Neurostimulation

Last updated
Neurostimulation
ICD-10-PCS 00H00MZ
OPS-301 code 8-631

Neurostimulation is the purposeful modulation of the nervous system's activity using invasive (e.g. microelectrodes) or non-invasive means (e.g. transcranial magnetic stimulation or transcranial electric stimulation, tES, such as tDCS or transcranial alternating current stimulation, tACS). Neurostimulation usually refers to the electromagnetic approaches to neuromodulation.

Contents

Neurostimulation technology can improve the life quality of those who are severely paralyzed or have profound losses to various sense organs, as well as for permanent reduction of severe, chronic pain which would otherwise require constant (around-the-clock), high-dose opioid therapy (such as neuropathic pain and spinal-cord injury). It serves as the key part of neural prosthetics for hearing aids, artificial vision, artificial limbs, and brain-machine interfaces. In the case of neural stimulation, mostly an electrical stimulation is utilized and charge-balanced biphasic constant current waveforms or capacitively coupled charge injection approaches are adopted. Alternatively, transcranial magnetic stimulation and transcranial electric stimulation have been proposed as non-invasive methods in which either a magnetic field or transcranially applied electric currents cause neurostimulation. [1] [2]

Brain stimulation

Brain stimulation has potentials to treat some disorders such as epilepsy. In this method, scheduled stimulation is applied to specific cortical or subcortical targets. There are available commercial devices [3] that can deliver an electrical pulse at scheduled time intervals. Scheduled stimulation is hypothesized to alter the intrinsic neurophysiologic properties of epileptic networks. The most explored targets for scheduled stimulation are the anterior nucleus of the thalamus and the hippocampus. The anterior nucleus of the thalamus has been studied, which has shown a significant seizure reduction with the stimulator on versus off during several months after stimulator implantation. [4] Moreover, the cluster headache (CH) can be treated by using a temporary stimulating electrode at sphenopalatine ganglion (SPG). Pain relief is reported within several minutes of stimulation in this method. [5] To avoid use of implanted electrodes, researchers have engineered ways to inscribe a "window" made of zirconia that has been modified to be transparent and implanted in mice skulls, to allow optical waves to penetrate more deeply, as in optogenetics, to stimulate or inhibit individual neurons. [6]

Deep brain stimulation

Deep brain stimulation (DBS) has shown benefits for movement disorders such as Parkinson's disease, tremor and dystonia and other neuropsychiatric disorders such as depression, obsessive-compulsive disorder, Tourette syndrome, chronic pain and cluster headache. DBS can directly change the brain activity in a controlled manner and is hence used to map fundamental mechanisms of brain functions along with neuroimaging methods.

A DBS system consists of three components: the implanted pulse generator (IPG), the lead, and an extension. The implantable pulse generator (PG) generates stimulation pulses, which are sent to intracranial leads at the target via an extension. The simulation pulses interfere with neural activity at the target site.

The application and effects of DBS, on both normal and diseased brains, involves many parameters. These include the physiological properties of the brain tissue, which may change with disease state. Also important are the stimulation parameters, such as amplitude and temporal characteristics, and the geometric configuration of the electrode and the tissue that surrounds it.

In spite of a huge number of studies on DBS, its mechanism of action is still not well understood. Developing DBS microelectrodes is still challenging. [7]

Non-invasive brain stimulation

rTMS in a rodent. From Oscar Arias-Carrion, 2008 Repetitive transcranial magnetic stimulation (rTMS) is a technique for noninvasive stimulation of the adult brain.jpg
rTMS in a rodent. From Oscar Arias-Carrión, 2008

Transcranial magnetic stimulation

Compared to electrical stimulation that utilizes brief, high-voltage electric shock to activate neurons, which can potentially activate pain fibers, transcranial magnetic stimulation (TMS) was developed by Baker in 1985. TMS uses a magnetic wire above the scalp, which carries a sharp and high current pulse. A time variant magnetic field is induced perpendicular to the coil due to the applied pulse which consequently generates an electric field based on Maxwell's law. The electric field provides the necessary current for a non-invasive and much less painful stimulation. There are two TMS devices called single pulse TMS and repetitive pulse TMS (rTMS) while the latter has greater effect but potential to cause seizure. TMS can be used for therapy particularly in psychiatry, as a tool to measure central motor conduction and a research tool to study different aspects of human brain physiology such as motor function, vision, and language. The rTMS method has been used to treat epilepsy with rates of 8–25 Hz for 10 seconds. The other therapeutic uses of rTMS include parkinson diseases, dystonia and mood diseases. Also, TMS can be used to determine the contribution of cortical networks to specific cognitive functions by disrupting activity in the focal brain region. [1] Early, inconclusive, results have been obtained in recovery from coma (persistent vegetative state) by Pape et al. (2009). [8]

Transcranial electrical stimulation of techniques. While tDCS uses constant current intensity, tRNS and tACS use oscillating current. The vertical axis represents the current intensity in milliamp (mA), while the horizontal axis illustrates the time-course. Fnhum-07-00435-g001.jpg
Transcranial electrical stimulation of techniques. While tDCS uses constant current intensity, tRNS and tACS use oscillating current. The vertical axis represents the current intensity in milliamp (mA), while the horizontal axis illustrates the time-course.

