Hybrid mass spectrometer

Last updated

A hybrid mass spectrometer is a device for tandem mass spectrometry that consists of a combination of two or more m/z separation devices of different types.

Contents

Notation

Schematic of a quadrupole time-of-flight (QTOF) mass spectrometer. QTOF Schematic.gif
Schematic of a quadrupole time-of-flight (QTOF) mass spectrometer.

The different m/z separation elements of a hybrid mass spectrometer can be represented by a shorthand notation. The symbol Q represents a quadrupole mass analyzer, q is a radio frequency collision quadrupole, TOF is a time-of-flight mass spectrometer, B is a magnetic sector and E is an electric sector.

Sector quadrupole

A sector instrument can be combined with a collision quadrupole and quadrupole mass analyzer to form a hybrid instrument. [1] A BEqQ configuration with a magnetic sector (B), electric sector (E), collision quadrupole (q) and m/z selection quadrupole (Q) have been constructed [2] [3] and an instrument with two electric sectors (BEEQ) has been described. [4]

Quadrupole time-of-flight

Hybrid quadrupole time-of-flight mass spectrometer. Q-TOF.jpg
Hybrid quadrupole time-of-flight mass spectrometer.

A triple quadrupole mass spectrometer with the final quadrupole replaced by a time-of-flight device is known as a quadrupole time-of-flight instrument. [5] [6] Such an instrument can be represented as QqTOF.

Ion trap time-of-flight

In an ion trap instrument, ions are trapped in a quadrupole ion trap and then injected into the TOF. The trap can be 3-D [7] or a linear trap. [8]

Linear ion trap and Fourier transform mass analyzers

Hybrid linear ion trap Fourier transform ion cyclotron resonance mass spectrometer LTQ-FTICR.jpg
Hybrid linear ion trap Fourier transform ion cyclotron resonance mass spectrometer

A linear ion trap combined with a Fourier transform ion cyclotron resonance [9] or Orbitrap [10] [11] [12] mass spectrometer is marketed by Thermo Scientific as the LTQ FT and LTQ Orbitrap, respectively.

Related Research Articles

Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a mass spectrum, a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures.

<span class="mw-page-title-main">Tandem mass spectrometry</span> Type of mass spectrometry

Tandem mass spectrometry, also known as MS/MS or MS2, is a technique in instrumental analysis where two or more mass analyzers are coupled together using an additional reaction step to increase their abilities to analyse chemical samples. A common use of tandem MS is the analysis of biomolecules, such as proteins and peptides.

<span class="mw-page-title-main">Quadrupole mass analyzer</span>

The quadrupole mass analyzer, originally conceived by Nobel Laureate Wolfgang Paul and his student Helmut Steinwedel, also known as quadrupole mass filter, is one type of mass analyzer used in mass spectrometry. As the name implies, it consists of four cylindrical rods, set parallel to each other. In a quadrupole mass spectrometer (QMS) the quadrupole is the mass analyzer - the component of the instrument responsible for selecting sample ions based on their mass-to-charge ratio (m/z). Ions are separated in a quadrupole based on the stability of their trajectories in the oscillating electric fields that are applied to the rods.

<span class="mw-page-title-main">Matrix-assisted laser desorption/ionization</span> Ionization technique

In mass spectrometry, matrix-assisted laser desorption/ionization (MALDI) is an ionization technique that uses a laser energy absorbing matrix to create ions from large molecules with minimal fragmentation. It has been applied to the analysis of biomolecules and various organic molecules, which tend to be fragile and fragment when ionized by more conventional ionization methods. It is similar in character to electrospray ionization (ESI) in that both techniques are relatively soft ways of obtaining ions of large molecules in the gas phase, though MALDI typically produces far fewer multi-charged ions.

<span class="mw-page-title-main">History of mass spectrometry</span>

The history of mass spectrometry has its roots in physical and chemical studies regarding the nature of matter. The study of gas discharges in the mid 19th century led to the discovery of anode and cathode rays, which turned out to be positive ions and electrons. Improved capabilities in the separation of these positive ions enabled the discovery of stable isotopes of the elements. The first such discovery was with the element neon, which was shown by mass spectrometry to have at least two stable isotopes: 20Ne and 22Ne. Mass spectrometers were used in the Manhattan Project for the separation of isotopes of uranium necessary to create the atomic bomb.

<span class="mw-page-title-main">Orbitrap</span>

In mass spectrometry, Orbitrap is an ion trap mass analyzer consisting of an outer barrel-like electrode and a coaxial inner spindle-like electrode that traps ions in an orbital motion around the spindle. The image current from the trapped ions is detected and converted to a mass spectrum using the Fourier transform of the frequency signal.

Soft laser desorption (SLD) is laser desorption of large molecules that results in ionization without fragmentation. "Soft" in the context of ion formation means forming ions without breaking chemical bonds. "Hard" ionization is the formation of ions with the breaking of bonds and the formation of fragment ions.

