Piezoresponse force microscopy

Last updated
PFM of BaTiO3 single crystal showing simultaneously acquired topography (top) and domain structure (bottom). The scale bar is 10 mm Bulk BTO PFM scan.png
PFM of BaTiO3 single crystal showing simultaneously acquired topography (top) and domain structure (bottom). The scale bar is 10 μm

Piezoresponse force microscopy (PFM) is a variant of atomic force microscopy (AFM) that allows imaging and manipulation of piezoelectric/ferroelectric materials domains. This is achieved by bringing a sharp conductive probe into contact with a ferroelectric surface (or piezoelectric material) and applying an alternating current (AC) bias to the probe tip in order to excite deformation of the sample through the converse piezoelectric effect (CPE). The resulting deflection of the probe cantilever is detected through standard split photodiode detector methods and then demodulated by use of a lock-in amplifier (LiA). In this way topography and ferroelectric domains can be imaged simultaneously with high resolution.

Contents

Basic principles

General overview

Piezoresponse force microscopy is a technique which since its inception and first implementation by Güthner and Dransfeld [1] has steadily attracted more and more interest. This is due in large part to the many benefits and few drawbacks that PFM offers researchers in varying fields from ferroelectrics, semiconductors and even biology. [2] In its most common format PFM allows for identification of domains from relatively large scale e.g. 100×100 µm2 scans right down to the nanoscale with the added advantage of simultaneous imaging of sample surface topography. Also possible is the ability to switch regions of ferroelectric domains with the application of a sufficiently high bias to the probe which opens up the opportunity of investigating domain formation on nanometre length scales with nanosecond time resolution. [3] Many recent advances have expanded the list of applications for PFM and further increased this powerful technique. Indeed what started as a user modified AFM has now attracted the attention of the major SPM manufacturers so much so that in fact many now supply ‘ready-made’ systems specifically for PFM each with novel features for research. This is testament to the growth of the field and reflects the numbers of users throughout the scientific world who are at the forefront of scientific research.

Top line shows an in-phase piezoresponse to the driving voltage and the bottom line shows a 180deg out-of-phase piezoresponse to driving voltage. Alignment of electric field and polarisation orientation (top right) results in an expansion of the domain, giving a positive deflection as measured by the photodiode. When the bias is negative the domain contracts giving a negative deflection as measured by the photodiode meaning that the piezoresponse will always be in-phase with the driving voltage. For anti-alignment of electric field and polarisation orientation (bottom right) a positive bias results in a contraction of the domain and so gives a negative deflection as measured by the photodiode therefore the piezoresponse is 180deg out-of-phase with the driving voltage. In this way the orientation of polarisation within a domain can be observed. Piezoresponse of parallel and antiparallel domains.png
Top line shows an in-phase piezoresponse to the driving voltage and the bottom line shows a 180° out-of-phase piezoresponse to driving voltage. Alignment of electric field and polarisation orientation (top right) results in an expansion of the domain, giving a positive deflection as measured by the photodiode. When the bias is negative the domain contracts giving a negative deflection as measured by the photodiode meaning that the piezoresponse will always be in-phase with the driving voltage. For anti-alignment of electric field and polarisation orientation (bottom right) a positive bias results in a contraction of the domain and so gives a negative deflection as measured by the photodiode therefore the piezoresponse is 180° out-of-phase with the driving voltage. In this way the orientation of polarisation within a domain can be observed.

Consider that a static or DC voltage applied to a piezoelectric surface will produce a displacement but as applied fields are quite low and the piezoelectric tensor coefficients are relatively small then the physical displacement will also be small such that it is below the level of possible detection of the system. Take as an example, the d33 piezoelectric tensor coefficient of BaTiO3, it has a value of 85.6 pm V−1 meaning that applying 1 V across the material results in a displacement of 85.6 pm or 0.0856 nm, a minute cantilever displacement even for the high precision of AFM deflection detection. In order to separate this low level signal from random noise a lock-in technique is used wherein a modulated voltage reference signal,

of frequency ω and amplitude Vac is applied to the tip giving rise to an oscillatory deformation of the sample surface,

from the equilibrium position d0 with amplitude D, and an associated phase difference φ. The resulting movement of the cantilever is detected by the photodiode and so an oscillating surface displacement is converted into an oscillating voltage. A lock-in-amplifier (LiA) is then able to retrieve the amplitude and phase of the CPE induced surface deformation by the process outlined below.