Transcranial electrical stimulation

Spinal cord stimulation

Spinal cord stimulation (SCS) is an effective therapy for the treatment of chronic and intractable pain including diabetic neuropathy, failed back surgery syndrome, complex regional pain syndrome, phantom limb pain, ischemic limb pain, refractory unilateral limb pain syndrome, postherpetic neuralgia and acute herpes zoster pain. Another pain condition that is a potential candidate for SCS treatment is Charcot-Marie-Tooth (CMT) disease, which is associated with moderate to severe chronic extremity pain. [9] SCS therapy consists of the electrical stimulation of the spinal cord to 'mask' pain. The gate theory proposed in 1965 by Melzack and Wall [10] provided a theoretical construct to attempt SCS as a clinical treatment for chronic pain. This theory postulates that activation of large diameter, myelinated primary afferent fibers suppresses the response of dorsal horn neurons to input from small, unmyelinated primary afferents. A simple SCS system consists of three different parts. First, microelectrodes are implanted in the epidural space to deliver stimulation pulses to the tissue. Second, an electrical pulse generator implanted in the lower abdominal area or gluteal region while is connected to the electrodes via wires, and third a remote control to adjust the stimulus parameters such as pulse width and pulse rate in the PG. Improvements have been made in both the clinical aspects of SCS such as transition from subdural placement of contacts to epidural placement, which reduces the risk and morbidity of SCS implantation, and also technical aspects of SCS such as improving percutaneous leads, and fully implantable multi-channel stimulators. However, there are many parameters that need to be optimized including number of implanted contacts, contact size and spacing, and electrical sources for stimulation. The stimulus pulse width and pulse rate are important parameters that need to be adjusted in SCS, which are typically 400 us and 8–200 Hz respectively. [11]

Spinal cord stimulation for movement disorders

Spinal cord stimulation has shown promising results in spinal cord injury [12] [13] and other movement disorders, such as multiple sclerosis. [14] The stimulation, applied over the lumbar spinal cord, works by activating large diameter afferent fibers entering the spinal cord, [15] [16] which then transsynaptically activate and engage spinal neuronal networks. [17] The same target structures can also be activated by transcutaneous electrodes placed over the lower thoracic spine and abdomen. [18] Transcutaneous spinal cord stimulation is completely non-invasive and, as it uses TENS electrodes and stimulators, can be applied at low cost. Yet, in comparison to the implanted epidural variant, the efficacy of transcutaneous spinal cord stimulation depends more strongly on the body position and spinal alignment, [19] [20] which could lead to inconsistent result if the body position and posture isn't controlled during the application.

Transcutaneous supraorbital nerve stimulation

Tentative evidence supports transcutaneous supraorbital nerve stimulation. [21] Side effects are few. [22]

Cochlear implants

Cochlear implant Cochlear implant.jpg
Cochlear implant

Cochlear implants have provided partial hearing to more than 120,000 persons worldwide as of 2008. The electrical stimulation is used in a cochlear implant to provide functional hearing in totally deafened persons. Cochlear implants include several subsystem components from the external speech processor and radio frequency (RF) transmission link to the internal receiver, stimulator, and electrode arrays. Modern cochlear implant research started in the 1960s and 1970s. In 1961, a crude single electrode device was implanted in two deaf patients and useful hearing with electric stimulation was reported. The first FDA approved complete single channel device was released in 1984. [23] In cochlear implants, the sound is picked up by a microphone and transmitted to the behind-the-ear external processor to be converted to the digital data. The digitized data is then modulated on a radio frequency signal and transmitted to an antenna inside a headpiece. The data and power carrier are transmitted through a pair of coupled coils to the hermetically sealed internal unit. By extracting the power and demodulating the data, electric current commands are sent to the cochlea to stimulate the auditory nerve through microelectrodes. [24] The key point is that the internal unit does not have a battery and it should be able to extract the required energy. To reduce the risk of infection, data is transmitted wirelessly along with power. Inductively coupled coils are good candidates for power and data telemetry, although radio-frequency transmission could provide better efficiency and data rates. [25] Parameters needed by the internal unit include the pulse amplitude, pulse duration, pulse gap, active electrode, and return electrode that are used to define a biphasic pulse and the stimulation mode. An example of the commercial devices include Nucleus 22 device that utilized a carrier frequency of 2.5 MHz and later in the newer revision called Nucleus 24 device, the carrier frequency was increased to 5 MHz. [26] The internal unit in the cochlear implants is an ASIC (application-specific integrated circuit) chip that is responsible to ensure safe and reliable electric stimulation. Inside the ASIC chip, there is a forward pathway, a backward pathway, and control units. The forward pathway recovers digital information from the RF signal which includes stimulation parameters and some handshaking bits to reduce the communication error. The backward pathway usually includes a back telemetry voltage sampler that reads the voltage over a period of time on the recording electrode. The stimulator block is responsible to deliver predetermined current by external unit to the microelectrodes. This block includes a reference current and a digital-to-analog converter to transform digital commands to an analog current. [27]

