A deuterium arc lamp (or simply deuterium lamp) is a low-pressure gas-discharge light source often used in spectroscopy when a continuous spectrum in the ultraviolet region is needed.

Contents

Plasma "arc" or discharge lamps using hydrogen are notable for their high output in the ultraviolet, with comparatively little output in the visible and infrared. This is similar to the situation in a hydrogen flame. Arc lamps made with ordinary light-hydrogen (hydrogen-1) provide a very similar UV spectrum to deuterium, and have been used in UV spectroscopes. However, lamps using deuterium have a longer life span and an emissivity (intensity) at the far end of their UV range which is three to five times that of an ordinary hydrogen arc bulb, at the same temperature. Deuterium arc lamps, therefore, despite being several times more expensive, are considered a superior light source to light-hydrogen arc lamps, for the shortwave UV range.

Principle of operation

Emission spectrum of an ultraviolet deuterium arc lamp showing characteristic hydrogen Balmer lines (sharp peaks at 486 nm and 656 nm labeled Db and Da from left to right respectively), continuum emission in the ~160-400 nm region and Fulcher band emission between around 560 to 640 nm. The emission spectrum of deuterium differs slightly from that of protium (hydrogen-1) due to the influence of hyperfine interactions, though these effects alter the wavelength of the lines by mere fractions of a nanometer and are too fine to be discerned by the spectrometer used here. Deuterium is used rather than normal hydrogen because of its greater intensity of UV emission in the molecular band. Deuterium lamp 1.png
Emission spectrum of an ultraviolet deuterium arc lamp showing characteristic hydrogen Balmer lines (sharp peaks at 486 nm and 656 nm labeled Dβ and Dα from left to right respectively), continuum emission in the ~160-400 nm region and Fulcher band emission between around 560 to 640 nm. The emission spectrum of deuterium differs slightly from that of protium (hydrogen-1) due to the influence of hyperfine interactions, though these effects alter the wavelength of the lines by mere fractions of a nanometer and are too fine to be discerned by the spectrometer used here. Deuterium is used rather than normal hydrogen because of its greater intensity of UV emission in the molecular band.

A deuterium lamp uses a tungsten filament and anode placed on opposite sides of a nickel box structure designed to produce the best output spectrum. Unlike an incandescent bulb, the filament is not the source of light in deuterium lamps. Instead an arc is created from the filament to the anode, a similar process to arc lamps. Because the filament must be very hot before it can operate, it is heated for approximately 20 seconds before use. Because the discharge process produces its own heat, the heater is turned down after discharge begins. Although firing voltages are 300 to 500 volts, once the arc is created voltages drop to around 100 to 200 volts. [1]

The arc created excites the molecular deuterium contained within the bulb to a higher energy state. The deuterium then emits light as it transitions back to its initial state. This continuous cycle is the origin of the continuous UV radiation. This process is not the same as the process of decay of atomic excited states (atomic emission), where electrons are excited and then emit radiation. Instead, a molecular emission process, where radiative decay of excited states in molecular deuterium (D2 or 2H2), causes the effect.

The spectral line structure of deuterium does not differ noticeably from that of light hydrogen, but deuterium has a slightly stronger molecular bond (439.5 vs. 432 kJ/mol) and is less-well ionized at the temperature of the arc. This causes a larger population of molecules and a greater emissivity (light output) of UV in the molecular part of the spectrum that is farthest into the ultraviolet. [2]

Because the lamp operates at high temperatures, normal glass housings cannot be used for a casing. They would also block UV radiation. Instead, a fused quartz, UV glass, or magnesium fluoride envelope is used depending on the specific function of the lamp. [3]

The typical lifetime of a deuterium lamp is approximately 2000 hours (Most manufacturers guarantee 2000 hours, but newer lamps are consistently performing well at 5000 hours and more). [ citation needed ]

Deuterium lamp spectra

The deuterium lamp emits radiation extending from 112 nm to 900 nm, though its continuous spectrum is only from 180 nm to 370 nm. The spectrum intensity does not actually decrease from 250 nm to 200 nm as shown in the spectrum plot above. The decrease in the plot is due to decreased efficiency at low wavelengths of the photo detector used to measure the lamp intensity. The deuterium lamp's continuous spectrum is useful as both a reference in UV radiometric work and to generate a signal in various photometric devices.

Safety

Due to the high intensity of UV radiation emitted by the bulb, eye protection is suggested when using a deuterium bulb. Care must also be taken to avoid touching the bulb directly to avoid burns due to high operating temperatures. Touching the bulb directly even when cool could deposit impurities onto the casing that strongly absorb the short wavelength UV and therefore reduce output intensity.

