Collision-induced dissociation

Last updated
Collision cell from a Waters Xevo TQ-S triple quadrupole mass spectrometer. Collision Cell.jpg
Collision cell from a Waters Xevo TQ-S triple quadrupole mass spectrometer.

Collision-induced dissociation (CID), also known as collisionally activated dissociation (CAD), is a mass spectrometry technique to induce fragmentation of selected ions in the gas phase. [1] [2] The selected ions (typically molecular ions or protonated molecules) are usually accelerated by applying an electrical potential to increase the ion kinetic energy and then allowed to collide with neutral molecules (often helium, nitrogen, or argon). In the collision, some of the kinetic energy is converted into internal energy which results in bond breakage and the fragmentation of the molecular ion into smaller fragments. These fragment ions can then be analyzed by tandem mass spectrometry.

Contents

CID and the fragment ions produced by CID are used for several purposes. Partial or complete structural determination can be achieved. In some cases, identity can be established based on previous knowledge without determining structure. Another use is in simply achieving more sensitive and specific detection. By detecting a unique fragment ion, the precursor ion can be detected in the presence of other ions of the same m/z value (mass-to-charge ratio), reducing the background and increasing the limit of detection.

Low-energy CID and high-energy CID

Low-energy CID is typically carried out with ion kinetic energies less than approximately 1 kiloelectron volt (1 keV). Low-energy CID is highly efficient in fragmenting the selected precursor ions, but the type of fragment ions observed in low-energy CID is strongly dependent on the ion kinetic energy. Very low collision energies favor ion structure rearrangement, and the probability of direct bond cleavage increases as ion kinetic energy increases, leading to higher ion internal energies. High-energy CID (HECID) is carried out in magnetic sector mass spectrometers or tandem magnetic sector mass spectrometers and in tandem time-of-flight mass spectrometers (TOF/TOF). High-energy CID involves ion kinetic energies in the kilovolt range (typically 1 keV to 20 keV). High-energy CID can produce some types of fragment ions that are not formed in low-energy CID, such as charge-remote fragmentation in molecules with hydrocarbon substructures or sidechain fragmentation in peptides.

Triple quadrupole mass spectrometers

In a triple quadrupole mass spectrometer there are three quadrupoles. The first quadrupole termed "Q1" can act as a mass filter and transmits a selected ion and accelerates it towards "Q2" which is termed a collision cell. The pressure in Q2 is higher and the ions collides with neutral gas in the collision cell and are fragmented by CID. The fragments are then accelerated out of the collision cell and enter Q3 which scans through the mass range, analyzing the resulting fragments (as they hit a detector). This produces a mass spectrum of the CID fragments from which structural information or identity can be gained. Many other experiments using CID on a triple quadrupole exist such as precursor ion scans that determine where a specific fragment came from rather than what fragments are produced by a given molecule.

Fourier transform ion cyclotron resonance

Ions trapped in the ICR cell can be excited by applying pulsed electric fields at their resonant frequency to increase their kinetic energy. [3] [4] The duration and amplitude of the pulse determines the ion kinetic energy. Because a collision gas present at low pressure requires a long time for excited ions to collide with neutral molecules, a pulsed valve can be used to introduce a short burst of collision gas. Trapped fragment ions or their ion-molecule reaction products can be re-excited for multistage mass spectrometry (MSn). [5] If the excitation is not applied on the resonant frequency, but at a slightly off-resonant frequency, the ions will alternately be excited and de-excited, permitting multiple collisions at low collision energy. Sustained off-resonance irradiation collision-induced dissociation (SORI-CID) [6] is a CID technique used in Fourier transform ion cyclotron resonance mass spectrometry which involves accelerating the ions in cyclotron motion (in a circle inside of an ion trap) in the presence of a collision gas. [7]

Higher-energy collisional dissociation

Higher-energy collisional dissociation (HCD) is a CID technique specific to the orbitrap mass spectrometer in which fragmentation takes place external to the trap. [8] HCD was formerly known as higher-energy C-trap dissociation. In HCD, the ions pass through the C-trap and into the HCD cell, an added multipole collision cell, where dissociation takes place. The ions are then returned to the C-trap before injection into the orbitrap for mass analysis. HCD does not suffer from the low mass cutoff of resonant-excitation (CID) and therefore is useful for isobaric tag–based quantification as reporter ions can be observed. Despite the name, the collision energy of HCD is typically in the regime of low energy collision induced dissociation (less than 100 eV). [8] [9]

