Ionization chamber

Last updated

The ionization chamber is the simplest type of gaseous ionisation detector, and is widely used for the detection and measurement of many types of ionizing radiation, including X-rays, gamma rays, alpha particles and beta particles. Conventionally, the term "ionization chamber" refers exclusively to those detectors which collect all the charges created by direct ionization within the gas through the application of an electric field. [1] It uses the discrete charges created by each interaction between the incident radiation and the gas to produce an output in the form of a small direct current. This means individual ionising events cannot be measured, so the energy of different types of radiation cannot be differentiated, but it gives a very good measurement of overall ionising effect.

Contents

It has a good uniform response to radiation over a wide range of energies and is the preferred means of measuring high levels of gamma radiation, such as in a radiation hot cell as they can tolerate prolonged periods in high radiation fields without degradation. They are widely used in the nuclear power industry, research labs, fire detection, radiation protection, and environmental monitoring.

Principle of operation

Schematic diagram of a parallel plate ion chamber, showing creation of ion pairs, and drift of ions due to electric field. Electrons typically drift 1000 times faster than positive ions due to their smaller mass. Ion chamber operation.gif
Schematic diagram of a parallel plate ion chamber, showing creation of ion pairs, and drift of ions due to electric field. Electrons typically drift 1000 times faster than positive ions due to their smaller mass.
Plot of ion current against voltage for a conceptual wire cylinder gaseous radiation detector. Ion chambers use the lowest voltage plateau. Detector regions.gif
Plot of ion current against voltage for a conceptual wire cylinder gaseous radiation detector. Ion chambers use the lowest voltage plateau.

A gas ionization chamber measures the charge from the number of ion pairs created within a gas caused by incident radiation. It consists of a gas-filled chamber with two electrodes; known as anode and cathode. The electrodes may be in the form of parallel plates (Parallel Plate Ionization Chambers: PPIC), or a cylinder with a coaxially located internal anode wire.

A voltage potential is applied between the electrodes to create an electric field in the fill gas. When gas atoms or molecules between the electrodes are ionized by incident ionizing radiation, ion-pairs are created and the resultant positive ions and dissociated electrons move to the electrodes of the opposite polarity under the influence of the electric field. This generates an ionization current which is measured by an electrometer circuit in the region of femtoamperes to picoamperes, depending on the chamber design, and is proportional to the radiation dose. [1]

The electric field has to be sufficiently strong to prevent the recombination of ion pairs which would diminish the ion current, and build-up of positive ions is prevented by their recombination with electrons when they reach the cathode. This mode of operation is referred to as "current" mode, meaning that the output signal is a continuous current, and not a pulse output as in the cases of the Geiger–Müller tube or the proportional counter. [1]

Referring to the accompanying ion-pair collection graph, it can be seen that in the ion chamber operating region the charge of a collected ion pair is effectively constant over a range of applied voltage, as due to its relatively low electric field strength the ion chamber does not have any multiplication effect. This is in distinction to the Geiger–Müller tube or the proportional counter whereby secondary electrons, and ultimately multiple avalanches, greatly amplify the original ionisation charges to produce measurable pulses. [1]

Chamber types and construction

The following chamber types are commonly used.

Free-air chamber

This is a chamber freely open to atmosphere, where the fill gas is ambient air. The domestic smoke detector is a good example of this, where a natural flow of air through the chamber is necessary so that smoke particles can be detected by the change in ion current. Other examples are applications where the ions are created outside the chamber but are carried in by a forced flow of air or gas.

Vented chamber

These chambers are normally cylindrical and operate at atmospheric pressure, but to prevent ingress of moisture a filter containing a desiccant is installed in the vent line. [2] This is to stop moisture building up in the interior of the chamber, which would otherwise be introduced by the "pump" effect of changing atmospheric air pressure. These chambers have a cylindrical body made of aluminium or plastic a few millimetres thick. The material is selected to have an atomic number similar to that of air so that the wall is said to be "air equivalent" over a range of radiation beam energies. [1] [3] [4] This has the effect of ensuring the gas in the chamber is acting as though it were a portion of an infinitely large gas volume, and increases the accuracy by reducing interactions of gamma with the wall material. The higher the atomic number of the wall material, the greater the chance of interaction. The wall thickness is a trade-off between maintaining the air effect with a thicker wall, and increasing sensitivity by using a thinner wall. These chambers often have an end window made of material thin enough, such as mylar, so that beta particles can enter the gas volume. Gamma radiation enters both through the end window and the side walls. For hand-held instruments the wall thickness is made as uniform as possible to reduce photon directionality though any beta window response is obviously highly directional. Vented chambers are susceptible to small changes in efficiency with air pressure [2] and correction factors can be applied for very accurate measurement applications.