Visual prosthesis

The Visual Cortical Implant ImplantSawan.JPG
The Visual Cortical Implant

Theoretical and experimental clinical evidences suggest that direct electrical stimulation of the retina might be able to provide some vision to subjects who have lost the photoreceptive elements of their retina. [28] Therefore, visual prostheses are developed to restore vision for the blind by using the stimulation. Depending upon which visual pathway location is targeted for neural stimulation, different approaches have been considered. Visual pathway consists mainly of the eye, optic nerve, lateral geniculate nucleus (LGN), and visual cortex. Therefore, retinal, optic nerve and visual cortex stimulation are the three different methods used in visual prostheses. [29] Retinal degenerative diseases, such as retinitis pigmentosa (RP) and age-related macular degeneration (AMD), are two likely candidate diseases in which retinal stimulation may be helpful. Three approaches called intraocular epiretinal, subretinal and extraocular transretinal stimulation are pursued in retinal devices that stimulate remaining retinal neural cells to bypass lost photoreceptors and allow the visual signal to reach the brain via the normal visual pathway. In epiretinal approach, electrodes are placed on the top side of the retina near ganglion cells, [30] whereas the electrodes are placed under the retina in subretinal approaches. [31] Finally, the posterior scleral surface of the eye is the place in which extraocular approach electrodes are positioned. Second Sight and the Humayun group at USC are the most active groups in the design of intraocular retinal prostheses. The ArgusTM 16 retinal implant is an intraocular retinal prosthesis utilizing video processing technologies. Regarding to the visual cortex stimulation, Brindley, and Dobelle were the first ones who did the experiments and demonstrated that by stimulating the top side of the visual cortex most of the electrodes can produce visual percept. [11] More recently Sawan built a complete implant for intracortical stimulation and validated the operation in rats. [32]

A pacemaker, scale in centimeters Pacemaker GuidantMeridianSR.jpg
A pacemaker, scale in centimeters

LGN, which is located in the midbrain to relay signals from the retina to the visual cortex, is another potential area that can be used for stimulation. But this area has limited access due to surgical difficulty. The recent success of deep brain stimulation techniques targeting the midbrain has encouraged research to pursue the approach of LGN stimulation for a visual prosthesis. [33]

Cardiac electrostimulation devices

Implantable pacemakers were proposed for the first time in 1959 and became more sophisticated since then. The therapeutic application of pacemakers consists of numerous rhythm disturbances including some forms of tachycardia (too fast a heart beat), heart failure, and even stroke. Early implantable pacemakers worked only a short time and needed periodic recharging by an inductive link. These implantable pacemakers needed a pulse generator to stimulate heart muscles with a certain rate in addition to electrodes. [34] Today, modern pulse generators are programmed non-invasively by sophisticated computerized machines using RF, obtaining information about the patient's and device's status by telemetry. Also they use a single hermetically sealed lithium iodide (LiI) cell as the battery. The pacemaker circuitry includes sense amplifiers to detect the heart's intrinsic electrical signals, which are used to track heart activity, rate adaptive circuitry, which determine the need for increased or reduced pacing rate, a microprocessor, memory to store the parameters, telemetry control for communication protocol and power supplies to provide regulated voltage. [35]

Stimulation microelectrode technologies

Utah microelectrode array Utah array pat5215088.jpg
Utah microelectrode array

Microelectrodes are one of the key components of the neurostimulation, which deliver the current to neurons. Typical microelectrodes have three main components: a substrate (the carrier), a conductive metal layer, and an insulation material. In cochlear implants, microelectrodes are formed from platinum-iridium alloy. State-of-the-art electrodes include deeper insertion to better match the tonotopic place of stimulation to the frequency band assigned to each electrode channel, improving efficiency of stimulation, and reducing insertion related trauma. These cochlear implant electrodes are either straight or spiral such as Med-El Combi 40+ and Advanced Bionics Helix microelectrodes respectively. In visual implants, there are two types of electrode arrays called planar type or three dimensional needle or pillar type, where needle type array such as Utah array is mostly used for cortical and optic nerve stimulations and rarely used in retinal implants due to the possible damage of retina. However, a pillar-shaped gold electrode array on thin-film polyimide has been used in an extraocular implant. On the other hand, planar stretchable microelectrode arrays are formed from flexible polymers, such as silicone, polyimide, and Parylene as candidates for retinal implants. Regarding to DBS microelectrodes an array, which can be controlled independently, distributed throughout the target nucleus would permit precise control of the spatial distribution of the stimulation, and thus, allow better personalized DBS. There are several requirements for DBS microelectrodes that include long lifetime without injury to the tissue or degradation of the electrodes, customized for different brain sites, long-term biocompatibility of the material, mechanically durable in order to reach the target without being damaged during handling by the implant surgeon, and finally uniformity of performance across the microelectrodes in a particular array. Tungsten microwire, iridium microwires, and sputtered or electrodeposited [36] Platinum-iridium alloy microelectrodes are the examples of microelectrode used in DBS. [11] Silicon carbide is a potential interesting material for realizing biocompatible semiconductor devices. [37]

Limitations

Brain tissue stimulation using non-invasive electrical and magnetic field methods raises several concerns, including the following:

The first issue is the uncertain dose (time and technical field parameters) for correct and healthy stimulation. [38] While neurophysiology lacks knowledge about the nature of such a treatment of nervous diseases at the cellular level, [39] many non-invasive electrical and magnetic therapeutic methods involve excessive exposure of the patient to an intense field, which is several times and even orders of magnitude higher than natural currents and electromagnetic fields in the brain. [40] [41]

Another significant challenge of non-invasive electrical and magnetic field methods is the impossibility of localizing the effect of stimulation on tissues in the relevant neural networks. [42] [43] We still need to gain knowledge about mental processes at the cellular level. The relationship between neural activity and cognitive processes continues to be an intriguing research question and challenge for treatment selection. Therefore, no one can be sure that electrical and magnetic fields reach only those neural structures of the brain that need treatment. An undefined dose and target of radiation can destroy healthy cells during a treatment procedure. Non-invasive brain tissue stimulation targets a large area of poorly characterized tissue. The inability to localize the effect of stimulation makes it challenging to target stimulation only to the desired neural networks. [42] [43]

Additionally, these methods are not generalizable to all patients because of more inter-individual variability in response to brain stimulation. [44]