Fragmentation mechanisms

Homolytic fragmentation Homolysis (Chemistry) V.1.svg
Homolytic fragmentation

Homolytic fragmentation is bond dissociation where each of the fragments retains one of the originally-bonded electrons. [10]

Heterolytic fragmentation Heterolysis (chemistry).svg
Heterolytic fragmentation

Heterolytic fragmentation is bond cleavage where the bonding electrons remain with only one of the fragment species. [11]

In CID, charge remote fragmentation is a type of covalent bond breaking that occurs in a gas phase ion in which the cleaved bond is not adjacent to the location of the charge. [12] [13] This fragmentation can be observed using tandem mass spectrometry. [14]

See also

Related Research Articles

<span class="mw-page-title-main">Mass spectrometry</span> Analytical technique based on determining mass to charge ratio of ions

Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a mass spectrum, a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures.

<span class="mw-page-title-main">Tandem mass spectrometry</span> Type of mass spectrometry

Tandem mass spectrometry, also known as MS/MS or MS2, is a technique in instrumental analysis where two or more stages of analysis using one or more mass analyzer are performed with an additional reaction step in between these analyses to increase their abilities to analyse chemical samples. A common use of tandem MS is the analysis of biomolecules, such as proteins and peptides.

<span class="mw-page-title-main">Gas chromatography–mass spectrometry</span> Analytical method

Gas chromatography–mass spectrometry (GC–MS) is an analytical method that combines the features of gas-chromatography and mass spectrometry to identify different substances within a test sample. Applications of GC–MS include drug detection, fire investigation, environmental analysis, explosives investigation, food and flavor analysis, and identification of unknown samples, including that of material samples obtained from planet Mars during probe missions as early as the 1970s. GC–MS can also be used in airport security to detect substances in luggage or on human beings. Additionally, it can identify trace elements in materials that were previously thought to have disintegrated beyond identification. Like liquid chromatography–mass spectrometry, it allows analysis and detection even of tiny amounts of a substance.

<span class="mw-page-title-main">Quadrupole mass analyzer</span> Type of mass spectrometer

In mass spectrometry, the quadrupole mass analyzer is a type of mass analyzer originally conceived by Nobel laureate Wolfgang Paul and his student Helmut Steinwedel. As the name implies, it consists of four cylindrical rods, set parallel to each other. In a quadrupole mass spectrometer (QMS) the quadrupole is the mass analyzer – the component of the instrument responsible for selecting sample ions based on their mass-to-charge ratio (m/z). Ions are separated in a quadrupole based on the stability of their trajectories in the oscillating electric fields that are applied to the rods.

Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or mass spectrometer) for determining the mass-to-charge ratio (m/z) of ions based on the cyclotron frequency of the ions in a fixed magnetic field. The ions are trapped in a Penning trap (a magnetic field with electric trapping plates), where they are excited (at their resonant cyclotron frequencies) to a larger cyclotron radius by an oscillating electric field orthogonal to the magnetic field. After the excitation field is removed, the ions are rotating at their cyclotron frequency in phase (as a "packet" of ions). These ions induce a charge (detected as an image current) on a pair of electrodes as the packets of ions pass close to them. The resulting signal is called a free induction decay (FID), transient or interferogram that consists of a superposition of sine waves. The useful signal is extracted from this data by performing a Fourier transform to give a mass spectrum.

Infrared multiple photon dissociation (IRMPD) is a technique used in mass spectrometry to fragment molecules in the gas phase usually for structural analysis of the original (parent) molecule.

<span class="mw-page-title-main">Electron-capture dissociation</span> Method in mass spectrometry

Electron-capture dissociation (ECD) is a method of fragmenting gas-phase ions for structure elucidation of peptides and proteins in tandem mass spectrometry. It is one of the most widely used techniques for activation and dissociation of mass selected precursor ion in MS/MS. It involves the direct introduction of low-energy electrons to trapped gas-phase ions.

Gas phase ion chemistry is a field of science encompassed within both chemistry and physics. It is the science that studies ions and molecules in the gas phase, most often enabled by some form of mass spectrometry. By far the most important applications for this science is in studying the thermodynamics and kinetics of reactions. For example, one application is in studying the thermodynamics of the solvation of ions. Ions with small solvation spheres of 1, 2, 3... solvent molecules can be studied in the gas phase and then extrapolated to bulk solution.

