Safety of magnetic resonance imaging

Last updated
Medical MRI scanner MRI-Philips.JPG
Medical MRI scanner

Magnetic resonance imaging (MRI) is in general a safe technique, although injuries may occur as a result of failed safety procedures or human error. [1] During the last 150 years, thousands of papers focusing on the effects or side effects of magnetic or radiofrequency fields have been published. They can be categorized as incidental and physiological. [2] Contraindications to MRI include most cochlear implants and cardiac pacemakers, shrapnel and metallic foreign bodies in the eyes. The safety of MRI during the first trimester of pregnancy is uncertain, but it may be preferable to other options. [3] Since MRI does not use any ionizing radiation, its use generally is favored in preference to CT when either modality could yield the same information. [4] (In certain cases, MRI is not preferred as it may be more expensive, time-consuming and claustrophobia-exacerbating.)

Contents

Structure and certification

In an effort to standardize the roles and responsibilities of MRI professionals, an international consensus document, written and endorsed by major MRI and medical physics professional societies from around the globe, has been published formally. [5] The document outlines specific responsibilities for the following positions:

The American Board of Magnetic Resonance Safety (ABMRS) provides testing and board certification for each of the three positions, MRMD, MRSO, and MRSE. As most MRI accidents and injuries are directly attributable to decisions at the point of care, testing and certification of MRI professionals seeks to reduce the rates of MRI accidents and improve patient safety through the establishment of safety competency levels for MRI professionals.

Implants

MR safe sign.svg
MR-Safe sign
MR conditional sign.svg
MR-Conditional sign
MR unsafe sign.svg
MR-Unsafe sign

All patients are reviewed for contraindications prior to MRI scanning. Medical devices and implants are categorized as MR Safe, MR Conditional or MR Unsafe: [6]

The MRI environment may cause harm in patients with MR-Unsafe devices such as cochlear implants, aneurysm clips, and many permanent pacemakers. In November 1992, a patient with an undisclosed cerebral aneurysm clip was reported to have died shortly after an MRI exam. [7] Several deaths have been reported in patients with pacemakers who have undergone MRI scanning without appropriate precautions. [8] Increasingly, MR-conditional pacemakers are available for selected patients. [9]

Ferromagnetic foreign bodies such as shell fragments, or metallic implants such as surgical prostheses and ferromagnetic aneurysm clips also are potential risks. Interaction of the magnetic and radio frequency fields with such objects may lead to heating or torque of the object during an MRI. [10] MRI is contraindicated in those suspected with metallic foreign body in the eye. MRI may be considered if there is strong suspicion of non-metallic foreign body. [11]

Titanium and its alloys are safe from attraction and torque forces produced by the magnetic field, although there may be some risks associated with Lenz effect forces acting on titanium implants in sensitive areas within the subject, such as stapes implants in the inner ear. [12]

Intrauterine devices with copper are generally safe in MRI, but may become dislodged or even expelled, and it is therefore recommended to check the location of the IUD both before and after MRI. [13]

Other implants that are contraindicated in MRI includes: magnetic dental implants, tissue expander, artificial limb, hearing aid, catheters with metallic components such as Swan-Ganz catheter and piercing. [14] However, tooth amalgam is not contraindicated in MRI. [15]

Risk of implant heating under MRI

Titanium and its alloys can heat from the radiofrequency field, as well as the switched gradient field (due to Faraday's law of magnetic induction).

Implants with metallic or electrically conductive parts may interact with the switched gradient and/or RF fields used in MRI, causing trauma or burns. [16]

Following Faraday’s law, the change of the magnetic flux through such a device induces eddy currents in the device and the metal subsequently converts electric energy into thermal energy... Furthermore, under specific conditions… gradient switching induced heating of other conductive material such as titanium, nitinol, or 316 stainless steel must be expected.

Hansjörg Graf, Günter Steidle, Fritz Schick, Heating of metallic implants and instruments induced by gradient switching in a 1.5‐Tesla whole‐body unit [17]

The amount of heating that takes place has a number of contributing factors:

A primary safety concern related to MRI is metallic medical implant heating by absorbing radiofrequency (RF) energy. This risk depends on the metal type, shape and orientation, the static magnetic field strength and the pulse sequence type and parameters.

M Hasegawa, K Miyata, Y Abe, and T Ishigami, Radiofrequency heating of metallic dental devices during 3.0 T MRI [18]

Injuries have been reported by this heating of metallic implants:

Event Description: It was reported to Siemens that a patient suffered a second degree burn on her right forearm after examination on the magnetom trio system... The patient has a titanium rod and screws placed in the right humerus. Patient has no feeling in right arm. Approximately five hours after the examination, the patient reported redness and a second degree blister, approximately 8 cm in length and 1,5 cm in width on the upper right forearm... According to the investigation from our experts, the RF burn was likely caused due to the presence of a titanium rod and screws placed in the right humerus.