Sealed low-pressure chamber

These are similar in construction to the vented chamber, but are sealed and operate at or around atmospheric pressure. These chambers also have the advantage of not requiring a vent and desiccant. To improve detection efficiency, they are filled with a noble gas because the highly electronegative oxygen in air easily captures free electrons, forming negative ions. The strength of the beta window limits the differential pressure from atmospheric pressure that can be tolerated, and common materials are stainless steel or titanium with a typical thickness of 25 μm. [5]

High-pressure chamber

Two high pressure cylindrical ion chambers in an enclosure. Ionizacni komory v dozimetru.jpg
Two high pressure cylindrical ion chambers in an enclosure.

The efficiency of the chamber can be further increased by the use of a high-pressure gas. Typically a pressure of 8-10 atmospheres can be used, and various noble gases are employed. The higher pressure results in a greater gas density and thereby a greater chance of collision with the fill gas and ion-pair creation by incident radiation. Because of the increased wall thickness required to withstand this high pressure, only gamma radiation can be detected. These detectors are used in survey meters and for environmental monitoring. [2]

Chamber geometry

Most commonly used for radiation therapy measurements is a cylindrical or "thimble" chamber. The active volume is housed within a thimble shaped cavity with an inner conductive surface (cathode) and a central anode. A bias voltage applied across the cavity collects ions and produces a current which can be measured with an electrometer.

Parallel-plate chambers (PPIC) are shaped like a small disc, with circular collecting electrodes separated by a small gap, typically 2mm or less. The upper disc is extremely thin, allowing for much more accurate near-surface dose measurements than are possible with a cylindrical chamber. Monitor chambers are typically PPICs which are used to continuously measure such as a radiation beam's intensity. For example, within the head of linear accelerators used for radiotherapy. Multi-cavity ionization chambers can measure the intensity of the radiation beam in several different regions, providing beam symmetry and flatness information.

Research and calibration chambers

Ionization chamber made by Pierre Curie, c 1895-1900 Ionisation chamber made by Pierre Curie, c 1895-1900. (9660571297).jpg
Ionization chamber made by Pierre Curie, c 1895-1900

Early versions of the ion chamber were used by Marie and Pierre Curie in their original work in isolating radioactive materials. Since then the ion chamber has been a widely used tool in the laboratory for research and calibration purposes.

Historical chambers

Condenser chamber

The condenser chamber has a secondary cavity within the stem which acts as a capacitor. When this capacitor is fully charged, any ionization within the thimble counteracts this charge, and the change in charge can be measured. They are only practical for beams with energy of 2 MeV or less, and high stem leakage makes them unsuited to precise dosimetry.

Extrapolation chamber

Similar in design to a parallel plate chamber, the upper plate of an extrapolation chamber can be lowered using micrometer screws. Measurements can be taken with different plate spacing and extrapolated to a plate spacing of zero, i.e. the dose without the chamber.

Instrument types

Hand-held

Hand-held integral ion chamber survey meter in use RTG radiation measurement.jpg
Hand-held integral ion chamber survey meter in use
View of sliding beta shield on integral hand held instrument CivilDefenseVictoreen720 underside.jpg
View of sliding beta shield on integral hand held instrument

Ion chambers are widely used in hand held radiation survey meters to measure beta and gamma radiation. They are particularly preferred for high dose rate measurements and for gamma radiation they give good accuracy for energies above 50-100 keV. [1]

There are two basic configurations; the "integral" unit with the chamber and electronics in the same case, and the "two-piece" instrument which has a separate ion chamber probe attached to the electronics module by a flexible cable.

The chamber of the integral instrument is generally at the front of the case facing downwards, and for beta/gamma instruments there is a window in the bottom of the casing. This usually has a sliding shield which enables discrimination between gamma and beta radiation. The operator closes the shield to exclude beta, and can thereby calculate the rate of each radiation type.

Some hand held instruments generate audible clicks similar to that produced by a G-M counter to assist operators, who use the audio feedback in radiation survey and contamination checks. As the ion chamber works in current mode, not pulse mode, this is synthesised from the radiation rate.

Installed

For industrial process measurements and interlocks with sustained high radiation levels, the ion chamber is the preferred detector. In these applications only the chamber is situated in the measurement area, and the electronics are remotely situated to protect them from radiation and connected by a cable. Installed instruments can be used for measuring ambient gamma for personnel protection and normally sound an alarm above a preset rate, though the Geiger–Müller tube instrument is generally preferred where high levels of accuracy are not required.