Converse piezoelectric effect

The converse piezoelectric effect (CPE) describes how an applied electric field will create a resultant strain which in turn leads to a physical deformation of the material. This effect can be described through the constitutive equations. [4] The CPE can be written as

where Xi is the strain tensor, dki is the piezoelectric tensor, and Ek is the electric field. If the piezoelectric tensor is considered to be that of the tetragonal crystal system (that of BaTiO3) then it is

such that the equation will lead to the strain components for an applied field. If the field is applied exclusively in one direction i.e. E3 for example, then the resulting strain components are: d31E3, d32E3, d33E3

Thus for an electric field applied along the c-axis of BaTiO3 i.e. E3, then the resulting deformation of the crystal will be an elongation along the c-axis and an axially symmetric contraction along the other orthogonal directions. PFM uses the effect of this deformation to detect domains and also to determine their orientation.

Conductive probe

The most important property of the probe for use in PFM is that it should be conducting. This is generally required in order to provide a means of applying a bias to the sample, and can be achieved through manufacturing standard silicon probes and coating them in a conductive material. Common coatings are platinum, gold, tungsten and even conductive diamond.

Scanning electron microscopy images of a PtIr5 coated scanning probe. From left to right shows images of increasing magnification where the scale bar in the first image is 50 mm and in the third is 200 nm. The first image shows the substrate, cantilever and the tip whereas the second image shows the tip geometry whilst the last image shows the tip apex and demonstrates the fine point that is achieved e.g. radius of curvature of less than 40 nm. Conductive probe SEM images.png
Scanning electron microscopy images of a PtIr5 coated scanning probe. From left to right shows images of increasing magnification where the scale bar in the first image is 50 μm and in the third is 200 nm. The first image shows the substrate, cantilever and the tip whereas the second image shows the tip geometry whilst the last image shows the tip apex and demonstrates the fine point that is achieved e.g. radius of curvature of less than 40 nm.

Lock-in amplifier

In the general case a lock-in amplifier (LiA) ‘compares’ an input signal against that of a reference signal (either generated internally or supplied by an external function generator) in order to separate the information contained in the input signal at the frequency of the reference signal. This is called demodulation and is done in a number of easy steps. The reference signal , and input signal, , are multiplied together to give the demodulator output,

where A is the input signal Amplitude and B is the reference signal Amplitude, ω is the frequency of both the reference and input signals, and φ is any phase shift between the two signals.

The above equation has an AC component at twice the frequency of the original signals (second term) and a DC component (first term) whose value is related to both the amplitude and phase of the input signal. The demodulator output is sent through a low-pass filter to remove the 2ω component and leave the DC component then the signal is integrated over a period of time defined as the Time Constant, τLiA which is a user-definable parameter. Several different outputs are commonly available from a LiA: X output is the demodulator output and Y is the second demodulator output which is shifted by 90° in reference to the first output, together they hold both the phase, θ, and magnitude, R, information and are given by

and

However, phase and amplitude of the input signal can also be calculated and made output from the LiA if desired, so that the full amount of information is available. The phase output can be determined from the following equation:

The magnitude is then given by:

This allows R to be calculated even if the input signal differs in phase from the reference signal.

