Sound level meter

Last updated

An integrating-averaging Cirrus Research's Optimus sound level meter which complies with IEC 61672-1:2002 Optimus Sound Level Meter.jpg
An integrating-averaging Cirrus Research's Optimus sound level meter which complies with IEC 61672-1:2002

A sound level meter (also called sound pressure level meter (SPL)) is used for acoustic measurements. It is commonly a hand-held instrument with a microphone. The best type of microphone for sound level meters is the condenser microphone, which combines precision with stability and reliability. [1] The diaphragm of the microphone responds to changes in air pressure caused by sound waves. That is why the instrument is sometimes referred to as a sound pressure level meter (SPL). This movement of the diaphragm, i.e. the sound pressure (unit pascal, Pa), is converted into an electrical signal (unit volt, V). While describing sound in terms of sound pressure, a logarithmic conversion is usually applied and the sound pressure level is stated instead, in decibels (dB), with 0 dB SPL equal to 20 micropascals.

Contents

A microphone is distinguishable by the voltage value produced when a known, constant root mean square sound pressure is applied. This is known as microphone sensitivity. The instrument needs to know the sensitivity of the particular microphone being used. Using this information, the instrument is able to accurately convert the electrical signal back to sound pressure, and display the resulting sound pressure level (unit decibel, dB).

Sound level meters are commonly used in noise pollution studies for the quantification of different kinds of noise, especially for industrial, environmental, mining and aircraft noise. [2] [3] The current international standard that specifies sound level meter functionality and performances is the IEC 61672-1:2013. However, the reading from a sound level meter does not correlate well to human-perceived loudness, which is better measured by a loudness meter. Specific loudness is a compressive nonlinearity and varies at certain levels and at certain frequencies. These metrics can also be calculated in a number of different ways. [4] [ example needed ]

The world's first hand-held and transistorized sound level meter, was released in 1960 and developed by the Danish company Brüel & Kjær. [5] In 1969, a group of University researchers from California founded Pulsar Instruments Inc. which became the first company to display sound exposure times on the scale of a sound level meter, as well as the sound level. This was to comply with the 1969 Walsh-Healey Act, which demanded that the noise in US workplaces should be controlled. [6] In 1980, Britain's Cirrus Research introduced the world's first handheld sound level meter to provide integrated Leq and sound exposure level (SEL) measurements. [7]

Classification

Types

The IEC 61672-1 specifies "three kinds of sound measuring instruments". [8] They are the "conventional" sound level meter, the integrating-averaging sound level meter, and the integrating sound level meter.

The standard sound level meter [9] can be called an exponentially averaging sound level meter as the AC signal from the microphone is converted to DC by a root-mean-square (RMS) circuit and thus it must have a time constant of integration; today referred to as the time-weighting. Three of these time-weightings have been internationally standardized, 'S' (1 s) originally called Slow, 'F' (125 ms) originally called Fast, and 'I' (35 ms) originally called Impulse. Their names were changed in the 1980s to be the same in any language. I-time-weighting is no longer in the body of the standard because it has little real correlation with the impulsive character of noise events.

The output of the RMS circuit is linear in voltage and is passed through a logarithmic circuit to give a readout linear in decibels (dB). This is 20 times the base 10 logarithm of the ratio of given root-mean-square sound pressure to the reference sound pressure. Root-mean-square sound pressure being obtained with a standard frequency weighting and standard time weighting. The reference pressure is set by the International agreement to be 20 micropascals for airborne sound. It follows that the decibel is, in a sense, not a unit, it is simply a dimensionless ratio; in this case the ratio of two pressures.

An exponentially averaging sound level meter, which gives a snapshot of the current noise level, is of limited use for hearing damage risk measurements; an integrating or integrating-averaging meter is usually mandated. An integrating meter simply integrates—or in other words 'sums'—the frequency-weighted noise to give sound exposure and the metric used is pressure squared times time, often Pa²·s, but Pa²·h is also used. However, because the unit of sound was historically described in decibels, the exposure is most often described in terms of sound exposure level (SEL), the logarithmic conversion of sound exposure into decibels.

Personal noise dosimeter

A common variant of the sound level meter is a noise dosemeter (dosimeter in American English). However, this is now formally known as a personal sound exposure meter (PSEM) and has its own international standard IEC 61252:1993.

A noise dosimeter (American) or noise dosemeter (British) is a specialized sound level meter intended specifically to measure the noise exposure of a person integrated over a period of time; usually to comply with Health and Safety regulations such as the Occupational Safety and Health (OSHA) 29 CFR 1910.95 Occupational Noise Exposure Standard [10] or EU Directive 2003–10/EC.

This is normally intended to be a body-worn instrument and thus has a relaxed technical requirement, as a body-worn instrument—because of the presence of the body—has a poorer overall acoustic performance. A PSEM gives a read-out based on sound exposure, usually Pa²·h, and the older 'classic' dosimeters giving the metric of 'percentage dose' are no longer used in most countries. The problem with "%dose" is that it relates to the political situation and thus any device can become obsolete if the "100%" value is changed by local laws.