Related Research Articles

<span class="mw-page-title-main">Atomic absorption spectroscopy</span> Type of spectroanalytical ciao

Atomic absorption spectroscopy (AAS) and atomic emission spectroscopy (AES) is a spectroanalytical procedure for the quantitative determination of chemical elements by free atoms in the gaseous state. Atomic absorption spectroscopy is based on absorption of light by free metallic ions.

<span class="mw-page-title-main">Ultraviolet</span> Energetic, invisible light energy range

Ultraviolet (UV) light is electromagnetic radiation of wavelengths of 10–400 nanometers, shorter than that of visible light, but longer than X-rays. UV radiation is present in sunlight, and constitutes about 10% of the total electromagnetic radiation output from the Sun. It is also produced by electric arcs, Cherenkov radiation, and specialized lights, such as mercury-vapor lamps, tanning lamps, and black lights.

<span class="mw-page-title-main">Arc lamp</span> Lamp that produces light by an electric arc

An arc lamp or arc light is a lamp that produces light by an electric arc.

<span class="mw-page-title-main">Fluorescent lamp</span> Lamp using fluorescence to produce light

A fluorescent lamp, or fluorescent tube, is a low-pressure mercury-vapor gas-discharge lamp that uses fluorescence to produce visible light. An electric current in the gas excites mercury vapor, which produces short-wave ultraviolet light that then causes a phosphor coating on the inside of the lamp to glow. A fluorescent lamp converts electrical energy into useful light much more efficiently than an incandescent lamp, but is less efficient than most LED lamps. The typical luminous efficacy of fluorescent lighting systems is 50–100 lumens per watt, several times the efficacy of incandescent bulbs with comparable light output. For comparison, the luminous efficiency of an incandescent bulb may only be 16 lumens per watt.

<span class="mw-page-title-main">Blacklight</span> Light fixture that emits long-wave ultraviolet light and very little visible light

A blacklight, also called a UV-A light, Wood's lamp, or ultraviolet light, is a lamp that emits long-wave (UV-A) ultraviolet light and very little visible light. One type of lamp has a violet filter material, either on the bulb or in a separate glass filter in the lamp housing, which blocks most visible light and allows through UV, so the lamp has a dim violet glow when operating. Blacklight lamps which have this filter have a lighting industry designation that includes the letters "BLB". This stands for "blacklight blue". A second type of lamp produces ultraviolet but does not have the filter material, so it produces more visible light and has a blue color when operating. These tubes are made for use in "bug zapper" insect traps, and are identified by the industry designation "BL". This stands for "blacklight".

<span class="mw-page-title-main">Flashtube</span> Incoherent light source

A flashtube (flashlamp) produces an electrostatic discharge with an extremely intense, incoherent, full-spectrum white light for a very short time. A flashtube is a glass tube with an electrode at each end and is filled with a gas that, when triggered, ionizes and conducts a high-voltage pulse to make light. Flashtubes are used most in photography; they also are used in science, medicine, industry, and entertainment.

<span class="mw-page-title-main">Fluorescence spectroscopy</span> Type of electromagnetic spectroscopy

Fluorescence spectroscopy is a type of electromagnetic spectroscopy that analyzes fluorescence from a sample. It involves using a beam of light, usually ultraviolet light, that excites the electrons in molecules of certain compounds and causes them to emit light; typically, but not necessarily, visible light. A complementary technique is absorption spectroscopy. In the special case of single molecule fluorescence spectroscopy, intensity fluctuations from the emitted light are measured from either single fluorophores, or pairs of fluorophores.

<span class="mw-page-title-main">Gas-filled tube</span> Assembly of electrodes at either end of an insulated tube filled with gas

A gas-filled tube, also commonly known as a discharge tube or formerly as a Plücker tube, is an arrangement of electrodes in a gas within an insulating, temperature-resistant envelope. Gas-filled tubes exploit phenomena related to electric discharge in gases, and operate by ionizing the gas with an applied voltage sufficient to cause electrical conduction by the underlying phenomena of the Townsend discharge. A gas-discharge lamp is an electric light using a gas-filled tube; these include fluorescent lamps, metal-halide lamps, sodium-vapor lamps, and neon lights. Specialized gas-filled tubes such as krytrons, thyratrons, and ignitrons are used as switching devices in electric devices.

<span class="mw-page-title-main">High-intensity discharge lamp</span> Type of electric lamp/bulb

High-intensity discharge lamps are a type of electrical gas-discharge lamp which produces light by means of an electric arc between tungsten electrodes housed inside a translucent or transparent fused quartz or fused alumina arc tube. This tube is filled with noble gas and often also contains suitable metal or metal salts. The noble gas enables the arc's initial strike. Once the arc is started, it heats and evaporates the metallic admixture. Its presence in the arc plasma greatly increases the intensity of visible light produced by the arc for a given power input, as the metals have many emission spectral lines in the visible part of the spectrum. High-intensity discharge lamps are a type of arc lamp.

