Oudin coil

Last updated
Oudin coil used for medical electrotherapy, 1907, and a schematic diagram of its circuit (left). Oudin coil and circuit diagram.png
Oudin coil used for medical electrotherapy, 1907, and a schematic diagram of its circuit (left).

An Oudin coil, also called an Oudin oscillator or Oudin resonator, is a resonant transformer circuit that generates very high voltage, high frequency alternating current (AC) electricity at low current levels, [1] [2] [3] used in the obsolete forms of electrotherapy around the turn of the 20th century. [4] It is very similar to the Tesla coil, with the difference being that the Oudin coil was connected as an autotransformer. [2] [5] It was invented in 1893 by French physician Paul Marie Oudin [6] as a modification of physician Jacques Arsene d'Arsonval's electrotherapy equipment [7] [8] and used in medical diathermy therapy as well as quack medicine until perhaps 1940. The high voltage output terminal of the coil was connected to an insulated handheld electrode which produced luminous brush discharges, which were applied to the patient's body to treat various medical conditions in electrotherapy. [4]

Contents

How it works

Oudin and Tesla coils are spark-excited air-core double-tuned transformer circuits that use resonance to generate very high voltages at low currents. [9] [10] They produce alternating current in the radio frequency (RF) range. The medical coils of the early 20th century produced potentials of 50,000 up to a million volts, at frequencies in the range 200 kHz to 5 MHz. [4] The primary circuit of the coil has Leyden jar capacitors (C) which in combination with the primary winding of the coil (L1) make a resonant circuit (tuned circuit). [4] In medical coils usually two capacitors were used for safety, one in each side of the primary circuit, to isolate the patient completely from the potentially lethal low frequency primary current. The primary circuit also has a spark gap (SG) that acts as a switch to excite oscillations in the primary. The primary circuit is powered by a high voltage transformer or induction coil [4] (T) at a potential of 2 - 15 kV. The transformer repeatedly charges the capacitors, which then discharge through the spark gap and the primary winding (a detailed description of the operation cycle in the Tesla coil article also applies to the Oudin coil). This cycle is repeated many times per second. During each spark, the charge moves rapidly back and forth between the capacitor plates through the primary coil, creating a damped RF oscillating current in the primary tuned circuit which induced the high voltage in the secondary.

The secondary winding (L2) is open-circuited, and connected to the output electrode of the device. In the Oudin coil, one side of the primary winding (L1) is grounded and the other side is connected to the secondary, so the primary and secondary are in series. There were two versions of the Oudin coil: [3] [11] [12]

Oudin coil circuit - early type.svg
Early Oudin coil.png
Early type Oudin coil, around 1904, with separate primary and secondary coils. In the drawing the primary (L1) is the small horizontal coil visible between the Leyden jars and the large vertical secondary resonator coil. The transformer (T) which powers the circuit is not shown.
Oudin coil circuit - later type.svg
Oudin coil Turpain.png
Later type with primary and secondary wound on the same form to make an autotransformer. The primary is the coarse winding at the bottom. In both pictures the spark gap (SG) is enclosed in a small box (S) at right to reduce noise.

Although it doesn't include a capacitor, the secondary winding is also a resonant circuit (electrical resonator); the parasitic capacitance between the ends of the secondary coil resonates with the large inductance of the secondary at a particular resonant frequency. When it is excited at this frequency by the primary, large oscillating voltages are induced in the secondary. The number of turns in the primary winding, and thus the resonant frequency of the primary, could be adjusted with a tap on the coil. When the two tuned circuits are adjusted to resonate at the same frequency, the large turns ratio of the coil, aided by the high Q of the tuned circuits, steps up the primary voltage to hundreds of thousands to millions of volts at the secondary.

