X-ray absorption near edge structure

Last updated

X-ray absorption near edge structure (XANES), also known as near edge X-ray absorption fine structure (NEXAFS), is a type of absorption spectroscopy that indicates the features in the X-ray absorption spectra (XAS) of condensed matter due to the photoabsorption cross section for electronic transitions from an atomic core level to final states in the energy region of 50–100 eV above the selected atomic core level ionization energy, where the wavelength of the photoelectron is larger than the interatomic distance between the absorbing atom and its first neighbour atoms.

Contents

Terminology

Both XANES and NEXAFS are acceptable terms for the same technique. XANES name was invented in 1980 by Antonio Bianconi to indicate strong absorption peaks in X-ray absorption spectra in condensed matter due to multiple scattering resonances above the ionization energy. [1] The name NEXAFS was introduced in 1983 by Jo Stohr and is synonymous with XANES, but is generally used when applied to surface and molecular science.

Theory

The fundamental processes which contribute to XANES spectra: 1) photoabsorption of an x-ray into a core level followed by photoelectron emission, followed by either 2) (left) filling of the core hole by an electron in another level, accompanied by fluorescence; or (right) filling of the core hole by an electron in another level followed by emission of an Auger electron. Photoabsorb modified.png
The fundamental processes which contribute to XANES spectra: 1) photoabsorption of an x-ray into a core level followed by photoelectron emission, followed by either 2) (left) filling of the core hole by an electron in another level, accompanied by fluorescence; or (right) filling of the core hole by an electron in another level followed by emission of an Auger electron.

The fundamental phenomenon underlying XANES is the absorption of an x-ray photon by condensed matter with the formation of many body excited states characterized by a core hole in a selected atomic core level (refer to the first Figure). In the single-particle theory approximation, the system is separated into one electron in the core levels of the selected atomic species of the system and N-1 passive electrons. In this approximation the final state is described by a core hole in the atomic core level and an excited photoelectron. The final state has a very short life time because of the short life-time of the core hole and the short mean free path of the excited photoelectron with kinetic energy in the range around 20-50 eV. The core hole is filled either via an Auger process or by capture of an electron from another shell followed by emission of a fluorescent photon. The difference between NEXAFS and traditional photoemission experiments is that in photoemission, the initial photoelectron itself is measured, while in NEXAFS the fluorescent photon or Auger electron or an inelastically scattered photoelectron may also be measured. The distinction sounds trivial but is actually significant: in photoemission the final state of the emitted electron captured in the detector must be an extended, free-electron state. By contrast, in NEXAFS the final state of the photoelectron may be a bound state such as an exciton since the photoelectron itself need not be detected. The effect of measuring fluorescent photons, Auger electrons, and directly emitted electrons is to sum over all possible final states of the photoelectrons, meaning that what NEXAFS measures is the total joint density of states of the initial core level with all final states, consistent with conservation rules. The distinction is critical because in spectroscopy final states are more susceptible to many-body effects than initial states, meaning that NEXAFS spectra are more easily calculable than photoemission spectra. Due to the summation over final states, various sum rules are helpful in the interpretation of NEXAFS spectra. When the x-ray photon energy resonantly connects a core level with a narrow final state in a solid, such as an exciton, readily identifiable characteristic peaks will appear in the spectrum. These narrow characteristic spectral peaks give the NEXAFS technique a lot of its analytical power as illustrated by the B 1s π* exciton shown in the second Figure.

Synchrotron radiation has a natural polarization that can be utilized to great advantage in NEXAFS studies. The commonly studied molecular adsorbates have sigma and pi bonds that may have a particular orientation on a surface. The angle dependence of the x-ray absorption tracks the orientation of resonant bonds due to dipole selection rules.

Experimental considerations

Normal-incidence boron 1s x-ray absorption spectra for two types of BN powder. The cubic phase shows only s-bonding while the hexagonal phase shows both p and s bonding. BnB.png
Normal-incidence boron 1s x-ray absorption spectra for two types of BN powder. The cubic phase shows only σ-bonding while the hexagonal phase shows both π and σ bonding.

Soft x-ray absorption spectra are usually measured either through the fluorescent yield, in which emitted photons are monitored, or total electron yield, in which the sample is connected to ground through an ammeter and the neutralization current is monitored. Because NEXAFS measurements require an intense tunable source of soft x-rays, they are performed at synchrotrons. Because soft x-rays are absorbed by air, the synchrotron radiation travels from the ring in an evacuated beam-line to the end-station where the specimen to be studied is mounted. Specialized beam-lines intended for NEXAFS studies often have additional capabilities such as heating a sample or exposing it to a dose of reactive gas.