Traditionally, noise dosemeters were relatively large devices with a microphone mounted near the ear and having a cable going to the instrument body, itself usually belt worn. These devices had several issues, mainly the reliability of the cable and the disturbance to the user's normal work mode, caused by the presence of the cable. In 1997 following a UK research grant an EU patent was issued for the first of a range of devices that were so small that they resembled a radiation badge and no cable was needed as the whole unit could be fitted near the ear. UK designer and manufacturer, Cirrus Research, introduced the doseBadge personal noise dosimeter, which was the world's first truly wireless noise dosimeter. [7] Today these devices measure not only simple noise dose, but some even have four separate dosemeters, each with many of the functions of a full-sized sound level meter, including in the latest models full octave band analysis.

Classes

IEC standards divide sound level meters into two "classes". Sound level meters of the two classes have the same functionality, but different tolerances for error. Class 1 instruments have a wider frequency range and a tighter tolerance than a lower cost Class 2 unit. This applies to both the sound level meter itself as well as the associated calibrator. Most national standards permit the use of "at least a Class 2 instrument". For many measurements, it is not necessary to use a Class 1 unit; these are best employed for research and law enforcement.

Similarly, the American National Standards Institute (ANSI) specifies sound level meters as three different Types 0, 1 and 2. These are described, as follows, in the Occupational Safety and Health OSHA Technical Manual TED01-00-015, Chapter 5, OSHA Noise and Hearing Conservation, Appendix III:A, [11] "These ANSI standards set performance and accuracy tolerances according to three levels of precision: Types 0, 1, and 2. Type 0 is used in laboratories, Type 1 is used for precision measurements in the field, and Type 2 is used for general-purpose measurements. For compliance purposes, readings with an ANSI Type 2 sound level meter and dosimeter are considered to have an accuracy of ±2 dBA, while a Type 1 instrument has an accuracy of ±1 dBA. A Type 2 meter is the minimum requirement by OSHA for noise measurements and is usually sufficient for general-purpose noise surveys. The Type 1 meter is preferred for the design of cost-effective noise controls. For unusual measurement situations, refer to the manufacturer's instructions and appropriate ANSI standards for guidance in interpreting instrument accuracy."

Measurements

Sound level label examples
Description
Label
 Level A-weighted equivalent  LAeq
 Level A-weighted fast maximum  LAFmax
 Level C-weighted slow minimum  LCSmin
 Level Z-weighted impulse maximum  LZImax

Labels used to describe sound and noise level values are defined in the IEC Standard 61672-1:2013 [12] For labels, the first letter is always an L. This stands for Level, as in the sound pressure level measured through a microphone or the electronic signal level measured at the output from an audio component, such as a mixing desk. Measurement results depend on the frequency weighting (how the sound level meter responds to different sound frequencies), and time weighting (how the sound level meter reacts to changes in sound pressure with time) applied. [1]

Frequency weighting

A, C and Z frequency weightings for sound Sound-Pressure-Level-Frequency-Weightings-wiki.jpg
A, C and Z frequency weightings for sound

The second letter indicates the frequency weighting. "Pattern approved" sound level meters typically offer noise measurements with A, C and Z frequency weighting. [13]

Z-weighting represents the sound pressure equally at all frequencies. A-weighting, weights lower and higher frequencies much less, and has a slight boost in the mid-range, representing the sensitivity of normal human hearing at low (quiet) levels. C-Weighting, more sensitive to the lower frequencies, represents what humans hear when the sound is loud (near 100 dB SPL).

The IEC 61672-1:2013 mandates the inclusion of an A-weighting filter in all sound level meters, and also describes C and Z (zero) frequency weightings. The older B and D frequency weightings are now obsolete and are no longer described in the standard.

In almost all countries, the use of A-weighting is mandated to be used for the protection of workers against noise-induced hearing loss. The A-weighting curve was based on the historical equal-loudness contours and while arguably A-weighting is no longer the ideal frequency weighting on purely scientific grounds, it is nonetheless the legally required standard for almost all such measurements and has the huge practical advantage that old data can be compared with new measurements. It is for these reasons that A-weighting is the only weighting mandated by the international standard, the frequency weightings 'C' and 'Z' being options.

Originally, the A-weighting was only meant for quiet sounds in the region of 40 dB sound pressure level (SPL), but is now mandated for all levels. C-weighting is however still used in the measurement of the peak value of a noise in some legislation, but B-weighting – a halfway house between 'A' and 'C' has almost no practical use. D-weighting was designed for use in measuring aircraft noise when non-bypass jets were being measured; after the demise of Concord, these are all military types. For all civil aircraft noise measurements, A-weighting is used, as is mandated by the ISO and ICAO standards.

Time weighting

F-S-time-weighting-1200px.png
I-time-weighting-1200px.png
Graphs of fast, slow, and impulse time weightings applied so that sound levels measured are easier to read on a sound level meter.

If the third letter is F, S or I, this represents the time weighting, with F = fast, S = slow, I = impulse. [14] Time weighting is applied so that levels measured are easier to read on a sound level meter. The time weighting damps sudden changes in level, thus creating a smoother display.