History

The primary findings about neurostimulation originated from the idea to stimulate nerves for therapeutic purposes. The 1st recorded use of electrical stimulation for pain relief goes back to 46 AD, when Scribonius Largus used torpedo fish (electric ray) for relieving headaches. [45] In the late 18th century, Luigi Galvani discovered that the muscles of dead frog legs twitched when struck by direct current on the nervous system. [46] The modulation of the brain activity by electrical stimulation of the motor cortex in dogs was shown in 1870 that resulted in limb movement. [47] From the late 18th century to today many milestones have been developed. Nowadays, sensory prosthetic devices, such as visual implants, cochlear implants, auditory midbrain implants, and spinal cord stimulators and also motor prosthetic devices, such as deep brain stimulators, Bion microstimulators, the brain control and sensing interface, and cardiac electro-stimulation devices are widely used. [11]

In 2013 the British pharmaceutical company GlaxoSmithKline (GSK) coined the term "electroceutical" to broadly encompass medical devices that use electrical, mechanical, or light stimulation to affect electrical signaling in relevant tissue types. [48] [49] Clinical neural implants such as cochlear implants to restore hearing, retinal implants to restore sight, spinal cord stimulators for pain relief or cardiac pacemakers and implantable defibrillators are proposed examples of electroceuticals. [48] GSK formed a venture fund and said it would host a conference in 2013 to lay out a research agenda for the field. [50] A 2016 review of research on interactions between the nervous and immune systems in autoimmune disorders mentioned "electroceuticals" in passing and quotation marks, referring to neurostimulation devices in development for conditions like arthritis. [51]

Research

In addition to the enormous usage of neurostimulation for clinical applications, it is also used widely in laboratories started dates back to 1920s by people like Delgado who used stimulation as an experimental manipulation to study basics of how the brain works. The primary works were on the reward center of the brain in which stimulation of those structures led to pleasure that requested more stimulation. Another most recent example is the electrical stimulation of the MT area of primary visual cortex to bias perception. In particular, the directionality of motion is represented in a regular way in the MT area. They presented monkeys with moving images on screen and monkey throughput was to determine what the direction is. They found that by systematically introducing some errors to the monkey's responses, by stimulating the MT area which is responsible for perceiving the motion in another direction, the monkey responded to somewhere in between the actual motion and the stimulated one. This was an elegant use of stimulation to show that MT area is essential in the actual perception of motion. Within the memory field, stimulation is used very frequently to test the strength of the connection between one bundle of cells to another by applying a small current in one cell which results in the release of neurotransmitters and measuring the postsynaptic potential.[ citation needed ]

Generally, a short but high-frequency current in the range of 100 Hz helps strengthening the connection known as long-term potentiation. However, longer but low-frequency current tends to weaken the connections known as long-term depression. [52]

See also

Related Research Articles

<span class="mw-page-title-main">Transcranial magnetic stimulation</span> Brain stimulation using magnetic fields

Transcranial magnetic stimulation (TMS) is a noninvasive form of brain stimulation in which a changing magnetic field is used to induce an electric current at a specific area of the brain through electromagnetic induction. An electric pulse generator, or stimulator, is connected to a magnetic coil connected to the scalp. The stimulator generates a changing electric current within the coil which creates a varying magnetic field, inducing a current within a region in the brain itself.

<span class="mw-page-title-main">Transcutaneous electrical nerve stimulation</span> Therapeutic technique

A transcutaneous electrical nerve stimulation is a device that produces mild electric current to stimulate the nerves for therapeutic purposes. TENS, by definition, covers the complete range of transcutaneously applied currents used for nerve excitation, but the term is often used with a more restrictive intent, namely, to describe the kind of pulses produced by portable stimulators used to reduce pain. The unit is usually connected to the skin using two or more electrodes which are typically conductive gel pads. A typical battery-operated TENS unit is able to modulate pulse width, frequency, and intensity. Generally, TENS is applied at high frequency (>50 Hz) with an intensity below motor contraction or low frequency (<10 Hz) with an intensity that produces motor contraction. More recently, many TENS units use a mixed frequency mode which alleviates tolerance to repeated use. Intensity of stimulation should be strong but comfortable with greater intensities, regardless of frequency, producing the greatest analgesia. While the use of TENS has proved effective in clinical studies, there is controversy over which conditions the device should be used to treat.

<span class="mw-page-title-main">Phosphene</span> Visual illusion

A phosphene is the phenomenon of seeing light without light entering the eye. The word phosphene comes from the Greek words phos (light) and phainein. Phosphenes that are induced by movement or sound may be associated with optic neuritis.

<span class="mw-page-title-main">Functional electrical stimulation</span> Technique that uses low-energy electrical pulses

Functional electrical stimulation (FES) is a technique that uses low-energy electrical pulses to artificially generate body movements in individuals who have been paralyzed due to injury to the central nervous system. More specifically, FES can be used to generate muscle contraction in otherwise paralyzed limbs to produce functions such as grasping, walking, bladder voiding and standing. This technology was originally used to develop neuroprostheses that were implemented to permanently substitute impaired functions in individuals with spinal cord injury (SCI), head injury, stroke and other neurological disorders. In other words, a person would use the device each time he or she wanted to generate a desired function. FES is sometimes also referred to as neuromuscular electrical stimulation (NMES).

Neurotechnology encompasses any method or electronic device which interfaces with the nervous system to monitor or modulate neural activity.

Neuroprosthetics is a discipline related to neuroscience and biomedical engineering concerned with developing neural prostheses. They are sometimes contrasted with a brain–computer interface, which connects the brain to a computer rather than a device meant to replace missing biological functionality.

Neural engineering is a discipline within biomedical engineering that uses engineering techniques to understand, repair, replace, or enhance neural systems. Neural engineers are uniquely qualified to solve design problems at the interface of living neural tissue and non-living constructs.

