandera, P. & Fischer, J. (1996-03-29). "Chromatographic behaviour in reversed-phase high-performance liquid chromatography with micellar and submicellar mobile phases". Journal of Chromatography A. 728 (1–2): 279–298. doi:10.1016/0021-9673(95)00955-8.[ dead link ]
  11. 1 2 3 Foley, J.P. Analytica Chimica Acta, 1990, 231, 237-247
  12. 1 2 Marina, M. L. & Garcia, M. A. (1997-09-12). "Evaluation of distribution coefficients in micellar liquid chromatography". Journal of Chromatography A. 780 (1–2): 103–116. doi:10.1016/S0021-9673(97)00329-4.[ dead link ]
  13. 1 2 3 4 Garcia-Alvarez-Coque, M.C.; Torres-Lapasio, J.R.; Baeza-Baeza, J.J.; Journal of Chromatography A, 1997, 780, 129-148
  14. 1 2 Rukhadze, M.; Bezarashvili, G.; Sebiskveradze, M.; Meyer, V. (1998-05-01). "Separation of barbiturates with micellar liquid chromatography and optimization by a second order mathematical design". Journal of Chromatography A. 805 (1–2): 45–53. doi:10.1016/S0021-9673(97)01301-0.[ dead link ]
  15. 1 2 Jandera, P.; Fischer, J.; Effenberger, H. (1998-05-20). "Characterisation of retention in micellar high-performance liquid chromatography and in micellar electrokinetic chromatography using lipophilicity and polarity indices". Journal of Chromatography A. 807 (1): 57–70. doi:10.1016/S0021-9673(98)00067-3.[ dead link ]
  16. 1 2 Cuenca-Benito, M.; Sagrado, S.; Villanueva-Camanas, R.; Medina-Hernandez, M; Journal of Chromatography A, 1998, 814, 121-132
  17. 1 2 Quiñones-Torrelo, C.; Sagrado, S.; Villanueva-Camañas, R. M.; Medina-Hernández, M. J. (1999-07-24). "Development of Predictive Retention−Activity Relationship Models of Tricyclic Antidepressants by Micellar Liquid Chromatography". Journal of Medicinal Chemistry. 42 (16): 3154–3162. doi:10.1021/jm9910369. PMID   10447960.
  18. 1 2 Escuder-Gilabert, L.; Sanchis-Mallols, J. M.; Sagrado, S.; Medina-Hernández, M. J.; Villanueva-Camañas, R. M. (1998-10-09). "Chromatographic quantitation of the hydrophobicity of ionic compounds by the use of micellar mobile phases". Journal of Chromatography A. 823 (1–2): 549–559. doi:10.1016/S0021-9673(98)00456-7.[ dead link ]
  19. 1 2 3 4 Bethod, Alain (1997-09-12). "Causes and remediation of reduced efficiency in micellar liquid chromatography". Journal of Chromatography A. 780 (1–2): 191–206. doi:10.1016/S0021-9673(97)00195-7.[ dead link ]
  20. 1 2 Rapado Martínez, I.; Villanueva Camañas, R. M.; García Alvarez-Coque, M. C. (1998-12-05). "Micellar Liquid Chromatography: A Worthy Technique for the Determination of β-Antagonists in Urine Samples". Analytical Chemistry. 71 (2): 319–326. doi:10.1021/ac980472k. PMID   9949726.
  21. 1 2 Menéndez-Fraga, P.; Blanco-Gonzalez, E.; Sanz-Medel, A.; Cannata-Andía, J. B. (1997-12-12). "Micellar versus reversed phase liquid chromatography for the determination of desferrioxamine and its chelates with aluminium and iron in uremic serum". Talanta. 45 (1): 25–33. doi:10.1016/S0039-9140(97)00097-0. PMID   18966977.[ dead link ]
  22. 1 2 Nishi, H. (1997-09-12). "Pharmaceutical applications of micelles in chromatography and electrophoresis". Journal of Chromatography A. 780 (1–2): 243–264. doi:10.1016/S0021-9673(97)00347-6. PMID   9335130.[ dead link ]
  23. 1 2 Okada, Tetsuo (1997-09-12). "Micellar chromatography of inorganic compounds". Journal of Chromatography A. 780 (1–2): 343–360. doi:10.1016/S0021-9673(97)00291-4.[ dead link ]