<span class="mw-page-title-main">Electron-transfer dissociation</span>

Electron-transfer dissociation (ETD) is a method of fragmenting multiply-charged gaseous macromolecules in a mass spectrometer between the stages of tandem mass spectrometry (MS/MS). Similar to electron-capture dissociation, ETD induces fragmentation of large, multiply-charged cations by transferring electrons to them. ETD is used extensively with polymers and biological molecules such as proteins and peptides for sequence analysis. Transferring an electron causes peptide backbone cleavage into c- and z-ions while leaving labile post translational modifications (PTM) intact. The technique only works well for higher charge state peptide or polymer ions (z>2). However, relative to collision-induced dissociation (CID), ETD is advantageous for the fragmentation of longer peptides or even entire proteins. This makes the technique important for top-down proteomics. The method was developed by Hunt and coworkers at the University of Virginia.

<span class="mw-page-title-main">Time-of-flight mass spectrometry</span> Method of mass spectrometry

Time-of-flight mass spectrometry (TOFMS) is a method of mass spectrometry in which an ion's mass-to-charge ratio is determined by a time of flight measurement. Ions are accelerated by an electric field of known strength. This acceleration results in an ion having the same kinetic energy as any other ion that has the same charge. The velocity of the ion depends on the mass-to-charge ratio. The time that it subsequently takes for the ion to reach a detector at a known distance is measured. This time will depend on the velocity of the ion, and therefore is a measure of its mass-to-charge ratio. From this ratio and known experimental parameters, one can identify the ion.

Robert Graham Cooks is the Henry Bohn Hass Distinguished Professor of Chemistry in the Aston Laboratories for Mass Spectrometry at Purdue University. He is an ISI Highly Cited Chemist, with over 1,000 publications and an H-index of 144.

<span class="mw-page-title-main">Christie G. Enke</span> American chemist

Christie G. Enke is a United States academic chemist who made pioneering contributions to the field of analytical chemistry.

<span class="mw-page-title-main">Mass-analyzed ion-kinetic-energy spectrometry</span>

Mass-analyzed ion kinetic-energy spectrometry (MIKES) is a mass spectrometry technique by which mass spectra are obtained from a sector instrument that incorporates at least one magnetic sector plus one electric sector in reverse geometry. The accelerating voltage V, and the magnetic field B, are set to select the precursor ions of a particular m/z. The precursor ions then dissociate or react in an electric field-free region between the two sectors. The ratio of the kinetic energy to charge of the product ions are analyzed by scanning the electric sector field E. The width of the product ion spectrum peaks is related to the kinetic energy release distribution for the dissociation process.

<span class="mw-page-title-main">Ion-mobility spectrometry–mass spectrometry</span>

Ion mobility spectrometry–mass spectrometry (IMS-MS) is an analytical chemistry method that separates gas phase ions based on their interaction with a collision gas and their masses. In the first step, the ions are separated according to their mobility through a buffer gas on a millisecond timescale using an ion mobility spectrometer. The separated ions are then introduced into a mass analyzer in a second step where their mass-to-charge ratios can be determined on a microsecond timescale. The effective separation of analytes achieved with this method makes it widely applicable in the analysis of complex samples such as in proteomics and metabolomics.

<span class="mw-page-title-main">Triple quadrupole mass spectrometer</span>

A triple quadrupole mass spectrometer (TQMS), is a tandem mass spectrometer consisting of two quadrupole mass analyzers in series, with a (non-mass-resolving) radio frequency (RF)–only quadrupole between them to act as a cell for collision-induced dissociation. This configuration is often abbreviated QqQ, here Q1q2Q3.

A hybrid mass spectrometer is a device for tandem mass spectrometry that consists of a combination of two or more m/z separation devices of different types.

<span class="mw-page-title-main">Fragmentation (mass spectrometry)</span>

In mass spectrometry, fragmentation is the dissociation of energetically unstable molecular ions formed from passing the molecules mass spectrum. These reactions are well documented over the decades and fragmentation patterns are useful to determine the molar weight and structural information of unknown molecules. Fragmentation that occurs in tandem mass spectrometry experiments has been a recent focus of research, because this data helps facilitate the identification of molecules.

In mass spectrometry, de novo peptide sequencing is the method in which a peptide amino acid sequence is determined from tandem mass spectrometry.