General precautions in use

Moisture is the main problem that affects the accuracy of ion chambers. The chamber's internal volume must be kept completely dry, and the vented type uses a desiccant to help with this. [2] Because of the very low currents generated, any stray leakage current must be kept to a minimum in order to preserve accuracy. Invisible hygroscopic moisture on the surface of cable dielectrics and connectors can be sufficient to cause a leakage current which will swamp any radiation-induced ion current. This requires scrupulous cleaning of the chamber, its terminations and cables, and subsequent drying in an oven. "Guard rings" are generally used as a design feature on higher voltage tubes to reduce leakage through or along the surface of tube connection insulators, which can require a resistance in the order of 1013 Ω. [6]

For industrial applications with remote electronics, the ion chamber is housed in a separate enclosure which provides mechanical protection and contains a desiccant to remove moisture which could affect the termination resistance.

In installations where the chamber is a long distance from the measuring electronics, readings can be affected by external electromagnetic radiation acting on the cable. To overcome this a local converter module is often used to translate the very low ion chamber currents to a pulse train or data signal related to the incident radiation. These are immune to electromagnetic effects.

Applications

Nuclear industry

Ionization chambers are widely used in the nuclear industry as they provide an output that is proportional to radiation dose. They find wide use in situations where a constant high dose rate is being measured as they have a greater operating lifetime than standard Geiger–Müller tubes, which suffer from gas break down and are generally limited to a life of about 1011 count events. [1] Additionally, the Geiger–Müller tube cannot operate above about 104 counts per second, due to dead-time effects, whereas there is no similar limitation on the ion chamber.

Smoke detectors

The ionization chamber has found wide and beneficial use in smoke detectors. In an ionisation type smoke detector, ambient air is allowed to freely enter the ionization chamber. The chamber contains a small amount of americium-241, which is an emitter of alpha particles which produce a constant ion current. If smoke enters the detector, it disrupts this current because ions strike smoke particles and are neutralized. This drop in current triggers the alarm. The detector also has a reference chamber which is sealed but is ionized in the same way. Comparison of the ion currents in the two chambers allows compensation for changes due to air pressure, temperature, or the ageing of the source. [7]

Medical radiation measurement

Diagram of a nuclear medicine dose calibrator or radionuclide calibrator that uses a "well-type" ionization chamber. The dipper is used to give a reproducible source position. The radioactive substance in this example is liquid. Dose Calibrator Diagram.png
Diagram of a nuclear medicine dose calibrator or radionuclide calibrator that uses a "well-type" ionization chamber. The dipper is used to give a reproducible source position. The radioactive substance in this example is liquid.

In medical physics and radiotherapy, ionization chambers are used to ensure that the dose delivered from a therapy unit [8] or radiopharmaceutical is what is intended. The devices used for radiotherapy are called "reference dosimeters", while those used for radiopharmaceuticals are called radioisotope dose calibrators - an inexact name for radionuclide radioactivity calibrators, which are used for measurement of radioactivity but not absorbed dose. [9] A chamber will have a calibration factor established by a national standards laboratory such as ARPANSA in Australia or the NPL in the UK, or will have a factor determined by comparison against a transfer standard chamber traceable to national standards at the user's site. [4] [10]

Guidance on application use

In the United Kingdom the HSE has issued a user guide on selecting the correct radiation measurement instrument for the particular application concerned. [11] This covers all radiation instrument technologies, and is a useful comparative guide to the use of ion chamber instruments.

See also

Notes

    Related Research Articles

    <span class="mw-page-title-main">Geiger counter</span> Instrument used for measuring ionizing radiation

    A Geiger counter is an electronic instrument used for detecting and measuring ionizing radiation. It is widely used in applications such as radiation dosimetry, radiological protection, experimental physics and the nuclear industry.

    <span class="mw-page-title-main">Beta particle</span> Ionizing radiation

    A beta particle, also called beta ray or beta radiation, is a high-energy, high-speed electron or positron emitted by the radioactive decay of an atomic nucleus during the process of beta decay. There are two forms of beta decay, β decay and β+ decay, which produce electrons and positrons respectively.