FDA, MAUDE Adverse Event Report [19]

Projectile risk

The very high strength of the magnetic field may cause projectile effect (or "missile-effect") accidents, where ferromagnetic objects are attracted to the center of the magnet. Pennsylvania reported 27 cases of objects becoming projectiles in the MRI environment between 2004 and 2008. [20] There have been incidents of injury and death. [21] [22] In one case, a six-year-old boy died in July 2001, during an MRI exam at the Westchester Medical Center, New York, after a metal oxygen tank was pulled across the room and crushed the child's head. [23] [24] To reduce the risk of projectile accidents, ferromagnetic objects and devices are typically prohibited near the MRI scanner, and patients undergoing MRI examinations must remove all metallic objects, often by changing into a gown or scrubs. Some radiology departments use ferromagnetic detection devices to ensure that no ferromagnetic objects enter the scanner room. [25] [26]

MRI-EEG

In research settings, structural MRI or functional MRI (fMRI) may be combined with EEG (electroencephalography) under the condition that the EEG equipment is MR-compatible. Although EEG equipment (electrodes, amplifiers, and peripherals) are either approved for research or clinical use, the same MR Safe, MR Conditional and MR Unsafe terminology applies. With the growth of the use of MR technology, the U.S. Food & Drug Administration [FDA] recognized the need for a consensus on standards of practice, and the FDA sought out ASTM International [ASTM] to achieve them. Committee F04 [27] of ASTM developed F2503, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment. [28]

Genotoxic effects

There is no proven risk of biological harm from any aspect of an MRI scan, including very powerful static magnetic fields, gradient magnetic fields, or radio frequency waves. [29] [23] Some studies have suggested possible genotoxic (i.e., potentially carcinogenic) effects of MRI scanning through micronuclei induction and DNA double strand breaks in vivo and in vitro, [30] [31] [32] however, in most, if not all cases, others have been unable to repeat or validate the results of these studies, [29] [23] and the majority of research shows no genotoxic, or otherwise harmful, effects caused by any part of MRI. [29] A recent study confirmed that MRI using some of the most potentially-risky parameters tested to date (7-tesla static magnetic field, 70 mT/m gradient magnetic field, and maximum strength radio frequency waves) did not cause any DNA damage in vitro. [33]

Peripheral nerve stimulation

The rapid switching on and off of the magnetic field gradients is capable of causing nerve stimulation. Volunteers report a twitching sensation when exposed to rapidly switched fields, particularly in their extremities. [34] [35] The reason the peripheral nerves are stimulated is that the changing field increases with distance from the center of the gradient coils (which more or less coincides with the center of the magnet). [36] Although PNS was not a problem for the slow, weak gradients used in the early days of MRI, the strong, rapidly switched gradients used in techniques such as EPI, fMRI, diffusion MRI, etc. are capable of inducing PNS. American and European regulatory agencies insist that manufacturers stay below specified dB/dt limits (dB/dt is the change in magnetic field strength per unit time), or else prove that no PNS is induced for any imaging sequence. As a result of dB/dt limitation, commercial MRI systems cannot use the full rated power of their gradient amplifiers.

Heating caused by absorption of radio waves

Every MRI scanner has a powerful radio transmitter that generates the electromagnetic field that excites the spins. If the body absorbs the energy, heating occurs. For this reason, the transmitter rate at which energy is absorbed by the body must be limited (see Specific absorption rate). It has been claimed that tattoos made with iron-containing dyes may lead to burns on the subject's body. [37] [38] Cosmetics are very unlikely to undergo heating, as well as body lotions, since the outcome of the reactions between those with the radio waves is unknown. The best option for clothing is 100% cotton.

There are several positions strictly forbidden during measurement such as crossing arms and legs, and the patient's body may not create loops of any kind for the RF during the measurement.

Acoustic noise

Switching of field gradients causes a change in the Lorentz force experienced by the gradient coils, producing minute expansions and contractions of the coil. As the switching typically is in the audible frequency range, the resulting vibration produces loud noises (clicking, banging or beeping). This behaviour, of sound being generated by the vibration of the conducting components, is described as a coupled acousto-magneto-mechanical system, solutions to which provide useful insight to the behaviour of the scanners. [39] This is most marked with high-field machines, [40] and rapid-imaging techniques in which sound pressure levels may reach 120 dB(A) (equivalent to a jet engine at take-off), [41] and therefore, appropriate ear protection is essential for anyone inside the MRI scanner room during the examination. [42]

Radio frequency in itself does not cause audible noises (at least for human beings), since modern systems are using frequencies of 8.5 MHz (0.2 T system) or higher. [43]

Cryogens

As described in the Physics of magnetic resonance imaging article, many MRI scanners rely on cryogenic liquids to enable the superconducting capabilities of the electromagnetic coils within. Although the cryogenic liquids used are non-toxic, their physical properties present specific hazards. [44]

An unintentional shut-down of a superconducting electromagnet, an event known as "quench", involves the rapid boiling of liquid helium from the device. If the rapidly expanding helium cannot be dissipated through an external vent, sometimes referred to as a 'quench pipe', it may be released into the scanner room where it may cause displacement of the oxygen and present a risk of asphyxiation. [45]

Oxygen deficiency monitors usually are used as a safety precaution. Liquid helium, the most commonly used cryogen in MRI, undergoes near explosive expansion as it changes from a liquid to gaseous state. The use of an oxygen monitor is important to ensure that oxygen levels are safe for patients and physicians. Rooms built for superconducting MRI equipment should be equipped with pressure relief mechanisms [46] and an exhaust fan, in addition to the required quench pipe.

Because a quench results in rapid loss of cryogens from the magnet, recommissioning the magnet is expensive and time-consuming. Spontaneous quenches are uncommon, but a quench also may be triggered by an equipment malfunction, an improper cryogen fill technique, contaminants inside the cryostat, or extreme magnetic or vibrational disturbances. [47] [48]

Pregnancy

No effects of MRI on the fetus have been demonstrated. [49] As opposed to many other forms of medical imaging in pregnancy, MRI avoids the use of ionizing radiation, to which the fetus is particularly sensitive. As a precaution, however, many guidelines recommend pregnant women only undergo MRI when essential, especially during the first trimester. [50]

The concerns in pregnancy are the same as for MRI in general, but the fetus may be more sensitive to the effects—particularly to heating and to noise. The use of gadolinium-based contrast media in pregnancy is an off-label indication and may be administered only in the lowest dose required to provide essential diagnostic information. [51]