Neurohacking is a subclass of biohacking, focused specifically on the brain. Neurohackers seek to better themselves or others by “hacking the brain” to improve reflexes, learn faster, or treat psychological disorders. The modern neurohacking movement has been around since the 1980s. However, herbal supplements have been used to increase brain function for hundreds of years. After a brief period marked by a lack of research in the area, neurohacking started regaining interest in the early 2000s. Currently, most neurohacking is performed via do-it-yourself (DIY) methods by in-home users.

<span class="mw-page-title-main">Neuromodulation</span> Regulation of neurons by neurotransmitters

Neuromodulation is the physiological process by which a given neuron uses one or more chemicals to regulate diverse populations of neurons. Neuromodulators typically bind to metabotropic, G-protein coupled receptors (GPCRs) to initiate a second messenger signaling cascade that induces a broad, long-lasting signal. This modulation can last for hundreds of milliseconds to several minutes. Some of the effects of neuromodulators include altering intrinsic firing activity, increasing or decreasing voltage-dependent currents, altering synaptic efficacy, increasing bursting activity and reconfiguring synaptic connectivity.

<span class="mw-page-title-main">Spinal cord stimulator</span> SCS TREATMENT

A spinal cord stimulator (SCS) or dorsal column stimulator (DCS) is a type of implantable neuromodulation device that is used to send electrical signals to select areas of the spinal cord for the treatment of certain pain conditions. SCS is a consideration for people who have a pain condition that has not responded to more conservative therapy. There are also spinal cord stimulators under research and development that could enable patients with spinal cord injury to walk again via epidural electrical stimulation (EES).

<span class="mw-page-title-main">Transcranial direct-current stimulation</span> Technique of brain electric stimulation therapy

Transcranial direct current stimulation (tDCS) is a form of neuromodulation that uses constant, low direct current delivered via electrodes on the head. It was originally developed to help patients with brain injuries or neuropsychiatric conditions such as major depressive disorder. It can be contrasted with cranial electrotherapy stimulation, which generally uses alternating current the same way, as well as transcranial magnetic stimulation.

Sacral nerve stimulation, also termed sacral neuromodulation, is a type of medical electrical stimulation therapy.

<span class="mw-page-title-main">Responsive neurostimulation device</span> Category of medical devices that respond to signals in a patients body to treat disease

Responsive neurostimulation device is a medical device that senses changes in a person's body and uses neurostimulation to respond in the treatment of disease. The FDA has approved devices for use in the United States in the treatment of epileptic seizures and chronic pain conditions. Devices are being studied for use in the treatment of essential tremor, Parkinson's disease, Tourette's syndrome, depression, obesity, and post-traumatic stress disorder.

Microelectrode arrays (MEAs) are devices that contain multiple microelectrodes through which neural signals are obtained or delivered, essentially serving as neural interfaces that connect neurons to electronic circuitry. There are two general classes of MEAs: implantable MEAs, used in vivo, and non-implantable MEAs, used in vitro. In each class, there are rigid, flexible, and stretchable microelectrode array.

Electroanalgesia is a form of analgesia, or pain relief, that uses electricity to ease pain. Electrical devices can be internal or external, at the site of pain (local) or delocalized throughout the whole body. It works by interfering with the electric currents of pain signals, inhibiting them from reaching the brain and inducing a response; different from traditional analgesics, such as opiates which mimic natural endorphins and NSAIDs that help relieve inflammation and stop pain at the source. Electroanalgesia has a lower addictive potential and poses less health threats to the general public, but can cause serious health problems, even death, in people with other electrical devices such as pacemakers or internal hearing aids, or with heart problems.

Neuromodulation is "the alteration of nerve activity through targeted delivery of a stimulus, such as electrical stimulation or chemical agents, to specific neurological sites in the body". It is carried out to normalize – or modulate – nervous tissue function. Neuromodulation is an evolving therapy that can involve a range of electromagnetic stimuli such as a magnetic field (rTMS), an electric current, or a drug instilled directly in the subdural space. Emerging applications involve targeted introduction of genes or gene regulators and light (optogenetics), and by 2014, these had been at minimum demonstrated in mammalian models, or first-in-human data had been acquired. The most clinical experience has been with electrical stimulation.

A peripheral nerve interface is the bridge between the peripheral nervous system and a computer interface which serves as a bi‐directional information transducer recording and sending signals between the human body and a machine processor. Interfaces to the nervous system usually take the form of electrodes for stimulation and recording, though chemical stimulation and sensing are possible. Research in this area is focused on developing peripheral nerve interfaces for the restoration of function following disease or injury to minimize associated losses. Peripheral nerve interfaces also enable electrical stimulation and recording of the peripheral nervous system to study the form and function of the peripheral nervous system. For example, recent animal studies have demonstrated high accuracy in tracking physiological meaningful measures, like joint angle. Many researchers also focus in the area of neuroprosthesis, linking the human nervous system to bionics in order to mimic natural sensorimotor control and function. Successful implantation of peripheral nerve interfaces depend on a number of factors which include appropriate indication, perioperative testing, differentiated planning, and functional training. Typically microelectrode devices are implanted adjacent to, around or within the nerve trunk to establish contact with the peripheral nervous system. Different approaches may be used depending on the type of signal desired and attainable.

<span class="mw-page-title-main">Shannon Criteria</span>

The Shannon criteria constitute an empirical rule in neural engineering that is used for evaluation of possibility of damage from electrical stimulation to nervous tissue.