The graph indicates how this works. In this example, the input signal suddenly increases from 50 dB to 80 dB, stays there for 6 seconds, then drops back suddenly to the initial level.

A slow measurement (yellow line) will take approximately 5 seconds (attack time) to reach 80 dB and around 6 seconds (decay time) to drop back down to 50 dB. S is appropriate when measuring a signal that fluctuates a lot.

A fast measurement (green line) is quicker to react. It will take approximately 0.6 seconds to reach 80 dB and just under 1 second to drop back down to 50 dB. F may be more suitable where the signal is less impulsive.

The decision to use fast or slow is often reached by what is prescribed in a standard or a law. However, the following can be used as a guideline: The slow characteristic is mainly used in situations where the reading with the fast response fluctuates too much (more than about 4 dB) to give a reasonably well-defined value. Modern digital displays largely overcome the problem of fluctuating analogue meters by indicating the maximum r.m.s. value for the preceding second. [15]

An impulse measurement (blue line) will take approximately 0.3 seconds to reach 80 dB and over 9 seconds to drop back down to 50 dB. The impulse response, I can be used in situations where there are sharp impulsive noises to be measured, such as fireworks or gunshots.

LAT or Leq: Equivalent continuous sound level

Graph of LAeq sound level measurement calculated over 5 minutes LAeq 5 minute.jpg
Graph of LAeq sound level measurement calculated over 5 minutes
Graph of a continuous LAeq sound level measurement LAeq.jpg
Graph of a continuous LAeq sound level measurement

eq = equivalent. Equivalent values are a form of time weighting that is easier to read on a display than the instantaneous sound level.

If you look at these graphs of sound level over time, the area under the blue curve represents the energy. The horizontal red line drawn to represent the same area under the blue curve, gives us the LAeq. That is the equivalent value or average of the energy over the entire graph.

LAeq is not always a straight line. If the LAeq is plotted as the equivalent from the beginning of the graph to each of the measurement points, the plot is shown in the second graph.

Sound exposure level—in decibels—is not much used in industrial noise measurement. Instead, the time-averaged value is used. This is the time average sound level or as it is usually called the 'equivalent continuous sound level' has the formal symbol LAT as described in paragraph 3,9 "Definitions" of IEC 61672-1 where many correct formal symbols and their common abbreviations are given. These mainly follow the formal ISO acoustic definitions. However, for mainly historical reasons, LAT is commonly referred to as Leq. [16]

Formally, LAT is 10 times the base 10 logarithm of the ratio of a root-mean-square A-weighted sound pressure during a stated time interval to the reference sound pressure and there is no time constant involved. To measure LAT an integrating-averaging meter is needed; this in concept takes the sound exposure, divides it by time, and then takes the logarithm of the result.

Short Leq

An important variant of overall LAT is "short Leq" where very short Leq values are taken in succession, say at 1/8 second intervals, each being stored in a digital memory. These data elements can either be transmitted to another unit or be recovered from the memory and re-constituted into almost any conventional metric long after the data has been acquired. This can be done using either dedicated programs or standard spreadsheets. Short Leq has the advantage that as regulations change, old data can be re-processed to check if a new regulation is met. It also permits data to be converted from one metric to another in some cases. Today almost all fixed airport noise monitoring systems, which are in concept just complex sound level meters, use short Leq as their metric, as a steady stream of the digital one second Leq values can be transmitted via telephone lines or the Internet to a central display and processing unit. Short Leq is a feature of most commercial integrating sound level meters—although some manufacturers give it many different names.

Short Leq is a very valuable method for acoustic data storage; initially, a concept of the French Government's Laboratoire National d'Essais (ref 1), it has now become the most common method of storing and displaying a true time history of the noise in professional commercial sound level meters. The alternative method, which is to generate a time history by storing and displaying samples of the exponential sound level, displays too many artifacts of the sound level meter to be as valuable and such sampled data cannot be readily combined to form an overall set of data.

Until 2003 there were separate standards for exponential and linear integrating sound level meters, (IEC 60651 and IEC 60804—both now withdrawn), but since then the combined standard IEC 61672 has described both types of meter. For short Leq to be valuable the manufacturer must ensure that each separate Leq element fully complies with IEC 61672.

Graph of an LAFmax sound level measurement calculated every minute LAFmax 1 minute.jpg
Graph of an LAFmax sound level measurement calculated every minute

Lmax and Lmin

If the words max or min appear in the label, this simply represents the maximum or minimum value measured over a certain period of time.

LCpk: peak sound pressure level

Most national regulations also call for the absolute peak value to be measured to protect workers hearing against sudden large pressure peaks, using either 'C' or 'Z' frequency weighting.[ citation needed ] 'Peak sound pressure level' should not be confused with 'MAX sound pressure level'. 'Max sound pressure level' is simply the highest RMS reading a conventional sound level meter gives over a stated period for a given time-weighting (S, F, or I) and can be many decibels less than the peak value.[ citation needed ] In the European Union, the maximum permitted value of the peak sound level is 140 dB(C)[ citation needed ] and this equates to 200 Pa pressure. The symbol for the A-frequency and S-time weighted maximum sound level is LASmax. For the C-frequency weighted peak it is LCpk or LC,peak.