Energy range

Edge energy range

In the absorption edge region of metals, the photoelectron is excited to the first unoccupied level above the Fermi level. Therefore, its mean free path in a pure single crystal at zero temperature is as large as infinite, and it remains very large, increasing the energy of the final state up to about 5 eV above the Fermi level. Beyond the role of the unoccupied density of states and matrix elements in single electron excitations, many-body effects appear as an "infrared singularity" at the absorption threshold in metals.

In the absorption edge region of insulators the photoelectron is excited to the first unoccupied level above the chemical potential but the unscreened core hole forms a localized bound state called core exciton.

EXAFS energy range

Pictorial view of photoelectron scattering processes in the single-scattering regime, EXAFS (this assumes the single scattering approximation... multiple scattering can be considered with EXAFS), and in the multiple scattering regime, XANES. In EXAFS the photoelectron is scattered only by a single neighbour atom, in XANES all the scattering pathways, classified according to the number of scattering event (3), (4), (5) etc. contribute to the absorption cross section. Single multiple scattering.jpg
Pictorial view of photoelectron scattering processes in the single-scattering regime, EXAFS (this assumes the single scattering approximation... multiple scattering can be considered with EXAFS), and in the multiple scattering regime, XANES. In EXAFS the photoelectron is scattered only by a single neighbour atom, in XANES all the scattering pathways, classified according to the number of scattering event (3), (4), (5) etc. contribute to the absorption cross section.

The fine structure in the x-ray absorption spectra in the high energy range extending from about 150 eV beyond the ionization potential is a powerful tool to determine the atomic pair distribution (i.e. interatomic distances) with a time scale of about 10−15 s. In fact the final state of the excited photoelectron in the high kinetic energy range (150-2000 eV ) is determined only by single backscattering events due to the low amplitude photoelectron scattering.

NEXAFS energy range

In the NEXAFS region, starting about 5 eV beyond the absorption threshold, because of the low kinetic energy range (5-150 eV) the photoelectron backscattering amplitude by neighbor atoms is very large so that multiple scattering events become dominant in the NEXAFS spectra.

The different energy range between NEXAFS and EXAFS can be also explained in a very simple manner by the comparison between the photoelectron wavelength and the interatomic distance of the photoabsorber-backscatterer pair. The photoelectron kinetic energy is connected with the wavelength by the following relation:

which means that for high energy the wavelength is shorter than interatomic distances and hence the EXAFS region corresponds to a single scattering regime; while for lower E, is larger than interatomic distances and the XANES region is associated with a multiple scattering regime.

Final states

The absorption peaks of NEXAFS spectra are determined by multiple scattering resonances of the photoelectron excited at the atomic absorption site and scattered by neighbor atoms. The local character of the final states is determined by the short photoelectron mean free path, that is strongly reduced (down to about 0.3 nm at 50 eV) in this energy range because of inelastic scattering of the photoelectron by electron-hole excitations (excitons) and collective electronic oscillations of the valence electrons called plasmons.

Applications

The great power of NEXAFS derives from its elemental specificity. Because the various elements have different core level energies, NEXAFS permits extraction of the signal from a surface monolayer or even a single buried layer in the presence of a huge background signal. Buried layers are very important in engineering applications, such as magnetic recording media buried beneath a surface lubricant or dopants below an electrode in an integrated circuit. Because NEXAFS can also determine the chemical state of elements which are present in bulk in minute quantities, it has found widespread use in environmental chemistry and geochemistry. The ability of NEXAFS to study buried atoms is due to its integration over all final states including inelastically scattered electrons, as opposed to photoemission and Auger spectroscopy, which study atoms only with a layer or two of the surface.

Much chemical information can be extracted from the NEXAFS region: formal valence (very difficult to experimentally determine in a nondestructive way); coordination environment (e.g., octahedral, tetrahedral coordination) and subtle geometrical distortions of it.

Transitions to bound vacant states just above the Fermi level can be seen. Thus NEXAFS spectra can be used as a probe of the unoccupied band structure of a material.

The near-edge structure is characteristic of an environment and valence state hence one of its more common uses is in fingerprinting: if you have a mixture of sites/compounds in a sample you can fit the measured spectra with a linear combinations of NEXAFS spectra of known species and determine the proportion of each site/compound in the sample. One example of such a use is the determination of the oxidation state of the plutonium in the soil at Rocky Flats.

The XANES experiments done on plutonium in soil, concrete and standards of the different oxidation states. PuXANES.jpg
The XANES experiments done on plutonium in soil, concrete and standards of the different oxidation states.

History

The acronym XANES was first used in 1980 during interpretation of multiple scattering resonances spectra measured at the Stanford Synchrotron Radiation Laboratory (SSRL) by A. Bianconi. In 1982 the first paper on the application of XANES for determination of local structural geometrical distortions using multiple scattering theory was published by A. Bianconi, P. J. Durham and J. B. Pendry. In 1983 the first NEXAFS paper examining molecules adsorbed on surfaces appeared. The first XAFS paper, describing the intermediate region between EXAFS and XANES, appeared in 1987.