A cortical implant is a subset of neuroprosthetics that is in direct connection with the cerebral cortex of the brain. By directly interfacing with different regions of the cortex, the cortical implant can provide stimulation to an immediate area and provide different benefits, depending on its design and placement. A typical cortical implant is an implantable microelectrode array, which is a small device through which a neural signal can be received or transmitted.

Neural dust is a hypothetical class of nanometer-sized devices operated as wirelessly powered nerve sensors; it is a type of brain–computer interface. The sensors may be used to study, monitor, or control the nerves and muscles and to remotely monitor neural activity. In practice, a medical treatment could introduce thousands of neural dust devices into human brains. The term is derived from "smart dust", as the sensors used as neural dust may also be defined by this concept.

References

  1. 1 2 Hallett M (July 2000). "Transcranial magnetic stimulation and the human brain". Nature. 406 (6792): 147–150. Bibcode:2000Natur.406..147H. doi:10.1038/35018000. PMID   10910346. S2CID   4413567.
  2. Nitsche, Michael A.; Cohen, Leonardo G.; Wassermann, Eric M.; Priori, Alberto; Lang, Nicolas; Antal, Andrea; Paulus, Walter; Hummel, Friedhelm; Boggio, Paulo S.; Fregni, Felipe; Pascual-Leone, Alvaro (2008). "Transcranial direct current stimulation: State of the art 2008". Brain Stimulation 1 (3): 206–223.
  3. Medtronic, Minneapolis, MN, U.S.A.
  4. Jobst BC, Darcey TM, Thadani VM, Roberts DW (July 2010). "Brain stimulation for the treatment of epilepsy". Epilepsia. 51 (Suppl 3): 88–92. doi: 10.1111/j.1528-1167.2010.02618.x . PMID   20618409.
  5. Ansarinia M, Rezai A, Tepper SJ, et al. (July 2010). "Electrical stimulation of sphenopalatine ganglion for acute treatment of cluster headaches". Headache. 50 (7): 1164–1174. doi:10.1111/j.1526-4610.2010.01661.x. PMID   20438584. S2CID   205683727.
  6. Damestani, Yasaman (2013). "Transparent nanocrystalline yttria-stabilized-zirconia calvarium prosthesis". Nanomedicine. 9 (8): 1135–1138. doi:10.1016/j.nano.2013.08.002. PMID   23969102. S2CID   14212180. Explained by Mohan, Geoffrey (September 4, 2013). "A window to the brain? It's here, says UC Riverside team". Los Angeles Times.
  7. Kringelbach ML, Jenkinson N, Owen SL, Aziz TZ (August 2007). "Translational principles of deep brain stimulation". Nat. Rev. Neurosci. 8 (8): 623–635. doi:10.1038/nrn2196. PMID   17637800. S2CID   147427108.
  8. Louise-Bender Pape, Theresa; Rosenow, Joshua; Lewis, Gwyn; Ahmed, Ghada; Walker, Matthew; Guernon, Ann; Roth, Heidi; Patil, Vijaya (January 13, 2009). "Repetitive transcranial magnetic stimulation-associated neurobehavioral gains during coma recovery". Brain Stimulation. 2 (1): 22–35. doi:10.1016/j.brs.2008.09.004. PMID   20633400. S2CID   41662030 via PubMed.
  9. Skaribas I.M.; Washburn S.N. (January 2010). "Successful treatment of charcot-marie-tooth chronic pain with spinal cord stimulation: A case study". Neuromodulation. 13 (3): 224–228. doi:10.1111/j.1525-1403.2009.00272.x. PMID   21992836. S2CID   8035147.
  10. Melzack R, Wall PD (November 1965). "Pain mechanisms: a new theory". Science. 150 (3699): 971–979. Bibcode:1965Sci...150..971M. doi:10.1126/science.150.3699.971. PMID   5320816.
  11. 1 2 3 4 Greenbaum, Elias S.; David Zhou (2009). Implantable Neural Prostheses 1: Devices and Applications. Biological and Medical Physics, Biomedical Engineering. Berlin: Springer. ISBN   978-0387772608.
  12. Wagner, Fabien B.; Mignardot, Jean-Baptiste; Le Goff-Mignardot, Camille G.; Demesmaeker, Robin; Komi, Salif; Capogrosso, Marco; Rowald, Andreas; Seáñez, Ismael; Caban, Miroslav; Pirondini, Elvira; Vat, Molywan; McCracken, Laura A.; Heimgartner, Roman; Fodor, Isabelle; Watrin, Anne; Seguin, Perrine; Paoles, Edoardo; Van Den Keybus, Katrien; Eberle, Grégoire; Schurch, Brigitte; Pralong, Etienne; Becce, Fabio; Prior, John; Buse, Nicholas; Buschman, Rik; Neufeld, Esra; Kuster, Niels; Carda, Stefano; von Zitzewitz, Joachim; Delattre, Vincent; Denison, Tim; Lambert, Hendrik; Minassian, Karen; Bloch, Jocelyne; Courtine, Grégoire (1 November 2018). "Targeted neurotechnology restores walking in humans with spinal cord injury". Nature. 563 (7729): 65–71. Bibcode:2018Natur.563...65W. doi:10.1038/s41586-018-0649-2. PMID   30382197. S2CID   53148162.