The secondary is directly connected to the primary circuit, which carries lethal low frequency 50/60 Hz currents at thousands of volts from the power transformer. Since the Oudin coil was a medical device, with the secondary current applied directly to a person's body, for safety the Oudin circuit has two capacitors (C), one in each leg of the primary, to completely isolate the coil and output electrode from the supply transformer at the mains frequency. [13] Because two identical capacitors in series have half the capacitance of a single capacitor, the resonant frequency of the Oudin circuit is

Oudin coil - Treatment by the effiuvation method.jpg
Treatment of knee by effluvium method using Oudin coil, 1907. The pointed electrode with insulated handle (B), attached to top of the coil, is directed at knee. The patient holds a "ground" electrode (A) attached to the coil bottom. Note operator's other hand (D) adjusting coil.
Oudin electrotherapy outfit.jpg
Oudin electrotherapy outfit. Includes the Oudin coil (right rear), induction coil which powers it (center rear), power panel with mercury interrupter (left rear), and condensing bed (foreground)
The treatment of cancer by cytolsis. Wellcome M0015160.jpg
Treatment of cancer by cytolisis with Oudin coil (left of the sitting patient). The induction coil and interrupter which powers the Oudin coil is behind patient's head
Oudin coil Fleming.png
Another Oudin coil of early type

Use

Electrodes used for electrotherapy. All attach to the black insulating handle (45, bottom) which is connected by wire to the Oudin terminal. Electrotherapy electrodes 1919.png
Electrodes used for electrotherapy. All attach to the black insulating handle (45, bottom) which is connected by wire to the Oudin terminal.
A vacuum electrode. Vacuum electrotherapy electrode.JPG
A vacuum electrode.

The high voltage terminal of the coil was attached through a wire to various types of handheld electrode which the physician used to apply the high voltage to the patient's body. The treatment was not painful for the patient, because alternating current in the radio frequency (RF) range, above 10 kHz in frequency, does not generally cause the sensation of electric shock. The Oudin coil was a "unipolar" generator, with the lower end of the coil grounded, so sometimes only one electrode was applied to the patient and the return path for the currents was through the ground. However usually a ground wire from the bottom of the coil was used; attached to a ground electrode which the patient held. A drawback of the Oudin coil was that movement of the electrode and wire during use changed the capacitance of the top terminal of the secondary coil, and thus its resonant frequency. [14] This threw the secondary coil out of resonance with the primary, causing a reduction in voltage. So the tap point on the primary coil had to be constantly adjusted during use to keep the primary and secondary in "tune" (seen in left image above).

Many specialized types of electrodes were used to apply the current to various parts of the patient's body. These generally fell into two types. To apply brush discharges (called "effluves") to the outside of the patient's body, electrodes consisting of one or more metal points on an insulating handle were used. Care had to be taken to keep these far enough from the body to prevent a continuous arc to the skin, which could cause painful RF burns. To apply current directly to the body surface, as well as to tissues inside the patient's body through the mouth, rectum, or vagina, a vacuum tube "condensing" electrode was used. This consisted of a partially evacuated glass tube of various shapes, with an electrode sealed inside, attached to the high voltage wire. This produced a dramatic violet glow when energized. The glass envelope of the tube formed a capacitor with the patient's body through which the current had to pass, limiting it to safe values.

To apply current to the whole body, a "condensing couch" was used. This was a bed or couch with a metal back under its mattress, connected by a wire to the high voltage terminal. Metal handrest electrodes at the sides, which the patient grasped during treatment, served as the "ground" return path and were attached to the bottom of the coil. Thus the couch formed a capacitor, with the patient's body as one electrode.

History

During the 1800s, experiments in applying electric currents to the human body grew into a Victorian era medical field, part legitimate experimental medicine and part quack medicine, called electrotherapy, in which currents were applied to treat many medical conditions. The discovery of radio waves by Heinrich Hertz in 1886 and subsequent development of radio by Oliver Lodge, Guglielmo Marconi sparked interest in radio frequency currents and circuits for generating them. "High frequency" currents meant any frequency above the audio range, > 20 kHz, and the resonant coils which generated them were generically called "oscillation transformers". During the 1890s doctors began to experiment with applying these high voltage and high frequency currents to the human body (ethical standards in the medical profession were looser then and physicians could experiment on their patients). In 1890 French physician Jacques Arsene d'Arsonval founded the field of high frequency electrotherapy, performing the first experiments applying high frequency currents to the human body. He discovered that currents above 10 kHz do not cause muscle contraction or activate nerves to cause the sensation of electric shock, so that extremely high voltages could be applied to a patient without discomfort. In 1891 in America, engineer Nikola Tesla independently discovered the same thing. Three types of apparatus were used, developed by three pioneers in the field, D'Arsonval, Tesla, and Oudin, and separate bodies of clinical technique grew up around them: [3] [4]

D'Arsonval apparatus Tesla high-frequency apparatus (Rankin Kennedy, Electrical Installations, Vol V, 1903).jpg
D'Arsonval apparatus
Small medical Tesla coil. The coil is immersed in oil to provide insulation. Unlike the Oudin, the Tesla coil is bipolar. Ducretet Tesla coil.jpg
Small medical Tesla coil. The coil is immersed in oil to provide insulation. Unlike the Oudin, the Tesla coil is bipolar.