Despite these concerns, MRI is rapidly growing in importance as a way of diagnosing and monitoring congenital defects of the fetus because it is able to provide more diagnostic information than ultrasound and it lacks the ionizing radiation of CT. MRI without contrast agents is the imaging mode of choice for pre-surgical, in-utero diagnosis and evaluation of fetal tumors, primarily teratomas, facilitating open fetal surgery, other fetal interventions, and planning for procedures (such as the EXIT procedure) to safely deliver and treat babies whose defects would otherwise be fatal. [52] [53]

Claustrophobia and discomfort

Although painless, MRI scans may be unpleasant for those who are claustrophobic or otherwise uncomfortable with the imaging device surrounding them. Older closed bore MRI systems have a fairly long tube or tunnel. The part of the body being imaged must lie at the center of the magnet, which is at the absolute center of the tunnel. Because scan times on these older scanners may be long (occasionally up to 40 minutes for the entire procedure), people with even mild claustrophobia are sometimes unable to tolerate an MRI scan without management. Some modern scanners have larger bores (up to 70 cm) and scan times are shorter. A 1.5 T wide short bore scanner increases the examination success rate in patients with claustrophobia and substantially reduces the need for anesthesia-assisted MRI examinations even when claustrophobia is severe. [54]

Alternative scanner designs, such as open or upright systems, may be helpful where these are available. Although open scanners have increased in popularity, they produce inferior scan quality because they operate at lower magnetic fields than closed scanners. Commercial 1.5-tesla open systems have become available recently, however, providing much better image quality than previous lower field strength open models. [55]

Mirror glasses may be used to help create the illusion of openness. The mirrors are angled at 45 degrees, allowing the patient to look down their body and out the end of the imaging area. The appearance is of an open tube pointing upward (as seen when lying in the imaging area). Even though one is able to see around the glasses and the proximity of the device is very evident, this illusion is quite persuasive and relieves the claustrophobic feeling.

For young children who cannot hold still or would be frightened during the examination, chemical sedation or general anesthesia are the norm. Some hospitals encourage children to pretend the MRI machine is a spaceship or other adventure. [56] Certain hospitals with Children's wards have decorated scanners for this purpose, such as that at the Boston Children's Hospital, which operates a scanner with a special casing designed to resemble a sandcastle. [57]

Obese patients and pregnant women may find the MRI machine a tight fit. Pregnant women in the third trimester also may have difficulty lying on their backs for an hour or more without moving.

MRI versus CT

MRI and computed tomography (CT) are complementary imaging technologies and each has advantages and limitations for particular applications. CT is more widely used than MRI in OECD countries with a mean of 132 vs. 46 exams per 1000 population performed respectively. [58] A concern is the potential for CT to contribute to radiation-induced cancer and in 2007 it was estimated that 0.4% of current cancers in the United States were due to CTs performed in the past, and that in the future this figure may rise to 1.5–2% based on historical rates of CT usage. [59] An Australian study found that one in every 1800 CT scans was associated with an excess cancer. [60] An advantage of MRI is that no ionizing radiation is used and so it is recommended over CT when either approach could yield the same diagnostic information. [4] Although the cost of MRI has fallen, making it more competitive with CT, there are not many common imaging scenarios in which MRI can simply replace CT, however, this substitution has been suggested for the imaging of liver disease. [61] The effect of low doses of radiation on carcinogenesis also are disputed. [62] Although MRI is associated with biological effects, these have not been proven to cause measurable harm. [29]

Iodinated contrast medium is routinely used in CT and the main adverse events are anaphylactoid reactions and nephrotoxicity. [63] Commonly used MRI contrast agents have a good safety profile, but linear non-ionic agents in particular have been implicated in nephrogenic systemic fibrosis in patients with severely impaired renal function. [64]

MRI is contraindicated in the presence of MR-unsafe implants, and although these patients may be imaged with CT, beam hardening artefact from metallic devices, such as pacemakers and implantable cardioverter-defibrillators, also may affect image quality. [65] MRI is a longer investigation than CT and an exam may take between 20 and 40 minutes depending on complexity. [66]

Guidance

Safety issues, including the potential for biostimulation device interference, movement of ferromagnetic bodies, and incidental localized heating, have been addressed in the American College of Radiology's White Paper on MR Safety, which originally was published in 2002 and expanded in 2004. The ACR White Paper on MR Safety has been rewritten and was released early in 2007 under the new title ACR Guidance Document for Safe MR Practices.

In December 2007, the Medicines and Healthcare products Regulatory Agency (MHRA), a UK healthcare regulatory body, issued their Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use. In February 2008, the Joint Commission, a U.S. healthcare accrediting organization, issued a Sentinel Event Alert #38, their highest patient safety advisory, on MRI safety issues. In July 2008, the United States Veterans Administration, a federal governmental agency serving the healthcare needs of former military personnel, issued a substantial revision to their MRI Design Guide, [67] that includes physical and facility safety considerations.

The European Directive on electromagnetic fields

This Directive (2013/35/EU – electromagnetic fields) [68] covers all known direct biophysical effects and indirect effects caused by electromagnetic fields within the EU and repealed the 2004/40/EC directive. The deadline for implementation of the new directive was 1 July 2016. Article 10 of the directive sets out the scope of the derogation for MRI, stating that the exposure limits may be exceeded during "the installation, testing, use, development, maintenance of or research related to magnetic resonance imaging (MRI) equipment for patients in the health sector, provided that certain conditions are met." Uncertainties remain regarding the scope and conditions of this derogation. [69]

Related Research Articles

<span class="mw-page-title-main">Magnetic resonance imaging</span> Medical imaging technique

Magnetic resonance imaging (MRI) is a medical imaging technique used in radiology to form pictures of the anatomy and the physiological processes inside the body. MRI scanners use strong magnetic fields, magnetic field gradients, and radio waves to generate images of the organs in the body. MRI does not involve X-rays or the use of ionizing radiation, which distinguishes it from computed tomography (CT) and positron emission tomography (PET) scans. MRI is a medical application of nuclear magnetic resonance (NMR) which can also be used for imaging in other NMR applications, such as NMR spectroscopy.