  13. Hofstoetter, Ursula S.; Freundl, Brigitta; Danner, Simon M.; Krenn, Matthias J.; Mayr, Winfried; Binder, Heinrich; Minassian, Karen (1 February 2020). "Transcutaneous Spinal Cord Stimulation Induces Temporary Attenuation of Spasticity in Individuals with Spinal Cord Injury". Journal of Neurotrauma. 37 (3): 481–493. doi:10.1089/neu.2019.6588. PMID   31333064. S2CID   198172141.
  14. Hofstoetter, Ursula S.; Freundl, Brigitta; Lackner, Peter; Binder, Heinrich (8 April 2021). "Transcutaneous Spinal Cord Stimulation Enhances Walking Performance and Reduces Spasticity in Individuals with Multiple Sclerosis". Brain Sciences. 11 (4): 472. doi: 10.3390/brainsci11040472 . PMC   8068213 . PMID   33917893.
  15. Ladenbauer, Josef; Minassian, Karen; Hofstoetter, Ursula S.; Dimitrijevic, Milan R.; Rattay, Frank (December 2010). "Stimulation of the Human Lumbar Spinal Cord With Implanted and Surface Electrodes: A Computer Simulation Study". IEEE Transactions on Neural Systems and Rehabilitation Engineering. 18 (6): 637–645. doi:10.1109/TNSRE.2010.2054112. PMID   21138794. S2CID   20180127.
  16. Danner, Simon M.; Hofstoetter, Ursula S.; Ladenbauer, Josef; Rattay, Frank; Minassian, Karen (March 2011). "Can the Human Lumbar Posterior Columns Be Stimulated by Transcutaneous Spinal Cord Stimulation? A Modeling Study". Artificial Organs. 35 (3): 257–262. doi:10.1111/j.1525-1594.2011.01213.x. PMC   4217151 . PMID   21401670.
  17. Danner, Simon M.; Hofstoetter, Ursula S.; Freundl, Brigitta; Binder, Heinrich; Mayr, Winfried; Rattay, Frank; Minassian, Karen (March 2015). "Human spinal locomotor control is based on flexibly organized burst generators". Brain. 138 (3): 577–588. doi:10.1093/brain/awu372. PMC   4408427 . PMID   25582580.
  18. Minassian, Karen; Persy, Ilse; Rattay, Frank; Dimitrijevic, Milan R.; Hofer, Christian; Kern, Helmut (March 2007). "Posterior root–muscle reflexes elicited by transcutaneous stimulation of the human lumbosacral cord". Muscle & Nerve. 35 (3): 327–336. doi:10.1002/mus.20700. PMID   17117411. S2CID   26116191.
  19. Danner, Simon M.; Krenn, Matthias; Hofstoetter, Ursula S.; Toth, Andrea; Mayr, Winfried; Minassian, Karen (21 January 2016). "Body Position Influences Which Neural Structures Are Recruited by Lumbar Transcutaneous Spinal Cord Stimulation". PLOS ONE. 11 (1): e0147479. Bibcode:2016PLoSO..1147479D. doi: 10.1371/journal.pone.0147479 . PMC   4721643 . PMID   26797502.
  20. Binder, Veronika E.; Hofstoetter, Ursula S.; Rienmüller, Anna; Száva, Zoltán; Krenn, Matthias J.; Minassian, Karen; Danner, Simon M. (26 November 2021). "Influence of Spine Curvature on the Efficacy of Transcutaneous Lumbar Spinal Cord Stimulation". Journal of Clinical Medicine. 10 (23): 5543. doi: 10.3390/jcm10235543 . PMC   8658162 . PMID   34884249.
  21. Jürgens, TP; Leone, M (Jun 2013). "Pearls and pitfalls: neurostimulation in headache". Cephalalgia: An International Journal of Headache. 33 (8): 512–525. doi:10.1177/0333102413483933. PMID   23671249. S2CID   42537455.
  22. Schoenen, J; Roberta, B; Magis, D; Coppola, G (29 March 2016). "Noninvasive neurostimulation methods for migraine therapy: The available evidence". Cephalalgia: An International Journal of Headache. 36 (12): 1170–1180. doi:10.1177/0333102416636022. PMID   27026674. S2CID   6812366.
  23. House WF, Urban J (1973). "Long term results of electrode implantation and electronic stimulation of the cochlea in man". Ann. Otol. Rhinol. Laryngol. 82 (4): 504–517. doi:10.1177/000348947308200408. PMID   4721186. S2CID   19339967.
  24. An SK, Park SI, Jun SB, et al. (June 2007). "Design for a simplified cochlear implant system". IEEE Trans Biomed Eng. 54 (6 Pt 1): 973–382. doi:10.1109/TBME.2007.895372. hdl: 10371/7911 . PMID   17554817. S2CID   7979564.
  25. Nikolayev, Denys; Joseph, Wout; Zhadobov, Maxim; Sauleau, Ronan; Martens, Luc (13 March 2019). "Optimal Radiation of Body-Implanted Capsules". Physical Review Letters. 122 (10): 108101. Bibcode:2019PhRvL.122j8101N. doi:10.1103/PhysRevLett.122.108101. hdl: 1854/LU-8611129 . PMID   30932680. S2CID   89621750.
  26. P. Crosby, C. Daly, D. Money, and et al., Aug. 1985, " Cochlear implant system for an auditory prosthesis," United States Patent 4532930.
  27. Ghovanloo M.; Najafi K. (December 2004). "A modular 32-site wireless neural stimulation microsystem". IEEE J. Solid-State Circuits. 39 (12): 2457–2466. Bibcode:2004IJSSC..39.2457G. CiteSeerX   10.1.1.681.6677 . doi:10.1109/jssc.2004.837026. S2CID   7525679.