Standardization

Sound level meters

Octave filters

Personal noise dosimeters

Measurement microphones

Room acoustics

Equipment safety

IEC61010-1 Ed. 2.0 (2001–02)

International standards

The following International standards define sound level meters, PSEM and associated devices. Most countries' national standards follow these very closely, the exception being the US. In many cases the equivalent European standard, agreed by the EU, is designated for example EN 61672 and the UK national standard then becomes BS. EN 61672.

These International Standards were prepared by IEC technical committee 29:Electroacoustics, in cooperation with the International Organization of Legal Metrology (OIML).

Until 2003 there were separate standards for exponential and linear integrating sound level meters, but since then IEC 61672 has described both types. The classic exponential meter was originally described in IEC 123 for 'industrial' meters followed by IEC 179 for 'precision' meters. Both of these were replaced by IEC 651, later renamed IEC 60651, while the linear integrating meters were initially described by IEC 804, later renamed IEC 60804. Both IEC 60651 and 60804 included four accuracy classes, called "types". In IEC 61672 these were reduced to just two accuracy classes 1 and 2. New in the standard IEC 61672 is a minimum 60 dB linear span requirement and Z-frequency-weighting, with a general tightening of limit tolerances, as well as the inclusion of maximum allowable measurement uncertainties for each described periodic test. The periodic testing part of the standard (IEC61672.3) also requires that manufacturers provide the testing laboratory with correction factors to allow laboratory electrical and acoustic testing to better mimic Free field (acoustics) responses. Each correction used should be provided with uncertainties, [17] that need to be accounted for in the testing laboratory final Measurement uncertainty budget. This makes it unlikely that a sound level meter designed to the older 60651 and 60804 standards will meet the requirements of IEC 61672 : 2013. These 'withdrawn' standards should no longer be used, especially for any official purchasing requirements, as they have significantly poorer accuracy requirements than IEC 61672.

Military standards

Combatants in every branch of the United States' military are at risk for auditory impairments from steady state or impulse noises. While applying double hearing protection helps prevent auditory damage, it may compromise effectiveness by isolating the user from his or her environment. With hearing protection on, a soldier is less likely to be aware of his or her movements, alerting the enemy to their presence. Hearing protection devices (HPD) could also require higher volume levels for communication, negating their purpose. [18]

Organizations

Pattern approval and periodic testing

A problem in selecting a sound level meter is "How do you know if it complies with its claimed standard?" This is a difficult question and IEC 61672 part 2 [24] tries to answer this by the concept of "pattern approval". A manufacturer has to supply instruments to a national laboratory which tests one of them and if it meets its claims issue a formal Pattern Approval certificate. [25] In Europe, the most common approval is often considered to be that from the PTB in Germany (Physikalisch-Technische Bundesanstalt). If a manufacturer cannot show at least one model in his range that has such approval, it is reasonable to be wary, but the cost of this approval militates against any manufacturer having all his range approved. Inexpensive sound level meters (under $200) are unlikely to have a Pattern Approval and may produce incorrect measurement results.

Even the most accurate approved sound level meter must be regularly checked for sensitivity—what most people loosely call 'calibration'. The procedures for periodic testing are defined within IEC61672.3-2013. To ensure accuracy in periodic testing, procedures should be carried out by a facility that can produce results traceable to International Laboratory Accreditation Cooperation, or other local International Laboratory Accreditation Cooperation signatories.

For a simple single level and frequency check, units consisting of a computer controlled generator with additional sensors to correct for humidity, temperature, battery voltage and static pressure can be used. The output of the generator is fed to a transducer in a half-inch cavity into which the sound level meter microphone is inserted. The acoustic level generated is 94 dB which is 1 pascal and is at a frequency of 1 kHz where all the frequency weightings have the same sensitivity.

For a complete sound level meter check, periodic testing outlined in IEC61672.3-2013 should be carried out. These tests excite the sound level meter across the entire frequency and dynamic range ensuring compliance with expected design goals defined in IEC61672.1-2013.

ANSI/IEC: the Atlantic divide

Sound level meters are also divided into two types in "the Atlantic divide". Sound level meters meeting the US American National Standards Institute (ANSI) specifications [26] cannot usually meet the corresponding International Electrotechnical Commission (IEC) specifications [27] at the same time, as the ANSI standard describes instruments that are calibrated to a randomly incident wave, i.e. a diffuse sound field, while internationally meters are calibrated to a free field wave, that is sound coming from a single direction. Further, US dosimeters have an exchange rate of level against time where every 5 dB increase in level halves the permitted exposure time; whereas in the rest of the world a 3 dB increase in level halves the permitted exposure time. The 3 dB doubling method is called the "equal energy" rule and there is no possible way of converting data taken under one rule to be used under the other. Despite these differences, many developing countries refer to both US and international specifications within one instrument in their national regulations. Because of this, many commercial PSEM have dual channels with 3 and 5 dB doubling, some even having 4 dB for the U.S. Air Force.