Differentiating vertical and lateral PFM signals

Diagrams showing the effect of cantilever movement with the photodetector represented by the square with quadrants labelled A, B, C and D. Torsional bending of the cantilever (left) leads to a change in lateral deflection and (right) vertical displacement of the cantilever leads to a change in vertical deflection Cantilever movements and optical deflections.png
Diagrams showing the effect of cantilever movement with the photodetector represented by the square with quadrants labelled A, B, C and D. Torsional bending of the cantilever (left) leads to a change in lateral deflection and (right) vertical displacement of the cantilever leads to a change in vertical deflection

A basic interpretation of PFM (which is generally accepted) identifies that two modes of imaging are possible, one that is sensitive to out-of-plane and one to in-plane piezoresponse, termed, vertical and lateral PFM (VPFM and LPFM) respectively. [5] The separation of these components is possible through the use of a split photodiode detector, standard to all optical detection AFM systems. In this setup the detector is split into quadrants, nominally A, B, C and D. The centre of the entire detector outputs 0 V but as the laser spot moves a radial distance from this centre point, the magnitude of the voltage in output will increase linearly. A vertical deflection can be defined as {(A+B)-(C+D)}/(ABCD) so that now positive and negative voltages are ascribed to positive and negative cantilever vertical displacements. Similarly a lateral deflection is defined as {(B+D)-(A+C)}/(ABCD) to describe positive and negative torsional movements of the cantilever. So VPFM will utilise the vertical deflection signal from the photodiode detector so will only be sensitive to out-of-plane polar components and LPFM will utilise the lateral deflection signal from the photodiode and will only be sensitive to in-plane polar components.

For polar components orientated such that they are parallel to the electric field the resulting oscillating movement will be entirely in-phase with the modulated electric field but for an anti-parallel alignment the motion will be 180° out-of-phase. In this way it is possible to determine the orientation of the vertical components of polarisation from analysis of the phase information, φ, contained in the input signal, readily available after demodulation in the LiA, when using the VPFM mode. In a similar sense the orientations of in-plane polar components can also be determined from the phase difference when using the LPFM mode. The amplitude of the piezoresponse of either VPFM or LPFM is also given by the LiA, in the form of the magnitude, R.

Examples of PFM imaging

180deg ferroelectric domains in KTP as imaged by PFM. Below are the associated line profiles across the domains 180 degree ferroelectric domains with profiles.png
180° ferroelectric domains in KTP as imaged by PFM. Below are the associated line profiles across the domains

The image shows periodically poled 180° domains in potassium titanyl phosphate (KTP) as imaged by VPFM. In the image piezoresponse amplitude can be seen where dark areas represent the zero amplitude that is expected at domain boundaries where the unit cell is cubic i.e. centrosymmetric and so therefore not ferroelectric. On the left hand side piezoresponse phase can be seen where the measured phase changes to show the out-of-plane components that are pointing out of the screen, white areas, and into the screen, dark areas. The scan area is 20×10 µm2. Below each scan is the relevant cross-section that shows in arbitrary units the PR amplitude and phase.

PFM applied to biological materials

PFM has been successfully applied to a range of biological materials such as teeth, [6] bone, lung, [7] and single collagen fibrils. [8] It has been hypothesized that the endogenous piezoelectricity in these materials may be relevant in their mechanobiology. For example, using PFM it has been shown that a single collagen fibril as small as 100 nm behaves predominantly as a shear piezoelectric material with an effective piezoelectric constant of ~1 pm/V.

Advanced PFM modes

Several additions have been made to PFM that substantially increase the flexibility of the technique to probe nanoscale features.

Stroboscopic PFM

Stroboscopic PFM allows for time resolved imaging of switching in pseudo real-time. [9] A voltage pulse of amplitude much higher than the coercive voltage of the sample but shorter in duration than the characteristic switching time is applied to the sample and subsequently imaged. Further pulses with the same amplitude but longer in time are then applied with regular PFM imaging at the intervals. In this way a series of images showing the switching of the sample can be obtained. Typical pulses are of tens of nanoseconds in duration and are therefore capable of resolving the first nucleation sites of domain reversal and then observing how these sites evolve.