<span class="mw-page-title-main">Electron-transfer dissociation</span>

Electron-transfer dissociation (ETD) is a method of fragmenting multiply-charged gaseous macromolecules in a mass spectrometer between the stages of tandem mass spectrometry (MS/MS). Similar to electron-capture dissociation, ETD induces fragmentation of large, multiply-charged cations by transferring electrons to them. ETD is used extensively with polymers and biological molecules such as proteins and peptides for sequence analysis. Transferring an electron causes peptide backbone cleavage into c- and z-ions while leaving labile post translational modifications (PTM) intact. The technique only works well for higher charge state peptide or polymer ions (z>2). However, relative to collision-induced dissociation (CID), ETD is advantageous for the fragmentation of longer peptides or even entire proteins. This makes the technique important for top-down proteomics. The method was developed by Hunt and coworkers at the University of Virginia.

<span class="mw-page-title-main">Time-of-flight mass spectrometry</span> Method of mass spectrometry

Time-of-flight mass spectrometry (TOFMS) is a method of mass spectrometry in which an ion's mass-to-charge ratio is determined by a time of flight measurement. Ions are accelerated by an electric field of known strength. This acceleration results in an ion having the same kinetic energy as any other ion that has the same charge. The velocity of the ion depends on the mass-to-charge ratio. The time that it subsequently takes for the ion to reach a detector at a known distance is measured. This time will depend on the velocity of the ion, and therefore is a measure of its mass-to-charge ratio. From this ratio and known experimental parameters, one can identify the ion.

Robert Graham Cooks is the Henry Bohn Hass Distinguished Professor of Chemistry in the Aston Laboratories for Mass Spectrometry at Purdue University. He is an ISI Highly Cited Chemist, with over 1,000 publications and an H-index of 144.

<span class="mw-page-title-main">Ion-mobility spectrometry–mass spectrometry</span>

Ion mobility spectrometry–mass spectrometry (IMS-MS) is an analytical chemistry method that separates gas phase ions based on their interaction with a collision gas and their masses. In the first step, the ions are separated according to their mobility through a buffer gas on a millisecond timescale using an ion mobility spectrometer. The separated ions are then introduced into a mass analyzer in a second step where their mass-to-charge ratios can be determined on a microsecond timescale. The effective separation of analytes achieved with this method makes it widely applicable in the analysis of complex samples such as in proteomics and metabolomics.

<span class="mw-page-title-main">Triple quadrupole mass spectrometer</span>

A triple quadrupole mass spectrometer (TQMS), is a tandem mass spectrometer consisting of two quadrupole mass analyzers in series, with a (non-mass-resolving) radio frequency (RF)–only quadrupole between them to act as a cell for collision-induced dissociation. This configuration is often abbreviated QqQ, here Q1q2Q3.

<span class="mw-page-title-main">Reflectron</span> Time-of-flight mass spectrometer with an ion mirror

A reflectron is a type of time-of-flight mass spectrometer that comprises a pulsed ion source, field-free region, ion mirror, and ion detector and uses a static or time dependent electric field in the ion mirror to reverse the direction of travel of the ions entering it. Using the reflectron, one can substantially diminish a spread of flight times of the ions with the same mass-to-charge ratio (m/z) caused by spread in kinetic energy of these ions measured at the exit from the ion source.

<span class="mw-page-title-main">Linear ion trap</span>

The linear ion trap (LIT) is a type of ion trap mass spectrometer.

<span class="mw-page-title-main">Collision-induced dissociation</span> Mass spectrometry technique to induce fragmentation of selected ions in the gas phase

Collision-induced dissociation (CID), also known as collisionally activated dissociation (CAD), is a mass spectrometry technique to induce fragmentation of selected ions in the gas phase. The selected ions are usually accelerated by applying an electrical potential to increase the ion kinetic energy and then allowed to collide with neutral molecules. In the collision some of the kinetic energy is converted into internal energy which results in bond breakage and the fragmentation of the molecular ion into smaller fragments. These fragment ions can then be analyzed by tandem mass spectrometry.

<span class="mw-page-title-main">Robert J. Cotter</span>

Robert J. Cotter was an American chemist and mass spectrometrist. His research contributed to many early advances in the field of time-of-flight mass spectrometry. From 1998 to 2000 he was president of the American Society for Mass Spectrometry. Cotter was also a co-investigator on the Mars Organic Molecule Analyzer (MOMA) project, developing a miniaturized, low power consumption ion trap/time-of-flight mass spectrometer that was to be deployed with the ExoMars rover, now the Rosalind Franklin rover.

<span class="mw-page-title-main">Miniature mass spectrometer</span>

A miniature mass spectrometer (MMS) is a type of mass spectrometer (MS) which has small size and weight and can be understood as a portable or handheld device. Current lab-scale mass spectrometers however, usually weigh hundreds of pounds and can cost on the range from thousands to millions of dollars. One purpose of producing MMS is for in situ analysis. This in situ analysis can lead to much simpler mass spectrometer operation such that non-technical personnel like physicians at the bedside, firefighters in a burning factory, food safety inspectors in a warehouse, or airport security at airport checkpoints, etc. can analyze samples themselves saving the time, effort, and cost of having the sample run by a trained MS technician offsite. Although, reducing the size of MS can lead to a poorer performance of the instrument versus current analytical laboratory standards, MMS is designed to maintain sufficient resolutions, detection limits, accuracy, and especially the capability of automatic operation. These features are necessary for the specific in-situ applications of MMS mentioned above.

