Protein purification

Last updated

Protein purification is a series of processes intended to isolate one or a few proteins from a complex mixture, usually cells, tissues or whole organisms. Protein purification is vital for the specification of the function, structure and interactions of the protein of interest. The purification process may separate the protein and non-protein parts of the mixture, and finally separate the desired protein from all other proteins. Ideally, to study a protein of interest, it must be separated from other components of the cell so that contaminants will not interfere in the examination of the protein of interest's structure and function. [1] Separation of one protein from all others is typically the most laborious aspect of protein purification. Separation steps usually exploit differences in protein size, physico-chemical properties, binding affinity and biological activity. The pure result may be termed protein isolate.

Contents

Purpose

The protein manufacturing cost remains high and there is a growing demand to develop cost efficient and rapid protein purification methods. Understanding of the different protein purification methods and optimizing the downstream processing are critical to minimize production costs while maintaining the quality of acceptable standards of homogeneity. [2] Protein purification is either preparative or analytical. Preparative purifications aim to produce a relatively large quantity of purified proteins for subsequent use. Examples include the preparation of commercial products such as enzymes (e.g. lactase), nutritional proteins (e.g. soy protein isolate), and certain biopharmaceuticals (e.g. insulin). Several preparative purifications steps are often deployed to remove bi-products, such as host cell proteins, which poses as a potential threat to the patient's health. [3] Analytical purification produces a relatively small amount of a protein for a variety of research or analytical purposes, including identification, quantification, and studies of the protein's structure, post-translational modifications and function. Each step of a protein purification scheme is monitored and takes into consideration purification levels and yield. A high purification level and a poor yield leave hardly any protein with which to experiment. On the other hand, a high yield with low purification levels leave many contaminants (proteins other than the one interest) which interfere in research purposes. [1]

Recombinant bacteria can be grown in a flask containing growth media. Shakeflask.JPG
Recombinant bacteria can be grown in a flask containing growth media.

Preliminary steps

Extraction

If the protein of interest is not secreted by the organism into the surrounding solution, the first step of each purification process is the disruption of the cells containing the protein. Depending on how fragile the protein is and how stable the cells are, one could, for instance, use one of the following methods: i) repeated freezing and thawing, ii) sonication, iii) homogenization by high pressure (French press), iv) homogenization by grinding (bead mill), and v) permeabilization by detergents (e.g. Triton X-100) and/or enzymes (e.g. lysozyme). [4] Finally, the cell debris can be removed by differential centrifugation, which is a procedure where the homogenate is centrifuged at low speed, then again at a greater force to yield a pellet consisting of nuclei and supernatant. This yields several fractions of decreasing density where more discriminating purification techniques are applied to one fraction. [1]

Also proteases are released during cell lysis, which will start digesting the proteins in the solution. If the protein of interest is sensitive to proteolysis, it is recommended to proceed quickly, and to keep the extract cooled, to slow down the digestion. Alternatively, one or more protease inhibitors can be added to the lysis buffer immediately before cell disruption. Sometimes it is also necessary to add DNAse in order to reduce the viscosity of the cell lysate caused by a high DNA content.

Ultracentrifugation

Centrifugation is a process that uses centrifugal force to separate mixtures of particles of varying masses or densities suspended in a liquid. When a vessel (typically a tube or bottle) containing a mixture of proteins or other particulate matter, such as bacterial cells, is rotated at high speeds, the inertia of each particle yields a force in the direction of the particles velocity that is proportional to its mass. The tendency of a given particle to move through the liquid because of this force is offset by the resistance the liquid exerts on the particle. The net effect of "spinning" the sample in a centrifuge is that massive, small, and dense particles move outward faster than less massive particles or particles with more "drag" in the liquid. When suspensions of particles are "spun" in a centrifuge, a "pellet" may form at the bottom of the vessel that is enriched for the most massive particles with low drag in the liquid.