<span class="mw-page-title-main">Radiology</span> Branch of medicine

Radiology is the medical specialty that uses medical imaging to diagnose diseases and guide their treatment, within the bodies of humans and other animals. It began with radiography, but today it includes all imaging modalities, including those that use no ionizing electromagnetic radiation, as well as others that do, such as computed tomography (CT), fluoroscopy, and nuclear medicine including positron emission tomography (PET). Interventional radiology is the performance of usually minimally invasive medical procedures with the guidance of imaging technologies such as those mentioned above.

<span class="mw-page-title-main">Medical imaging</span> Technique and process of creating visual representations of the interior of a body

Medical imaging is the technique and process of imaging the interior of a body for clinical analysis and medical intervention, as well as visual representation of the function of some organs or tissues (physiology). Medical imaging seeks to reveal internal structures hidden by the skin and bones, as well as to diagnose and treat disease. Medical imaging also establishes a database of normal anatomy and physiology to make it possible to identify abnormalities. Although imaging of removed organs and tissues can be performed for medical reasons, such procedures are usually considered part of pathology instead of medical imaging.

Image-guided surgery (IGS) is any surgical procedure where the surgeon uses tracked surgical instruments in conjunction with preoperative or intraoperative images in order to directly or indirectly guide the procedure. Image guided surgery systems use cameras, ultrasonic, electromagnetic or a combination of fields to capture and relay the patient's anatomy and the surgeon's precise movements in relation to the patient, to computer monitors in the operating room or to augmented reality headsets. This is generally performed in real-time though there may be delays of seconds or minutes depending on the modality and application.

<span class="mw-page-title-main">Inferior vena cava filter</span>

An inferior vena cava filter is a medical device made of metal that is implanted by vascular surgeons or interventional radiologists into the inferior vena cava to prevent a life-threatening pulmonary embolism (PE) or venous thromboembolism (VTE).

<span class="mw-page-title-main">Magnetic resonance angiography</span> Group of techniques based on magnetic resonance imaging (MRI) to image blood vessels.

Magnetic resonance angiography (MRA) is a group of techniques based on magnetic resonance imaging (MRI) to image blood vessels. Magnetic resonance angiography is used to generate images of arteries in order to evaluate them for stenosis, occlusions, aneurysms or other abnormalities. MRA is often used to evaluate the arteries of the neck and brain, the thoracic and abdominal aorta, the renal arteries, and the legs.

<span class="mw-page-title-main">Neuroimaging</span> Set of techniques to measure and visualize aspects of the nervous system

Neuroimaging is the use of quantitative (computational) techniques to study the structure and function of the central nervous system, developed as an objective way of scientifically studying the healthy human brain in a non-invasive manner. Increasingly it is also being used for quantitative research studies of brain disease and psychiatric illness. Neuroimaging is highly multidisciplinary involving neuroscience, computer science, psychology and statistics, and is not a medical specialty. Neuroimaging is sometimes confused with neuroradiology.

<span class="mw-page-title-main">Interventional magnetic resonance imaging</span>

Interventional magnetic resonance imaging, also interventional MRI or IMRI, is the use of magnetic resonance imaging (MRI) to do interventional radiology procedures.

<span class="mw-page-title-main">Cardiac magnetic resonance imaging</span> Biomedical imaging technology

Cardiac magnetic resonance imaging, also known as cardiovascular MRI, is a magnetic resonance imaging (MRI) technology used for non-invasive assessment of the function and structure of the cardiovascular system. Conditions in which it is performed include congenital heart disease, cardiomyopathies and valvular heart disease, diseases of the aorta such as dissection, aneurysm and coarctation, coronary heart disease. It can also be used to look at pulmonary veins.

<span class="mw-page-title-main">Physics of magnetic resonance imaging</span> Overview article

Magnetic resonance imaging (MRI) is a medical imaging technique mostly used in radiology and nuclear medicine in order to investigate the anatomy and physiology of the body, and to detect pathologies including tumors, inflammation, neurological conditions such as stroke, disorders of muscles and joints, and abnormalities in the heart and blood vessels among other things. Contrast agents may be injected intravenously or into a joint to enhance the image and facilitate diagnosis. Unlike CT and X-ray, MRI uses no ionizing radiation and is, therefore, a safe procedure suitable for diagnosis in children and repeated runs. Patients with specific non-ferromagnetic metal implants, cochlear implants, and cardiac pacemakers nowadays may also have an MRI in spite of effects of the strong magnetic fields. This does not apply on older devices, and details for medical professionals are provided by the device's manufacturer.

<span class="mw-page-title-main">Real-time MRI</span> Type of MRI

Real-time magnetic resonance imaging (RT-MRI) refers to the continuous monitoring of moving objects in real time. Traditionally, real-time MRI was possible only with low image quality or low temporal resolution. An iterative reconstruction algorithm removed limitations. Radial FLASH MRI (real-time) yields a temporal resolution of 20 to 30 milliseconds for images with an in-plane resolution of 1.5 to 2.0 mm. Real-time MRI adds information about diseases of the joints and the heart. In many cases MRI examinations become easier and more comfortable for patients, especially for the patients who cannot calm their breathing or who have arrhythmia.