Other applications

Building acoustics, sound insulation and reverberation time

Some advanced sound level meters can also include reverberation time (RT60) (a measure of the time required for the sound to "fade away" in an enclosed area after the source of the sound has stopped) measurement capabilities. Measurements can be done using the integrated impulse response or interrupted noise methods. Such sound level meters should comply with latest ISO 3382-2 and ASTM E2235-04 measurement standards.

Required for measuring the acoustics in buildings is a signal generator that provides pink or white noise through an amplifier and omnidirectional speakers. In fact, the omnidirectional speaker, or sound source, should provide an equal dispersion of sound throughout the room. To achieve accurate measurements, sound should radiate evenly. This can be achieved using a spherical distribution aligning 12 speakers in a so-called dodecahedral configuration, as illustrated by Brüel & Kjær OmniPower Sound Source Type 4292. All speakers should be connected in a series–parallel network, to achieve in-phase operation and impedance matching to the amplifier.

The reverberation-time measurements are often used to calculate wall/partition sound insulation or to quantify and validate building acoustics. [28]

Noise monitoring stations

Noise monitoring station at Muir Woods National Monument in California Sound Level Monitor.jpg
Noise monitoring station at Muir Woods National Monument in California

Some applications require the ability to monitor noise continuously on a permanent or semi-permanent basis. Some manufacturers offer permanent and semi-permanent noise monitoring stations for this purpose. [30] [31] Such monitoring stations are typically based on a sound level meter at the heart and some added capabilities such as remote communication, GPS, and weather stations. These can often also be powered using solar power. Applications for such monitoring stations include airport noise, construction noise, mining noise, traffic noise, rail noise, community noise, wind farm noise, industrial noise, etc.

Modern monitoring stations can also offer remote communication capabilities using cellular modems, WiFi networks or direct LAN wires. Such devices allow for real-time alerts and notifications via email and text messages upon exceeding a certain dB level. Systems can also remotely email reports on a daily, weekly or monthly basis. Real-time data publication is often also desired, which can be achieved by pushing data to a website. [32] [33]

Smartphone applications

NIOSH sound level meter application (app) NIOSHSoundLevelMeter.jpg
NIOSH sound level meter application (app)

The ubiquity of smartphones, their constant network connectivity, the built-in geographic information system functionality and user-interactivity features present a great opportunity to revolutionize the way we look at noise, its measurement, and its effects on hearing and overall health. The ability to acquire and display real-time noise exposure data raises people's awareness about their work (and off-work) environment and allows them to make informed decisions about hearing hazards and overall well-being. The National Institute for Occupational Safety and Health (NIOSH) conducted a pilot study to select and characterize the functionality and accuracy of smartphone sound measurement applications (apps) as an initial step in a broader effort to determine whether these apps can be relied on to conduct participatory noise monitoring studies in the workplace. [35]

Researchers reported that challenges remain with using smartphones to collect and document noise exposure data due to encounters with privacy and collection of personal data, motivation to participate in such studies, corrupted or bad data, and the ability to store the data collected. Researchers concluded that smartphone sound apps can serve to empower workers and help them make educated decisions about their workplace environments. [36] Although most smartphone sound measurement apps are not accurate enough to be used for legally required measurements, the NIOSH Sound Level Meter app met the requirements of IEC 61672/ANSI S1.4 Sound Level Meter Standards (Electroacoustics - Sound Level Meters - Part 3: Periodic Tests). [37] Calibrated microphones greatly enhances the accuracy and precision of smartphone-based noise measurements. To calibrate the sound level meter apps one must use an acoustical calibrator rather than relying on the pre-defined profiles. This study indicated that the gap between professional instruments and smartphone-based apps are narrowing. [38]

Healthy Hearing, [39] an organization dedicated to hearing health, reported on the top smartphone sound level meter apps: [40] NIOSH Sound Level Meter, [41] Decibel X, [42] and Too Noisy Pro. [43]

See also

General:

Related Research Articles

The decibel is a relative unit of measurement equal to one tenth of a bel (B). It expresses the ratio of two values of a power or root-power quantity on a logarithmic scale. Two signals whose levels differ by one decibel have a power ratio of 101/10 or root-power ratio of 10120.

<span class="mw-page-title-main">Weighting filter</span>

A weighting filter is used to emphasize or suppress some aspects of a phenomenon compared to others, for measurement or other purposes.

<span class="mw-page-title-main">Noise</span> Unwanted sound

Noise is unwanted sound considered unpleasant, loud, or disruptive to hearing. From a physics standpoint, there is no distinction between noise and desired sound, as both are vibrations through a medium, such as air or water. The difference arises when the brain receives and perceives a sound.

<span class="mw-page-title-main">Microphone</span> Device that converts sound into an electrical signal

A microphone, colloquially called a mic, is a transducer that converts sound into an electrical signal. Microphones are used in many applications such as telephones, hearing aids, public address systems for concert halls and public events, motion picture production, live and recorded audio engineering, sound recording, two-way radios, megaphones, and radio and television broadcasting. They are also used in computers and other electronic devices, such as mobile phones, for recording sounds, speech recognition, VoIP, and other purposes, such as ultrasonic sensors or knock sensors.