Non-compacted particles remain mostly in the liquid called "supernatant" and can be removed from the vessel thereby separating the supernatant from the pellet. The rate of centrifugation is determined by the angular acceleration applied to the sample, typically measured in comparison to the g. If samples are centrifuged long enough, the particles in the vessel will reach equilibrium wherein the particles accumulate specifically at a point in the vessel where their buoyant density is balanced with centrifugal force. Such an "equilibrium" centrifugation can allow extensive purification of a given particle.

Sucrose gradient centrifugation—a linear concentration gradient of sugar (typically sucrose, glycerol, or a silica based density gradient media, like Percoll)—is generated in a tube such that the highest concentration is on the bottom and lowest on top. Percoll is a trademark owned by GE Healthcare companies. A protein sample is then layered on top of the gradient and spun at high speeds in an ultracentrifuge. This causes heavy macromolecules to migrate towards the bottom of the tube faster than lighter material. During centrifugation in the absence of sucrose, as particles move farther and farther from the center of rotation, they experience more and more centrifugal force (the further they move, the faster they move). The problem with this is that the useful separation range within the vessel is restricted to a small observable window. Spinning a sample twice as long does not mean the particle of interest will go twice as far; in fact, it will go significantly further. However, when the proteins are moving through a sucrose gradient, they encounter liquid of increasing density and viscosity. A properly designed sucrose gradient will counteract the increasing centrifugal force so the particles move in close proportion to the time they have been in the centrifugal field. Samples separated by these gradients are referred to as "rate zonal" centrifugations. After separating the protein/particles, the gradient is then fractionated and collected. In biochemistry, ultra centrifugation is valuable for separating biomolecules and analyzing their physical properties. [1]

Purification strategies

Chromatographic equipment. Here set up for a size exclusion chromatography. The buffer is pumped through the column (right) by a computer controlled device. Size exclusion.jpg
Chromatographic equipment. Here set up for a size exclusion chromatography. The buffer is pumped through the column (right) by a computer controlled device.

Choice of a starting material is key to the design of a purification process. In a plant or animal, a particular protein usually is not distributed homogeneously throughout the body; different organs or tissues have higher or lower concentrations of the protein. Use of only the tissues or organs with the highest concentration decreases the volumes needed to produce a given amount of purified protein. If the protein is present in low abundance, or if it has a high value, scientists may use recombinant DNA technology to develop cells that will produce large quantities of the desired protein (this is known as an expression system). Recombinant expression allows the protein to be tagged, e.g. by a His-tag [5] or Strep-tag [6] to facilitate purification, reducing the number of purification steps required.

An analytical purification generally utilizes three properties to separate proteins. First, proteins may be purified according to their isoelectric points by running them through a pH graded gel or an ion exchange column. Second, proteins can be separated according to their size or molecular weight via size exclusion chromatography or by SDS-PAGE (sodium dodecyl sulfate-polyacrylamide gel electrophoresis) analysis. Proteins are often purified by using 2D-PAGE and are then analysed by peptide mass fingerprinting to establish the protein identity. This is very useful for scientific purposes and the detection limits for protein are nowadays very low and nanogram amounts of protein are sufficient for their analysis. Thirdly, proteins may be separated by polarity/hydrophobicity via high performance liquid chromatography or reversed-phase chromatography.

Usually a protein purification protocol contains one or more chromatographic steps. The basic procedure in chromatography is to flow the solution containing the protein through a column packed with various materials. Different proteins interact differently with the column material, and can thus be separated by the time required to pass the column, or the conditions required to elute the protein from the column. Usually proteins are detected as they are coming off the column by their absorbance at 280 nm. Many different chromatographic methods exist:

Precipitation and differential solubilization

Most proteins require some salt to dissolve in water, a process called salting in. As the salt concentration is increased, proteins can precipitate, a process called salting out which involves changing protein solubility. [1] For example, in bulk protein purification, a common first step to isolate proteins is precipitation with ammonium sulfate (NH4)2SO4. [7] This is performed by adding increasing amounts of ammonium sulfate and collecting the different fractions of precipitated protein. Subsequently, ammonium sulfate can be removed using dialysis (separating proteins from small molecules through a semipermeable membrane). [1] During the ammonium sulfate precipitation step, hydrophobic groups present on the proteins are exposed to the atmosphere, attracting other hydrophobic groups; the result is formation of an aggregate of hydrophobic components. In this case, the protein precipitate will typically be visible to the naked eye. One advantage of this method is that it can be performed inexpensively, even with very large volumes.