Contact resonance PFM

Remembering that in PFM an AC bias of a certain frequency causes a deformation of the sample material at that same frequency the system can be considered as a driven harmonic oscillator. As such there exists a resonance as a function of driving frequency. This effect has been exploited in PFM to provide an enhancement in the PR signal, thus allowing for a higher signal-to-noise ratio or similar signal-to-noise ratio at lower driving bias amplitude. [10] Typically this contact resonance is in the kilo- to mega-hertz range which is several times higher in frequency than the first free harmonic in air of the cantilever used. However a drawback is that the contact resonance is dependent not only on the dynamic response of the cantilever but also on the elastic modulus of the sample material immediately in contact with the probe tip and so therefore can change during scanning over different areas. This leads to a change in the measured PR amplitude and so is undesirable. One method of bypassing the inherent disadvantages of contact resonance PFM is to change the driving frequency in order to shadow or track the changes in the frequency of the contact resonance. This feature as developed by Asylum Research called Dual AC™ Resonance Tracking (DART) uses two limit frequencies on either side of the contact resonance peak and so can sense changes in the peak position. It is then possible to adapt the AC bias driving frequency correspondingly in order to maintain the signal boost that results from the contact resonance.

Switching spectroscopy (SS) PFM

In this technique the area underneath the PFM tip is switched with simultaneous acquisition of a hysteresis loop that can be analysed to obtain information about the sample properties. [11] A series of hysteresis loops are acquired across the sample surface in order to map the switching characteristics as a function of position. In this way an image representing switching properties such as coercive voltage, remnant polarisation, imprint and work of switching amongst others can be displayed in which each pixel displays the desired data from the hysteresis loop acquired at that point. This allows spatial analysis of switching properties to be compared with sample topography.

Band Excitation PFM

The Band Excitation (BE) technique for scanning probe microscopy uses a precisely determined waveform that contains specific frequencies to excite the cantilever or sample in an atomic force microscope to extract more information, and more reliable information from a sample. [12] [13] There are a myriad of details and complexities associated with implementing the BE technique. There is therefore a need to have a user friendly interface that allows typical microscopists access to this methodology. This software enables users of atomic force microscopes to easily: build complex band-excitation waveforms, set up the microscope scanning conditions, configure the input and output electronics to generate the waveform as a voltage signal and capture the response of the system, perform analysis on the captured response, and display the results of the measurement.

Pin Point PFM

The conventional PFM operates in contact mode in which the AFM tip is in contact with the sample during the scanning. Contact mode is not suitable for samples with features susceptible to damage or displacement by the tip's drag. In Pin Point PFM, the AFM tip does not contact the surface. The tip is halted at a height at which a predefined force threshold (a threshold at which piezoelectric response is optimal) is reached. At this height, the piezoelectric response is recorded before moving to the next point. In Pin Point mode, tip wear off is reduced significantly.

Advantages and disadvantages

Advantages

Disadvantages

Related Research Articles

In signal processing, group delay and phase delay are delay times experienced by a signal's various frequency components when the signal passes through a linear time-invariant system (LTI), such as a microphone, coaxial cable, amplifier, loudspeaker, telecommunications system or ethernet cable. These delays are generally frequency dependent, which means that different frequency components experience different delays. As a result, the signal's waveform experiences distortion as it passes through the system. This distortion can cause problems such as poor fidelity in analog video and analog audio, or a high bit-error rate in a digital bit stream. For a modulation signal, the information carried by the signal is carried exclusively in the wave envelope. Group delay therefore operates only with the frequency components derived from the envelope.

<span class="mw-page-title-main">Resonance</span> Tendency to oscillate at certain frequencies

Resonance describes the phenomenon of increased amplitude that occurs when the frequency of an applied periodic force is equal or close to a natural frequency of the system on which it acts. When an oscillating force is applied at a resonant frequency of a dynamic system, the system will oscillate at a higher amplitude than when the same force is applied at other, non-resonant frequencies.

<span class="mw-page-title-main">Electrical impedance</span> Opposition of a circuit to a current when a voltage is applied

In electrical engineering, impedance is the opposition to alternating current presented by the combined effect of resistance and reactance in a circuit.