The D'Arsonval and Oudin apparatus became popular in Europe, while the Tesla-Thompson apparatus was mostly used in America. During the first decades of the 20th century there was a rivalry between these camps, and debate in the medical literature as to whether "Tesla currents" or "Oudin currents" were better for various conditions. By 1920 it was realized that the currents were very similar. Since the circuits were so similar, medical suppliers sold combination "high frequency" units that could be set up for Tesla, D'Arsonval, or Oudin therapy, often also combined with Rontgen ray (X-ray)

After Oudin combined the primary and "resonator" coil together on the same form, making them an air-core autotransformer, the only significant difference between the Tesla and Oudin apparatus was that the medical Tesla coil was "bipolar" while the Oudin coil was "unipolar", with one end grounded. As time went on the meaning of the terms changed, until (by perhaps 1920) the term Tesla coil meant a "bipolar" coil; any high voltage coil with an ungrounded balanced secondary with two output terminals, while the term Oudin coil meant a "unipolar" coil; any coil with a grounded secondary and a single output terminal. [11]

Around the 1930s vacuum tube oscillators replaced spark-excited circuits in high frequency medical equipment. The field of electrotherapy was replaced by the modern field of diathermy, and the Oudin coil became obsolete. Ironically modern day Tesla coil designs are unipolar, with a single high voltage terminal, and so are sometimes called Oudin coils. [11] [15]

See also

Related Research Articles

<span class="mw-page-title-main">Inductor</span> Passive two-terminal electrical component that stores energy in its magnetic field

An inductor, also called a coil, choke, or reactor, is a passive two-terminal electrical component that stores energy in a magnetic field when electric current flows through it. An inductor typically consists of an insulated wire wound into a coil.

<span class="mw-page-title-main">Transformer</span> Device to couple energy between circuits

A transformer is a passive component that transfers electrical energy from one electrical circuit to another circuit, or multiple circuits. A varying current in any coil of the transformer produces a varying magnetic flux in the transformer's core, which induces a varying electromotive force (EMF) across any other coils wound around the same core. Electrical energy can be transferred between separate coils without a metallic (conductive) connection between the two circuits. Faraday's law of induction, discovered in 1831, describes the induced voltage effect in any coil due to a changing magnetic flux encircled by the coil.

<span class="mw-page-title-main">Tesla coil</span> Electrical resonant transformer circuit invented by Nikola Tesla

A Tesla coil is an electrical resonant transformer circuit designed by inventor Nikola Tesla in 1891. It is used to produce high-voltage, low-current, high-frequency alternating-current electricity. Tesla experimented with a number of different configurations consisting of two, or sometimes three, coupled resonant electric circuits.

<span class="mw-page-title-main">Induction coil</span> Type of electrical transformer

An induction coil or "spark coil" is a type of electrical transformer used to produce high-voltage pulses from a low-voltage direct current (DC) supply. To create the flux changes necessary to induce voltage in the secondary coil, the direct current in the primary coil is repeatedly interrupted by a vibrating mechanical contact called an interrupter. Invented in 1836 by Nicholas Callan, with additional research by Charles Grafton Page and others, the induction coil was the first type of transformer. It was widely used in x-ray machines, spark-gap radio transmitters, arc lighting and quack medical electrotherapy devices from the 1880s to the 1920s. Today its only common use is as the ignition coils in internal combustion engines and in physics education to demonstrate induction.

Diathermy is electrically induced heat or the use of high-frequency electromagnetic currents as a form of physical therapy and in surgical procedures. The earliest observations on the reactions of high-frequency electromagnetic currents upon the human organism were made by Jacques Arsene d'Arsonval. The field was pioneered in 1907 by German physician Karl Franz Nagelschmidt, who coined the term diathermy from the Greek words dia and θέρμη therma, literally meaning "heating through".