The first proteins to be purified are water-soluble proteins. Purification of integral membrane proteins requires disruption of the cell membrane in order to isolate any one particular protein from others that are in the same membrane compartment. Sometimes a particular membrane fraction can be isolated first, such as isolating mitochondria from cells before purifying a protein located in a mitochondrial membrane. A detergent such as sodium dodecyl sulfate (SDS) can be used to dissolve cell membranes and keep membrane proteins in solution during purification; however, because SDS causes denaturation, milder detergents such as Triton X-100 or CHAPS can be used to retain the protein's native conformation during complete purification.

Size exclusion (gel-filtration chromatography)

Chromatography can be used to separate protein in solution or denaturing conditions by using porous gels. This technique is a more discriminating separation and is known as size exclusion chromatography. The principle is that smaller molecules have to traverse a larger volume in a porous matrix. Consequentially, proteins of a certain range in size will require a variable volume of eluent (solvent) before being collected at the other end of the column of gel.

In the context of protein purification, the eluent is usually pooled in different test tubes. All test tubes containing no measurable trace of the protein to purify are discarded. The remaining solution is thus made of the protein to purify and any other similarly-sized proteins.

Separation based on charge (ion-exchange chromatography)

One chromatography technique based on molecular properties is usually not sufficient in obtaining a protein of high purity. [1] In addition to size, ion exchange chromatography separates compounds according to the nature and degree of their ionic charge. The column to be used is selected according to its type and strength of charge. Anion exchange resins have a positive charge and are used to retain and separate negatively charged compounds (anions), while cation exchange resins have a negative charge and are used to separate positively charged molecules (cations).

Before the separation begins a buffer is pumped through the column to equilibrate the opposing charged ions. Upon injection of the sample, solute molecules will exchange with the buffer ions as each competes for the binding sites on the resin. The length of retention for each solute depends upon the strength of its charge. The most weakly charged compounds will elute first, followed by those with successively stronger charges. Because of the nature of the separating mechanism, pH, buffer type, buffer concentration, and temperature all play important roles in controlling the separation.

Ion exchange chromatography is a very powerful tool for use in protein purification and is frequently used in both analytical and preparative separations.

Nickel-affinity column. The resin is blue since it has bound nickel. Metal-affinity column.jpg
Nickel-affinity column. The resin is blue since it has bound nickel.

Free-flow-electrophoresis

Free-flow electrophoresis (FFE) is a carrier-free electrophoresis technique that allows preparative protein separation in a laminar buffer stream by using an orthogonal electric field. By making use of a pH-gradient, that can for example be induced by ampholytes, this technique allows to separate protein isoforms up to a resolution of < 0.02 delta-pI.

Separation based on hydrophobicity (hydrophobic interaction chromatography)

HIC media is amphiphilic, with both hydrophobic and hydrophilic regions, allowing for separation of proteins based on their surface hydrophobicity. Target proteins and their product aggregate species tend to have different hydrophobic properties and removing them via HIC further purifies the protein of interest. [8] Additionally, the environment used typically employs less harsh denaturing conditions than other chromatography techniques, thus helping to preserve the protein of interest in its native and functional state. In pure water, the interactions between the resin and the hydrophobic regions of protein would be very weak, but this interaction is enhanced by applying a protein sample to HIC resin in high ionic strength buffer. The ionic strength of the buffer is then reduced to elute proteins in order of decreasing hydrophobicity. [9]

Affinity chromatography

Affinity Chromatography is another powerful separation technique that is highly selective for the protein of interest based upon molecular conformation, which frequently utilizes application specific resins. These resins have ligands attached to their surfaces which are specific for the compounds to be separated. Most frequently, these ligands function in a fashion similar to that of antibody-antigen interactions. This "lock and key" fit between the ligand and its target compound makes it highly specific, frequently generating a single peak, while all else in the sample is unretained.