<span class="mw-page-title-main">Gary Glish</span>

Gary Glish is an American analytical chemist at the University of North Carolina at Chapel Hill. He is a leading researcher in the fields of mass spectrometry, ion chemistry, and biomolecule analysis.

<span class="mw-page-title-main">Digital ion trap</span> Scientific analytical tool

The digital ion trap (DIT) is an quadrupole ion trap driven by digital signals, typically in a rectangular waveform, generated by switching rapidly between discrete DC voltage levels. The digital ion trap has been mainly developed as a mass analyzer.

References

  1. Glish, G.; McLuckey, S; Ridley, T; Cooks, R (1982). "A new "hybrid" sector/quadrupole mass spectrometer for mass spectrometry/mass spectrometry". International Journal of Mass Spectrometry and Ion Physics. 41 (3): 157–177. Bibcode:1982IJMSI..41..157G. doi:10.1016/0020-7381(82)85032-8.
  2. Schoen, A.; Amy, J.W.; Ciupek, J.D.; Cooks, R.G.; Dobberstein, P.; Jung, G. (1985). "A hybrid BEQQ mass spectrometer". International Journal of Mass Spectrometry and Ion Processes. 65 (1–2): 125–140. Bibcode:1985IJMSI..65..125S. doi:10.1016/0168-1176(85)85059-X.
  3. Harrison, A. (1986). "A hybrid BEQQ mass spectrometer for studies in gaseous ion chemistry". International Journal of Mass Spectrometry and Ion Processes. 74 (1): 13–31. Bibcode:1986IJMSI..74...13H. doi:10.1016/0168-1176(86)85020-0.
  4. Winger, B. E.; Laue, H. -J.; Horning, S. R.; Julian, R. K.; Lammert, S. A.; Riederer, D. E.; Cooks, R. G. (1992). "Hybrid BEEQ tandem mass spectrometer for the study of ion/surface collision processes". Review of Scientific Instruments. 63 (12): 5613. Bibcode:1992RScI...63.5613W. doi:10.1063/1.1143391.
  5. Shevchenko A, Loboda A, Shevchenko A, Ens W, Standing KG (May 2000). "MALDI quadrupole time-of-flight mass spectrometry: a powerful tool for proteomic research". Anal. Chem. 72 (9): 2132–41. doi:10.1021/ac9913659. PMID   10815976.
  6. Steen H, Küster B, Mann M (July 2001). "Quadrupole time-of-flight versus triple-quadrupole mass spectrometry for the determination of phosphopeptides by precursor ion scanning". J Mass Spectrom. 36 (7): 782–90. Bibcode:2001JMSp...36..782S. doi:10.1002/jms.174. PMID   11473401.
  7. Fountain ST, Lee H, Lubman DM (May 1994). "Ion fragmentation activated by matrix-assisted laser desorption/ionization in an ion-trap/reflectron time-of-flight device". Rapid Commun. Mass Spectrom. 8 (5): 407–16. Bibcode:1994RCMS....8..407F. doi:10.1002/rcm.1290080514. PMID   7517726.
  8. Campbell, J. M.; Collings, B. A.; Douglas, D. J. (1998). "A new linear ion trap time-of-flight system with tandem mass spectrometry capabilities". Rapid Communications in Mass Spectrometry. 12 (20): 1463–1474. doi:10.1002/(SICI)1097-0231(19981030)12:20<1463::AID-RCM357>3.0.CO;2-H.
  9. Syka JE, Marto JA, Bai DL, et al. (2004). "Novel linear quadrupole ion trap/FT mass spectrometer: performance characterization and use in the comparative analysis of histone H3 post-translational modifications". J. Proteome Res. 3 (3): 621–6. doi:10.1021/pr0499794. PMID   15253445.
  10. Makarov A, Denisov E, Kholomeev A, Balschun W, Lange O, Strupat K, Horning S (2006). "Performance evaluation of a hybrid linear ion trap/orbitrap mass spectrometer". Anal. Chem. 78 (7): 2113–20. doi:10.1021/ac0518811. PMID   16579588.
  11. Olsen JV, de Godoy LM, Li G, et al. (December 2005). "Parts per million mass accuracy on an Orbitrap mass spectrometer via lock mass injection into a C-trap". Mol. Cell. Proteomics. 4 (12): 2010–21. doi: 10.1074/mcp.T500030-MCP200 . PMID   16249172.
  12. Yates JR, Cociorva D, Liao L, Zabrouskov V (January 2006). "Performance of a linear ion trap-Orbitrap hybrid for peptide analysis". Anal. Chem. 78 (2): 493–500. doi:10.1021/ac0514624. PMID   16408932.