<span class="mw-page-title-main">PET-MRI</span>

Positron emission tomography–magnetic resonance imaging (PET–MRI) is a hybrid imaging technology that incorporates magnetic resonance imaging (MRI) soft tissue morphological imaging and positron emission tomography (PET) functional imaging.

Cardiac imaging refers to minimally invasive imaging of the heart using ultrasound, magnetic resonance imaging (MRI), computed tomography (CT), or nuclear medicine (NM) imaging with PET or SPECT. These cardiac techniques are otherwise referred to as echocardiography, Cardiac MRI, Cardiac CT, Cardiac PET and Cardiac SPECT including myocardial perfusion imaging.

<span class="mw-page-title-main">Computed tomography of the head</span> Cross-sectional X-rays of the head

Computed tomography of the head uses a series of X-rays in a CT scan of the head taken from many different directions; the resulting data is transformed into a series of cross sections of the brain using a computer program. CT images of the head are used to investigate and diagnose brain injuries and other neurological conditions, as well as other conditions involving the skull or sinuses; it used to guide some brain surgery procedures as well. CT scans expose the person getting them to ionizing radiation which has a risk of eventually causing cancer; some people have allergic reactions to contrast agents that are used in some CT procedures.

Synthetic MRI is a simulation method in Magnetic Resonance Imaging (MRI), for generating contrast weighted images based on measurement of tissue properties. The synthetic (simulated) images are generated after an MR study, from parametric maps of tissue properties. It is thereby possible to generate several contrast weightings from the same acquisition. This is different from conventional MRI, where the signal acquired from the tissue is used to generate an image directly, often generating only one contrast weighting per acquisition. The synthetic images are similar in appearance to those normally acquired with an MRI scanner.

Magnetic resonance enterography is a magnetic resonance imaging technique used to evaluate bowel wall features of both upper and lower gastro-intestinal tract, although it is usually used for small bowel evaluation. It is a less invasive technique with the advantages of no ionizing radiation exposure, multiplanarity and high contrast resolution for soft tissue.

<span class="mw-page-title-main">History of magnetic resonance imaging</span>

The history of magnetic resonance imaging (MRI) includes the work of many researchers who contributed to the discovery of nuclear magnetic resonance (NMR) and described the underlying physics of magnetic resonance imaging, starting early in the twentieth century. One researcher was American physicist Isidor Isaac Rabi who won the Nobel Prize in Physics in 1944 for his discovery of nuclear magnetic resonance, which is used in magnetic resonance imaging. MR imaging was invented by Paul C. Lauterbur who developed a mechanism to encode spatial information into an NMR signal using magnetic field gradients in September 1971; he published the theory behind it in March 1973.

<span class="mw-page-title-main">MRI pulse sequence</span> A pulse sequence during a medical test

An MRI pulse sequence in magnetic resonance imaging (MRI) is a particular setting of pulse sequences and pulsed field gradients, resulting in a particular image appearance.

Magnetic resonance myelography is a noninvasive medical imaging technique that can provide anatomic information about the subarachnoid space. It is a type of MRI examination that uses a contrast medium and magnetic resonance imaging scanner to detect pathology of the spinal cord, including the location of a spinal cord injury, cysts, tumors and other abnormalities. The procedure involves the injection of a gadolinium based contrast media into the cervical or lumbar spine, followed by the MRI scan.

Hyperpolarized gas MRI, also known as hyperpolarized helium-3 MRI or HPHe-3 MRI, is a medical imaging technique that uses hyperpolarized gases to improve the sensitivity and spatial resolution of magnetic resonance imaging (MRI). This technique has many potential applications in medicine, including the imaging of the lungs and other areas of the body with low tissue density.