Many membrane proteins are glycoproteins and can be purified by lectin affinity chromatography. Detergent-solubilized proteins can be allowed to bind to a chromatography resin that has been modified to have a covalently attached lectin. Proteins that do not bind to the lectin are washed away and then specifically bound glycoproteins can be eluted by adding a high concentration of a sugar that competes with the bound glycoproteins at the lectin binding site. Some lectins have high affinity binding to oligosaccharides of glycoproteins that is hard to compete with sugars, and bound glycoproteins need to be released by denaturing the lectin.

Immunoaffinity chromatography

An HPLC. From left to right: A pumping device generating a gradient of two different solvents, a steel enforced column and an apparatus for measuring the absorbance. Hplc.JPG
An HPLC. From left to right: A pumping device generating a gradient of two different solvents, a steel enforced column and an apparatus for measuring the absorbance.

Immunoaffinity chromatography uses the specific binding of an antibody-antigen to selectively purify the target protein. The procedure involves immobilizing a protein to a solid substrate (e.g. a porous bead or a membrane), which then selectively binds the target, while everything else flows through. The target protein can be eluted by changing the pH or the salinity. The immobilized ligand can be an antibody (such as immunoglobulin G) or it can be a protein (such as protein A). Because this method does not involve engineering in a tag, it can be used for proteins from natural sources. [10]

HPLC

High performance liquid chromatography or high pressure liquid chromatography is a form of chromatography applying high pressure to drive the solutes through the column faster. This means that the diffusion is limited and the resolution is improved. The most common form is "reversed phase" HPLC, where the column material is hydrophobic. The proteins are eluted by a gradient of increasing amounts of an organic solvent, such as acetonitrile. The proteins elute according to their hydrophobicity. After purification by HPLC the protein is in a solution that only contains volatile compounds, and can easily be lyophilized. [11] HPLC purification frequently results in denaturation of the purified proteins and is thus not applicable to proteins that do not spontaneously refold.

Purification of a tagged protein

Another way to tag proteins is to engineer an antigen peptide tag onto the protein, and then purify the protein on a column or by incubating with a loose resin that is coated with an immobilized antibody. This particular procedure is known as immunoprecipitation. Immunoprecipitation is capable of generating an extremely specific interaction which usually results in binding only the desired protein. The purified tagged proteins can then easily be separated from the other proteins in solution and later eluted back into clean solution.

When the tags are not needed anymore, they can be cleaved off by a protease. This often involves engineering a protease cleavage site between the tag and the protein.

Note, that self-cleaving tag eliminates a need of using proteases to separate tag from target protein of interest during purification process (e.g.iCapTag™). [12] [13] The main component of the tag is an intein, which cleaves off simply after pH change. Tagless and pure target protein is then released into the elution buffer.

Concentration of the purified protein

A selectively permeable membrane can be mounted in a centrifuge tube. The buffer is forced through the membrane by centrifugation, leaving the protein in the upper chamber. Ultra filtration.JPG
A selectively permeable membrane can be mounted in a centrifuge tube. The buffer is forced through the membrane by centrifugation, leaving the protein in the upper chamber.

At the end of a protein purification, the protein often has to be concentrated. Different methods exist.

Lyophilization

If the solution doesn't contain any other soluble component than the protein in question the protein can be lyophilized (dried). This is commonly done after an HPLC run. This simply removes all volatile components, leaving the proteins behind.

Ultrafiltration

Ultrafiltration concentrates a protein solution using selective permeable membranes. The function of the membrane is to let the water and small molecules pass through while retaining the protein. The solution is forced against the membrane by mechanical pump, gas pressure, or centrifugation.