<span class="mw-page-title-main">Noise pollution</span> Excessive, displeasing human, animal, or machine-created environmental noise

Noise pollution, or sound pollution, is the propagation of noise or sound with ranging impacts on the activity of human or animal life, most of which are harmful to a degree. The source of outdoor noise worldwide is mainly caused by machines, transport and propagation systems. Poor urban planning may give rise to noise disintegration or pollution, side-by-side industrial and residential buildings can result in noise pollution in the residential areas. Some of the main sources of noise in residential areas include loud music, transportation, lawn care maintenance, construction, electrical generators, wind turbines, explosions and people.

<span class="mw-page-title-main">Loudness</span> Subjective perception of sound pressure

In acoustics, loudness is the subjective perception of sound pressure. More formally, it is defined as the "attribute of auditory sensation in terms of which sounds can be ordered on a scale extending from quiet to loud". The relation of physical attributes of sound to perceived loudness consists of physical, physiological and psychological components. The study of apparent loudness is included in the topic of psychoacoustics and employs methods of psychophysics.

Sound pressure or acoustic pressure is the local pressure deviation from the ambient atmospheric pressure, caused by a sound wave. In air, sound pressure can be measured using a microphone, and in water with a hydrophone. The SI unit of sound pressure is the pascal (Pa).

Occupational noise is the amount of acoustic energy received by an employee's auditory system when they are working in the industry. Occupational noise, or industrial noise, is often a term used in occupational safety and health, as sustained exposure can cause permanent hearing damage. Occupational noise is considered an occupational hazard traditionally linked to loud industries such as ship-building, mining, railroad work, welding, and construction, but can be present in any workplace where hazardous noise is present.

<span class="mw-page-title-main">Earplug</span> Device to protect ears from loud noises

An earplug is a device that is inserted in the ear canal to protect the user's ears from loud noises, intrusion of water, foreign bodies, dust or excessive wind. Since they reduce the sound volume, earplugs may prevent hearing loss and tinnitus, in some causes.

<span class="mw-page-title-main">Equal-loudness contour</span> Frequency characteristics of hearing and perceived volume

An equal-loudness contour is a measure of sound pressure level, over the frequency spectrum, for which a listener perceives a constant loudness when presented with pure steady tones. The unit of measurement for loudness levels is the phon and is arrived at by reference to equal-loudness contours. By definition, two sine waves of differing frequencies are said to have equal-loudness level measured in phons if they are perceived as equally loud by the average young person without significant hearing impairment.

A weighting curve is a graph of a set of factors, that are used to 'weight' measured values of a variable according to their importance in relation to some outcome. An important example is frequency weighting in sound level measurement where a specific set of weighting curves known as A-, B-, C- and D-weighting as defined in IEC 61672 are used. Unweighted measurements of sound pressure do not correspond to perceived loudness because the human ear is less sensitive at low and high frequencies, with the effect more pronounced at lower sound levels. The four curves are applied to the measured sound level, for example by the use of a weighting filter in a sound level meter, to arrive at readings of loudness in phons or in decibels (dB) above the threshold of hearing.

<span class="mw-page-title-main">Weighting</span>

The process of weighting involves emphasizing the contribution of particular aspects of a phenomenon over others to an outcome or result; thereby highlighting those aspects in comparison to others in the analysis. That is, rather than each variable in the data set contributing equally to the final result, some of the data is adjusted to make a greater contribution than others. This is analogous to the practice of adding (extra) weight to one side of a pair of scales in order to favour either the buyer or seller.

In acoustics, noise measurement can be for the purpose of measuring environmental noise or measuring noise in the workplace. Applications include monitoring of construction sites, aircraft noise, road traffic noise, entertainment venues and neighborhood noise. One of the definitions of noise covers all "unwanted sounds". When sound levels reach a high enough intensity, the sound, whether it is wanted or unwanted, may be damaging to hearing. Environmental noise monitoring is the measurement of noise in an outdoor environment caused by transport, industry and recreational activities. The laws and limits governing environmental noise monitoring differ from country to country.

Analysis of sound and acoustics plays a role in such engineering tasks as product design, production test, machine performance, and process control. For instance, product design can require modification of sound level or noise for compliance with standards from ANSI, IEC, and ISO. The work might also involve design fine-tuning to meet market expectations. Here, examples include tweaking an automobile door latching mechanism to impress a consumer with a satisfying click or modifying an exhaust manifold to change the tone of an engine's rumble. Aircraft designers are also using acoustic instrumentation to reduce the noise generated on takeoff and landing.