<span class="mw-page-title-main">Plasma globe</span> Decorative electrical device

A plasma ball, plasma globe, or plasma lamp is a clear glass container filled with noble gases, usually a mixture of neon, krypton, and xenon, that has a high-voltage electrode in the center of the container. When voltage is applied, a plasma is formed within the container. Plasma filaments extend from the inner electrode to the outer glass insulator, giving the appearance of multiple constant beams of colored light. Plasma balls were popular as novelty items in the 1980s.

<span class="mw-page-title-main">Spark-gap transmitter</span> Type of radio transmitter

A spark-gap transmitter is an obsolete type of radio transmitter which generates radio waves by means of an electric spark. Spark-gap transmitters were the first type of radio transmitter, and were the main type used during the wireless telegraphy or "spark" era, the first three decades of radio, from 1887 to the end of World War I. German physicist Heinrich Hertz built the first experimental spark-gap transmitters in 1887, with which he proved the existence of radio waves and studied their properties.

<span class="mw-page-title-main">Electronic component</span> Discrete device in an electronic system

An electronic component is any basic discrete electronic device or physical entity part of an electronic system used to affect electrons or their associated fields. Electronic components are mostly industrial products, available in a singular form and are not to be confused with electrical elements, which are conceptual abstractions representing idealized electronic components and elements. A datasheet for an electronic component is a technical document that provides detailed information about the component's specifications, characteristics, and performance.

<span class="mw-page-title-main">Bifilar coil</span> Electromagnetic coil containing two closely spaced parallel windings

A bifilar coil is an electromagnetic coil that contains two closely spaced, parallel windings. In electrical engineering, the word bifilar describes wire which is made of two filaments or strands. It is commonly used to denote special types of winding wire for transformers. Wire can be purchased in bifilar form, usually as different colored enameled wire bonded together. For three strands, the term trifilar coil is used.

<span class="mw-page-title-main">Ignition coil</span> Automobile fuel ignition system component

An ignition coil is used in the ignition system of a spark-ignition engine to transform the battery voltage to the much higher voltages required to operate the spark plug(s). The spark plugs then use this burst of high-voltage electricity to ignite the air-fuel mixture.

Paul Marie Oudin (1851–1923) was a French physician and medical researcher. He was born, and later died, in Épinal. He conducted research in the Victorian era medical field of high frequency electrotherapy, the application of radio frequency electric currents to the body, and collaborated with the founder of the field, pioneering physiologist and biophysicist Dr. Jacques Arsene d'Arsonval. In 1893 he modified d'Arsonval's electrotherapy equipment by the addition of a wire coil resonator to produce higher potentials, inventing the Oudin coil. This device, very similar to a Tesla coil, could produce very high voltages from several hundred thousand to a million volts. In use, the brush discharges from a pointed electrode attached to the high voltage terminal of the coil would be played over various parts of the body to treat a variety of medical conditions. The Oudin coil was used in electrotherapy and diathermy through the 1920s.

<span class="mw-page-title-main">Blocking oscillator</span>

A blocking oscillator is a simple configuration of discrete electronic components which can produce a free-running signal, requiring only a resistor, a transformer, and one amplifying element such as a transistor or vacuum tube. The name is derived from the fact that the amplifying element is cut-off or "blocked" for most of the duty cycle, producing periodic pulses on the principle of a relaxation oscillator. The non-sinusoidal output is not suitable for use as a radio-frequency local oscillator, but it can serve as a timing generator, to power lights, LEDs, EL wire, or small neon indicators. If the output is used as an audio signal, the simple tones are also sufficient for applications such as alarms or a Morse code practice device. Some cameras use a blocking oscillator to strobe the flash prior to a shot to reduce the red-eye effect.

<span class="mw-page-title-main">AC motor</span> Electric motor driven by an AC electrical input

An AC motor is an electric motor driven by an alternating current (AC). The AC motor commonly consists of two basic parts, an outside stator having coils supplied with alternating current to produce a rotating magnetic field, and an inside rotor attached to the output shaft producing a second rotating magnetic field. The rotor magnetic field may be produced by permanent magnets, reluctance saliency, or DC or AC electrical windings.

<span class="mw-page-title-main">Transformer types</span> Overview of electrical transformer types

A variety of types of electrical transformer are made for different purposes. Despite their design differences, the various types employ the same basic principle as discovered in 1831 by Michael Faraday, and share several key functional parts.