Evaluating purification yield

The most general method to monitor the purification process is by running a SDS-PAGE of the different steps. This method only gives a rough measure of the amounts of different proteins in the mixture, and it is not able to distinguish between proteins with similar apparent molecular weight.

If the protein has a distinguishing spectroscopic feature or an enzymatic activity, this property can be used to detect and quantify the specific protein, and thus to select the fractions of the separation, that contains the protein. If antibodies against the protein are available then western blotting and ELISA can specifically detect and quantify the amount of desired protein. Some proteins function as receptors and can be detected during purification steps by a ligand binding assay, often using a radioactive ligand.

In order to evaluate the process of multistep purification, the amount of the specific protein has to be compared to the amount of total protein. The latter can be determined by the Bradford total protein assay or by absorbance of light at 280 nm, however some reagents used during the purification process may interfere with the quantification. For example, imidazole (commonly used for purification of polyhistidine-tagged recombinant proteins) is an amino acid analogue and at low concentrations will interfere with the bicinchoninic acid (BCA) assay for total protein quantification. Impurities in low-grade imidazole will also absorb at 280 nm, resulting in an inaccurate reading of protein concentration from UV absorbance.

Another method to be considered is surface plasmon resonance (SPR). SPR can detect binding of label free molecules on the surface of a chip. If the desired protein is an antibody, binding can be translated directly to the activity of the protein. One can express the active concentration of the protein as the percent of the total protein. SPR can be a powerful method for quickly determining protein activity and overall yield. It is a powerful technology that requires an instrument to perform.

Analytical

Denaturing-condition electrophoresis

Gel electrophoresis is a common laboratory technique that can be used both as preparative and analytical method. The principle of electrophoresis relies on the movement of a charged ion in an electric field. In practice, the proteins are denatured in a solution containing a detergent (SDS). In these conditions, the proteins are unfolded and coated with negatively charged detergent molecules. The proteins in SDS-PAGE are separated on the sole basis of their size.

In analytical methods, the protein migrate as bands based on size. Each band can be detected using stains such as Coomassie blue dye or silver stain. Preparative methods to purify large amounts of protein, require the extraction of the protein from the electrophoretic gel. This extraction may involve excision of the gel containing a band, or eluting the band directly off the gel as it runs off the end of the gel.

In the context of a purification strategy, denaturing condition electrophoresis provides an improved resolution over size exclusion chromatography, but does not scale to large quantity of proteins in a sample as well as the late chromatography columns.

Non-denaturing-condition electrophoresis

A non-denaturing electrophoretic procedure for isolating bioactive metalloproteins in complex protein mixtures is preparative native PAGE. The intactness or the structural integrity of the isolated protein has to be confirmed by an independent method. [14]

See also

Related Research Articles

<span class="mw-page-title-main">Agarose</span> Heteropolysaccharide found in red algae

Agarose is a heteropolysaccharide, generally extracted from certain red seaweed. It is a linear polymer made up of the repeating unit of agarobiose, which is a disaccharide made up of D-galactose and 3,6-anhydro-L-galactopyranose. Agarose is one of the two principal components of agar, and is purified from agar by removing agar's other component, agaropectin.

In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent called the mobile phase, which carries it through a system on which a material called the stationary phase is fixed. Because the different constituents of the mixture tend to have different affinities for the stationary phase and are retained for different lengths of time depending on their interactions with its surface sites, the constituents travel at different apparent velocities in the mobile fluid, causing them to separate. The separation is based on the differential partitioning between the mobile and the stationary phases. Subtle differences in a compound's partition coefficient result in differential retention on the stationary phase and thus affect the separation.

<span class="mw-page-title-main">Size-exclusion chromatography</span> Chromatographic method in which dissolved molecules are separated by their size & molecular weight

Size-exclusion chromatography, also known as molecular sieve chromatography, is a chromatographic method in which molecules in solution are separated by their size, and in some cases molecular weight. It is usually applied to large molecules or macromolecular complexes such as proteins and industrial polymers. Typically, when an aqueous solution is used to transport the sample through the column, the technique is known as gel-filtration chromatography, versus the name gel permeation chromatography, which is used when an organic solvent is used as a mobile phase. The chromatography column is packed with fine, porous beads which are commonly composed of dextran, agarose, or polyacrylamide polymers. The pore sizes of these beads are used to estimate the dimensions of macromolecules. SEC is a widely used polymer characterization method because of its ability to provide good molar mass distribution (Mw) results for polymers.