<span class="mw-page-title-main">A-weighting</span> Frequency response curves used in sound pressure level measurement

A-weighting is the most commonly used of a family of curves defined in the International standard IEC 61672:2003 and various national standards relating to the measurement of sound pressure level. A-weighting is applied to instrument-measured sound levels in an effort to account for the relative loudness perceived by the human ear, as the ear is less sensitive to low audio frequencies. It is employed by arithmetically adding a table of values, listed by octave or third-octave bands, to the measured sound pressure levels in dB. The resulting octave band measurements are usually added to provide a single A-weighted value describing the sound; the units are written as dB(A). Other weighting sets of values – B, C, D and now Z – are discussed below.

<span class="mw-page-title-main">Noise dosimeter</span>

A noise dosimeter or noise dosemeter is a specialized sound level meter intended specifically to measure the noise exposure of a person integrated over a period of time; usually to comply with Health and Safety regulations such as the Occupational Safety and Health (OSHA) 29 CFR 1910.95 Occupational Noise Exposure Standard or EU Directive 2003/10/EC.

<span class="mw-page-title-main">Sound</span> Vibration that travels via pressure waves in matter

In physics, sound is a vibration that propagates as an acoustic wave through a transmission medium such as a gas, liquid or solid. In human physiology and psychology, sound is the reception of such waves and their perception by the brain. Only acoustic waves that have frequencies lying between about 20 Hz and 20 kHz, the audio frequency range, elicit an auditory percept in humans. In air at atmospheric pressure, these represent sound waves with wavelengths of 17 meters (56 ft) to 1.7 centimeters (0.67 in). Sound waves above 20 kHz are known as ultrasound and are not audible to humans. Sound waves below 20 Hz are known as infrasound. Different animal species have varying hearing ranges.

<span class="mw-page-title-main">Brüel & Kjær</span> Danish multinational company

Brüel & Kjær is a Danish multinational engineering and electronics company headquartered in Nærum, near Copenhagen. It was the largest producer in the world of equipment for acoustic and vibrational measurements. Brüel & Kjær is a subsidiary of Spectris.

Effective perceived noise in decibels (EPNdB) or Effective Perceived Noise Level (EPNL) is a measure of the relative noisiness of an individual aircraft pass-by event. It is used for aircraft noise certification and applies to an individual aircraft, not the noise exposure from an airport. Separate ratings are stated for takeoff, overflight and landing events, and represent the integrated power sum of noisiness during the event. Instantaneous value of noisiness is computed with the PNL or PNdB metric over the period within which the noise from the aircraft is within 10 dB of the maximum noise It is defined, with computational instructions, in Annex 16 of the Convention on International Civil Aviation and in Part 36 of the US Federal Aviation Regulations. The scaling is such that the EPNdB rating represents the integrated noisiness over a ten-second period; EPNdB of 100 dB means that the event has the same integrated noisiness as a 100 PNdB sound lasting ten seconds. Direct comparison with A-weighted sound pressure level, which is used for many other environmental sound measurements, is not possible because PNdB is a noisiness metric rather than a sound pressure metric.

<span class="mw-page-title-main">Occupational hearing loss</span> Form of hearing loss

Occupational hearing loss (OHL) is hearing loss that occurs as a result of occupational hazards, such as excessive noise and ototoxic chemicals. Noise is a common workplace hazard, and recognized as the risk factor for noise-induced hearing loss and tinnitus but it is not the only risk factor that can result in a work-related hearing loss. Also, noise-induced hearing loss can result from exposures that are not restricted to the occupational setting.