<span class="mw-page-title-main">Atomic force microscopy</span> Type of microscopy

Atomic force microscopy (AFM) or scanning force microscopy (SFM) is a very-high-resolution type of scanning probe microscopy (SPM), with demonstrated resolution on the order of fractions of a nanometer, more than 1000 times better than the optical diffraction limit.

Dynamic mechanical analysis is a technique used to study and characterize materials. It is most useful for studying the viscoelastic behavior of polymers. A sinusoidal stress is applied and the strain in the material is measured, allowing one to determine the complex modulus. The temperature of the sample or the frequency of the stress are often varied, leading to variations in the complex modulus; this approach can be used to locate the glass transition temperature of the material, as well as to identify transitions corresponding to other molecular motions.

<span class="mw-page-title-main">Envelope detector</span> Electronic circuit

An envelope detector is an electronic circuit that takes a (relatively) high-frequency amplitude modulated signal as input and provides an output, which is the demodulated envelope of the original signal.

<span class="mw-page-title-main">Frequency mixer</span> Circuit that creates new frequencies from two signals

In electronics, a mixer, or frequency mixer, is an electrical circuit that creates new frequencies from two signals applied to it. In its most common application, two signals are applied to a mixer, and it produces new signals at the sum and difference of the original frequencies. Other frequency components may also be produced in a practical frequency mixer.

Scanning probe microscopy (SPM) is a branch of microscopy that forms images of surfaces using a physical probe that scans the specimen. SPM was founded in 1981, with the invention of the scanning tunneling microscope, an instrument for imaging surfaces at the atomic level. The first successful scanning tunneling microscope experiment was done by Gerd Binnig and Heinrich Rohrer. The key to their success was using a feedback loop to regulate gap distance between the sample and the probe.

In mathematics and signal processing, an analytic signal is a complex-valued function that has no negative frequency components. The real and imaginary parts of an analytic signal are real-valued functions related to each other by the Hilbert transform.

<span class="mw-page-title-main">Magnetic force microscope</span>

Magnetic force microscopy (MFM) is a variety of atomic force microscopy, in which a sharp magnetized tip scans a magnetic sample; the tip-sample magnetic interactions are detected and used to reconstruct the magnetic structure of the sample surface. Many kinds of magnetic interactions are measured by MFM, including magnetic dipole–dipole interaction. MFM scanning often uses non-contact AFM (NC-AFM) mode.

<span class="mw-page-title-main">Kelvin probe force microscope</span> Noncontact variant of atomic force microscopy

Kelvin probe force microscopy (KPFM), also known as surface potential microscopy, is a noncontact variant of atomic force microscopy (AFM). By raster scanning in the x,y plane the work function of the sample can be locally mapped for correlation with sample features. When there is little or no magnification, this approach can be described as using a scanning Kelvin probe (SKP). These techniques are predominantly used to measure corrosion and coatings.

<span class="mw-page-title-main">Near-field scanning optical microscope</span>

Near-field scanning optical microscopy (NSOM) or scanning near-field optical microscopy (SNOM) is a microscopy technique for nanostructure investigation that breaks the far field resolution limit by exploiting the properties of evanescent waves. In SNOM, the excitation laser light is focused through an aperture with a diameter smaller than the excitation wavelength, resulting in an evanescent field on the far side of the aperture. When the sample is scanned at a small distance below the aperture, the optical resolution of transmitted or reflected light is limited only by the diameter of the aperture. In particular, lateral resolution of 6 nm and vertical resolution of 2–5 nm have been demonstrated.

<span class="mw-page-title-main">Parametric oscillator</span>

A parametric oscillator is a driven harmonic oscillator in which the oscillations are driven by varying some parameter of the system at some frequency, typically different from the natural frequency of the oscillator. A simple example of a parametric oscillator is a child pumping a playground swing by periodically standing and squatting to increase the size of the swing's oscillations. The child's motions vary the moment of inertia of the swing as a pendulum. The "pump" motions of the child must be at twice the frequency of the swing's oscillations. Examples of parameters that may be varied are the oscillator's resonance frequency and damping .