<span class="mw-page-title-main">High-performance liquid chromatography</span> Technique in analytical chemistry

High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify specific components in mixtures. The mixtures can originate from food, chemicals, pharmaceuticals, biological, environmental and agriculture, etc, which have been dissolved into liquid solutions.

<span class="mw-page-title-main">Centrifugation</span> Mechanical process

Centrifugation is a mechanical process which involves the use of the centrifugal force to separate particles from a solution according to their size, shape, density, medium viscosity and rotor speed. The denser components of the mixture migrate away from the axis of the centrifuge, while the less dense components of the mixture migrate towards the axis. Chemists and biologists may increase the effective gravitational force of the test tube so that the precipitate (pellet) will travel quickly and fully to the bottom of the tube. The remaining liquid that lies above the precipitate is called a supernatant or supernate.

<span class="mw-page-title-main">His-tag</span>

A polyhistidine-tag, best known by the trademarked name His-tag, is an amino acid motif in proteins that typically consists of at least six histidine (His) residues, often at the N- or C-terminus of the protein. It is also known as a hexa histidine-tag, 6xHis-tag, or His6 tag. The tag was invented by Roche, although the use of histidines and its vectors are distributed by Qiagen. Various purification kits for histidine-tagged proteins are commercially available from multiple companies.

The first isolation of deoxyribonucleic acid (DNA) was done in 1869 by Friedrich Miescher. DNA extraction is the process of isolating DNA from the cells of an organism isolated from a sample, typically a biological sample such as blood, saliva, or tissue. It involves breaking open the cells, removing proteins and other contaminants, and purifying the DNA so that it is free of other cellular components. The purified DNA can then be used for downstream applications such as PCR, sequencing, or cloning. Currently, it is a routine procedure in molecular biology or forensic analyses.

Affinity chromatography is a method of separating a biomolecule from a mixture, based on a highly specific macromolecular binding interaction between the biomolecule and another substance. The specific type of binding interaction depends on the biomolecule of interest; antigen and antibody, enzyme and substrate, receptor and ligand, or protein and nucleic acid binding interactions are frequently exploited for isolation of various biomolecules. Affinity chromatography is useful for its high selectivity and resolution of separation, compared to other chromatographic methods.

<span class="mw-page-title-main">Column chromatography</span> Method to isolate a compound in a mixture

Column chromatography in chemistry is a chromatography method used to isolate a single chemical compound from a mixture. Chromatography is able to separate substances based on differential adsorption of compounds to the adsorbent; compounds move through the column at different rates, allowing them to be separated into fractions. The technique is widely applicable, as many different adsorbents can be used with a wide range of solvents. The technique can be used on scales from micrograms up to kilograms. The main advantage of column chromatography is the relatively low cost and disposability of the stationary phase used in the process. The latter prevents cross-contamination and stationary phase degradation due to recycling. Column chromatography can be done using gravity to move the solvent, or using compressed gas to push the solvent through the column.

<span class="mw-page-title-main">Ion chromatography</span> Separates ions and polar molecules

Ion chromatography is a form of chromatography that separates ions and ionizable polar molecules based on their affinity to the ion exchanger. It works on almost any kind of charged molecule—including small inorganic anions, large proteins, small nucleotides, and amino acids. However, ion chromatography must be done in conditions that are one pH unit away from the isoelectric point of a protein.