References

  1. Watson, Robert E. (October 2015). "Lessons Learned from MRI Safety Events". Current Radiology Reports. 3 (10): 37. doi:10.1007/s40134-015-0122-z. S2CID   57880401.
  2. "Rinck, PA. Magnetic Resonance Imaging: Safety of Patients and Personnel. Free Offprint from Rinck PA. Magnetic Resonance in Medicine - A Critical Introduction. The Basic Textbook of the European Magnetic Resonance Forum. 12th edition, 2018/2020. BoD. ISBN 978-3-7460-9518-9".
  3. Wang, Page I.; Chong, Suzanne T.; Kielar, Ania Z.; Kelly, Aine M.; Knoepp, Ursula D.; Mazza, Michael B.; Goodsitt, Mitchell M. (April 2012). "Imaging of Pregnant and Lactating Patients: Part 1, Evidence-Based Review and Recommendations". American Journal of Roentgenology. 198 (4): 778–784. doi:10.2214/AJR.11.7405. PMID   22451541.
  4. 1 2 "iRefer". Royal College of Radiologists. Retrieved 10 November 2013.
  5. Calamante, Fernando; Ittermann, Bernd; Kanal, Emanuel; Norris, David (November 2016). "Recommended responsibilities for management of MR safety". Journal of Magnetic Resonance Imaging. 44 (5): 1067–1069. doi: 10.1002/jmri.25282 . PMID   27255437. S2CID   10815801.
  6. ASTM International (2005). "American Society for Testing and Materials (ASTM) International, Designation: F2503-05. Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment".{{cite journal}}: Cite journal requires |journal= (help)
  7. "MRI-Related Death of Patient With Aneurysm Clip". Food and Drug Administration. 25 November 1992. Retrieved 19 October 2016. FDA has learned of a fatal injury sustained by a patient with a cerebral aneurysm clip while she was being prepared for an MRI procedure. It was reported that upon exposure to the magnetic field in the room, the clip moved and lacerated the patient's middle cerebral artery. Subsequently, the explanted device was shown to be magnetically active. This particular style or clip, which was implanted in 1978, was listed in several articles and recent medical texts as non-deflecting in a magnetic field.
  8. "Physics of magnetic resonance imaging". My-MS.org. Retrieved 27 April 2012.
  9. Colletti, Patrick M.; Shinbane, Jerold S.; Shellock, Frank G. (September 2011). "'MR-Conditional' Pacemakers: The Radiologist's Role in Multidisciplinary Management". American Journal of Roentgenology. 197 (3): W457–W459. doi:10.2214/AJR.11.7120. PMID   21862773.
  10. "Magnetic resonance safety policy of ucsf". University of California, San Francisco. Retrieved 28 April 2012.
  11. "Identifying Intraocular Foreign Bodies - American Academy of Ophthalmology". October 2007.
  12. Ghadimi, Maryam; Sapra, Amit (2021). "Magnetic Resonance Imaging Contraindications". Treasure Island (FL): StatPearls Publishing. PMID   31869133.{{cite journal}}: Cite journal requires |journal= (help)
  13. Berger-Kulemann, Vanessa; Einspieler, Henrik; Hachemian, Nilouparak; Prayer, Daniela; Trattnig, Siegfried; Weber, Michael; Ba-Ssalamah, Ahmed (2013). "Magnetic Field Interactions of Copper-Containing Intrauterine Devices in 3.0-Tesla Magnetic Resonance Imaging: In Vivo Study". Korean Journal of Radiology. 14 (3): 416–422. doi:10.3348/kjr.2013.14.3.416. PMC   3655294 . PMID   23690707.
  14. Ghadimi, M.; Sapra, A. (2022). "Magnetic Resonance Imaging Contraindications". Magnetic Resonance Imaging Contraindications - StatPearls - NCBI Bookshelf. StatPearls. PMID   31869133.
  15. Yilmaz S, Misirlioglu M (2013). "The effect of 3 T MRI on microleakage of amalgam restorations". Dento Maxillo Facial Radiology. 42 (8): 20130072. doi:10.1259/dmfr.20130072. PMC   3756742 . PMID   23674614.
  16. Health Protection Agency (2008). Protection of Patients and Volunteers Undergoing MRI Procedures (PDF). Health Protection Agency. ISBN   978-0-85951-623-5.
  17. Graf, Hansjörg; Steidle, Günter; Schick, Fritz (November 2007). "Heating of metallic implants and instruments induced by gradient switching in a 1.5-Tesla whole-body unit". Journal of Magnetic Resonance Imaging. 26 (5): 1328–1333. doi: 10.1002/jmri.21157 . PMID   17969167. S2CID   30634381.
  18. Hasegawa, M; Miyata, K; Abe, Y; Ishigami, T (May 2013). "Radiofrequency heating of metallic dental devices during 3.0 T MRI". Dentomaxillofacial Radiology. 42 (5). doi:10.1259/dmfr.20120234. PMC   3635774 . PMID   23520391.
  19. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=6295871& pc=LNH
  20. "Safety in the MR Environment: Ferromagnetic Projectile Objects in the MRI Scanner Room". Pa Patient Saf Advis. 6 (2): 56–62. June 2009. Archived from the original on 4 February 2015. Retrieved 4 February 2015.
  21. Randal C. Archibold, "Hospital Details Failures Leading to M.R.I. Fatality, The New York Times , August 22, 2001
  22. Donald G. McNeil Jr, "M.R.I.'s Strong Magnets Cited in Accidents," The New York Times , August 19, 2005.
  23. 1 2 3 Hartwig, Valentina; Giovannetti, Giulio; Vanello, Nicola; Lombardi, Massimo; Landini, Luigi; Simi, Silvana (10 June 2009). "Biological Effects and Safety in Magnetic Resonance Imaging: A Review". International Journal of Environmental Research and Public Health. 6 (6): 1778–1798. doi: 10.3390/ijerph6061778 . PMC   2705217 . PMID   19578460.
  24. Chen, David W. (31 July 2001). "Boy, 6, Dies Of Skull Injury During M.R.I." The New York Times . Retrieved 24 October 2019.