Electrostatic force microscopy (EFM) is a type of dynamic non-contact atomic force microscopy where the electrostatic force is probed.. This force arises due to the attraction or repulsion of separated charges. It is a long-range force and can be detected 100 nm or more from the sample.

In electronics, complex gain is the effect that circuitry has on the amplitude and phase of a sine wave signal. The term complex is used because mathematically this effect can be expressed as a complex number.

In microscopy, scanning joule expansion microscopy (SJEM) is a form of scanning probe microscopy heavily based on atomic force microscopy (AFM) that maps the temperature distribution along a surface. Resolutions down to 10 nm have been achieved and 1 nm resolution is theoretically possible. Thermal measurements at the nanometer scale are of both academic and industrial interest, particularly in regards to nanomaterials and modern integrated circuits.

<span class="mw-page-title-main">Photoconductive atomic force microscopy</span>

Photoconductive atomic force microscopy (PC-AFM) is a variant of atomic force microscopy that measures photoconductivity in addition to surface forces.

<span class="mw-page-title-main">Atomic force acoustic microscopy</span>

Atomic force acoustic microscopy (AFAM) is a type of scanning probe microscopy (SPM). It is a combination of acoustics and atomic force microscopy. The principal difference between AFAM and other forms of SPM is the addition of a transducer at the bottom of the sample which induces longitudinal out-of-plane vibrations in the specimen. These vibrations are sensed by a cantilever and tip called a probe. The figure shown here is the clear schematic of AFAM principle here B is the magnified version of the tip and sample placed on the transducer and tip having some optical coating generally gold coating to reflect the laser light on to the photodiode.

<span class="mw-page-title-main">Non-contact atomic force microscopy</span>

Non-contact atomic force microscopy (nc-AFM), also known as dynamic force microscopy (DFM), is a mode of atomic force microscopy, which itself is a type of scanning probe microscopy. In nc-AFM a sharp probe is moved close to the surface under study, the probe is then raster scanned across the surface, the image is then constructed from the force interactions during the scan. The probe is connected to a resonator, usually a silicon cantilever or a quartz crystal resonator. During measurements the sensor is driven so that it oscillates. The force interactions are measured either by measuring the change in amplitude of the oscillation at a constant frequency just off resonance or by measuring the change in resonant frequency directly using a feedback circuit to always drive the sensor on resonance.

Bimodal Atomic Force Microscopy is an advanced atomic force microscopy technique characterized by generating high-spatial resolution maps of material properties. Topography, deformation, elastic modulus, viscosity coefficient or magnetic field maps might be generated. Bimodal AFM is based on the simultaneous excitation and detection of two eigenmodes (resonances) of a force microscope microcantilever.