<span class="mw-page-title-main">Resonant inductive coupling</span> Phenomenon with inductive coupling

Resonant inductive coupling or magnetic phase synchronous coupling is a phenomenon with inductive coupling in which the coupling becomes stronger when the "secondary" (load-bearing) side of the loosely coupled coil resonates. A resonant transformer of this type is often used in analog circuitry as a bandpass filter. Resonant inductive coupling is also used in wireless power systems for portable computers, phones, and vehicles.

A trigger transformer is a small, usually ferrite cored transformer used in applications requiring a high voltage pulse, typically to start ionization of a gas to allow a current to pass.

This glossary of electrical and electronics engineering is a list of definitions of terms and concepts related specifically to electrical engineering and electronics engineering. For terms related to engineering in general, see Glossary of engineering.

<span class="mw-page-title-main">History of the Tesla coil</span> An electric circuit which produces very high voltage alternating current

Nikola Tesla patented the Tesla coil circuit on April 25, 1891. and first publicly demonstrated it May 20, 1891 in his lecture "Experiments with Alternate Currents of Very High Frequency and Their Application to Methods of Artificial Illumination" before the American Institute of Electrical Engineers at Columbia College, New York. Although Tesla patented many similar circuits during this period, this was the first that contained all the elements of the Tesla coil: high voltage primary transformer, capacitor, spark gap, and air core "oscillation transformer".

Medical applications of radio frequency (RF) energy, in the form of electromagnetic waves or electrical currents, have existed for over 125 years, and now include diathermy, hyperthermy treatment of cancer, electrosurgery scalpels used to cut and cauterize in operations, and radiofrequency ablation. Magnetic resonance imaging (MRI) uses radio frequency waves to generate images of the human body.

References

  1. Strong, Frederick Finch (1908). High-frequency currents. Rebman Company. p. 41.
  2. 1 2 3 Curtis, Thomas Stanley (1916). High Frequency Apparatus: Its Construction and Practical Application. New York: Everyday Mechanics Co. Inc. pp.  45–47. Oudin coil.
  3. 1 2 3 4 5 Martin, James M. (1912). Practical electro-therapeutics and X-ray therapy. C.V. Mosby Co. pp.  187–192. Oudin coil.
  4. 1 2 3 4 5 6 Strong, Frederick Finch (1918). Essentials of modern electro-therapeutics (2 ed.). New York: Rebman Co. pp.  73–75. Oudin.
  5. Hickman, Bert (November 30, 2000). "Re: Oudin vs Tesla coils". Tesla Coil Mailing List. Chip Atkinson's website. Retrieved December 2, 2014.
  6. Manders, Horace (August 1, 1902). "Some phenomena of high frequency currents". Journal of Physical Therapeutics. London: John Bale, Sons, and Danielsson, Ltd. 3 (1): 226. Retrieved December 2, 2014.
  7. Filingeri, V.; Gravante, G.; Cassisa, D. (2005). "Physics of radiofrequency in proctology" (PDF). European Review for Medical and Pharmacological Sciences. 9 (6): 349. Retrieved April 11, 2015.
  8. Eberhart, Noble Murray (1919). A Working Manual of High Frequency Currents (5 ed.). Chicago: New Medicine Publishing Co. pp.  40. Oudin resonator.
  9. Brewster, Frank (June 1914). "High Frequency Current Apparatus Ch. 5: High Frequency Transformer". Modern Electrics and Magnetics. New York: Modern Publishing Co. 28 (6): 762–764. Retrieved December 2, 2014.
  10. Naidu, M. S.; Kamaraju, V. (2009). High Voltage Engineering. India: Tata McGraw-Hill Education. p. 170. ISBN   978-0070669284.
  11. 1 2 3 4 Behary, Jeff (July 1, 2007). "Re: Oudin coil". Tesla list (Mailing list). Retrieved April 10, 2015.
  12. Tousey, Sinclair (1921). Medical electricity Rontgen rays and Radium, 3rd Ed. W.B. Saunders Co. pp. 541, 548–550.
  13. Manders, 1902, Some phenomena of high frequency currents, p. 220-221
  14. Manders, 1902, Some phenomena of high frequency currents, p. 223
  15. Skelton, Ken. "Oudin Oscillator". Public Understanding of Science. Dept. of Physics and Astronomy, Univ. of Glasgow. Archived from the original on April 4, 2005. Retrieved April 10, 2015.