References

  1. 1 2 "What Is A Sound Level Meter?". Brüel & Kjær . Retrieved 24 February 2021.
  2. "Environment Noise / Noise Assessment". Brüel & Kjær . Retrieved 24 February 2021.
  3. Kanji A, Khoza-Shangase K, Ntlhakana L (June 2019). "Noise-induced hearing loss: what South African mineworkers know". International Journal of Occupational Safety and Ergonomics. 25 (2): 305–310. doi:10.1080/10803548.2017.1412122. PMID   29214904. S2CID   46754344.
  4. "Sharpness 2 Loudness calculation" (PDF). Psychoacoustic Analyses. HEAD acoustics GmbH.
  5. "History - Sound and Vibration". Brüel & Kjær . Retrieved 24 February 2021.
  6. "Our History". Pulsar Instruments Plc . 13 November 2021. Retrieved 7 March 2022.
  7. 1 2 "Our History". Cirrus Research . Retrieved 24 February 2021.
  8. "IEC 61672-1:2013 – IEC Webstore". webstore.iec.ch.
  9. rebecca (5 March 2019). "What is Sound Level Meter?". Scarlet Tech. Retrieved 25 October 2022.
  10. "OSHA 29 CFR 1910.95 Occupational Noise Exposure Standard". Occupational Health and Safety Administration. 3 March 2011. Retrieved 10 September 2012.
  11. "OSHA Noise and Hearing Conservation, Appendix III:A". Occupational Health and Safety Administration. 7 March 1996. Retrieved 9 April 2013.
  12. International Electrotechnical Commission, IEC. "IEC 61672-1:2013 Electroacoustics - Sound level meters - Part 1: Specifications". iec.ch. Retrieved 16 March 2018.
  13. "The Top 10 posts from our blog" (PDF). cms.esi.info. Cirrus Research PLC. Retrieved 13 March 2021.
  14. "fast slow impulse time weighting what do they mean". NTi Audio. Retrieved 16 March 2018.
  15. Noise Measurement Briefing, Product Technology Partners Ltd., archived from the original on 30 June 2008
  16. Kjær, Brüel. "Leq - LAE/SEL - When To Choose What?". Brüel & Kjær. Retrieved 25 April 2021.
  17. "IEC 62585:2012 – IEC Webstore". webstore.iec.ch.
  18. 1 2 Amrein BE (2015). Military standard 1474E: Design criteria for noise limits vs. operational effectiveness. Proceedings of Meetings on Acoustics. Acoustical Society of America. p. 040005. doi: 10.1121/2.0000207 .
  19. "Department of Defense Design Criteria, Noise Limits". Department of Defense. 12 February 1997. Retrieved 18 September 2012.
  20. "Department of Defense Design Criteria, Noise Limits" (PDF). Department of Defense. 15 April 2015. Retrieved 3 August 2018.
  21. 1 2 3 4 De Paolis A, Bikson M, Nelson JT, de Ru JA, Packer M, Cardoso L (June 2017). "Analytical and numerical modeling of the hearing system: Advances towards the assessment of hearing damage". Hearing Research. 349: 111–128. doi:10.1016/j.heares.2017.01.015. PMC   7000179 . PMID   28161584.
  22. 1 2 Nakashima A (November 2015). A comparison of metrics for impulse noise exposure (PDF) (Report). Defense Research and Development Canada. Retrieved 3 July 2018.
  23. "Test Operations Procedure:Sound Level Measurements". US Department of Defense. 1 January 2011. Retrieved 18 September 2012.
  24. "Electroacoustics - Sound level meters - Part 2: Pattern evaluation tests". International Electrotechnical Commission.
  25. Audio, NTi. "Sample Pattern Approval Certificate" (PDF). www.nti-audio.com.
  26. "American National Standard for Sound Level Meters". American National Standards Institute. 2006. Retrieved 29 April 2013.
  27. "IEC 61672-1, Electroacoustics – Sound level meters – Part 1: Specifications". International Electrotechnical Commission (IEC). May 2002. Retrieved 29 April 2013.
  28. "Environment Noise / Noise Assessment | Brüel & Kjær". www.bksv.com. Retrieved 19 May 2021.
  29. Barringer F (21 February 2011). "Shhh, and Not Because the Fauna Are Sleeping". The New York Times. Retrieved 2 August 2014.
  30. Audio, NTi. "Noise Monitoring Station". www.nti-audio.com.
  31. Ltd, Acoustic Research Labs Pty. "Acoustic Research Labs Pty Ltd". www.acousticresearch.com.au.
  32. Audio, NTi. "NoiseScout Unattended Noise Monitoring Solution". www.noisescout.com.
  33. "Acoustic Research Labs Pty Ltd". www.acousticresearch.com.au.
  34. "CDC – Noise and Hearing Loss Prevention – NIOSH Sound Level Meter App – NIOSH Workplace Safety and Health Topic". www.cdc.gov. Retrieved 30 January 2017.
  35. Kardous CA, Shaw PB (April 2014). "Evaluation of smartphone sound measurement applications". The Journal of the Acoustical Society of America. 135 (4): EL186-92. Bibcode:2014ASAJ..135L.186K. doi:10.1121/1.4865269. PMC   4659422 . PMID   25236152.
  36. "So How Accurate Are These Smartphone Sound Measurement Apps?". NIOSH Science Blog. Centers for Disease Control. Retrieved 30 January 2017.PD-icon.svg This article incorporates text from this source, which is in the public domain .
  37. Celestina M, Hrovat J, Kardous CA (2018). "Smartphone-based sound level measurement apps: Evaluation of compliance with international sound level meter standards". Applied Acoustics. 139: 119–128. doi:10.1016/j.apacoust.2018.04.011. ISSN   0003-682X. S2CID   116822722.
  38. Kardous CA, Shaw PB (October 2016). "Evaluation of smartphone sound measurement applications (apps) using external microphones-A follow-up study". The Journal of the Acoustical Society of America. 140 (4): EL327–EL333. Bibcode:2016ASAJ..140L.327K. doi:10.1121/1.4964639. PMC   5102154 . PMID   27794313.
  39. "Hearing aid and hearing clinic directory - Healthy Hearing". Healthy Hearing. Retrieved 4 December 2018.
  40. "The best smartphone decibel meter apps to measure noise levels". Healthy Hearing. 26 November 2014. Retrieved 4 December 2018.
  41. "CDC - Noise and Hearing Loss Prevention - NIOSH Sound Level Meter App - NIOSH Workplace Safety and Health Topic". www.cdc.gov. 5 October 2018. Retrieved 4 December 2018.
  42. "Decibel X: Pro dBA Noise Meter". skypaw.com. Retrieved 4 December 2018.
  43. "A Noise Level Meter built for the Classroom.| Too Noisy". toonoisyapp.com. Retrieved 4 December 2018.

Further reading