Software for NEXAFS analysis

Related Research Articles

<span class="mw-page-title-main">Photoelectric effect</span> Emission of electrons when light hits a material

The photoelectric effect is the emission of electrons when electromagnetic radiation, such as light, hits a material. Electrons emitted in this manner are called photoelectrons. The phenomenon is studied in condensed matter physics, and solid state and quantum chemistry to draw inferences about the properties of atoms, molecules and solids. The effect has found use in electronic devices specialized for light detection and precisely timed electron emission.

<span class="mw-page-title-main">X-ray photoelectron spectroscopy</span> Spectroscopic technique

X-ray photoelectron spectroscopy (XPS) is a surface-sensitive quantitative spectroscopic technique based on the photoelectric effect that can identify the elements that exist within a material or are covering its surface, as well as their chemical state, and the overall electronic structure and density of the electronic states in the material. XPS is a powerful measurement technique because it not only shows what elements are present, but also what other elements they are bonded to. The technique can be used in line profiling of the elemental composition across the surface, or in depth profiling when paired with ion-beam etching. It is often applied to study chemical processes in the materials in their as-received state or after cleavage, scraping, exposure to heat, reactive gasses or solutions, ultraviolet light, or during ion implantation.

<span class="mw-page-title-main">Synchrotron light source</span>

A synchrotron light source is a source of electromagnetic radiation (EM) usually produced by a storage ring, for scientific and technical purposes. First observed in synchrotrons, synchrotron light is now produced by storage rings and other specialized particle accelerators, typically accelerating electrons. Once the high-energy electron beam has been generated, it is directed into auxiliary components such as bending magnets and insertion devices in storage rings and free electron lasers. These supply the strong magnetic fields perpendicular to the beam which are needed to convert high energy electrons into photons.

<span class="mw-page-title-main">Photoemission spectroscopy</span> Examining a substance by measuring electrons emitted in the photoelectric effect

Photoemission spectroscopy (PES), also known as photoelectron spectroscopy, refers to energy measurement of electrons emitted from solids, gases or liquids by the photoelectric effect, in order to determine the binding energies of electrons in the substance. The term refers to various techniques, depending on whether the ionization energy is provided by X-ray, XUV or UV photons. Regardless of the incident photon beam, however, all photoelectron spectroscopy revolves around the general theme of surface analysis by measuring the ejected electrons.

Photoemission electron microscopy is a type of electron microscopy that utilizes local variations in electron emission to generate image contrast. The excitation is usually produced by ultraviolet light, synchrotron radiation or X-ray sources. PEEM measures the coefficient indirectly by collecting the emitted secondary electrons generated in the electron cascade that follows the creation of the primary core hole in the absorption process. PEEM is a surface sensitive technique because the emitted electrons originate from a shallow layer. In physics, this technique is referred to as PEEM, which goes together naturally with low-energy electron diffraction (LEED), and low-energy electron microscopy (LEEM). In biology, it is called photoelectron microscopy (PEM), which fits with photoelectron spectroscopy (PES), transmission electron microscopy (TEM), and scanning electron microscopy (SEM).

<span class="mw-page-title-main">X-ray absorption fine structure</span> Specific structure observed in X-ray absorption spectroscopy

X-ray absorption fine structure (XAFS) is a specific structure observed in X-ray absorption spectroscopy (XAS). By analyzing the XAFS, information can be acquired on the local structure and on the unoccupied local electronic states.

<span class="mw-page-title-main">Extended X-ray absorption fine structure</span> Measurement of X-ray absorption of a material as a function of energy

Extended X-ray absorption fine structure (EXAFS), along with X-ray absorption near edge structure (XANES), is a subset of X-ray absorption spectroscopy (XAS). Like other absorption spectroscopies, XAS techniques follow Beer's law. The X-ray absorption coefficient of a material as a function of energy is obtained using X-rays of a narrow energy resolution are directed at a sample and the incident and transmitted x-ray intensity is recorded as the incident x-ray energy is incremented.

<span class="mw-page-title-main">X-ray absorption spectroscopy</span> Panel of different types of X-ray absorption spectroscopy requiring a synchrotron radiation facility

X-ray absorption spectroscopy (XAS) is a widely used technique for determining the local geometric and/or electronic structure of matter. The experiment is usually performed at synchrotron radiation facilities, which provide intense and tunable X-ray beams. Samples can be in the gas phase, solutions, or solids.