  25. "ACR Guidance Document for Safe MR Practices: 2007" . Retrieved 2 August 2010.
  26. "MRI Design Guide" (PDF). Archived from the original (PDF) on 14 July 2011. Retrieved 2 August 2010.
  27. "Committee F04 on Medical and Surgical Materials and Devices." Committee F04
  28. "Specification for Concrete Aggregates". doi:10.1520/C0033-03.{{cite journal}}: Cite journal requires |journal= (help)
  29. 1 2 3 4 Formica, Domenico; Silvestri, Sergio (December 2004). "Biological effects of exposure to magnetic resonance imaging: an overview". BioMedical Engineering OnLine. 3 (1): 11. doi: 10.1186/1475-925X-3-11 . PMC   419710 . PMID   15104797.
  30. Lee, Joong Won; Kim, Myeong Seong; Kim, Yang Jee; Choi, Young Joo; Lee, Younghyun; Chung, Hai Won (October 2011). "Genotoxic effects of 3 T magnetic resonance imaging in cultured human lymphocytes". Bioelectromagnetics. 32 (7): 535–542. doi:10.1002/bem.20664. PMID   21412810. S2CID   205467617.
  31. Simi, Silvana; Ballardin, Michela; Casella, Marta; De Marchi, Daniele; Hartwig, Valentina; Giovannetti, Giulio; Vanello, Nicola; Gabbriellini, Sabrina; Landini, Luigi; Lombardi, Massimo (October 2008). "Is the genotoxic effect of magnetic resonance negligible? Low persistence of micronucleus frequency in lymphocytes of individuals after cardiac scan". Mutation Research/Fundamental and Molecular Mechanisms of Mutagenesis. 645 (1–2): 39–43. Bibcode:2008MRFMM.645...39S. doi:10.1016/j.mrfmmm.2008.08.011. PMID   18804118.
  32. Suzuki, Y; Ikehata, M; Nakamura, K; Nishioka, M; Asanuma, K; Koana, T; Shimizu, H (November 2001). "Induction of micronuclei in mice exposed to static magnetic fields". Mutagenesis. 16 (6): 499–501. doi: 10.1093/mutage/16.6.499 . PMID   11682641.
  33. Fatahi, Mahsa; Reddig, Annika; Friebe, Björn; Hartig, Roland; Prihoda, Thomas J.; Ricke, Jens; Roggenbuck, Dirk; Reinhold, Dirk; Speck, Oliver; Speck, O (June 2016). "DNA double-strand breaks and micronuclei in human blood lymphocytes after repeated whole body exposures to 7T Magnetic Resonance Imaging". NeuroImage. 133: 288–293. doi:10.1016/j.neuroimage.2016.03.023. PMID   26994830. S2CID   3538334.
  34. Cohen, Mark S.; Weisskoff, Robert M.; Rzedzian, Richard R.; Kantor, Howard L. (May 1990). "Sensory stimulation by time-varying magnetic fields". Magnetic Resonance in Medicine. 14 (2): 409–414. doi:10.1002/mrm.1910140226. PMID   2345521. S2CID   36042581.
  35. Budinger, Thomas F.; Fischer, Hubertus; Hentschel, Dietmar; Reinfelder, Hans-Erich; Schmitt, Franz (November 1991). "Physiological Effects of Fast Oscillating Magnetic Field Gradients". Journal of Computer Assisted Tomography. 15 (6): 909–914. doi:10.1097/00004728-199111000-00001. PMID   1939767.
  36. Reilly JP (March 1989). "Peripheral nerve stimulation by induced electric currents: exposure to time-varying magnetic fields". Med Biol Eng Comput. 27 (2): 101–10. doi:10.1007/BF02446217. PMID   2689806. S2CID   23722382.
  37. Ross, James R.; Matava, Matthew J. (September 2011). "Tattoo-Induced Skin 'Burn' During Magnetic Resonance Imaging in a Professional Football Player: A Case Report". Sports Health: A Multidisciplinary Approach. 3 (5): 431–434. doi:10.1177/1941738111411698. PMC   3445217 . PMID   23016039.
  38. Rose Eveleth (March 6, 2014). "Some Tattoo Inks Can Burn You During an MRI".
  39. Bagwell S., Ledger P.D., Gil A.J., Mallett M., Kruip M. (2017). "A linearised hp–finite element framework for acousto-magneto-mechanical coupling in axisymmetric MRI scanners". International Journal for Numerical Methods in Engineering. 112 (10): 1323–1352. Bibcode:2017IJNME.112.1323B. doi: 10.1002/nme.5559 .{{cite journal}}: CS1 maint: multiple names: authors list (link)
  40. "The Evolution of Magnetic Resonance Imaging: 3T MRI in Clinical Applications" Archived 2013-06-15 at the Wayback Machine , Terry Duggan-Jahns, www.eradimaging.com
  41. Price, David L.; Wilde, Janet P. De; Papadaki, Annie M.; Curran, Jane S.; Kitney, Richard I. (2001). "Investigation of acoustic noise on 15 MRI scanners from 0.2 T to 3 T". Journal of Magnetic Resonance Imaging. 13 (2): 288–293. doi: 10.1002/1522-2586(200102)13:2<288::aid-jmri1041>3.0.co;2-p . PMID   11169836.
  42. The Open University 2007: Understanding Cardiovascular Diseases, course book for the lesson SK121 Understanding cardiovascular diseases , printed by Cambridge University Press, ISBN   978-0-7492-2677-0 (can be found at OUW), pages 220 and 224.
  43. "Introduction to MRI Physics, Page 4". www.simplyphysics.com. Retrieved 2017-06-09.
  44. "Safety data sheet Nitrogen, refrigerated, liquid" (PDF). BOC. Archived from the original (PDF) on 2013-10-19. Retrieved 2014-09-11.
  45. Kanal, Emanuel; Barkovich, A. James; Bell, Charlotte; Borgstede, James P.; Bradley, William G.; Froelich, Jerry W.; Gilk, Tobias; Gimbel, J. Rod; Gosbee, John; Kuhni-Kaminski, Ellisa; Lester, James W.; Nyenhuis, John; Parag, Yoav; Schaefer, Daniel J.; Sebek-Scoumis, Elizabeth A.; Weinreb, Jeffrey; Zaremba, Loren A.; Wilcox, Pamela; Lucey, Leonard; Sass, Nancy (June 2007). "ACR Guidance Document for Safe MR Practices: 2007". American Journal of Roentgenology. 188 (6): 1447–1474. doi:10.2214/AJR.06.1616. PMID   17515363.