    <span class="mw-page-title-main">Dosimeter</span> Device measuring ionizing radiation exposure

    A radiation dosimeter is a device that measures dose uptake of external ionizing radiation. It is worn by the person being monitored when used as a personal dosimeter, and is a record of the radiation dose received. Modern electronic personal dosimeters can give a continuous readout of cumulative dose and current dose rate, and can warn the wearer with an audible alarm when a specified dose rate or a cumulative dose is exceeded. Other dosimeters, such as thermoluminescent or film types, require processing after use to reveal the cumulative dose received, and cannot give a current indication of dose while being worn.

    <span class="mw-page-title-main">Geiger–Müller tube</span> Part of a Geiger counter

    The Geiger–Müller tube or G–M tube is the sensing element of the Geiger counter instrument used for the detection of ionizing radiation. It is named after Hans Geiger, who invented the principle in 1908, and Walther Müller, who collaborated with Geiger in developing the technique further in 1928 to produce a practical tube that could detect a number of different radiation types.

    Ionizing radiation (US) (or ionising radiation [UK]), including nuclear radiation, consists of subatomic particles or electromagnetic waves that have sufficient energy to ionize atoms or molecules by detaching electrons from them. Some particles can travel up to 99% of the speed of light, and the electromagnetic waves are on the high-energy portion of the electromagnetic spectrum.

    The gray is the unit of ionizing radiation dose in the International System of Units (SI), defined as the absorption of one joule of radiation energy per kilogram of matter.

    <span class="mw-page-title-main">Dosimetry</span> Measurement of absorbed ionizing radiation

    Radiation dosimetry in the fields of health physics and radiation protection is the measurement, calculation and assessment of the ionizing radiation dose absorbed by an object, usually the human body. This applies both internally, due to ingested or inhaled radioactive substances, or externally due to irradiation by sources of radiation.

    Radiation protection, also known as radiological protection, is defined by the International Atomic Energy Agency (IAEA) as "The protection of people from harmful effects of exposure to ionizing radiation, and the means for achieving this". Exposure can be from a source of radiation external to the human body or due to internal irradiation caused by the ingestion of radioactive contamination.

    <span class="mw-page-title-main">Health physics</span>

    Health physics, also referred to as the science of radiation protection, is the profession devoted to protecting people and their environment from potential radiation hazards, while making it possible to enjoy the beneficial uses of radiation. Health physicists normally require a four-year bachelor’s degree and qualifying experience that demonstrates a professional knowledge of the theory and application of radiation protection principles and closely related sciences. Health physicists principally work at facilities where radionuclides or other sources of ionizing radiation are used or produced; these include research, industry, education, medical facilities, nuclear power, military, environmental protection, enforcement of government regulations, and decontamination and decommissioning—the combination of education and experience for health physicists depends on the specific field in which the health physicist is engaged.

    Absorbed dose is a dose quantity which is the measure of the energy deposited in matter by ionizing radiation per unit mass. Absorbed dose is used in the calculation of dose uptake in living tissue in both radiation protection, and radiology. It is also used to directly compare the effect of radiation on inanimate matter such as in radiation hardening.

    A semiconductor detector in ionizing radiation detection physics is a device that uses a semiconductor to measure the effect of incident charged particles or photons.

    The measurement of ionizing radiation is sometimes expressed as being a rate of counts per unit time as registered by a radiation monitoring instrument, for which counts per minute (cpm) and counts per second (cps) are commonly used quantities.

    <span class="mw-page-title-main">Proportional counter</span> Gaseous ionization detector

    The proportional counter is a type of gaseous ionization detector device used to measure particles of ionizing radiation. The key feature is its ability to measure the energy of incident radiation, by producing a detector output pulse that is proportional to the radiation energy absorbed by the detector due to an ionizing event; hence the detector's name. It is widely used where energy levels of incident radiation must be known, such as in the discrimination between alpha and beta particles, or accurate measurement of X-ray radiation dose.

    A wire chamber or multi-wire proportional chamber is a type of proportional counter that detects charged particles and photons and can give positional information on their trajectory, by tracking the trails of gaseous ionization.

    <span class="mw-page-title-main">Gaseous ionization detector</span> Radiation detector

    Gaseous ionization detectors are radiation detection instruments used in particle physics to detect the presence of ionizing particles, and in radiation protection applications to measure ionizing radiation.

    <span class="mw-page-title-main">Survey meter</span>

    Survey meters in radiation protection are hand-held ionising radiation measurement instruments used to check such as personnel, equipment and the environment for radioactive contamination and ambient radiation. The hand-held survey meter is probably the most familiar radiation measuring device owing to its wide and visible use.