<span class="mw-page-title-main">Fast protein liquid chromatography</span>

Fast protein liquid chromatography (FPLC) is a form of liquid chromatography that is often used to analyze or purify mixtures of proteins. As in other forms of chromatography, separation is possible because the different components of a mixture have different affinities for two materials, a moving fluid and a porous solid. In FPLC the mobile phase is an aqueous buffer solution. The buffer flow rate is controlled by a positive-displacement pump and is normally kept constant, while the composition of the buffer can be varied by drawing fluids in different proportions from two or more external reservoirs. The stationary phase is a resin composed of beads, usually of cross-linked agarose, packed into a cylindrical glass or plastic column. FPLC resins are available in a wide range of bead sizes and surface ligands depending on the application.

Reversed-phase liquid chromatography (RP-LC) is a mode of liquid chromatography in which non-polar stationary phase and polar mobile phases are used for the separation of organic compounds. The vast majority of separations and analyses using high-performance liquid chromatography (HPLC) in recent years are done using the reversed phase mode. In the reversed phase mode, the sample components are retained in the system the more hydrophobic they are.

DNA separation by silica adsorption is a method of DNA separation that is based on DNA molecules binding to silica surfaces in the presence of certain salts and under certain pH conditions.

Chromatography is a physical method of separation that distributes the components you want to separate between two phases, one stationary, the other moving in a definite direction. Cold ethanol precipitation, developed by Cohn in 1946, manipulates pH, ionic strength, ethanol concentration and temperature to precipitate different protein fractions from plasma. Chromatographic techniques utilise ion exchange, gel filtration and affinity resins to separate proteins. Since the 1980s it has emerged as an effective method of purifying blood components for therapeutic use.

Displacement chromatography is a chromatography technique in which a sample is placed onto the head of the column and is then displaced by a solute that is more strongly sorbed than the components of the original mixture. The result is that the components are resolved into consecutive "rectangular" zones of highly concentrated pure substances rather than solvent-separated "peaks". It is primarily a preparative technique; higher product concentration, higher purity, and increased throughput may be obtained compared to other modes of chromatography.

<span class="mw-page-title-main">Spin column-based nucleic acid purification</span>

Spin column-based nucleic acid purification is a solid phase extraction method to quickly purify nucleic acids. This method relies on the fact that nucleic acid will bind to the solid phase of silica under certain conditions.

Blood plasma fractionation are the general processes separating the various components of blood plasma, which in turn is a component of blood obtained through blood fractionation. Plasma-derived immunoglobulins are giving a new narrative to healthcare across a wide range of autoimmune inflammatory diseases. This widespread applicability is anticipated to leverage market prospects for plasma fractionation, pegged to witness a noteworthy 7% CAGR. COVID-19 pandemic is expected to generate growth opportunities for the plasma fractionation market.

<span class="mw-page-title-main">Desalting and buffer exchange</span>

Desalting and buffer exchange are methods to separate soluble macromolecules from smaller molecules (desalting) or replace the buffer system used for another one suitable for a downstream application. These methods are based on gel filtration chromatography, also called molecular sieve chromatography, which is a form of size-exclusion chromatography. Desalting and buffer exchange are two of the most common gel filtration chromatography applications, and they can be performed using the same resin.

Anion-exchange chromatography is a process that separates substances based on their charges using an ion-exchange resin containing positively charged groups, such as diethyl-aminoethyl groups (DEAE). In solution, the resin is coated with positively charged counter-ions (cations). Anion exchange resins will bind to negatively charged molecules, displacing the counter-ion. Anion exchange chromatography is commonly used to purify proteins, amino acids, sugars/carbohydrates and other acidic substances with a negative charge at higher pH levels. The tightness of the binding between the substance and the resin is based on the strength of the negative charge of the substance.

<span class="mw-page-title-main">SDS-PAGE</span> Biochemical technique

SDS-PAGE is a discontinuous electrophoretic system developed by Ulrich K. Laemmli which is commonly used as a method to separate proteins with molecular masses between 5 and 250 kDa. The combined use of sodium dodecyl sulfate and polyacrylamide gel eliminates the influence of structure and charge, and proteins are separated by differences in their size. At least up to 2012, the publication describing it was the most frequently cited paper by a single author, and the second most cited overall.