References

  1. Güthner, P.; Dransfeld, K. (1992). "Local poling of ferroelectric polymers by scanning force microscopy". Applied Physics Letters. 61 (9): 1137–1139. Bibcode:1992ApPhL..61.1137G. doi:10.1063/1.107693.
  2. Rodriguez, B.J.; Kalinin, S.V.; Shin, J.; Jesse, S.; Grichko, V.; Thundat, T.; Baddorf, A.P.; Gruverman, A. (2006). "Electromechanical imaging of biomaterials by scanning probe microscopy" (PDF). Journal of Structural Biology. 153 (2): 151–9. doi:10.1016/j.jsb.2005.10.008. PMID   16403652.
  3. Kalinin, Sergei V; Morozovska, Anna N; Chen, Long Qing; Rodriguez, Brian J (2010). "Local polarization dynamics in ferroelectric materials". Reports on Progress in Physics. 73 (5): 056502. Bibcode:2010RPPh...73e6502K. doi:10.1088/0034-4885/73/5/056502. S2CID   122479907.
  4. Rosen, C.Z., Hiremath, B.V., Newnham, R. (ed) "Piezoelectricity" American Institute of Physics, Key Papers in Physics, No 5, 227–283 (1992)
  5. Kalinin, SV; Rodriguez, BJ; Jesse, S; Shin, J; Baddorf, AP; Gupta, P; Jain, H; Williams, DB; Gruverman, A (2006). "Vector piezoresponse force microscopy". Microscopy and Microanalysis. 12 (3): 206–20. Bibcode:2006MiMic..12..206K. doi:10.1017/S1431927606060156. hdl: 10197/5514 . PMID   17481357. S2CID   17687498.
  6. Kalinin, Sergei V.; Rodriguez, B. J.; Jesse, S.; Thundat, T.; Gruverman, A. (2005). "Electromechanical imaging of biological systems with sub-10 nm resolution". Applied Physics Letters. 87 (5): 053901. arXiv: cond-mat/0504232 . Bibcode:2005ApPhL..87e3901K. doi:10.1063/1.2006984. S2CID   11761213.
  7. Jiang, Peng; Yan, Fei; Nasr Esfahani, Ehsan; Xie, Shuhong; Zou, Daifeng; Liu, Xiaoyan; Zheng, Hairong; Li, Jiangyu (2017-08-14). "Electromechanical Coupling of Murine Lung Tissues Probed by Piezoresponse Force Microscopy". ACS Biomaterials Science & Engineering. 3 (8): 1827–1835. doi:10.1021/acsbiomaterials.7b00107. PMID   33429664.
  8. Minary-Jolandan, Majid; Yu, Min-Feng (2009). "Uncovering Nanoscale Electromechanical Heterogeneity in the Subfibrillar Structure of Collagen Fibrils Responsible for the Piezoelectricity of Bone" (PDF). ACS Nano. 3 (7): 1859–63. doi:10.1021/nn900472n. PMID   19505115. Archived from the original (PDF) on 2012-12-19.
  9. Gruverman, A.; Rodriguez, B. J.; Dehoff, C.; Waldrep, J. D.; Kingon, A. I.; Nemanich, R. J.; Cross, J. S. (2005). "Direct studies of domain switching dynamics in thin film ferroelectric capacitors". Applied Physics Letters. 87 (8): 082902. Bibcode:2005ApPhL..87h2902G. doi:10.1063/1.2010605. hdl:10197/5333. S2CID   35719760.
  10. Harnagea, C.; Alexe, M.; Hesse, D.; Pignolet, A. (2003). "Contact resonances in voltage-modulated force microscopy" (PDF). Applied Physics Letters. 83 (2): 338. Bibcode:2003ApPhL..83..338H. doi:10.1063/1.1592307.
  11. Rodriguez, Brian J; Jesse, Stephen; Baddorf, Arthur P; Zhao, T; Chu, Y H; Ramesh, R; Eliseev, Eugene A; Morozovska, Anna N; Kalinin, Sergei V (2007). "Spatially resolved mapping of ferroelectric switching behavior in self-assembled multiferroic nanostructures: strain, size, and interface effects". Nanotechnology. 18 (40): 405701. Bibcode:2007Nanot..18N5701R. doi:10.1088/0957-4484/18/40/405701. S2CID   137039357.
  12. Jesse, Stephen (2017-01-02). "Band Excitation for Scanning Probe Microscopy". OSTI   1340998.{{cite journal}}: Cite journal requires |journal= (help)
  13. US 9097738,Jesse, Stephen&Kalinin, Sergei V.,"Band excitation method applicable to scanning probe microscopy",issued August 4, 2015
  14. Lei, Xiaojun; Li, Dongbo; Shao, Rui; Bonnell, Dawn A. (1 March 2005). "In situ deposition/positioning of magnetic nanoparticles with ferroelectric nanolithography". Journal of Materials Research. 20 (3): 712–718. Bibcode:2005JMatR..20..712L. doi:10.1557/JMR.2005.0093.