    <span class="mw-page-title-main">Radiation monitoring</span> Measurement of radiation doses or contamination

    Radiation monitoring involves the measurement of radiation dose or radionuclide contamination for reasons related to the assessment or control of exposure to radiation or radioactive substances, and the interpretation of the results.

    <span class="mw-page-title-main">Quartz fiber dosimeter</span> Type of radiation dosimeter

    A quartz fiber dosimeter, sometimes called a self indicating pocket dosimeter (SIPD) or self reading pocket dosimeter (SRPD) or quartz fibre electrometer (QFE), is a type of radiation dosimeter, a pen-like device that measures the cumulative dose of ionizing radiation received by the device, usually over one work period. It is clipped to a person's clothing, normally a breast pocket for whole body exposure, to measure the user's exposure to radiation.

    <span class="mw-page-title-main">Roentgen (unit)</span> Measurement of radiation exposure

    The roentgen or röntgen is a legacy unit of measurement for the exposure of X-rays and gamma rays, and is defined as the electric charge freed by such radiation in a specified volume of air divided by the mass of that air . In 1928, it was adopted as the first international measurement quantity for ionizing radiation to be defined for radiation protection, as it was then the most easily replicated method of measuring air ionization by using ion chambers. It is named after the German physicist Wilhelm Röntgen, who discovered X-rays and was awarded the first Nobel Prize in Physics for the discovery.

    <span class="mw-page-title-main">X-ray detector</span> Instrument that can measure properties of X-rays

    X-ray detectors are devices used to measure the flux, spatial distribution, spectrum, and/or other properties of X-rays.

    References

    1. 1 2 3 4 5 6 7 8 Knoll, Glenn F (1999). Radiation detection and measurement (3rd ed.). New York: Wiley. ISBN   978-0-471-07338-3.
    2. 1 2 3 4 Steinmeyer, Paul R. (2003). "Ion Chambers: Everything You've Wanted to Know (But Were Afraid to Ask)" (PDF). RSO Magazine. 8 (5). Archived from the original (PDF) on 2012-09-15. Retrieved 2013-08-18.
    3. Seco, Joao; Clasie, Ben; Partridge, Mike (21 October 2014). "Review on the characteristics of radiation detectors for dosimetry and imaging". Physics in Medicine and Biology. 59 (20): R303–R347. Bibcode:2014PMB....59R.303S. doi: 10.1088/0031-9155/59/20/R303 . PMID   25229250.
    4. 1 2 Hill, Robin; Healy, Brendan; Holloway, Lois; Kuncic, Zdenka; Thwaites, David; Baldock, Clive (21 March 2014). "Advances in kilovoltage x-ray beam dosimetry". Physics in Medicine and Biology. 59 (6): R183–R231. Bibcode:2014PMB....59R.183H. doi:10.1088/0031-9155/59/6/R183. PMID   24584183. S2CID   18082594.
    5. LND, Inc. ion chamber specification sheets
    6. Taylor, D.; Sharpe, J. (April 1951). "Nuclear particle and radiation detectors. Part 1: Ion chambers and ion-chamber instruments". Proceedings of the IEE - Part II: Power Engineering. 98 (62): 174–190. doi:10.1049/pi-2.1951.0058.
    7. Cote, Arthur; Bugbee, Percy (1988). "Ionization smoke detectors". Principles of fire protection. Quincy, MA: National Fire Protection Association. p. 249. ISBN 0-87765-345-3.
    8. Hill, R; Mo, Z; Haque, M; Baldock, C (2009). "An evaluation of ionization chambers for the relative dosimetry of kilovoltage x-ray beams". Medical Physics. 36 (9Part1): 3971–3981. Bibcode:2009MedPh..36.3971H. doi:10.1118/1.3183820. PMID   19810470.
    9. Mo, L.; Reinhard, M.I.; Davies, J.B.; Alexiev, D.; Baldock, C. (April 2006). "Calibration of the Capintec CRC-712M dose calibrator for 18F". Applied Radiation and Isotopes. 64 (4): 485–489. doi:10.1016/j.apradiso.2005.09.006. PMID   16293417.
    10. Seco, Joao; Clasie, Ben; Partridge, Mike (Oct 2014). "Review on the characteristics of radiation detectors for dosimetry and imaging". Physics in Medicine and Biology. 59 (20): R303–R347. Bibcode:2014PMB....59R.303S. doi:10.1088/0031-9155/59/20/R303. PMID   25229250. S2CID   4393848.
    11. "Selection, use and maintenance of portable monitoring instruments" (PDF). Health & Safety Executive. 2001. Archived from the original (PDF) on 8 October 2012. Retrieved 3 November 2012.