References

  1. 1 2 3 4 5 6 7 Berg, Jeremy M.; Tymoczko, John L.; Stryer, Lubert (2012). Biochemistry (7th ed.). New York: W. H. Freeman and Company. pp. 69–70. ISBN   978-1-4292-2936-4.
  2. Labrou, N. E. (2021). "Protein Purification Technologies". Protein Downstream Processing. Methods in Molecular Biology. Vol. 2178. pp. 3–10. doi:10.1007/978-1-0716-0775-6_1. ISBN   978-1-0716-0774-9. PMID   33128738. S2CID   226224596.
  3. Wang X, Hunter AK, Mozier NM (June 2009). "Host cell proteins in biologics development: Identification, quantitation and risk assessment". Biotechnology and Bioengineering. 103 (3): 446–58. doi: 10.1002/bit.22304 . PMID   19388135.
  4. Scopes RK (1994). Protein Purification - Springer. doi:10.1007/978-1-4757-2333-5. ISBN   978-1-4419-2833-7.
  5. Spriestersbach A, Kubicek J, Schäfer F, Block H, Maertens B (2015). "Purification of His-Tagged Proteins". Laboratory Methods in Enzymology: Protein Part D. Methods in Enzymology. Vol. 559. Elsevier. pp. 1–15. doi:10.1016/bs.mie.2014.11.003. ISBN   978-0-12-800279-7. PMID   26096499.
  6. Schmidt TG, Skerra A (2007). "The Strep-tag system for one-step purification and high-affinity detection or capturing of proteins". Nature Protocols. 2 (6): 1528–35. doi:10.1038/nprot.2007.209. PMID   17571060. S2CID   25209313.
  7. Wingfield P (May 2001). "Protein precipitation using ammonium sulfate". Current Protocols in Protein Science. Appendix 3: A.3F.1–A.3F.8. doi:10.1002/0471140864.psa03fs13. ISBN   978-0-471-14086-3. PMC   4817497 . PMID   18429073.
  8. McCue, Justin T. (2009). Theory and use of hydrophobic interaction chromatography in protein purification applications. Methods in Enzymology. Vol. 463. pp. 405–414. doi:10.1016/S0076-6879(09)63025-1. ISBN   978-0-12-374536-1. ISSN   1557-7988. PMID   19892185.
  9. Kennedy, RM (1990). Hydrophobic chromatography. Methods in Enzymology. Vol. 182. pp. 339–43. doi:10.1016/0076-6879(90)82029-2. ISBN   978-0-12-182083-1. PMID   2314246.
  10. Ehle H, Horn A (1990). "Immunoaffinity chromatography of enzymes". Bioseparation. 1 (2): 97–110. PMID   1368167.
  11. Regnier FE (October 1983). "High-performance liquid chromatography of biopolymers". Science. 222 (4621): 245–52. Bibcode:1983Sci...222..245R. doi:10.1126/science.6353575. PMID   6353575.
  12. Cooper, Merideth A.; Taris, Joseph E.; Shi, Changhua; Wood, David W. (2018). "A Convenient Split‐Intein Tag Method for the Purification of Tagless Target Proteins" . Current Protocols in Protein Science. 91 (1): 5.29.1–5.29.23. doi:10.1002/cpps.46. ISSN   1934-3655. PMID   29516483. S2CID   3749506.
  13. Prabhala, Sai Vivek; Gierach, Izabela; Wood, David W. (2022). "The Evolution of Intein-Based Affinity Methods as Reflected in 30 years of Patent History". Frontiers in Molecular Biosciences. 9: 857566. doi: 10.3389/fmolb.2022.857566 . ISSN   2296-889X. PMC   9033041 . PMID   35463948.
  14. Kastenholz B (2004). "Preparative native continuous polyacrylamide gel electrophoresis (PNC‐PAGE): an efficient method for isolating cadmium cofactors in biological systems". Analytical Letters. 37 (4): 657–665. doi:10.1081/AL-120029742. S2CID   97636537.