  28. Clausen J (1955). "Visual sensations (phosphenes) produced by AC sine wave stimulation". Acta Psychiatr Neurol Scand Suppl. 94: 1–101. PMID   13258326.
  29. Weiland J.D.; Humayun M.S. (July 2008). "Visual prosthesis". Proceedings of the IEEE. 96 (7): 1076–1084. doi:10.1109/JPROC.2008.922589. S2CID   21649550.
  30. Humayun MS, de Juan E, Dagnelie G, Greenberg RJ, Propst RH, Phillips DH (January 1996). "Visual perception elicited by electrical stimulation of retina in blind humans" (PDF). Arch. Ophthalmol. 114 (1): 40–46. doi:10.1001/archopht.1996.01100130038006. PMID   8540849. S2CID   29334227.
  31. Chow AY, Chow VY (March 1997). "Subretinal electrical stimulation of the rabbit retina". Neurosci. Lett. 225 (1): 13–16. doi:10.1016/S0304-3940(97)00185-7. PMID   9143006. S2CID   22119389.
  32. Sawan, Mohamad. "Curriculum Vitae".
  33. Pezaris JS, Reid RC (May 2007). "Demonstration of artificial visual percepts generated through thalamic microstimulation". Proc. Natl. Acad. Sci. USA. 104 (18): 7670–7675. Bibcode:2007PNAS..104.7670P. doi: 10.1073/pnas.0608563104 . PMC   1863473 . PMID   17452646.
  34. Elmvquist R.; Senning A. (1960). "Implantable pacemaker for the heart". In Smyth CN (ed.). Medical electronics. Paris: Iliffe & Sons.
  35. Warren J., Nelson J. (2000). "Pacemakers and ICD pulse generator circuitry". In Ellenbogen KA, Kay GN, Wilkoff BL (eds.). Clinical cardiac pacing and defibrillation (2nd ed.). Philadelphia: WB Saunders. pp. 194–216.
  36. "Microelectrodes".
  37. Saddow SE (2011). Silicon Carbide Biotechnology: A Biocompatible Semiconductor for Advanced Biomedical Devices and Applications. Elsevier LTD. ISBN   978-0123859068.
  38. Benussi A, Pascual-Leone A, Borroni B (2020). "Non-Invasive Cerebellar Stimulation in Neurodegenerative Ataxia: A Literature Review". ‘’International Journal of Molecular Sciences.’’ 21 (6): 1948. doi:10.3390/ijms21061948
  39. Rosa, MA; Lisanby, SH (2012). "Somatic treatments for mood disorders". ‘’Neuropsychopharmacology’’. 37 (1): 102–116. doi:10.1038/npp.2011.225
  40. Grimaldi G, Argyropoulos GP, Boehringer A, Celnik P, Edwards MJ, Ferrucci R, et al. (2014). "Non-invasive cerebellar stimulation--a consensus paper" (PDF). ‘’Cerebellum.’’ 13 (1): 121–138. doi:10.1007/s12311-013-0514-7
  41. Siebner HR, Hartwigsen G, Kassuba T, Rothwell JC (2009). "How does transcranial magnetic stimulation modify neuronal activity in the brain? Implications for studies of cognition". ‘’Cortex; A Journal Devoted to the Study of the Nervous System and Behavior.’’ 45 (9): 1035–1042. doi:10.1016/j.cortex.2009.02.007
  42. 1 2 Sparing R, Mottaghy FM (2008). "Noninvasive brain stimulation with transcranial magnetic or direct current stimulation (TMS/tDCS)-From insights into human memory to therapy of its dysfunction". ‘’Methods.’’ 44 (4): 329–337. doi:10.1016/j.ymeth.2007.02.001
  43. 1 2 Kirsch, D. L., & Nichols, F. (2013). “Cranial electrotherapy stimulation for treatment of anxiety, depression, and insomnia.” ‘’Psychiatric Clinics’’, 36(1), 169-176.
  44. Benussi A, Pascual-Leone A, Borroni B (March 2020). "Non-Invasive Cerebellar Stimulation in Neurodegenerative Ataxia: A Literature Review". International Journal of Molecular Sciences. 21 (6): 1948. doi:10.3390/ijms21061948
  45. Jensen JE, Conn RR, Hazelrigg G, Hewett JE (1985). "The use of transcutaneous neural stimulation and isokinetic testing in arthroscopic knee surgery". Am J Sports Med. 13 (1): 27–33. doi:10.1177/036354658501300105. PMID   3872082. S2CID   19217534.
  46. Weisstein, Eric W. (2002). "Galvani, Luigi (1737–1798)". Eric Weisstein's World of Scientific Biography. Wolfram Research.
  47. Fritsch G.; Hitzig E. (1870). "Uber die elektrische Erregbarkeit des Grosshirns". Arch. Anat. Physiol. 37: 300–332.
  48. 1 2 Moore, Samuel (29 May 2015). "The Vagus Nerve: A Back Door for Brain Hacking". IEEE Spectrum. Retrieved 4 June 2015.
  49. Famm, Kristoffer; Litt, Brian; Tracey, Kevin J.; Boyden, Edward S.; Slaoui, Moncef (10 April 2013). "Drug discovery: A jump-start for electroceuticals". Nature. 496 (7444): 159–161. Bibcode:2013Natur.496..159F. doi:10.1038/496159a. PMC   4179459 . PMID   23579662.
  50. Solon, Olivia (28 May 2013). "Electroceuticals: swapping drugs for devices". Wired UK.
  51. Reardon, Colin (October 2016). "Neuro-immune interactions in the cholinergic anti-inflammatory reflex". Immunology Letters. 178: 92–96. doi:10.1016/j.imlet.2016.08.006. PMID   27542331.
  52. Interview with Dr. J. Manns, Emory University, Oct. 2010