  46. International Electrotechnical Commission 2008: Medical Electrical Equipment – Part 2-33: Particular requirements for basic safety and essential performance of magnetic resonance equipment for medical diagnosis, manufacturers' trade standards , published by International Electrotechnical Commission, ISBN   2-8318-9626-6 (can be found for purchase at ).
  47. "Cryogen Awareness Quenching and MRI Safety Training". Falck Productions. Retrieved 10 July 2012.
  48. "GE Health Care" (PDF). GE. Archived from the original (PDF) on 15 January 2013. Retrieved 10 July 2012.
  49. Alorainy, Ibrahim A.; Albadr, Fahad B.; Abujamea, Abdullah H. (July 2006). "Attitude towards MRI safety during pregnancy". Annals of Saudi Medicine. 26 (4): 306–309. doi:10.5144/0256-4947.2006.306. PMC   6074503 . PMID   16885635.
  50. Coakley, Fergus V.; Glenn, Orit A.; Qayyum, Aliya; Barkovich, Anthony J.; Goldstein, Ruth; Filly, Roy A. (January 2004). "Fetal MRI: A Developing Technique for the Developing Patient". American Journal of Roentgenology. 182 (1): 243–252. doi:10.2214/ajr.182.1.1820243. PMID   14684546.
  51. Webb, Judith AW; Thomsen, Henrik S (July 2013). "Gadolinium contrast media during pregnancy and lactation". Acta Radiologica. 54 (6): 599–600. doi:10.1177/0284185113484894. PMID   23966544. S2CID   20005180.
  52. Kathary, Nishita; Bulas, D. I.; Newman, Kurt D.; Schonberg, Rhonda L. (October 2001). "MRI imaging of fetal neck masses with airway compromise: utility in delivery planning". Pediatric Radiology. 31 (10): 727–731. doi:10.1007/s002470100527. PMID   11685443. S2CID   22826630.
  53. Mota, Raquel; Ramalho, Carla; Monteiro, Joaquim; Correia-Pinto, Jorge; Rodrigues, Manuela; Guimarães, Hercília; Spratley, Jorge; Macedo, Filipe; Matias, Alexandra; Montenegro, Nuno (2007). "Evolving Indications for the EXIT Procedure: The Usefulness of Combining Ultrasound and Fetal MRI". Fetal Diagnosis and Therapy. 22 (2): 107–111. doi:10.1159/000097106. PMID   17135754. S2CID   34157494. Our two cases stress once more the importance of combining fetal ultrasound and magnetic resonance imaging in the characterization of cervical masses and its usefulness in programming the procedure with a multidisciplinary team.
  54. Hunt, C.H.; Wood, C.P.; Lane, J.I.; Bolster, B.D.; Bernstein, M.A.; Witte, R.J. (September 2011). "Wide, Short Bore Magnetic Resonance at 1.5 T: Reducing the Failure Rate in Claustrophobic Patients". Clinical Neuroradiology. 21 (3): 141–144. doi:10.1007/s00062-011-0075-4. PMID   21598040. S2CID   25557946.
  55. "Siemens Introduces First 1.5 Tesla Open Bore MRI". Medical.siemens.com. 2004-07-29. Retrieved 2010-08-02.
  56. Kelley, Tom Kelley, David (18 October 2013). "Kids Were Terrified of Getting MRIs. Then One Man Figured Out a Better Way". Slate Magazine.{{cite news}}: CS1 maint: multiple names: authors list (link)
  57. "Magnetic Resonance Imaging (MRI)". Boston Children's Hospital. 12 September 2018. Archived from the original on 17 April 2022. Retrieved 12 September 2018.
  58. Health at a Glance 2011 OECD Indicators. OECD Publishing. 2011. doi:10.1787/health_glance-2011-en. hdl:10818/6660. ISBN   978-92-64-11153-0.[ page needed ]
  59. Brenner, David J.; Hall, Eric J. (2007-11-29). "Computed Tomography — An Increasing Source of Radiation Exposure". New England Journal of Medicine. 357 (22): 2277–2284. doi:10.1056/NEJMra072149. PMID   18046031. S2CID   2760372.
  60. Mathews, J. D.; Forsythe, A. V.; Brady, Z.; Butler, M. W.; Goergen, S. K.; Byrnes, G. B.; Giles, G. G.; Wallace, A. B.; Anderson, P. R.; Guiver, T. A.; McGale, P.; Cain, T. M.; Dowty, J. G.; Bickerstaffe, A. C.; Darby, S. C. (21 May 2013). "Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians". BMJ. 346 (may21 1): f2360. doi:10.1136/bmj.f2360. PMC   3660619 . PMID   23694687.
  61. Semelka, Richard C.; Armao, Diane M.; Elias, Jorge; Huda, Walter (May 2007). "Imaging strategies to reduce the risk of radiation in CT studies, including selective substitution with MRI". Journal of Magnetic Resonance Imaging. 25 (5): 900–909. doi: 10.1002/jmri.20895 . PMID   17457809. S2CID   5788891.
  62. Health risks from exposure to low levels of ionizing radiation : BEIR VII Phase . Washington, D.C.: National Academies Press. 2006. ISBN   978-0-309-09156-5.
  63. Bettmann, Michael A. (October 2004). "Frequently Asked Questions: Iodinated Contrast Agents". RadioGraphics. 24 (Suppl 1): S3–S10. doi:10.1148/rg.24si045519. PMID   15486247.
  64. "Nephrogenic Systemic Fibrosis" (PDF). ACR Manual on Contrast Material. American College of Radiology. Retrieved 13 October 2012.
  65. Mak, Gary S.; Truong, Quynh A. (October 2012). "Cardiac CT: Imaging of and Through Cardiac Devices". Current Cardiovascular Imaging Reports. 5 (5): 328–336. doi:10.1007/s12410-012-9150-8. PMC   3636997 . PMID   23626865.
  66. "MRI procedure". Royal College of Radiologists. Archived from the original on 24 July 2003. Retrieved 17 November 2013.
  67. "MRI Design Guide". United States Department of Veterans Affairs. April 2008. Retrieved 12 October 2012.
  68. "Directive 2013/35/EU of the European Parliament and of the Council. Official Journal of the European Union 2004 L179/1".
  69. Keevil, S F; Lomas, D J (December 2013). "The European Union physical agents (electromagnetic fields) directive: an update for the MRI community". The British Journal of Radiology. 86 (1032): 20130492. doi:10.1259/bjr.20130492. PMC   3856543 . PMID   24096591.