<span class="mw-page-title-main">Mercury-vapor lamp</span> Light source using an electric arc through mercury vapor

A mercury-vapor lamp is a gas-discharge lamp that uses an electric arc through vaporized mercury to produce light. The arc discharge is generally confined to a small fused quartz arc tube mounted within a larger soda lime or borosilicate glass bulb. The outer bulb may be clear or coated with a phosphor; in either case, the outer bulb provides thermal insulation, protection from the ultraviolet radiation the light produces, and a convenient mounting for the fused quartz arc tube.

<span class="mw-page-title-main">Germicidal lamp</span> Ultraviolet C light-emitting device

A germicidal lamp is an electric light that produces ultraviolet C (UVC) light. This short-wave ultraviolet light disrupts DNA base pairing, causing formation of pyrimidine dimers, and leads to the inactivation of bacteria, viruses, and protozoans. It can also be used to produce ozone for water disinfection. They are used in ultraviolet germicidal irradiation (UVGI).

<span class="mw-page-title-main">Gas laser</span> Laser in which electricity is discharged through gas

A gas laser is a laser in which an electric current is discharged through a gas to produce coherent light. The gas laser was the first continuous-light laser and the first laser to operate on the principle of converting electrical energy to a laser light output. The first gas laser, the Helium–neon laser (HeNe), was co-invented by Iranian engineer and scientist Ali Javan and American physicist William R. Bennett, Jr., in 1960. It produced a coherent light beam in the infrared region of the spectrum at 1.15 micrometres.

<span class="mw-page-title-main">Hydrargyrum medium-arc iodide lamp</span>

Hydrargyrum medium-arc iodide (HMI) is the trademark name of Osram's brand of metal-halide gas discharge medium arc-length lamp, made specifically for film and entertainment applications. Hydrargyrum comes from the Greek name for the element mercury.

<span class="mw-page-title-main">Xenon arc lamp</span> Gas discharge lamp that produces intense white light

A xenon arc lamp is a highly specialized type of gas discharge lamp, an electric light that produces light by passing electricity through ionized xenon gas at high pressure. It produces a bright white light to simulate sunlight, with applications in movie projectors in theaters, in searchlights, and for specialized uses in industry and research. For example, Xenon arc lamps and mercury lamps are the two most common lamps used in wide-field fluorescence microscopes.

Luminous efficacy is a measure of how well a light source produces visible light. It is the ratio of luminous flux to power, measured in lumens per watt in the International System of Units (SI). Depending on context, the power can be either the radiant flux of the source's output, or it can be the total power consumed by the source. Which sense of the term is intended must usually be inferred from the context, and is sometimes unclear. The former sense is sometimes called luminous efficacy of radiation, and the latter luminous efficacy of a light source or overall luminous efficacy.

<span class="mw-page-title-main">Ultraviolet germicidal irradiation</span> Disinfection method using ultraviolet light

Ultraviolet germicidal irradiation (UVGI) is a disinfection technique employing ultraviolet (UV) light, particularly UV-C (180–280 nm), to kill or inactivate microorganisms. UVGI primarily inactivates microbes by damaging their genetic material, thereby inhibiting their capacity to carry out vital functions.

<span class="mw-page-title-main">Solar simulator</span> Device that approximates natural sunlight

A solar simulator is a device that provides illumination approximating natural sunlight. The purpose of the solar simulator is to provide a controllable indoor test facility under laboratory conditions. It can be used for the testing of any processes or materials that are photosensitive, including solar cells, sun screen, cosmetics, plastics, aerospace materials, skin cancer, bioluminescence, photosynthesis, water treatment, crude-oil degradation, and free radical formation. Solar simulators are used in a wide range of research areas including photobiology, photo-oxidation, photodegradation, photovoltaics, and photocatalysis.

<span class="mw-page-title-main">Dental curing light</span> Dental equipment

A dental curing light is a piece of dental equipment that is used for polymerization of light-cure resin-based composites. It can be used on several different dental materials that are curable by light. The light used falls under the visible blue light spectrum. This light is delivered over a range of wavelengths and varies for each type of device. There are four basic types of dental curing light sources: tungsten halogen, light-emitting diodes (LED), plasma arcs, and lasers. The two most common are halogen and LEDs.

An excimer lamp is a source of ultraviolet light based on spontaneous emission of excimer (exciplex) molecules.