<span class="mw-page-title-main">Extreme ultraviolet</span> Ultraviolet light with a wavelength of 10–121nm

Extreme ultraviolet radiation or high-energy ultraviolet radiation is electromagnetic radiation in the part of the electromagnetic spectrum spanning wavelengths from 124 nm down to 10 nm, and therefore having photons with energies from 10 eV up to 124 eV. EUV is naturally generated by the solar corona and artificially by plasma, high harmonic generation sources and synchrotron light sources. Since UVC extends to 100 nm, there is some overlap in the terms.

Rydberg ionization spectroscopy is a spectroscopy technique in which multiple photons are absorbed by an atom causing the removal of an electron to form an ion.

Electron spectroscopy refers to a group formed by techniques based on the analysis of the energies of emitted electrons such as photoelectrons and Auger electrons. This group includes X-ray photoelectron spectroscopy (XPS), which also known as Electron Spectroscopy for Chemical Analysis (ESCA), Electron energy loss spectroscopy (EELS), Ultraviolet photoelectron spectroscopy (UPS), and Auger electron spectroscopy (AES). These analytical techniques are used to identify and determine the elements and their electronic structures from the surface of a test sample. Samples can be solids, gases or liquids.

Ultraviolet photoelectron spectroscopy (UPS) refers to the measurement of kinetic energy spectra of photoelectrons emitted by molecules which have absorbed ultraviolet photons, in order to determine molecular orbital energies in the valence region.

In X-ray absorption spectroscopy, the K-edge is a sudden increase in x-ray absorption occurring when the energy of the X-rays is just above the binding energy of the innermost electron shell of the atoms interacting with the photons. The term is based on X-ray notation, where the innermost electron shell is known as the K-shell. Physically, this sudden increase in attenuation is caused by the photoelectric absorption of the photons. For this interaction to occur, the photons must have more energy than the binding energy of the K-shell electrons (K-edge). A photon having an energy just above the binding energy of the electron is therefore more likely to be absorbed than a photon having an energy just below this binding energy or significantly above it.

X-ray Raman scattering (XRS) is non-resonant inelastic scattering of X-rays from core electrons. It is analogous to vibrational Raman scattering, which is a widely used tool in optical spectroscopy, with the difference being that the wavelengths of the exciting photons fall in the X-ray regime and the corresponding excitations are from deep core electrons.

Laser-based angle-resolved photoemission spectroscopy is a form of angle-resolved photoemission spectroscopy that uses a laser as the light source. Photoemission spectroscopy is a powerful and sensitive experimental technique to study surface physics. It is based on the photoelectric effect originally observed by Heinrich Hertz in 1887 and later explained by Albert Einstein in 1905 that when a material is shone by light, the electrons can absorb photons and escape from the material with the kinetic energy: , where is the incident photon energy, the work function of the material. Since the kinetic energy of ejected electrons are highly associated with the internal electronic structure, by analyzing the photoelectron spectroscopy one can realize the fundamental physical and chemical properties of the material, such as the type and arrangement of local bonding, electronic structure and chemical composition.

Surface-extended X-ray absorption fine structure (SEXAFS) is the surface-sensitive equivalent of the EXAFS technique. This technique involves the illumination of the sample by high-intensity X-ray beams from a synchrotron and monitoring their photoabsorption by detecting in the intensity of Auger electrons as a function of the incident photon energy. Surface sensitivity is achieved by the interpretation of data depending on the intensity of the Auger electrons instead of looking at the relative absorption of the X-rays as in the parent method, EXAFS.

John J. Rehr is an American theoretical physicist, professor emeritus of physics at the University of Washington in Seattle. He has worked in the field of theoreticalX-ray and electron-spectroscopies.

<span class="mw-page-title-main">Two-photon photoelectron spectroscopy</span>

Time-resolved two-photon photoelectron (2PPE) spectroscopy is a time-resolved spectroscopy technique which is used to study electronic structure and electronic excitations at surfaces. The technique utilizes femtosecond to picosecond laser pulses in order to first photoexcite an electron. After a time delay, the excited electron is photoemitted into a free electron state by a second pulse. The kinetic energy and the emission angle of the photoelectron are measured in an electron energy analyzer. To facilitate investigations on the population and relaxation pathways of the excitation, this measurement is performed at different time delays.

<span class="mw-page-title-main">X-ray emission spectroscopy</span>

X-ray emission spectroscopy (XES) is a form of X-ray spectroscopy in which the X-ray line spectra are measured with a spectral resolution sufficient to analyze the impact of the chemical environment on the X-ray line energy and on branching ratios. This is done by exciting electrons out of their shell and then watching the emitted photons of the recombinating electrons.

References

  1. Bianconi, Antonio (1980). "Surface X-ray absorption spectroscopy: Surface EXAFS and surface XANES". Applications of Surface Science. 6 (3–4): 392–418. doi:10.1016/0378-5963(80)90024-0.

Bibliography