In mass spectrometry, data-independent acquisition (DIA) is a method of molecular structure determination in which all ions within a selected m/z range are fragmented and analyzed in a second stage of tandem mass spectrometry. Tandem mass spectra are acquired either by fragmenting all ions that enter the mass spectrometer at a given time or by sequentially isolating and fragmenting ranges of m/z. DIA is an alternative to data-dependent acquisition (DDA) where a fixed number of precursor ions are selected and analyzed by tandem mass spectrometry.

References

  1. Wells JM, McLuckey SA (2005). "Collision‐Induced Dissociation (CID) of Peptides and Proteins". Biological Mass Spectrometry. Methods in Enzymology. Vol. 402. pp. 148–85. doi:10.1016/S0076-6879(05)02005-7. ISBN   9780121828073. PMID   16401509.{{cite book}}: |journal= ignored (help)
  2. Sleno L, Volmer DA (2004). "Ion activation methods for tandem mass spectrometry". Journal of Mass Spectrometry. 39 (10): 1091–112. Bibcode:2004JMSp...39.1091S. doi:10.1002/jms.703. PMID   15481084.
  3. Cody, R.B.; Freiser, B.S. (1982). "Collision-induced dissociation in a fourier-transform mass spectrometer". International Journal of Mass Spectrometry and Ion Physics. 41 (3): 199–204. Bibcode:1982IJMSI..41..199C. doi:10.1016/0020-7381(82)85035-3. ISSN   0020-7381.
  4. Cody, R. B.; Burnier, R. C.; Freiser, B. S. (1982). "Collision-induced dissociation with Fourier transform mass spectrometry". Analytical Chemistry. 54 (1): 96–101. doi:10.1021/ac00238a029. ISSN   0003-2700.
  5. Cody, R. B.; Burnier, R. C.; Cassady, C. J.; Freiser, B. S. (1982). "Consecutive collision-induced dissociations in Fourier transform mass spectrometry". Analytical Chemistry. 54 (13): 2225–2228. doi:10.1021/ac00250a021. ISSN   0003-2700.
  6. Gauthier, J.W.; Trautman, T.R.; Jacobson, D.B. (1991). "Sustained off-resonance irradiation for collision-activated dissociation involving Fourier transform mass spectrometry. Collision-activated dissociation technique that emulates infrared multiphoton dissociation". Analytica Chimica Acta. 246 (1): 211–225. doi:10.1016/s0003-2670(00)80678-9. ISSN   0003-2670.
  7. Laskin, Julia; Futrell, Jean H. (2005). "Activation of large lons in FT-ICR mass spectrometry". Mass Spectrometry Reviews. 24 (2): 135–167. Bibcode:2005MSRv...24..135L. doi:10.1002/mas.20012. ISSN   0277-7037. PMID   15389858.
  8. 1 2 Olsen JV, Macek B, Lange O, Makarov A, Horning S, Mann M (September 2007). "Higher-energy C-trap dissociation for peptide modification analysis". Nat. Methods. 4 (9): 709–12. doi:10.1038/nmeth1060. PMID   17721543.
  9. Murray, Kermit K.; Boyd, Robert K.; Eberlin, Marcos N.; Langley, G. John; Li, Liang; Naito, Yasuhide (2013). "Definitions of terms relating to mass spectrometry (IUPAC Recommendations 2013)". Pure and Applied Chemistry. 85 (7): 1515. doi: 10.1351/PAC-REC-06-04-06 . ISSN   1365-3075.
  10. IUPAC , Compendium of Chemical Terminology , 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006) " homolysis (homolytic) ". doi : 10.1351/goldbook.H02851
  11. IUPAC , Compendium of Chemical Terminology , 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006) " heterolysis (heterolytic) ". doi : 10.1351/goldbook.H02809
  12. Cheng C, Gross ML (2000), "Applications and mechanisms of charge-remote fragmentation", Mass Spectrom Rev, 19 (6): 398–420, Bibcode:2000MSRv...19..398C, doi:10.1002/1098-2787(2000)19:6<398::AID-MAS3>3.0.CO;2-B, PMID   11199379.
  13. Gross, M. (2000), "Charge-remote fragmentation: an account of research on mechanisms and applications", International Journal of Mass Spectrometry, 200 (1–3): 611–624, Bibcode:2000IJMSp.200..611G, doi:10.1016/S1387-3806(00)00372-9
  14. "Remote-site (charge-remote) fragmentation", Rapid Communications in Mass Spectrometry, 2 (10): 214–217, 1988, Bibcode:1988RCMS....2..214., doi:10.1002/rcm.1290021009